Supplier controls and component testing in FDA Form 483 observations
Qualifying suppliers and contract labs, testing incoming components, and relying on a supplier's certificate of analysis. FDAgent finds 128 such observations in 89 Form 483s issued to 86 facilities, for inspections that ended between March 2010 and September 2026. Each is quoted below from the 483 FDA posted.
A Form 483 lists the investigators' observations at the end of an inspection. It is not FDA's final determination.
Source: Form 483 PDFs in FDA's Electronic Reading Room, last synced Nov 23, 2026, 04:33 UTC. Every quote is from a 483 whose reading passed FDAgent's checks against the scan. FDAgent is independent of FDA.
- Observations
- 128
- Each quoted below
- Form 483s
- 89
- Posted by FDA
- Facilities
- 86
- Distinct FEIs
- Most recent
- Sep 9, 2026
- Inspection end date
Observations by year
These observations by the year the inspection ended. They follow which 483s FDA has posted and FDAgent has read, so they show what is on record, not a trend in FDA's findings. 2026 runs to the latest sync (lighter bar).
Peak: 48 in 2026
| Year | Observations |
|---|---|
| 2010 | 1 |
| 2011 | 0 |
| 2012 | 1 |
| 2013 | 2 |
| 2014 | 2 |
| 2015 | 3 |
| 2016 | 8 |
| 2017 | 1 |
| 2018 | 3 |
| 2019 | 7 |
| 2020 | 1 |
| 2021 | 4 |
| 2022 | 7 |
| 2023 | 14 |
| 2024 | 3 |
| 2025 | 23 |
| 2026 | 48 |
How FDAgent sorts observations into this theme
Two fixed rules, applied by code to every observation FDAgent can quote. An observation counts when either holds. No model and no similarity score decides.
It is tied to one of these 21 CFR entries
The same entries the supplier controls topic counts. A 483 prints no regulation, so FDAgent ties an observation to a section only when it opens with FDA's own citation wording for that section, word for word.
Or its text contains one of these phrases
“supplier*” · “vendor qualif*” · “vendor audit*” · “vendor approv*” · “approved vendor*” · “certificate* of analys*” · “CoA*” · “COA*” · “purchasing control*” · “contract manufactur*” · “contract laborator*” · “contract test*” · “identity test*” · “component* … not … tested”
* matches any ending and … up to three words in between. Case is ignored unless a phrase has capitals.
The 128 observations, newest first
Quoted word for word from each Form 483, with only line breaks joined; … marks where the quote is cut. The highlight shows why the observation is in this theme. Each links to FDAgent's page for the 483.
Greenwich RxInspection ended Sep 9, 2026 · FEI 3043120213 · Observation 1
… The firm continues to receive this grade of API based on a review of supplier provided COAs for Glutathione (L) USP.
Contains “supplier*”Read the Form 483 →
Greenwich RxInspection ended Sep 9, 2026 · FEI 3043120213 · Observation 4
… The rubber stoppers were removed from a non-sterile (b)(4) bag that were originally stored in the approved supplier's sterile packaging.
Contains “supplier*”Read the Form 483 →
OPS International Inc dba Olympia PharmaceuticalsInspection ended Sep 3, 2026 · FEI 3009724085 · Observation 2
… to your Senior Quality Manager and Senior Director of Compliance, following the in-house incubation of the swab samples, the samples were sent to your contract laboratory for sterility testing; however, the associated records, including the third-party sterility test results, could not be located. (B) On August 21, 2026, during aseptic setup of the (b)(4) filling line for Biotin 0.5 mg/mL (Lot #H17A05-26), [illegible] Chamber (Material Chamber) …
Contains “contract laborator*”Read the Form 483 →
OPS International Inc dba Olympia PharmaceuticalsInspection ended Sep 3, 2026 · FEI 3009724085 · Observation 4
… were submitted for size characterization, the analysis was performed using the particle-size distribution method described in USP <429>, as documented in the Certificates of Analysis. The resulting data do not establish the size of the individual particulate defect contained in each qualification unit. (B) No written procedures or records exist documenting the preparation and contents of each particulate defect unit, including particulate type, …
Contains “certificate* of analys*”Read the Form 483 →
OPS International Inc dba Olympia PharmaceuticalsInspection ended Sep 3, 2026 · FEI 3009724085 · Observation 5
The written stability program for drug products does not include reliable, meaningful and specific test methods. Specifically, your firm assigns a (b)(4) beyond-use date (BUD) to Sermorelin Acetate 0.9 mg/mL Injection based on your contract laboratory stability study. However, your firm failed to provide adequate demonstration that the analytical procedure can detect relevant degradation-related changes. Your method validation report for …
Contains “contract laborator*”Read the Form 483 →
OPS International Inc dba Olympia PharmaceuticalsInspection ended Sep 3, 2026 · FEI 3009724085 · Observation 7
Component testing is deficient in that each component is not tested for conformity with all appropriate written specifications for purity, strength, and quality. Specifically, (A) SOP-035, Procurement, Receipt, and Inspection of Incoming Components and Materials, does not include a sampling plan specifying the number of containers to be sampled per shipment. According to the Senior Supply Chain Manager, (b)(4) sent from your (b)(4) warehouse to …
Opens with FDA's citation wording for 21 CFR 211.84(d)(2)Read the Form 483 →
Houston Rx LLCInspection ended Aug 28, 2026 · FEI 3045022201 · Observation 9
Use of ingredients not intended for pharmaceutical use in sterile drug production. Specifically, your firm is using active pharmaceutical ingredients and excipients that are not always USP/NF grade or are non-pharmaceutical grade. For example, a) The Certificate of Analysis (CoA) for lot #(b)(4) of Methylcobalamin states "For R&D only". Your firm has made approximately (b)(4) lots of Methylcobalamin 5mg/mL using lot #(b)(4). b) The CoA for lot …
Contains “certificate* of analys*”Read the Form 483 →
Houston Rx LLCInspection ended Aug 28, 2026 · FEI 3045022201 · Observation 14
Your firm is not storing active pharmaceutical ingredients (API) per the labeled storage conditions. All APIs are stored in a (b)(4) refrigerator. Examples of APIs used in the production of sterile drug products that are not stored according to the labeled storage conditions include the following. Sterile drug products made with these APIs have been released and distributed. a) The CoA for lot #(b)(4) of Semaglutide API states to store at (b)(4). …
Contains “CoA*”Read the Form 483 →
TAILSTORM HEALTH INCInspection ended Aug 28, 2026 · FEI 3015929581 · Observation 5
… Growth promotion on the manufacturer COA for Plate Count (b) (4) only lists Staphylococcus aureus, Escherichia coli, and Enterococcus faecalis. It does not include waterborne pathogenic microorganisms of interest such as B. cepacia or P. aeruginosa that are known to colonize high purity (b) (4) systems. B. The firm accepts Plate Count (b) (4) plates based on manufacturer COA and does not perform growth promotion before use. Nitrogen Your firm …
Contains “COA*”Read the Form 483 →
Cascade Specialty Pharmacy LLCInspection ended Jul 17, 2026 · FEI 3015133983 · Observation 5
Specific identification tests are not conducted on components that have been accepted based on the supplier's report of analysis. Specifically, Your firm accepts the material based on the supplier's certificate of analysis in lieu of testing each component received. However, your firm does not perform identity tests on each component received to ensure that the component received has the purported identity. The following components were accepted …
Opens with FDA's citation wording for 21 CFR 211.84(d)(2)Read the Form 483 →
Brookfield Medical Surgical Supplies IncInspection ended Jul 10, 2026 · FEI 3011430551 · Observation 1
… All OOS were reportedly attributed to the contract testing laboratory and resolved via retesting at the same lab. Table below shows all the potency OOS: Deviation Date Type # Initiated DV24-051 September 19, 2024 Potency OOS DV25-053 April 3, 2025 Potency Morphine Sulfate OOS DV25-054 April 7, 2025 Potency Methylprednisolone OOS DV25-055 June 9, 2025 Potency Betamethasone Sodium Phosphate OOS DV25-056 July 3, 2025 Potency Triamcinolone Acetonide …
Contains “contract test*”Read the Form 483 →
Brookfield Medical Surgical Supplies IncInspection ended Jul 10, 2026 · FEI 3011430551 · Observation 2
… (12) of these OOS results were invalidated and product was released without documented evidence that the quality control unit independently reviewed the contract laboratory's raw data, chromatograms, instrument logs and invalidation rationale. One result (Triamcinolone Acetonide 3 mL, Lot 051526-1) was confirmed and the contract lab's invalidation conclusions in 12 of 13 cases were not independently verified by the QCU. For DV25-058 …
Contains “contract laborator*”Read the Form 483 →
Glenmark Pharmaceuticals LimitedInspection ended Jun 30, 2026 · FEI 3004672766 · Observation 3
Each component is not tested for conformity with all written specifications for purity, strength, and quality. Specifically, Your firm failed to perform individual container testing, including identity testing with limits for (b)(4) (NMT (b)(4)%), on each individual container of each shipment of each lot of high-risk drug components prior to their use in manufacturing of drug products distributed to the United States market. Your procedure for …
Contains “component* … not … tested”Read the Form 483 →
Cytovance Biologics Inc.Inspection ended Jun 17, 2026 · FEI 3006341547 · Observation 5
There is a failure to justify defined intervals for conducting periodic full testing (verifying supplier's Certificate of Analysis). Specifically, per OKC-SOP-1379 Raw Material Test Strategy, revision 03 effective 18 Feb 2026, periodic full testing of raw materials is performed (b) (4) (or the next receipt (b) (4) after the last full testing) for common stock products and at every (b) (4) receipt for project specific raw materials. Your firm does …
Contains “supplier*”Read the Form 483 →
Cytovance Biologics Inc.Inspection ended Jun 17, 2026 · FEI 3006341547 · Observation 6
… Only (b) (4) sampled for identity testing. This lot has been used to manufacture several lots of (b) (4), including lot #(b) (4) released on March 19, 2026.
Contains “identity test*”Read the Form 483 →
Jubilant HollisterStier LLCInspection ended Jun 17, 2026 · FEI 3010477 · Observation 1
Component testing is deficient in that each component is not tested for conformity with all appropriate written specifications for purity, strength, and quality. Specifically, your firm does not always perform complete identification testing of raw materials (b)(4) that is used as an ingredient in the manufacture of your (b)(4) injection and (b)(4) solution (b)(4) products. For (b)(4) raw materials supplied to you by your client used in the …
Opens with FDA's citation wording for 21 CFR 211.84(d)(2)Read the Form 483 →
PQ Pharmacy LLCInspection ended Jun 1, 2026 · FEI 3017374013 · Observation 7
… B) You do not have a scientific rationale for not performing identity testing for glycerin and propylene glycol, which includes evaluating acceptable levels of diethylene glycol and ethylene glycol. These components used in part for the manufacture of sterile drug products.
Contains “identity test*”Read the Form 483 →
PETNET Solutions Inc.Inspection ended May 22, 2026 · FEI 3036193770 · Observation 1
You did not use a reliable supplier as a source of each lot of component. Specifically, Your firm failed to follow procedure SOP # D0011306 titled "Supplier Qualification, Approval, and Maintenance," Revision H (Effective Date: 11/1/2025), which establishes requirements for vendor selection, approval, and oversight. For example: A. Your firm failed to show (b) (4) as an approved supplier of the precursor (b) (4) mg/vial), a critical component …
Contains “supplier*”Read the Form 483 →
Shanghai AmbioPharm, Inc.Inspection ended May 22, 2026 · FEI 3006494603 · Observation 1
The responsibilities and procedures applicable to the quality control unit are not in writing or fully followed. Specifically, Your firm utilizes third-party testing from multiple contract testing laboratories. Your firm has not established a Quality Agreement with the third-party laboratories or provided approval for any subcontracted testing not performed by your headquarters.
Contains “contract test*”Read the Form 483 →
USV Private LimitedInspection ended Apr 17, 2026 · FEI 3004086192 · Observation 6
Written procedures for qualifying, controlling, and evaluating suppliers and materials used in aseptic processing areas are not established, written or followed. Specifically, your firm lacks a comprehensive, risk-based program for qualifying and monitoring suppliers of critical consumables used in ISO 5 (Grade A) areas, including garments, (b) (4), wipes, and disinfectants. Your firm relies solely on supplier Certificates of Analysis without …
Contains “supplier*”Read the Form 483 →
21 CFR sections these observations open with
The sections FDAgent ties the most of these observations to by FDA's wording, with how many of this theme's observations each covers. Each links to its page of FDA citations, warning letters and the regulation text.
- 21 CFR 211.84Testing and approval or rejection of components, drug product containers, and closuresIn this theme's rule15
- 21 CFR 211.22Responsibilities of quality control unit8
- 21 CFR 211.192Production record review6
- 21 CFR 211.160General requirements5
- 21 CFR 211.100Written procedures; deviations4
- 21 CFR 211.113Control of microbiological contamination3
- 21 CFR 211.166Stability testing3
- 21 CFR 211.194Laboratory records3
- 21 CFR 820.75Process validation2
- 21 CFR 1.512What FSVP may I have if I am a very small importer or I am importing certain food from certain small foreign suppliers?1
Ask FDAgent about supplier controls in Form 483s
FDAgent reads FDA's Form 483s, inspections, citations and warning letters and answers in plain English, with links to the records.
Related
About this data
- The observations are the text FDAgent read by OCR from Form 483 PDFs in FDA's Electronic Reading Room, shown only from 483s whose every printed observation passed FDAgent's checks against the scanned text, as on each 483's own page.
- Which theme an observation belongs to is decided by the fixed rules above; every count on this page is computed from them.
- A 483 addressed to an individual is listed without the name, and an observation that names a person is left out.
- FDAgent is independent of FDA.