Supplier controls and component testing in FDA Form 483 observations

Qualifying suppliers and contract labs, testing incoming components, and relying on a supplier's certificate of analysis. FDAgent finds 128 such observations in 89 Form 483s issued to 86 facilities, for inspections that ended between March 2010 and September 2026. Each is quoted below from the 483 FDA posted.

A Form 483 lists the investigators' observations at the end of an inspection. It is not FDA's final determination.

Source: Form 483 PDFs in FDA's Electronic Reading Room, last synced Nov 23, 2026, 04:33 UTC. Every quote is from a 483 whose reading passed FDAgent's checks against the scan. FDAgent is independent of FDA.

Observations
128
Each quoted below
Form 483s
89
Posted by FDA
Facilities
86
Distinct FEIs
Most recent
Sep 9, 2026
Inspection end date

Observations by year

These observations by the year the inspection ended. They follow which 483s FDA has posted and FDAgent has read, so they show what is on record, not a trend in FDA's findings. 2026 runs to the latest sync (lighter bar).

Peak: 48 in 2026

Supplier controls observations in Form 483s by inspection year
YearObservations
20101
20110
20121
20132
20142
20153
20168
20171
20183
20197
20201
20214
20227
202314
20243
202523
202648

How FDAgent sorts observations into this theme

Two fixed rules, applied by code to every observation FDAgent can quote. An observation counts when either holds. No model and no similarity score decides.

It is tied to one of these 21 CFR entries

21 CFR 211.8421 CFR 820.50

The same entries the supplier controls topic counts. A 483 prints no regulation, so FDAgent ties an observation to a section only when it opens with FDA's own citation wording for that section, word for word.

Or its text contains one of these phrases

“supplier*” · “vendor qualif*” · “vendor audit*” · “vendor approv*” · “approved vendor*” · “certificate* of analys*” · “CoA*” · “COA*” · “purchasing control*” · “contract manufactur*” · “contract laborator*” · “contract test*” · “identity test*” · “component* … not … tested”

* matches any ending and … up to three words in between. Case is ignored unless a phrase has capitals.

The 128 observations, newest first

Quoted word for word from each Form 483, with only line breaks joined; … marks where the quote is cut. The highlight shows why the observation is in this theme. Each links to FDAgent's page for the 483.

  1. Jiangsu Intco Medical Products Co., LtdInspection ended Nov 17, 2023 · FEI 3005260401 · Observation 4

    Establishment of the reliability of the component supplier's report of analyses is deficient in that the test results are not appropriately validated at appropriate intervals. Specifically, a) [illegible] has never fully tested a drug component supplier’s certificate of analysis for an [illegible] drug component. b) Your firm qualified suppliers despite not having fully performed testing of drug components per USP.

    Opens with FDA's citation wording for 21 CFR 211.84(d)(2)Read the Form 483 →

  2. Catalent Indiana, LLCInspection ended Nov 15, 2023 · FEI 3005949964 · Observation 2

    … (6) 4 queries for the lots of stoppers 6 4 ( ) and lots of syringes 6 4 ) were limited to supplier complaints, as opposed to any b 4 record (deviation, LIR, CAPA, change control, etc.) that includes the aforementioned lots of components. 2. Deviation #749638, opened 09 AUG 2023, included a query to determine if other records included the words "unfilled" and "syringes". This query was limited to customer complaint records, as opposed to searching …

    Contains “supplier*”Read the Form 483 →

  3. Sumitomo Osaka Cement Co., LtdInspection ended Oct 27, 2023 · FEI 3003943727 · Observation 3

    Reports of analysis from component suppliers are accepted without establishing the reliability of the supplier's analyses through appropriate validation of the supplier's test results at appropriate intervals. Specifically, (b)(4) the raw material utilized in the manufacturing of Raw material suppliers of (b)(4) API (b)(4) were qualified without establishing for US (b)(4) and complete reliability of supplier's certificates of analysis (COA) …

    Contains “supplier*”Read the Form 483 →

  4. Sumitomo Osaka Cement Co., LtdInspection ended Oct 27, 2023 · FEI 3003943727 · Observation 4

    … The firm does not perform testing of assay and instead confirm assay with the COA of supplier to ensure if the results meet the firms' specification. 6. The operating (b)(4) of the (b)(4) Id # APK-002 utilized in (b)(4) of raw material to manufacture (b)(4) (b)(4) and (b)(4) API is (b)(4) to (b)(4). However, the calibration of the old (b)(4) was performed between (b)(4) to (b)(4) (Certificate# FK 610003-1-8) and between (b)(4) to (b)(4) …

    Contains “COA*”Read the Form 483 →

  5. Seqirus Vaccines LimitedInspection ended May 26, 2023 · FEI 3002806949 · Observation 2

    … The root cause (QE-000436) for the (b)(4) in the monovalent bulk drug (b)(4) (b)(4) sample has been attributed to external supplier, specifically the manufacturer of the (b)(4) with (b)(4) containers. The supplier investigation is ongoing, however there is no assurance that the Fluad Monovalent bulk is representative of the Quality Control samples taken for release testing of the monovalent drug substance batches. The firm continues to use the …

    Contains “supplier*”Read the Form 483 →

  6. Apollo Care LLCInspection ended Apr 5, 2023 · FEI 3013927023 · Observation 1

    … during review of your closed deviation investigation, number D-22-067, into contaminated media plates, lot number (b) (4), received from your media plate supplier, which you initiated on 09/06/2022 and listed the duration of the event as 08/15/2022 to 09/15/2022: -Deviation report number D-22-067 states "There has been a noticeable increase in growth on samples taken. Any alert and/or action limits exceeded will be disregarded for the stated …

    Contains “supplier*”Read the Form 483 →

  7. FRUEH VERPACKUNGSTECHNIK AGInspection ended Mar 24, 2023 · FEI 3010363503 · Observation 2

    Requirements that must be met by suppliers have not been adequately established. Specifically, components of (b)(4) drug product are not adequately controlled for bioburden levels. Specifications for (b)(4) do not include a bioburden requirement and incoming certificates of conformity do not certify that components meet a bioburden requirement.

    Opens with FDA's citation wording for 21 CFR 820.50(a)Read the Form 483 →

  8. Genlbet Biopbarmaceuticals SAInspection ended Jan 31, 2023 · FEI 3011160795 · Observation 1

    … The supplier for (b)(4) respectively), used allows the [illegible] for the (b)(4) steps during the manufacturing of [illegible]. A toxicological assessment was not performed to evaluate the safety of the presence of particles which are not particulate-free. [illegible] while using the (b)(4). f. There are no equipment logs for the (b)(4) and the usage and cleaning history of these pieces of equipment are not readily available.

    Contains “supplier*”Read the Form 483 →

  9. Catalent Indiana LLCInspection ended Sep 1, 2022 · FEI 3005949964 · Observation 1

    … The root cause of the bent (b)(4) was not investigated (i.e., supplier was not notified). F. Deviation 390307, dated 7/19/21, regarded a fiber found in the (b)(4) stopper of a retain, product (b)(4) lot (b)(4). The retains were reviewed as a response to complaints regarding lack of effect. The fiber was identified on 7/28/21 as (b)(4) organized, (b)(4) fiber (not product related). The investigation states the fiber has the characteristics of …

    Contains “supplier*”Read the Form 483 →

  10. Catalent Indiana LLCInspection ended Sep 1, 2022 · FEI 3005949964 · Observation 2

    … As provided by your firm, a total of 52 supplier complaints for stopper related issues were submitted between 01 Jul 2019 to 05 Aug 2022. Out of these, 17 supplier complaints were related to Stopper item# 402832. Three of these investigations are still open (SCR 555104, SCR 540975, SCR 534591 opened on 04 Jul 2022, 05 June 2022, and 22 May 2022, respectively). Your firm did not perform an adequate investigation into the repeat stopper defect …

    Contains “supplier*”Read the Form 483 →

  11. Catalent Indiana LLCInspection ended Sep 1, 2022 · FEI 3005949964 · Observation 11

    Records are not maintained so that data therein can be reviewed at least annually to evaluate the quality standards of each drug product to determine the need for changes in specifications or manufacturing or control procedures. Your firm did not establish criteria during supplier qualification and requalification based on recurring supplier complaints in A-SOP-03-03-003, Supplier Qualification. In addition, your firm has not been tracking and …

    Contains “supplier*”Read the Form 483 →

  12. AnazaoHealth CorporationInspection ended Jul 29, 2022 · FEI 3011152407 · Observation 1

    The quality unit lacks authority to review production records to assure that no errors have occurred and fully investigate errors that have occurred. Specifically, the firm has not been reviewing sterility test data packets from their contract testing laboratory for all drug products produced as part of their batch release process for the (b) (4) Suite since January 20, 2015 and for the (b) (4) Suite built in August 2020. Examples of drug …

    Contains “contract test*”Read the Form 483 →

  13. AnazaoHealth CorporationInspection ended Jul 29, 2022 · FEI 3011152407 · Observation 4

    The accuracy, sensitivity, specificity, and reproducibility of test methods have not been established and documented. Specifically, on May 19, 2022, your contract testing laboratory detected visual turbidity in the sterility testing (b) (4) for your (b) (4) Nutrient Cocktail lot (b) (4), and confirmed that this turbidity was caused by microbial contamination with Staphylococcus epidermidis. However, your contract testing laboratory failed to …

    Contains “contract test*”Read the Form 483 →

  14. ID Biomedical Corporation of QuebecInspection ended May 20, 2022 · FEI 3004835647 · Observation 2

    … probable root cause of "mishandling by the healthcare professional"; and the third was a most probable cause "the defect (LLA crack) was originated in the LLA suppliers process". No "internal root cause at GSK level". A crack in the LLA (luer lock adaptor) on the syringe was confirmed by the supplier for multiple lots of syringes, one of which was provided to your company. FDA was not notified with regards to the defective devices that were …

    Contains “supplier*”Read the Form 483 →

  15. (b)(4)Inspection ended Apr 12, 2022 · FEI 3008461022 · Observation 1

    As a very small importer, you did not obtain written assurance, before importing a food, that your foreign supplier produced the food in compliance with process and procedures that provide the required level of public health protection. This is a repeat observation from the previous inspection conducted on January 6, 2020. Specifically, you did not obtain written assurance, before importing a food, for (b) (4) from (b) (4) (b) (4) and (b) (4) and …

    Contains “supplier*”Read the Form 483 →

  16. Engage Uni LLCInspection ended Nov 12, 2021 · FEI 3016249968 · Observation 1

    Procedures for acceptance activities have not been adequately established. Specifically, Your firm's machining and finishing supplier for (b) (4), (b) (4), and (b) (4) failed to identify results of in-process inspection that deviated from the specifications documented on the inspection form, and these out of specification measurements were subsequently unidentified during DHR review conducted by your firm upon acceptance of finished devices. All …

    Contains “supplier*”Read the Form 483 →

  17. Dr. Reddy's Laboratories Ltd.Inspection ended Oct 29, 2021 · FEI 3006549835 · Observation 5

    Establishment of the reliability of the component supplier's report results are not appropriately validated at appropriate. Specifically, results are taken from the manufacturer an appropriate interval. Your procedures do include the requirement for Packaging Material (SOP~C-0094) at a specified Materials (SOP~C-0113) and Sampling procedures allow for test results to be taken from availability of in-house testing facility. There is no the …

    Contains “supplier*”Read the Form 483 →

  18. Med Shop Total Care Inc.Inspection ended Jul 27, 2021 · FEI 3002835459 · Observation 7

    Your firm released drug product in which the strength differs from, or its purity or quality falls below, that which it purports or is represented to possess. Specifically, after notification from your contract laboratory of a sterility failure for bulk intermediate vial of Bupivacaine [illegible], lot (b)(4), you recalled drug products that contained this lot.

    Contains “contract laborator*”Read the Form 483 →

  19. Imprimis NJOF, -LLCInspection ended Jan 29, 2021 · FEI 3013024146 · Observation 2

    … The (b) (4) PV lot (b) (4) failed specification limits according to the COA provided by the external testing lab when results for sample representing (b) (4) the batch for Bromfenac potency failed the test with a value of 110.4% against a specification limit of (b) (4)%. The firm accepted the failing results and counted the batch as an acceptable PV lot. In addition, the firm rejected the (b) (4) PV batch (b) (4) when potency values (for …

    Contains “COA*”Read the Form 483 →

  20. Vital Care of Dickson, LLC.Inspection ended Mar 11, 2020 · FEI 3015826069 · Observation 1

    Compounding with APIs that have not been verified to assure that they do not contribute endotoxin contamination that may be objectionable given the product's intended use. Specifically, the Certificate of Analysis for the APIs that the firm uses do not document if endotoxin tests have been performed and the firm has not performed any endotoxin testing on APIs or finished sterile drug products which are for intrathecal administration.

    Contains “certificate* of analys*”Read the Form 483 →

21 CFR sections these observations open with

The sections FDAgent ties the most of these observations to by FDA's wording, with how many of this theme's observations each covers. Each links to its page of FDA citations, warning letters and the regulation text.

Ask FDAgent about supplier controls in Form 483s

FDAgent reads FDA's Form 483s, inspections, citations and warning letters and answers in plain English, with links to the records.

Edit it or ask as written. 3 questions free, no card needed.

About this data

  • The observations are the text FDAgent read by OCR from Form 483 PDFs in FDA's Electronic Reading Room, shown only from 483s whose every printed observation passed FDAgent's checks against the scanned text, as on each 483's own page.
  • Which theme an observation belongs to is decided by the fixed rules above; every count on this page is computed from them.
  • A 483 addressed to an individual is listed without the name, and an observation that names a person is left out.
  • FDAgent is independent of FDA.