Complaint handling in FDA Form 483 observations
How firms record, review and investigate complaints from customers, patients and health professionals. FDAgent finds 75 such observations in 56 Form 483s issued to 54 facilities, for inspections that ended between November 2008 and June 2026. Each is quoted below from the 483 FDA posted.
A Form 483 lists the investigators' observations at the end of an inspection. It is not FDA's final determination.
Source: Form 483 PDFs in FDA's Electronic Reading Room, last synced Nov 23, 2026, 04:33 UTC. Every quote is from a 483 whose reading passed FDAgent's checks against the scan. FDAgent is independent of FDA.
- Observations
- 75
- Each quoted below
- Form 483s
- 56
- Posted by FDA
- Facilities
- 54
- Distinct FEIs
- Most recent
- Jun 30, 2026
- Inspection end date
Observations by year
These observations by the year the inspection ended. They follow which 483s FDA has posted and FDAgent has read, so they show what is on record, not a trend in FDA's findings. 2026 runs to the latest sync (lighter bar).
Peak: 20 in 2025
| Year | Observations |
|---|---|
| 2008 | 1 |
| 2009 | 0 |
| 2010 | 0 |
| 2011 | 0 |
| 2012 | 1 |
| 2013 | 0 |
| 2014 | 0 |
| 2015 | 4 |
| 2016 | 5 |
| 2017 | 8 |
| 2018 | 3 |
| 2019 | 2 |
| 2020 | 0 |
| 2021 | 2 |
| 2022 | 9 |
| 2023 | 4 |
| 2024 | 4 |
| 2025 | 20 |
| 2026 | 12 |
How FDAgent sorts observations into this theme
Two fixed rules, applied by code to every observation FDAgent can quote. An observation counts when either holds. No model and no similarity score decides.
It is tied to one of these 21 CFR entries
A 483 prints no regulation, so FDAgent ties an observation to a section only when it opens with FDA's own citation wording for that section, word for word.
Or its text contains one of these phrases
“complaint*”
* matches any ending and … up to three words in between. Case is ignored unless a phrase has capitals.
The 75 observations, newest first
Quoted word for word from each Form 483, with only line breaks joined; … marks where the quote is cut. The highlight shows why the observation is in this theme. Each links to FDAgent's page for the 483.
Meridian Medical Technologies, Inc. a Pfizer CompanyInspection ended Mar 24, 2017 · FEI 1950222 · Observation 3
Procedures describing the handling of all written and oral complaints regarding a drug product are not established, written and followed. Specifically, A. Your firm has not documented rationale why (b) (4) complaints of a similar nature on the same lot are necessary to identify a trend per your procedure Product Complaint Handling SOP-QLC-QLE-00702. B. Complaint classifications, listed in GPB-QS1073 Prioritization of Pfizer Product Quality …
Opens with FDA's citation wording for 21 CFR 211.198(a)Read the Form 483 →
Meridian Medical Technologies, Inc. a Pfizer CompanyInspection ended Mar 24, 2017 · FEI 1950222 · Observation 5
Records are not maintained so that data therein can be reviewed at least annually to evaluate the quality standards of each drug product to determine the need for changes in specifications or manufacturing or control procedures. Specifically, your analysis of the quality data used to identify existing and recurring quality problems does not include an analysis of complaints based on finished product lots, manufacturing dates or component lots to …
Contains “complaint*”Read the Form 483 →
Meridian Medical Technologies, Inc. a Pfizer CompanyInspection ended Mar 24, 2017 · FEI 1950222 · Observation 13
… (4)) and therefore (b) (4) (b) (4) (all units were (b) (4) This batch was subsequently released by the firm; and during commercial distribution received a complaint related to Liquids Atypical Color / Cloudy. The subsequent compliant investigation (MMT Complaint Number 2016-07-0037004-US) did not consider the atypical nature of the inspection process nor the potential (b) (4) as a root cause factor. B. "SOP-QLA-GEN-00802, version 10.0, Management …
Contains “complaint*”Read the Form 483 →
Bayer AG PharmaInspection ended Jan 20, 2017 · FEI 3002806462 · Observation 5
Failure to conduct thorough complaint investigations and extend the investigation to related batches. 1. Complaint 2015-015988 received 30 April 2015 and complaint 2015-016372 received 05 May 2015 documented leakage from (b) (4) bags where the (b) (4) connected to the bag on batch (b) (4) of (b) (4). Investigation determined a packaging material defect could be the root cause. The investigation failed to identify other batches which used the same …
Contains “complaint*”Read the Form 483 →
Aurobindo Pharma Limited Unit 4Inspection ended Sep 16, 2016 · FEI 3008461619 · Observation 1
An Field Alert Report was not submitted within three working days of receipt of information concerning significant chemical, physical, or other change or deterioration in a distributed drug product. Specifically, your firm has received six complaints pertaining to four separate batches of the sterile human drug (b) (4) Injectable (b) (4) mg (b) (4) mL that describe the (b) (4) stopper pushing into the vial upon spiking with infusion sets. …
Contains “complaint*”Read the Form 483 →
Indoco Remedies LimitedInspection ended Sep 4, 2016 · FEI 3005124189 · Observation 1
Investigations of an unexplained discrepancy and a failure of a batch or any of its components to meet any of its specifications did not extend to other batches of the same drug product and other drug products that may have been associated with the specific failure or discrepancy. Specifically, between January and August 2016, approximately 459 complaints describing leaking or empty bottles for sterile ©; oducts have been received. Although a …
Contains “complaint*”Read the Form 483 →
Indoco Remedies LimitedInspection ended Sep 4, 2016 · FEI 3005124189 · Observation 6
An| “Field Alert Report was not submitted within three working days of receipt of information concerning bacteriological contamination and significant chemical, physical, or other change or deterioration in a distributed drug product. Specifically, between January and August 2016, approximately 459 complaints of leaking or empty bottles of various ®@ products have been received, which span multiple lots. There have been no © “’Field Alert Reports …
Contains “complaint*”Read the Form 483 →
Cipla LimitedInspection ended Aug 30, 2016 · FEI 3004081307 · Observation 1
Investigations of an unexplained discrepancy and a failure of a batch or any of its components to meet any of its specifications did not extend to other batches of the same drug product. Specifically. a. Regarding LL ®I¥S olution, approximately 7 complaints have been received for lot OE exp. ®@ 1 between April 2015 and June 2016, describing empty containers or similar issues. Approximately 9 complaints have been received for this issue across …
Contains “complaint*”Read the Form 483 →
Cipla LimitedInspection ended Aug 30, 2016 · FEI 3004081307 · Observation 4
An Field Alert Report was not submitted within three working days of receipt of information concerning significant chemical, physical, or other change or deterioration in a distributed drug product. Specifically. an| Field Alert Report (FAR) was filed only after 7 complaints for empty container in sterile product pd Solution, lot ®' were received, between April 2015 and June 2016. This product is contract manufactured for ®® who holds the product …
Contains “complaint*”Read the Form 483 →
UCLA Ronald Reagan Medical CenterInspection ended Dec 18, 2015 · FEI 3007967308 · Observation 1
The user facility failed to provide all information concerning individual adverse event reports that is reasonably known to them, including information found in documents in possession of the user facility. Specifically, a medical device complaint was submitted under MAUDE (Manufacturer and User Facility Device Experience), [illegible]. In which the report included information where there was a [illegible] involved (all involved reportable …
Contains “complaint*”Read the Form 483 →
Alphatec Spine, Inc.Inspection ended Mar 13, 2015 · FEI 2027467 · Observation 1
… Specifically, Between 1/2009 and 4/25/2014, the firm has on file 32 complaint reports for the Illico Screw Extender (Part No.'s 73724, 73772 and 73779) for dimples/tabs partially or fully breaking off during the implant surgical procedure, which were not reported to the FDA as MDR malfunction reports. The firm's rationale for not reporting these events as MDR reportable was because there was no associated harm to the patient or delay in the …
Contains “complaint*”Read the Form 483 →
Alphatec Spine, Inc.Inspection ended Mar 13, 2015 · FEI 2027467 · Observation 2
… In addition, post the release of CO No.: (b) (4), the firm's number of complaints for "broken tabs," has increased (i.e., prior to the release of CO No.: (b) (4) the firm had five (5) complaints on file from 1/2009 through 1/2013; post the release of CO No. (b) (4) (the first set built date was (b) (4)) the firm has approximately 28 complaint reports on file ranging from 5/2013 through 4/25/2014, which were MDR reportable events.
Contains “complaint*”Read the Form 483 →
Alphatec Spine, Inc.Inspection ended Mar 13, 2015 · FEI 2027467 · Observation 7
Procedures for corrective and preventive action have not been adequately established. Specifically, a) Procedure No.: SOP-025 - Corrective and Preventive Action (CAPA) System, Revision L is not defined in that the procedure does not provide instructions or a linkage to an action or alert level for the (b) (4) complaint trending data, which is reviewed at the (b) (4) trend review meetings. For example, the (b) (4) complaint rate by product line is …
Contains “complaint*”Read the Form 483 →
Boston Analytical, Inc.Inspection ended Nov 9, 2012 · FEI 1221426 · Observation 1
… e.) SOP P-1130, Version 11 Client Complaint Program states all client complaints are to be closed out within (b)(4) of the opening of the complaint. For CO-12-0029, date of contact for complaint is 02/10/12, complaint closed on 03/14/12.
Contains “complaint*”Read the Form 483 →
American Red Cross Blood ServicesInspection ended Nov 24, 2008 · FEI 1273014 · Observation 2
Written procedure was not followed regarding handling of "Management of Suspect Products". Specifically, [illegible] 7/17/08 [illegible] and [illegible] Hospital Services Supervisor (JO) instructed [illegible] to destroy units, instead of referring the customer complaints to ARC, [illegible] follow up, as it's written apply the appropriate [illegible] both cases, Hospital Service [illegible]
Contains “complaint*”Read the Form 483 →
21 CFR sections these observations open with
The sections FDAgent ties the most of these observations to by FDA's wording, with how many of this theme's observations each covers. Each links to its page of FDA citations, warning letters and the regulation text.
- 21 CFR 211.192Production record review17
- 21 CFR 211.198Complaint filesIn this theme's rule5
- 21 CFR 211.100Written procedures; deviations3
- 21 CFR 211.22Responsibilities of quality control unit2
- 21 CFR 211.180General requirements2
- 21 CFR 820.198Complaint filesIn this theme's rule2
- 21 CFR 211.42Design and construction features1
- 21 CFR 211.67Equipment cleaning and maintenance1
- 21 CFR 211.94Drug product containers and closures1
- 21 CFR 211.113Control of microbiological contamination1
Ask FDAgent about complaint handling in Form 483s
FDAgent reads FDA's Form 483s, inspections, citations and warning letters and answers in plain English, with links to the records.
Related
About this data
- The observations are the text FDAgent read by OCR from Form 483 PDFs in FDA's Electronic Reading Room, shown only from 483s whose every printed observation passed FDAgent's checks against the scanned text, as on each 483's own page.
- Which theme an observation belongs to is decided by the fixed rules above; every count on this page is computed from them.
- A 483 addressed to an individual is listed without the name, and an observation that names a person is left out.