Batch records and documentation in FDA Form 483 observations
Master and batch production records, device history records, logbooks and good documentation practice. FDAgent finds 241 such observations in 143 Form 483s issued to 137 facilities, for inspections that ended between October 2008 and September 2026. Each is quoted below from the 483 FDA posted.
A Form 483 lists the investigators' observations at the end of an inspection. It is not FDA's final determination.
Source: Form 483 PDFs in FDA's Electronic Reading Room, last synced Nov 23, 2026, 04:33 UTC. Every quote is from a 483 whose reading passed FDAgent's checks against the scan. FDAgent is independent of FDA.
- Observations
- 241
- Each quoted below
- Form 483s
- 143
- Posted by FDA
- Facilities
- 137
- Distinct FEIs
- Most recent
- Sep 28, 2026
- Inspection end date
Observations by year
These observations by the year the inspection ended. They follow which 483s FDA has posted and FDAgent has read, so they show what is on record, not a trend in FDA's findings. 2026 runs to the latest sync (lighter bar).
Peak: 80 in 2026
| Year | Observations |
|---|---|
| 2008 | 3 |
| 2009 | 0 |
| 2010 | 1 |
| 2011 | 0 |
| 2012 | 4 |
| 2013 | 4 |
| 2014 | 0 |
| 2015 | 3 |
| 2016 | 16 |
| 2017 | 12 |
| 2018 | 15 |
| 2019 | 13 |
| 2020 | 0 |
| 2021 | 7 |
| 2022 | 18 |
| 2023 | 22 |
| 2024 | 15 |
| 2025 | 28 |
| 2026 | 80 |
How FDAgent sorts observations into this theme
Two fixed rules, applied by code to every observation FDAgent can quote. An observation counts when either holds. No model and no similarity score decides.
It is tied to one of these 21 CFR entries
21 CFR 211.18021 CFR 211.18421 CFR 211.18621 CFR 211.18821 CFR 820.18121 CFR 820.184
A 483 prints no regulation, so FDAgent ties an observation to a section only when it opens with FDA's own citation wording for that section, word for word.
Or its text contains one of these phrases
“batch record*” · “batch production … record*” · “master production … record*” · “master batch record*” · “device history record*” · “device master record*” · “documentation practice*” · “logbook*” · “log book*” · “BPR*” · “MPR*” · “BMR*”
* matches any ending and … up to three words in between. Case is ignored unless a phrase has capitals.
The 241 observations, newest first
Quoted word for word from each Form 483, with only line breaks joined; … marks where the quote is cut. The highlight shows why the observation is in this theme. Each links to FDAgent's page for the 483.
Cadila Healthcare LimitedInspection ended Mar 10, 2022 · FEI 3013712903 · Observation 2
OBSERVATION 2 Written procedures are not followed for evaluations conducted at least annually to review records associated with a representative number of batches, whether approved or rejected. Specifically, written procedure pertaining to the Annual Product Quality Review of the commercial drug products manufactured at the site is not followed. The firm performs Annual Product Quality Reviews (APQR) as per SOP, "Annual Product Quality Review", …
Opens with FDA's citation wording for 21 CFR 211.180(e)(1)Read the Form 483 →
Zhejiang Tianyu Pharmaceutical Co. Ltd.Inspection ended Mar 4, 2022 · FEI 3010972581 · Observation 3
… A review of your usage logbook identifies cleaning occurred on the following dates, outside of the specified time-frame: [illegible] identified as located in the Workshop II used for the synthesis of [illegible] were observed with an apparent brownish rust like (rouge) spots on the top walls of the [illegible] on product contact surface areas. In addition, layers of thick black spots are above the brownish rust like (rouge) spots around the walls …
Contains “logbook*”Read the Form 483 →
Davis City Pharmacy Inc.Inspection ended Feb 18, 2022 · FEI 3016774626 · Observation 5
Batch production and control records do not include complete information relating to the production and control of each batch. Specifically, Your firm's production records (formula worksheets) for Benzocaine 20%, Lidocaine 10%, Tetracaine 10% cream, Lot #09302021 and Lot #20101002012022 are deficient because they do not contain the following information: e Order of mixing ingredients e Step by step instructions ® Procedures to ensure proper …
Opens with FDA's citation wording for 21 CFR 211.188Read the Form 483 →
Late Night Concepts Inc. dba Brux Night GuardInspection ended Oct 28, 2021 · FEI 3015547648 · Observation 2
Procedures for device history records have not been established. Specifically, a Device History Record has not been maintained for Brux Night Guard.
Opens with FDA's citation wording for 21 CFR 820.184Read the Form 483 →
Late Night Concepts Inc. dba Brux Night GuardInspection ended Oct 28, 2021 · FEI 3015547648 · Observation 3
A device master record has not been maintained. Specifically, a Device Master Record has not been maintained for Brux Night Guard.
Opens with FDA's citation wording for 21 CFR 820.181Read the Form 483 →
Fujifilm Diosynth Biotechnologies USAInspection ended Apr 21, 2021 · FEI 3000718852 · Observation 2
… Installation of the valve to remove excess air is not specified in the batch record for this unit operation.
Contains “batch record*”Read the Form 483 →
Emergent Manufacturing Operations Baltimore, LLC.Inspection ended Apr 20, 2021 · FEI 3015448605 · Observation 1
… The deviation did not include consideration of operator (b) (6) who is recorded on the batch record as weighing and dispensing the raw materials for media batch (b) (4) used in the manufacture of (b) (4) batch (b) (4) on (b) (4). This batch of media is implicated by your firm in the deviation as the most probable cause of the cross-contamination event. Operator (b) (6) also entered both manufacturing areas where client (b) (4) viral vaccine drug …
Contains “batch record*”Read the Form 483 →
Emergent Manufacturing Operations Baltimore, LLC.Inspection ended Apr 20, 2021 · FEI 3015448605 · Observation 4
… According to the security badge access log, security camera footage, and batch record (b) (4) on (b) (4), a manufacturing associate (Operator upstream MFG) was observed entering manufacturing Area (b) (4) when manufacturing for client (b) (4) was taking place, then (b) (4) for raw materials for client (b) (4), and then loading of materials into the (b) (4) in manufacturing Area (b) (4) for client (b) (4) without documenting de-gowning and …
Contains “batch record*”Read the Form 483 →
Imprimis NJOF, -LLCInspection ended Jan 29, 2021 · FEI 3013024146 · Observation 1
(REPEAT OBSERVATION) There is a failure to thoroughly review any unexplained discrepancy, the failure of a batch or any of its components to meet any of its specifications whether or not the batch has been already distributed. Specifically, A) During the review of Adverse Event Investigation ADE2020081 and associated batch record for Dexamethasone/Moxifloxacin, 1/5 mg/ml, PF Injectable solution, lot #(b) (4) it was discovered that the product lot …
Contains “batch record*”Read the Form 483 →
Imprimis NJOF, -LLCInspection ended Jan 29, 2021 · FEI 3013024146 · Observation 2
… Batch records used to document bulk hold time (using the same PV protocol) incorrectly referenced Moxifloxacin and Bromfenac Phase and wrong lot number (b) (4) throughout the batch records for Phase A (Bromfenac). This was neither caught nor corrected during quality review of the batch records. PV batches (b) (4) were released for commercial distribution. In addition, approximately more commercial batches were released for commercial distribution …
Contains “batch record*”Read the Form 483 →
Abilene Nuclear LLC dba National Central PharmacyInspection ended Nov 7, 2019 · FEI 3004819860 · Observation 8
OBSERVATION 8 The batch production and control records are deficient in that they do not include identification of persons performing and checking each significant step in the operation. Specifically, your firm's production worksheets failed to include a second verifier for critical steps in your pharmacy's production process. For example, in selected production batch records, I found only your firm's pharmacist (the only firm employee) who …
Opens with FDA's citation wording for 21 CFR 211.188(b)(11)Read the Form 483 →
SCA Pharmaceuticals, Inc.Inspection ended Nov 7, 2019 · FEI 3010683157 · Observation 3
Appropriate controls are not exercised over computers or related systems to assure that changes in master production and control records or other records are instituted only by authorized personnel. Specifically, you have never conducted an audit trail review of the SCANRDI system since you began using it in 2013 for sterility analysis nor do you have a procedure for conducting an audit trail review. Additionally, an employee who performs most of …
Contains “master production … record*”Read the Form 483 →
SCA Pharmaceuticals, Inc.Inspection ended Nov 7, 2019 · FEI 3010683157 · Observation 4
The results of the examination of the packaged and labeled products were not documented in the batch production or control records. Specifically, I observed during the visual inspection by employees rejects from the 100% and the AQL were placed in the same bin, thus not allowing for accurate data to be entered in the batch records to determine if both passed your specification for visual inspection as defined in SOP COM-013-LR. For the AQL, …
Contains “batch production … record*”Read the Form 483 →
GPC Medical LimitedInspection ended Oct 24, 2019 · FEI 3006513362 · Observation 3
Procedures for Corrective and Preventive Action have not been adequately established. CAPA procedure / QSP / 12 identifies product non conformance (receiving and finished product) as a Quality Data source for CAPA. The firm log book of product non conformance is not analyzed to identify recurring non conformance for Corrective actions. There have been no CAPA actions for product non conformance in the last three years.
Contains “log book*”Read the Form 483 →
Isorx, Corp.Inspection ended Sep 18, 2019 · FEI 3006624492 · Observation 4
OBSERVATION 4 Batch production and control records are not prepared for each batch of drug product produced and do not include complete information relating to the production and control of each batch. Specifically, You do not maintain production records for the manufacture of drug products. Also your production records do not include a quality verification that the steps of production have been performed correctly. One example includes 1-123 …
Opens with FDA's citation wording for 21 CFR 211.188Read the Form 483 →
Rapid Equine Solutions, LLCInspection ended Aug 9, 2019 · FEI 3006708880 · Observation 6
The batch production and control records are deficient in that they do not include documentation of the accomplishment of each significant step in manufacturing, processing, packing, and holding. Specifically, your firm uses the (b) (4) computer system to create and maintain production records. Batch templates are printed out and each page is individually placed into plastic zip closure bags for reference during sterile production procedures. …
Opens with FDA's citation wording for 21 CFR 211.188(b)Read the Form 483 →
Cipla Ltd. Virgonagar SiteInspection ended Jul 19, 2019 · FEI 3002806691 · Observation 2
… parameter’ rate ( L/hr) for the during the manufacture o API intermediate, which is further manufactured into APIs and Currently you use a system for in the BMR for both E4®)#)-04 and E3. -19, but have never demonstrated that the the®@) 1 is consistent with the rate by validating it against a| Note that this is (0) (4) and can easily form undesired side ®)@) "| products if the ®@) | rate deviates from (b) L/hr. Note that during the inspection we …
Contains “BMR*”Read the Form 483 →
Cipla Ltd. Virgonagar SiteInspection ended Jul 19, 2019 · FEI 3002806691 · Observation 3
… goods issue slips (Proposed work order for API), which your warehouse personnel use to pick the specific raw materials in your batch production order, to the batch production records. As such, your QA department has no way to verify during final batch record review if the raw materials listed on the goods issue slip match the materials which were picked from the warehouse and weighed/dispensed for batch production. * You do not perform raw …
Contains “batch production … record*”Read the Form 483 →
Cipla Ltd. Virgonagar SiteInspection ended Jul 19, 2019 · FEI 3002806691 · Observation 5
5. Stability studies performed to determine the expiry dates for finished APIs are not conducted in the same container closure system as the packaging procedure specified in the batch records. Specifically, your stability protocol for (BY) USP API (Protocol: SP/BL/DSA? 9001) states that the (6)@) USP stability samples are to be packaged in (6) @) and ®@ | bags which contain (6)@) (6) (4) and (b) (4) and placed in(®) + drums. This is not in …
Contains “batch record*”Read the Form 483 →
Baxalta Belgium Manufacturing S.A.Inspection ended May 31, 2019 · FEI 3002806503 · Observation 5
… Minimal investigational activities specified in SOP-2024, Adverse Event Quality Investigations, 23Mar2018, included a review of the following: batch records, events during production, product release testing results (including sterility), environmental and bioburden testing results, pre and post filter integrity results, visual inspection of retention samples (where applicable), stability review, and NAT/PCR and/or serology testing.
Contains “batch record*”Read the Form 483 →
21 CFR sections these observations open with
The sections FDAgent ties the most of these observations to by FDA's wording, with how many of this theme's observations each covers. Each links to its page of FDA citations, warning letters and the regulation text.
- 21 CFR 211.188Batch production and control recordsIn this theme's rule17
- 21 CFR 211.100Written procedures; deviations14
- 21 CFR 211.22Responsibilities of quality control unit13
- 21 CFR 211.68Automatic, mechanical, and electronic equipment13
- 21 CFR 211.113Control of microbiological contamination11
- 21 CFR 211.180General requirementsIn this theme's rule8
- 21 CFR 211.186Master production and control recordsIn this theme's rule5
- 21 CFR 211.42Design and construction features4
- 21 CFR 211.160General requirements4
- 21 CFR 211.192Production record review4
Ask FDAgent about batch records and documentation in Form 483s
FDAgent reads FDA's Form 483s, inspections, citations and warning letters and answers in plain English, with links to the records.
Related
About this data
- The observations are the text FDAgent read by OCR from Form 483 PDFs in FDA's Electronic Reading Room, shown only from 483s whose every printed observation passed FDAgent's checks against the scanned text, as on each 483's own page.
- Which theme an observation belongs to is decided by the fixed rules above; every count on this page is computed from them.
- A 483 addressed to an individual is listed without the name, and an observation that names a person is left out.