FDA has 8 inspections on file for Boston Analytical, Inc. in Salem, New Hampshire (FEI 1221426), from 2009 to 2023. The latest ended on and was classified NAI, no action indicated.

NAI means FDA found no objectionable conditions or practices, or found none that justified further action.

No warning letter, court action or import refusal is filed under this FEI.

As of Nov 21, 2026, the site is registered with FDA as a drug establishment for analysis, through Dec 31, 2027. Its registrant is Boston Analytical Inc dba. BA Sciences (D-U-N-S 181194176).

Sources:FDA firm profile ↗FDA drug establishment registrations ↗Synced Nov 23, 2026

Ask FDAgent about this site

Free account: 3 questions, no card needed. Each answer links to its FDA sources.

Email me when FDA posts a new inspection, Form 483 or warning letter for this site

Weekly, on a free account. Pause it any time.

Watch this facility

Comparing suppliers? Check this site beside 2 others, free

Latest inspection
NAI
Inspections on file
8, 2009–2023
0 OAI · 7 VAI · 1 NAI
Inspections with 483 observations shown
5 of 8
1 posted 483 not matched to an inspection
Form 483s posted by FDA
↗
Warning letters
None on file
Import refusals
None under this FEI
Import alerts
Not linked by FEI
FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗
Drug establishment registration
Through

Inspection history

8 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC · FDA reading room, last synced Nov 23, 2026, 04:33 UTC

  1. NAINo objectionable conditions that need follow-upDrugs · Drug Quality AssuranceNone
  2. VAIObjectionable conditions found; FDA took no regulatory actionDrugs · Drug Quality Assurance1 citation published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Sampling and testing plans not followed

      Written procedures for sampling and testing plans are not followed for each drug product.

  3. VAIObjectionable conditions found; FDA took no regulatory actionVeterinary · Monitoring of Marketed Animal Drugs, Feed, and DevicesCitations not published
  4. VAIObjectionable conditions found; FDA took no regulatory actionDrugs · Drug Quality Assurance2 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Written program not followed

      The written stability testing program is not followed.

    2. Laboratory Test Method Verification

      Verification of the suitability of the testing methods is deficient in that they are not performed under actual conditions of use.

  5. Not classified yetFDA posted this Form 483; FDAgent has not matched it to a Data Dashboard inspectionNo inspection matchedForm 483 pagePDF
  6. VAIObjectionable conditions found; FDA took no regulatory actionDrugs · Drug Quality Assurance4 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Written procedures fail to include

      Written procedures for cleaning and maintenance fail to include description in sufficient detail of the methods of disassembling and reassembling equipment as necessary to assure proper cleaning and maintenance .

    2. Complaint Handling Procedure

      Procedures describing the handling of written and oral complaints related to drug products are not written or followed .

    3. Test method modification records do not include

      Records maintained of any modification of an established method employed in testing do not include the reason for the modification .

    4. Personnel dating/signing equipment log

      The persons double-checking the cleaning and maintenance are not dating and signing or initialing the equipment cleaning and use log.

  7. VAIObjectionable conditions found; FDA took no regulatory actionDrugs · Drug Quality Assurance3 citations published

    Read from the FDA-posted Form 483 PDF by OCR. FDAgent keeps an observation only when it matches the scanned text; check the PDF before quoting. The short titles are FDAgent’s.

    1. Observation 1

      Production procedures not followed

      Written production and process control procedures are not followed in the execution of production and process control functions. Specifically, the following SOPs were not followed: a.) SOP P-1050, Version 16 Laboratory Investigation Policy, states all unexpected, out of trend or out of specification results are to be investigated however this was not followed for the following samples analyzed: * Monosodium Glutamate Project #(b)(4), Sample #(b)(4) Assay tested on 07/12/13 reported an OOS result of [illegible] 78.46. Analyst did not follow SOP P-1050 and conduct a Laboratory Investigation. The chemist re-assayed the sample by performing a step not described in test procedure and reported the retest results of 99.84%. The C of A issued to the customer reported [illegible]. * L-aspartic Monosodium Salt (Project #(b)(4), Sample #(b)(4)) Identification Test reported an unexpected result of inconclusive color change. Analyst did not follow SOP P-1050 and conduct a Laboratory Investigation. The chemist increased the concentration using a procedure not described in test procedure, re-analyzed sample and reported the retest results of 99.84%. The C of A issued to the customer reported [illegible] and did not indicate the results were reported from a retest as indicated in SOP. b.) SOP P-1050, Version 16, Laboratory Investigation Policy, is not being followed which states testing into compliance without regards for scientific justification is not acceptable. For analysis of Levofloxacin (Project Number (b)(4), Sample #(b)(4)), Assay and Impurities test reported OOS results on 10/06/11, the sample was tested three times and the third analytical findings were reported on C of A. c.) SOP P-510, Version Number: 12 Log-In of Samples for Testing describes procedure for logging in samples into computer system; however it does not include logging in of all samples as they are received the Sample Log-In Area. There is no traceability of samples received and stored in Sample Log-In Area. During inspection, samples were observed stored in the Sample Log-In Area with no sample submission forms and employee stated that all samples are not logging into computer upon receipt. d.) Firm not completing testing of stability of samples prior to the "Test by date" as stated in SOP P-645, Version 11, Initiating and Removing Stability Studies. Morphine Sulfate ER 60mg Tablets, Project #(b)(4), Sample #(b)(4), Sample pull date 10/01/2012 and test on 11/01/2012. e.) SOP P-1130, Version 11 Client Complaint Program states all client complaints are to be closed out within (b)(4) of the opening of the complaint. For CO-12-0029, date of contact for complaint is 02/10/12, complaint closed on 03/14/12.

    2. Observation 2

      Inadequate laboratory controls and procedures

      Laboratory controls do not include the establishment of scientifically sound and appropriate test procedures designed to assure that drug products conform to appropriate standards of identity, strength, quality and purity. a.) Specifically, described in SOP P-900 General Analytical Procedures not following test procedures for analysis of the following samples: Levofloxacin (Project Number (b)(4), Sample (b)(4)), Assay test and Impurities test not being followed according to USP 34. * Sodium Bicarbonate (Project Number (b)(4), Sample Number (b)(4)), the Identification test of Bicarbonate was not performed as stated in USP 34 <191>. * Sodium Bicarbonate (Project #(b)(4), Sample #(b)(4), Project #(b)(4) Sample #(b)(4), Project #(b)(4), Sample #(b)(4)); the Limit of Arsenic test was not performed as stated in USP 34 <211> Method I. b.) SOP P-1913 How to Use Electronic Laboratory Notebooks, does not specify when data should be entered into electronic laboratory notebooks. During inspection it was observed while performing Assay of Tramadol HCl tablets 50 mg (Sample Number (b)(4)) using Procedure (b)(4) the analyst was recording the raw data for the standard weight on the volumetric flask. c.) Inadequate Deviation Report prepared and followed for the following samples analyzed: * Deviation Report, 11-0107 for Project Number (b)(4) Bleach Solution and Deviation 11-0109 Project Number (b)(4), Bleach Solutions and future Bleach Solution Samples tested post 6/2/11. Modification was made to the USP 34 Monograph for Sodium Hypochlorite Topical Solution. However, one reagent, Bleach was modified and subsequent reagents acetic acid and potassium iodide ratio were not correspondingly changed. d.) Inadequate Technical Review and Quality Assurance review according to SOP P-1100 Version 10, Technical Review and Quality Assurance Release for the following records; * Project Number (b)(4) Sample Number (b)(4), Sodium bicarbonate For Sample (b)(4), Limit of Arsenic test USP 34 <211> Method I * Deviation Report 12-0221 and Deviation 12-0190 * Instrument Maintenance Form, Instrument (b)(4) date 30 Oct 2012, and Instrument (b)(4) 069

    Show observation 3
    1. Observation 3

      Employees lack required training

      Employees engaged in the processing of a drug product lack the training required to perform their assigned functions. Specifically, a.) Introduction to cGMPs training records for Director of Chemistry Laboratories not signed by trainer. b.) According to training records analyst received training in HPLC Analysis dated 11/05/12, however while observing analyst conducting HPLC analysis the analyst also could not identify the precipitate that was present in the sample preparation. c.) Two new analyst in-training stated they could independently analyze samples using Dissolution, HPLC, and UV-Visible Spectrophotometer equipment; however training records received on 11/7/12 did not indicate they were trained.

  8. VAIObjectionable conditions found; FDA took no regulatory actionDrugs · Drug Quality AssuranceCitations not published
  9. VAIObjectionable conditions found; FDA took no regulatory actionDrugs · Drug Quality Assurance5 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Complete Test Data

      Laboratory records are deficient in that they do not include a complete record of all data obtained during testing.

    2. Failing drug products not rejected

      Drug products failing to meet established specifications are not rejected.

    3. Instruments, apparatus, et. al. not meeting specs

      The use of instruments and apparatus not meeting established specifications was observed.

    4. Stability testing records not included

      Laboratory records do not include complete records of all stability testing performed.

    5. Following/documenting laboratory controls

      Established standards are not followed.

Regulations FDA cited here

The 7 sections of 21 CFR in FDA’s published citations for 4 of this facility’s 8 inspections, with how many of them cited each.

Observations read from posted 483 PDFs carry no CFR section of FDA’s, so they are not counted here.

FDA drug establishment registration

FDA registration search
Registered operations
Analysis
Registration runs through
Dec 31, 2027
Registrant
Boston Analytical Inc dba. BA SciencesD-U-N-S 181194176

From FDA’s Drug Establishments Current Registration Site, as of Nov 21, 2026. A registration says what a site told FDA it does. It is not an inspection result. FDA’s registration search looks firms up by name.

Go further than this page

FDAgent researches FDA’s records and links every answer to its source.

Free account: 3 questions, no card needed.

About this data. Every record here is filed under FEI 1221426 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
  • Every record on this page is filed under FEI 1221426. Records filed under another FEI or name, including other sites of the same company, are not shown here.
  • FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
  • Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
  • “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
  • Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
  • FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.