Abilene Nuclear LLC dba National Central Pharmacy Form 483, November 2019

FDA issued this Form 483 to Abilene Nuclear LLC dba National Central Pharmacy in Abilene, Texas (FEI 3004819860) at the end of a drug inspection that ended November 7, 2019. It lists 8 observations. FDA classified the inspection Voluntary Action Indicated (VAI).

A Form 483 lists the investigators' observations at the end of an inspection. It is not FDA's final determination.

Source: FDA's Electronic Reading Room, posted Jan 7, 2020. Form 483 PDF on fda.gov. FDAgent is independent of FDA.

Inspection ended
Nov 7, 2019
Drugs · Drug Quality Assurance
FDA's outcome
VAI
Voluntary Action Indicated
Posted by FDA
Jan 7, 2020
61 days after the inspection
Facility
FEI 3004819860
Abilene, Texas

The 8 observations on this Form 483

Read from the FDA-posted Form 483 PDF by OCR. FDAgent keeps an observation only when it matches the scanned text; check the PDF before quoting. The short titles are FDAgent's.

  1. 1

    Exposed skin during aseptic processing

    OBSERVATION 1 Personnel engaging in aseptic processing was observed with exposed skin. Specifically, your firm's pharmacy manager was observed during the production of the non-patient specific sterile drug product, EPI-Shugarcaine Su/F Ophth 1 ml/vial, (b)(6), as having his head, partially exposed facial skin, inside LAF Hood.

  2. 2

    Materials not disinfected before entry

    OBSERVATION 2 Materials or supplies were not disinfected prior to entering the ISO 5 LAF Hood. Specifically, your pharmacy manager failed to adequately disinfect materials transferred from the ISO 7 Cleanroom into LAF Hood which was later used in production. For example, on 10/30/2019, I observed your pharmacy [illegible] 'Office Use Only', EPI-Shugarcaine Su/F Ophth 1 ml/vial, 11/13/2019. While producing this lot of sterile drug product, I observed the following items were not disinfected prior to entering the ISO 5 LAF Hood. A. During setup for production, your pharmacy placed ampules of Lidocaine 4% Pres-Free [illegible] removed from the non-disinfected tray and disinfected once inside the LAF Hood. B. Your firm's pharmacy manager was observed placing a non-disinfected light green plastic container used to hold the vials into the ISO 5 LAF hood for the purpose of [illegible].

  3. 3

    Gloved hands not disinfected/replaced

    OBSERVATION 3 Personnel touched equipment or other surfaces located outside the ISO 5 area with gloved hands and then engaged in aseptic processing without changing or sanitizing gloves. Specifically, your firm's pharmacy failed to disinfect sterile gloves prior to re-entry into ISO 5 LAF hood while aseptically processing the sterile drug product, EPI-Shugarcaine Su/F Ophth 1 ml/vial. After collecting additional supplies in the ISO 7 Cleanroom too many times to count, during aseptic processing, I observed your firm's pharmacy manager pick paper from a trashcan, and return into the ISO 5 LAF hood without disinfecting and/or changing the sterile gloves prior to re-entry into the ISO 5 area.

  4. 4

    Supplies blocking first-pass airflow

    OBSERVATION 4 Personnel placed equipment/supplies in an area that blocked the movement of first pass airflow within the LAF hood, either before or after it was filled with sterile product. Specifically, your firm's pharmacy manager failed to place supplies to maintain first pass airflow within LAF Hood during aseptic processing. For example, during the aseptic processing of the sterile drug product, EPI-Shugarcaine Su/F Ophth 1 ml/vial, empty/filled vials and Lidocaine 4% Pres-Free 4.5 mls ampules were arranged in a manner that caused a disruption in first pass airflow within the ISO 5 LAF hood.

  5. 5

    Inadequate HEPA filter coverage/recertification

    OBSERVATION 5 The facility design was observed to allow the influx of poor quality air from entering into the ISO 7 Cleanroom/high quality air area. Specifically, your firm failed to have adequate HEPA filter coverage to prevent poor air quality from entering into the ISO 7 Cleanroom. For example, your facility is designed with (15)(4) [illegible] classified radio-pharmaceutical inspection/packaging areas that do not have adequate HEPA filter coverage to ensure poor air quality does not enter into the ISO 7 Cleanroom. Additionally, your firm failed to include the (b)(4) recertification to verify air quality.

  6. 6

    Inappropriate cleaning agents used

    OBSERVATION 6 Disinfecting agents and cleaning pads or wipes used are not appropriate for use in the ISO 5 LAF hood designated for the production of sterile drug product. Specifically, your firm is using the non-sterile cleaning product, (b)(4), in the LAF Hood designated for the production of sterile drug product.

  7. 7

    Media fills not worst-case representative

    OBSERVATION 7 Media fills were not performed that closely simulate aseptic production operations incorporating, as appropriate, worst-case activities and conditions that provide a challenge to aseptic operations. Specifically, your firm's procedure, Media Fill Challenge, Document #6.36, uses a (15)(4) [illegible]. (15)(4) Your firm's PIC reported the manipulations and complexity performed using the kit is representative of those performed during radio-pharmaceutical drug processing. The same kit components were modified to simulate non-radio-pharma aseptic transfers that uses a (6)(4-1) manipulation into syringes. The simulated media fill failed to be representative of non-radio-pharma aseptic processing worst-case activities and conditions.

  8. 8

    Batch records lack second verifier

    OBSERVATION 8 The batch production and control records are deficient in that they do not include identification of persons performing and checking each significant step in the operation. Specifically, your firm's production worksheets failed to include a second verifier for critical steps in your pharmacy's production process. For example, in selected production batch records, I found only your firm's pharmacist (the only firm employee) who reviewed and approved the production batch/lot. Your pharmacy failed to have a second person to verify defined critical steps within the production process that affect finished drug product identity, strength, quality, and purity. (b)(6) A. Rx, Cefuroxime 7.5 mg/ml Ophth Soln 2.5 mls, Expiry 1/18/2019 B. Rx(b)(6), EPI-Shugarcaine Su/F Ophth 1 ml/vial, Expiry [illegible] C. Rx(b)(6), Joint Mix Ropiv/Morph 10 mg, Exp [illegible] D. Rx(b)(6), Joint Mix #2 Bupiv/Mo, HAB 1-0.5-25 per 0.2 ml Lipo, Exp [illegible] E. Rx(b)(6) [illegible]

Ask FDAgent about this Form 483

FDAgent reads the 483, the facility's inspections, warning letters and FDA's citations and answers in plain English, with links to the records.

Edit it or ask as written. 3 questions free, no card needed.

About this data

  • The PDF and its dates come from FDA's Electronic Reading Room, last synced Nov 23, 2026, 04:33 UTC; the inspection, its outcome and FDA's citations from FDA's Data Dashboard.
  • The observations are the text FDAgent read from this PDF by OCR, shown only when every printed observation passed FDAgent’s checks against the scanned text.
  • FDAgent leaves out the names of the people the Form 483 is addressed to and of the investigators.
  • FDAgent is independent of FDA.