Batch records and documentation in FDA Form 483 observations
Master and batch production records, device history records, logbooks and good documentation practice. FDAgent finds 241 such observations in 143 Form 483s issued to 137 facilities, for inspections that ended between October 2008 and September 2026. Each is quoted below from the 483 FDA posted.
A Form 483 lists the investigators' observations at the end of an inspection. It is not FDA's final determination.
Source: Form 483 PDFs in FDA's Electronic Reading Room, last synced Nov 23, 2026, 04:33 UTC. Every quote is from a 483 whose reading passed FDAgent's checks against the scan. FDAgent is independent of FDA.
- Observations
- 241
- Each quoted below
- Form 483s
- 143
- Posted by FDA
- Facilities
- 137
- Distinct FEIs
- Most recent
- Sep 28, 2026
- Inspection end date
Observations by year
These observations by the year the inspection ended. They follow which 483s FDA has posted and FDAgent has read, so they show what is on record, not a trend in FDA's findings. 2026 runs to the latest sync (lighter bar).
Peak: 80 in 2026
| Year | Observations |
|---|---|
| 2008 | 3 |
| 2009 | 0 |
| 2010 | 1 |
| 2011 | 0 |
| 2012 | 4 |
| 2013 | 4 |
| 2014 | 0 |
| 2015 | 3 |
| 2016 | 16 |
| 2017 | 12 |
| 2018 | 15 |
| 2019 | 13 |
| 2020 | 0 |
| 2021 | 7 |
| 2022 | 18 |
| 2023 | 22 |
| 2024 | 15 |
| 2025 | 28 |
| 2026 | 80 |
How FDAgent sorts observations into this theme
Two fixed rules, applied by code to every observation FDAgent can quote. An observation counts when either holds. No model and no similarity score decides.
It is tied to one of these 21 CFR entries
21 CFR 211.18021 CFR 211.18421 CFR 211.18621 CFR 211.18821 CFR 820.18121 CFR 820.184
A 483 prints no regulation, so FDAgent ties an observation to a section only when it opens with FDA's own citation wording for that section, word for word.
Or its text contains one of these phrases
“batch record*” · “batch production … record*” · “master production … record*” · “master batch record*” · “device history record*” · “device master record*” · “documentation practice*” · “logbook*” · “log book*” · “BPR*” · “MPR*” · “BMR*”
* matches any ending and … up to three words in between. Case is ignored unless a phrase has capitals.
The 241 observations, newest first
Quoted word for word from each Form 483, with only line breaks joined; … marks where the quote is cut. The highlight shows why the observation is in this theme. Each links to FDAgent's page for the 483.
Absolute Pharmacy LLCInspection ended Apr 11, 2016 · FEI 3011116100 · Observation 6
6. Your firm failed to follow SOP No. 9.014 Procedures in [Chemicals]. As a result, your firm failed to conduct an [illegible] following: a) During the visual inspection of Testosterone Enanthate 210 mg/mL, pharmacist G.O.H observed two vials containing particulate. [illegible] distribution and it is unclear if the two vials containing [illegible] In addition, the batch record does not state that a (b)(4) sterilization. b) During the current FDA …
Contains “batch record*”Read the Form 483 →
Cosma SpAInspection ended Jan 15, 2016 · FEI 3002806754 · Observation 5
… Logbook pages can be copied from the procedures by production personnel as needed. 4. Electronic chromatograms are not reviewed during the release of analytical data. In addition, not all analytical data sheets used by QC analysts during the testing of APIs are documented as reviewed.
Contains “logbook*”Read the Form 483 →
Cosma SpAInspection ended Jan 15, 2016 · FEI 3002806754 · Observation 6
… The [illegible] was released based on a review of the cleaning batch record and visual inspection as the reviewer was unaware swab and rinse samples were taken and needed to be acceptable prior to the release of the equipment. 3. Operators do not document the amount of solvents or water used during the execution of cleaning records. 4. Justification that the length of the production campaign performed prior to cleaning did not result in the …
Contains “batch record*”Read the Form 483 →
Cosma SpAInspection ended Jan 15, 2016 · FEI 3002806754 · Observation 9
… Documentation is not performed contemporaneously during the execution of batch records. Production operators document time requirements to the nearest 5 minutes.
Contains “batch record*”Read the Form 483 →
Cosma SpAInspection ended Jan 15, 2016 · FEI 3002806754 · Observation 11
… The version of the procedure being trained on is not documented and training is not provided to operators on revisions to master production batch records and master cleaning records.
Contains “master production … record*”Read the Form 483 →
Cosma SpAInspection ended Jan 15, 2016 · FEI 3002806754 · Observation 12
Observation 12 Working copies of cleaning batch records are not exact copies of the master records. Specifically, the header templates used in the printing of master and working copies of cleaning batch records contain different information. The header template used in the working copies is not approved during the approval of the master cleaning record.
Contains “batch record*”Read the Form 483 →
Master Pharm LLCInspection ended May 6, 2015 · FEI 3005734706 · Observation 1
… sterile solutions and sterile equipment were being manipulated, there was clutter including several plastic bins either containing alcohol wipes or the paper batch record contained within a plastic cover.
Contains “batch record*”Read the Form 483 →
Alphatec Spine, Inc.Inspection ended Mar 13, 2015 · FEI 2027467 · Observation 4
A device history record has not been adequately maintained. Specifically, the firm acquired the (b) (4) (spinal implant device) from another medical device company in (b) (4). However, the only device history records (DHRs) on file for Part No.: (b) (4)-(b) (4), Final Assembly, (b) (4), Revision B/Catalog No.: (b) (4) are dated (b) (4) and (b) (4) and for Revision C of Part No.: (b) (4)/Catalog No.: (b) (4) they are dated (b) (4) through (b) (4). …
Opens with FDA's citation wording for 21 CFR 820.184Read the Form 483 →
Alphatec Spine, Inc.Inspection ended Mar 13, 2015 · FEI 2027467 · Observation 6
Quality system procedures and instructions have not been established. Specifically, the firm acquired the (b) (4) (i.e., spinal implant (b) (4)) medical device family from another medical device company in (b) (4). However, the firm has no procedures in place to ensure the quality system records associated with the acquired device to include device history records, drawings, specifications, procedures, etc., were received at the time of the …
Contains “device history record*”Read the Form 483 →
Qingdao Kangyuan Pharmaceutical Co., LTDInspection ended Jan 25, 2013 · FEI 3006512854 · Observation 10
The BPR does not document the duration of the (b)(4) step of the (b)(4) during production; for example, at (b)(4) (b)(4) the (b)(4) stage the (b)(4) (b)(4) is (b)(4)
Contains “BPR*”Read the Form 483 →
Qingdao Kangyuan Pharmaceutical Co., LTDInspection ended Jan 25, 2013 · FEI 3006512854 · Observation 11
An entry in page (b)(4) of the BPR for the (b)(4) processing step documents that (b)(4) of the (b)(4) (b)(4) were collected and measured. However, while questioning this BPR entry, it was reported (b)(4) (b)(4) that the operator measures the amount of The (b)(4) measurement is done The actual steps as to how the operator came up with the value, documented in the BPR, are not documented in the (b)(4) BPR or explained in an operating instruction …
Contains “BPR*”Read the Form 483 →
Qingdao Kangyuan Pharmaceutical Co., LTDInspection ended Jan 25, 2013 · FEI 3006512854 · Observation 15
There was no identification, calibration record, or usage log on a scale found in the raw material weighing and dispensing room. Reportedly, this scale is used to weigh (b)(4) used in the product formulation and recorded in the BPR.
Contains “BPR*”Read the Form 483 →
Qingdao Kangyuan Pharmaceutical Co., LTDInspection ended Jan 25, 2013 · FEI 3006512854 · Observation 16
There were several undated loose pages of partially completed batch production records for processing stages (b)(4) steps) in a drawer of a desk in the office used by production operators and management.
Contains “batch production … record*”Read the Form 483 →
Alaunus Pharmaceutical LLCInspection ended Nov 16, 2012 · FEI 3008170448 · Observation 3
The master production records are deficient in that not all critical steps are defined in the master record. For example, after the API is filled into the (b) (4) container, the containers are labeled, and placed into a (b) (4)r (b) (4) bag and sealed. After the bag is sealed, another label is placed on the outside of this (b) (4) bag. The operation to label this outer bag is not a documented step in the master batch record.
Contains “master production … record*”Read the Form 483 →
Alaunus Pharmaceutical LLCInspection ended Nov 16, 2012 · FEI 3008170448 · Observation 4
The batch production records are deficient in that: • There is no documentation to indicate which operator added the outer label to the (b) (4) bag and there is no documentation that this step was verified. • The verification step/witness step of critical operations occurred approximately 2 weeks after lot 12-AC-001 for Hydromorphone HCL was repackaged.
Contains “batch production … record*”Read the Form 483 →
Cangene CorporationInspection ended Jun 21, 2012 · FEI 3003153579 · Observation 6
… for these lots was adequately addressed in that these deviations were classified as (b)(4) Deviations" that are used (b)(4) Per SOP 10.060.0023.20.19, QA Batch Record Review, (b)(4) Deviations require (b)(4) (b)(4)
Contains “batch record*”Read the Form 483 →
Cangene CorporationInspection ended Jun 21, 2012 · FEI 3003153579 · Observation 8
Written procedures are not always followed. The (b)(4) treatment module of the batch record (Document number 24.5002.08, step 17, page 24 of 35) specifies a (b)(4) As described in deviation 211915, the (b)(4) was terminated early during manufacture of VIGIV lot (b)(4)
Contains “batch record*”Read the Form 483 →
ISE America Inc.Inspection ended Oct 15, 2010 · FEI 3002954898 · Observation 2
Complex and contractor logbooks are in need of update with currently approved forms and required information. For example, Pesticide Usage Forms/Pest Control Logs, Fly Monitoring documents differ among contractors.
Contains “logbook*”Read the Form 483 →
Genzyme CorporationInspection ended Oct 10, 2008 · FEI 1000305672 · Observation 3
The batch production record for drug substance is not adequately controlled. For example, the production record is printed on green paper and is issued by Quality Assurance. Attached to the production record are white manufacturing records. Operators frequently record the performance of manufacturing activities, test results or meter readings in the manufacturing records. The manufacturing records are not issued by Quality Assurance. There is no …
Contains “batch production … record*”Read the Form 483 →
Genzyme CorporationInspection ended Oct 10, 2008 · FEI 1000305672 · Observation 4
A. Activities performed during drug substance manufacture are not adequately documented. For example: When performance of an activity is optional [illegible] there is frequently no place in the batch record to record whether the activity was performed. Dated and signed crossing-out of optional activities that are not performed is not used consistently in the batch record. Additional activities may be performed during manufacture of some …
Contains “batch record*”Read the Form 483 →
21 CFR sections these observations open with
The sections FDAgent ties the most of these observations to by FDA's wording, with how many of this theme's observations each covers. Each links to its page of FDA citations, warning letters and the regulation text.
- 21 CFR 211.188Batch production and control recordsIn this theme's rule17
- 21 CFR 211.100Written procedures; deviations14
- 21 CFR 211.22Responsibilities of quality control unit13
- 21 CFR 211.68Automatic, mechanical, and electronic equipment13
- 21 CFR 211.113Control of microbiological contamination11
- 21 CFR 211.180General requirementsIn this theme's rule8
- 21 CFR 211.186Master production and control recordsIn this theme's rule5
- 21 CFR 211.42Design and construction features4
- 21 CFR 211.160General requirements4
- 21 CFR 211.192Production record review4
Ask FDAgent about batch records and documentation in Form 483s
FDAgent reads FDA's Form 483s, inspections, citations and warning letters and answers in plain English, with links to the records.
Related
About this data
- The observations are the text FDAgent read by OCR from Form 483 PDFs in FDA's Electronic Reading Room, shown only from 483s whose every printed observation passed FDAgent's checks against the scanned text, as on each 483's own page.
- Which theme an observation belongs to is decided by the fixed rules above; every count on this page is computed from them.
- A 483 addressed to an individual is listed without the name, and an observation that names a person is left out.
- FDAgent is independent of FDA.