Complaint handling in FDA Form 483 observations

How firms record, review and investigate complaints from customers, patients and health professionals. FDAgent finds 75 such observations in 56 Form 483s issued to 54 facilities, for inspections that ended between November 2008 and June 2026. Each is quoted below from the 483 FDA posted.

A Form 483 lists the investigators' observations at the end of an inspection. It is not FDA's final determination.

Source: Form 483 PDFs in FDA's Electronic Reading Room, last synced Nov 23, 2026, 04:33 UTC. Every quote is from a 483 whose reading passed FDAgent's checks against the scan. FDAgent is independent of FDA.

Observations
75
Each quoted below
Form 483s
56
Posted by FDA
Facilities
54
Distinct FEIs
Most recent
Jun 30, 2026
Inspection end date

Observations by year

These observations by the year the inspection ended. They follow which 483s FDA has posted and FDAgent has read, so they show what is on record, not a trend in FDA's findings. 2026 runs to the latest sync (lighter bar).

Peak: 20 in 2025

Complaint handling observations in Form 483s by inspection year
YearObservations
20081
20090
20100
20110
20121
20130
20140
20154
20165
20178
20183
20192
20200
20212
20229
20234
20244
202520
202612

How FDAgent sorts observations into this theme

Two fixed rules, applied by code to every observation FDAgent can quote. An observation counts when either holds. No model and no similarity score decides.

It is tied to one of these 21 CFR entries

21 CFR 211.19821 CFR 820.198

A 483 prints no regulation, so FDAgent ties an observation to a section only when it opens with FDA's own citation wording for that section, word for word.

Or its text contains one of these phrases

“complaint*”

* matches any ending and … up to three words in between. Case is ignored unless a phrase has capitals.

The 75 observations, newest first

Quoted word for word from each Form 483, with only line breaks joined; … marks where the quote is cut. The highlight shows why the observation is in this theme. Each links to FDAgent's page for the 483.

  1. Farmakeio Outsourcing LLCInspection ended Sep 5, 2025 · FEI 3014982757 · Observation 1

    … Your firm failed to adequately investigate received customer complaints. For example: 1. COMP-2025-0023 dated 6/20/2025 and COMP-2025-0018 dated 5/23/2025, documents received complaints for cracked pellets. Your firm's failed to adequately investigate the source of cracked pellets within your firm's manufacturing process and expand investigation into all area including distribution. 2. COMP-2025-0002 dated 1/04/2025 - documents received complaint …

    Contains “complaint*”Read the Form 483 →

  2. Creative Testing SolutionsInspection ended Aug 29, 2025 · FEI 3000204058 · Observation 2

    Failure to perform a thorough investigation and make a record of the conclusions and follow-up of an unexplained discrepancy. Specifically, A. None of your firm’s standard operating procedures require investigation into all unexplained discrepancies. Your SOP titled, Error Management, only requires investigation into customer complaints and errors involving released test results. B. Your firm failed to thoroughly investigate and document improper …

    Contains “complaint*”Read the Form 483 →

  3. Aptuit LtdInspection ended Aug 22, 2025 · FEI 3003444315 · Observation 2

    There is a failure to thoroughly review any unexplained discrepancy and the failure to [illegible] of a batch or any of its components to meet any of its specifications whether or not the batch has been already distributed. Specifically, The quality unit failed to perform a thorough customer complaint investigation regarding [illegible] batches of [illegible] API that obtained [illegible] contents above the specifications limits. The quality unit …

    Contains “complaint*”Read the Form 483 →

  4. Foremount Enterprise Co., LtdInspection ended Jun 19, 2025 · FEI 3010402691 · Observation 2

    Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been established. Specifically, your firm is a Manufacturer of Class II Non-Rebreathing Valves, Product Code CBP and Emergency Manual (Resuscitator) Ventilators, Product Code BTM, medical devices exported and intended for the US market, and your firm's Customer Complaint Handling Procedure, QP824, V.05, dated May 27, 2025, has no requirement to …

    Opens with FDA's citation wording for 21 CFR 820.198(a)Read the Form 483 →

  5. Foremount Enterprise Co., LtdInspection ended Jun 19, 2025 · FEI 3010402691 · Observation 3

    … QP752, V.10, dated December 20, 2024, is not adequately established as to date your firm has not established procedure(s) to ensure your firm is receiving complaints for your medical devices exported and intended for the US market from your US market distributors to ensure your firm reviews and evaluates all complaints in a uniform and timely manner to determine whether an investigation is necessary and are evaluated to determine whether the …

    Contains “complaint*”Read the Form 483 →

  6. Performance Orthopedics Associates IncInspection ended Jun 13, 2025 · FEI 3014678321 · Observation 2

    The quality program has not ensured that procedures exist for information relating to core CGTP requirements, including complaints. Specifically, 1. There is no established procedure or records addressing complaint handling. 2. There are no established procedures or records addressing investigation of HCT/P deviations relating to core CGTP requirements or adverse reactions.

    Contains “complaint*”Read the Form 483 →

  7. Piramal, Piramal Pharma LimitedInspection ended Feb 17, 2025 · FEI 3008763768 · Observation 3

    OBSERVATION 3 Written records of investigation of an API complaint do not include adequate follow-up. Specifically, Customer Complaint CC-22-06 was initiated on May 21, 2022, for foreign particles being observed for [illegible] Batch No. [illegible], Retest Date: May 2024. Samples from another API batch, Batch No. (b)(4), was used to test traces of potential foreign/metallic particles. However, there was no documented attempt to retrieve and test …

    Contains “complaint*”Read the Form 483 →

  8. Glenmark Pharmaceuticals Ltd.Inspection ended Feb 14, 2025 · FEI 3008565058 · Observation 1

    … You received two consumer complaints on 10/18/2023 and 12/14/2024 respectively in which patients reported an adverse event while using (Cb)l4 tablets related to facilities, equipment, and air supply that is used for (CbTr4 tablets). Your investigations for both complaints were inadequate as you failed to assess potential cross-contamination that may occur when (Cbfl"l drugs are manufactured using shared facilities and equipment.

    Contains “complaint*”Read the Form 483 →

  9. Zhejiang Huahai Pharmaceutical Co., Ltd.Inspection ended Jan 24, 2025 · FEI 3003999190 · Observation 4

    … Complaint investigation CF-23274 was opened when L.L. customer obtained an OOS for dissolution on 4 4 :I -l•JC1 and l.,_______00l___.rJ able~..___(b_ _H} ng batch. 1.,_____l.l>TI__,.i (EU market). The OOS was continued during testing of retain samples. The investigation identified higher main compression force during tableting, which had no established limits, as well as increased disintegration times during storage. The cause of the higher …

    Contains “complaint*”Read the Form 483 →

  10. Philips Medical Systems Nederland B.V.Inspection ended Jan 17, 2025 · FEI 3000976525 · Observation 1

    … Your firm subsequently received 12 complaints (dated July 2023 - November 2023) from customers reporting issues related to patient studies unable to be archived or transferred, including Complaint #4061517 (dated 13-Nov-2023) alleging that a patient expired due to a delay in transmission of study images possibly resulting in a delay in treatment. Complaint #4061517 states that the "risk was not previously assessed in the RMM [Risk Management …

    Contains “complaint*”Read the Form 483 →

  11. Philips Medical Systems Nederland B.V.Inspection ended Jan 17, 2025 · FEI 3000976525 · Observation 2

    … Your firm received 12 complaints (dated July 2023 - November 2023) from customers reporting issues with the IntelliSpace Cardiovascular (ISCV) software product related to patient studies unable to be archived or transferred, including Complaint #4061517 (dated 13-Nov-2023) alleging that a patient expired due to a delay in transmission of study images possibly resulting in a delay in treatment. Your firm concluded this Complaint #4061517 presents …

    Contains “complaint*”Read the Form 483 →

  12. Philips Medical Systems Nederland B.V.Inspection ended Jan 17, 2025 · FEI 3000976525 · Observation 3

    Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been adequately established. Specifically, A. Review of data downloaded from your (b)(4) reporting IntelliSpace Cardiovascular malfunctions (with no deaths or serious injuries reported) were identified as not reportable as Medical Device Reports ("No" entered under "Potentially Reportable Event"); however, there is no valid justification …

    Opens with FDA's citation wording for 21 CFR 820.198(a)Read the Form 483 →

  13. Janssen Vaccines CorporationInspection ended Nov 19, 2024 · FEI 3012637764 · Observation 1

    Responsibilities and procedures for quality units are not in writing and/or followed to ensure the FARS (BPDR) are submitted for the distributed products that are implicated with product quality issues. Specifically, You failed to follow your procedure, TV-WI-21853 Version 29.0 Product Quality Complaint Investigation (Effective date 5/25/2024) Section 7.11.1.2 that sets the requirement for FAR (BPDR) as, "***Two or more complaints with the same …

    Contains “complaint*”Read the Form 483 →

  14. Janssen Vaccines CorporationInspection ended Nov 19, 2024 · FEI 3012637764 · Observation 4

    … You have recorded a complaint # 90000338418 for Uncharacteristic Color. This complaint was received for (b) (4) Vial (b) (4) mg (b) (4) mL Batch No. (b) (4) Mfg. date 2/7/2024, Expiry date (b) (4) for the drug product being (b) (4) color. No root cause was determined.

    Contains “complaint*”Read the Form 483 →

  15. AnazaoHealth CorporationInspection ended Sep 23, 2024 · FEI 3011152407 · Observation 2

    … Between October 31, 2022 and June 19, 2024, your firm received 8 complaints related to glass fragments found inside the pellet container closure system from the broken vial inserts. Glass fragments introduce potential risks to patients if the glass contamination is undetected.

    Contains “complaint*”Read the Form 483 →

  16. Hikma Pharmaceuticals USA Inc.Inspection ended Jun 12, 2024 · FEI 2220525 · Observation 3

    … According to section 7.4.2.4 of SOP CHH15028, "when performing an investigation, if a review of previous lot data, stability and retain samples, customer complaints, Annual Product Reviews, and other identifiable data is performed, this documentation of reviews should be included in the investigation". However, the documentation of the following deviations QE-021941, QE-018568, and QE-007589 did not include the results and data of the additional …

    Contains “complaint*”Read the Form 483 →

  17. Catalent Indiana, LLCInspection ended Nov 15, 2023 · FEI 3005949964 · Observation 2

    … (6) 4 queries for the lots of stoppers 6 4 ( ) and lots of syringes 6 4 ) were limited to supplier complaints, as opposed to any b 4 record (deviation, LIR, CAPA, change control, etc.) that includes the aforementioned lots of components. 2. Deviation #749638, opened 09 AUG 2023, included a query to determine if other records included the words "unfilled" and "syringes". This query was limited to customer complaint records, as opposed to searching …

    Contains “complaint*”Read the Form 483 →

  18. Sanofi PasteurInspection ended Sep 29, 2023 · FEI 3002972083 · Observation 2

    … that two vials of (b)(4) (US Market) [illegible] were observed to contain cellulose particles during visual inspection of retain samples related to a complaint of vial breakage (complaint no. 200255622). This lot had been released and distributed within the US market and was reported to the agency as a Biological Product Deviation Report (BPDR) on 09/15/2022. While reviewing the firm's CAPA investigation report MLED22A2877 pertaining to deviation …

    Contains “complaint*”Read the Form 483 →

  19. Linhai Tianyu Pharmaceutical Co., Ltd.Inspection ended Sep 15, 2023 · FEI 3013345425 · Observation 3

    Written records of investigation into unexplained discrepancies and the failure of a batch or any of its components to meet specification do not always include the conclusions and follow-up. Specifically, Investigations initiated and performed by your Quality Unit in response to Out of Specification results, laboratory events, and market complaints related to intermediates and API manufactured by your firm are not always scientifically sound or …

    Contains “complaint*”Read the Form 483 →

  20. Ipca Laboratories LimitedInspection ended Jun 23, 2023 · FEI 3007574780 · Observation 8

    Procedures describing the handling of all written and oral complaints regarding a drug product are not written and followed. Specifically, Your firm’s Corporate Standard Operating Procedure for Handling of Product Complaint (CSOP/2013/004; Number: CSOP/2013/004/R09, Effective Date: June 02, 2022; CSOP/2013/004/R08, Effective Date: Feb 03, 2022; CSOP/2013/004/R07, Effective Date: Sept 16, 2020) requires that all CAPAs initiated in reference to …

    Opens with FDA's citation wording for 21 CFR 211.198(a)Read the Form 483 →

21 CFR sections these observations open with

The sections FDAgent ties the most of these observations to by FDA's wording, with how many of this theme's observations each covers. Each links to its page of FDA citations, warning letters and the regulation text.

Ask FDAgent about complaint handling in Form 483s

FDAgent reads FDA's Form 483s, inspections, citations and warning letters and answers in plain English, with links to the records.

Edit it or ask as written. 3 questions free, no card needed.

About this data

  • The observations are the text FDAgent read by OCR from Form 483 PDFs in FDA's Electronic Reading Room, shown only from 483s whose every printed observation passed FDAgent's checks against the scanned text, as on each 483's own page.
  • Which theme an observation belongs to is decided by the fixed rules above; every count on this page is computed from them.
  • A 483 addressed to an individual is listed without the name, and an observation that names a person is left out.
  • FDAgent is independent of FDA.