Supplier controls and component testing in FDA Form 483 observations
Qualifying suppliers and contract labs, testing incoming components, and relying on a supplier's certificate of analysis. FDAgent finds 128 such observations in 89 Form 483s issued to 86 facilities, for inspections that ended between March 2010 and September 2026. Each is quoted below from the 483 FDA posted.
A Form 483 lists the investigators' observations at the end of an inspection. It is not FDA's final determination.
Source: Form 483 PDFs in FDA's Electronic Reading Room, last synced Nov 23, 2026, 04:33 UTC. Every quote is from a 483 whose reading passed FDAgent's checks against the scan. FDAgent is independent of FDA.
- Observations
- 128
- Each quoted below
- Form 483s
- 89
- Posted by FDA
- Facilities
- 86
- Distinct FEIs
- Most recent
- Sep 9, 2026
- Inspection end date
Observations by year
These observations by the year the inspection ended. They follow which 483s FDA has posted and FDAgent has read, so they show what is on record, not a trend in FDA's findings. 2026 runs to the latest sync (lighter bar).
Peak: 48 in 2026
| Year | Observations |
|---|---|
| 2010 | 1 |
| 2011 | 0 |
| 2012 | 1 |
| 2013 | 2 |
| 2014 | 2 |
| 2015 | 3 |
| 2016 | 8 |
| 2017 | 1 |
| 2018 | 3 |
| 2019 | 7 |
| 2020 | 1 |
| 2021 | 4 |
| 2022 | 7 |
| 2023 | 14 |
| 2024 | 3 |
| 2025 | 23 |
| 2026 | 48 |
How FDAgent sorts observations into this theme
Two fixed rules, applied by code to every observation FDAgent can quote. An observation counts when either holds. No model and no similarity score decides.
It is tied to one of these 21 CFR entries
The same entries the supplier controls topic counts. A 483 prints no regulation, so FDAgent ties an observation to a section only when it opens with FDA's own citation wording for that section, word for word.
Or its text contains one of these phrases
“supplier*” · “vendor qualif*” · “vendor audit*” · “vendor approv*” · “approved vendor*” · “certificate* of analys*” · “CoA*” · “COA*” · “purchasing control*” · “contract manufactur*” · “contract laborator*” · “contract test*” · “identity test*” · “component* … not … tested”
* matches any ending and … up to three words in between. Case is ignored unless a phrase has capitals.
The 128 observations, newest first
Quoted word for word from each Form 483, with only line breaks joined; … marks where the quote is cut. The highlight shows why the observation is in this theme. Each links to FDAgent's page for the 483.
USV Private LimitedInspection ended Apr 17, 2026 · FEI 3004086192 · Observation 7
Each component is not tested for conformity with all written specifications for purity, strength, and quality. Specifically, your firm lacks appropriate testing of components used in the manufacture of your drug product. For example, your firm lacks a specific identity test to detect (b) (4) in all shipments, containers, and lots of (b) (4) before use in the manufacturing of (b) (4) Tablets. In the last two years your firm has manufactured over …
Contains “component* … not … tested”Read the Form 483 →
Conroe Welding Supply, IncInspection ended Apr 3, 2026 · FEI 1625647 · Observation 1
… Your firm failed to conduct supplier audits and failed to establish a supplier audit management SOP. Your firm has never performed supplier audits. K. Your recall SOP is inadequate. It does not include important elements, including but not limited to investigation procedures. Additionally, the SOP does not have an effective date, version number, or reviewer/approver signature. L. Your firm failed to maintain records of medical gas distribution. …
Contains “supplier*”Read the Form 483 →
Conroe Welding Supply, IncInspection ended Apr 3, 2026 · FEI 1625647 · Observation 4
Your firm failed to perform required tests on bulk medical (b) (4) USP received for production. Specifically, No testing was performed on any bulk lot of medical (b) (4) USP received from your supplier. Your firm has not established and maintained a program to ensure the reliability of the supplier's capabilities through appropriate assessment and testing procedures.
Contains “supplier*”Read the Form 483 →
Adimmune CorporationInspection ended Apr 2, 2026 · FEI 3008919998 · Observation 2
Your firm failed to establish the reliability of supplier's analyses through appropriate validation of the supplier's test results at appropriate intervals. Specifically, (b) (4) and (b) (4) utilized for drug product overlay have not been conducted to validate supplier analyses.
Contains “supplier*”Read the Form 483 →
Dendreon Pharmaceuticals LLCInspection ended Mar 25, 2026 · FEI 3008821247 · Observation 1
… Despite these findings, the lot was approved for use by Supplier Quality Assurance (SQA). This lot was subsequently used in the manufacturing of (b) (4) final product batches of Sipuleucel-T and remains the current lot in use for finished product. b) For critical material, (b) (4), Lot # (b) (4) were rejected during the (b) (4) packaging operation post-sterilization due to defects such as cracked (b) (4) and (b) (4). However, Quality Unit did not …
Contains “supplier*”Read the Form 483 →
Dendreon Pharmaceuticals LLCInspection ended Mar 25, 2026 · FEI 3008821247 · Observation 4
The responsibilities and procedures applicable to the quality control unit are not fully followed. A. Your firm uses a contract manufacturer to (b) (4) (b) (4) of your (b) (4). Your firm's quality control unit does not ensure that the final visual inspection for the (b) (4) solution is completed and acceptable for release. For example, your quality control unit did not document the reason for rejection provided by the contract manufacturer for …
Contains “contract manufactur*”Read the Form 483 →
Dendreon Pharmaceuticals LLCInspection ended Mar 25, 2026 · FEI 3008821247 · Observation 6
… A comparison study was conducted to demonstrate the comparability between your in-house manufacturing process and the prior contract manufacturer's process. Your current manufacturing process significantly differs from the contract manufacturer's process in the following parameters: I. Lot Size II. Mixing Conditions III. Hold times IV. Elimination of (b) (4) step V. Filling Environment ISO (b) (4) to ISO (b) (4) VI. Method of (b) (4) …
Contains “contract manufactur*”Read the Form 483 →
SGS North America Inc.Inspection ended Mar 20, 2026 · FEI 1418028 · Observation 1
… Instead, according to SOP-000898, Reference Standards, Version 4, your firm relies on a review of the Certificate of Analysis of incoming lots to assess the quality of the reference standard. Examples are as follows: A. Your firm used (b) (4) Reference Standard (Lot (DB) (4) supplied by (b) (4) received 04Jun2025) to conduct method transfer activities without performing verification testing to confirm the identity, quality, and suitability of the …
Contains “certificate* of analys*”Read the Form 483 →
Accura-Pac, Inc.Inspection ended Mar 5, 2026 · FEI 1818911 · Observation 4
‘Written procedures are not followed for the approval of components. Specifically, Your firm failed to follow your written procedures for qualifying new or existing raw material (active ingredient and excipient) suppliers. According to your written procedures CORP-QA-SOP-1887, titled “Supplier Selection, Qualification, and Maintenance”, your firm failed to perform site audit for Risk Level I suppliers (active ingredient suppliers), and failed to …
Contains “supplier*”Read the Form 483 →
Unilab Chemicals and Pharmaceuticals Private Ltd.Inspection ended Feb 27, 2026 · FEI 3004985767 · Observation 1
There is a failure to establish the reliability of the component supplier's report of analyses. Specifically, A) There is no supplier qualification or quality agreement established for the following non-USP solvents used for manufacturing APIs: (b) (4) B) Certificate of Analysis (COA) test results provided by the supplier of non-USP solvents have not been validated. This includes the lack of verification for Identity test results, (b) (4) Limit …
Contains “supplier*”Read the Form 483 →
Hillyard GMPInspection ended Feb 27, 2026 · FEI 3009690111 · Observation 1
OBSERVATION 1 (QUALITY SYSTEM) Written procedures for the qualification and approval implemented. Specifically, Your firm's SOP QA200-07.02 Vendor Audits effective date 8/19/2020, requires the performance paper audits as part of supplier qualification. However, paper audits were not conducted for suppliers components used in the manufacture of drug products. a. There was no documentation demonstrating that suppliers approved by the Quality Unit …
Contains “vendor audit*”Read the Form 483 →
RayzeBio, IncInspection ended Feb 24, 2026 · FEI 3030530655 · Observation 10
Your firm failed to test samples of each component for identity and conformity with all appropriate written specifications for purity, strength, and quality. Your firm also failed to validate and establish the reliability of your component supplier's test analyses at appropriate intervals. Specifically, 1. Your firm does not conduct identity testing of raw materials used in the production of (b)(4). 2. Your firm does not conduct testing on …
Contains “supplier*”Read the Form 483 →
Orean Personal Care LtdInspection ended Feb 20, 2026 · FEI 3010999645 · Observation 2
… You do not have quality agreements with the suppliers of API (b)(4). The release test and stability study of your drug products is carried out at a third-party lab (b)(4). You do not have quality agreement with this (b)(4) lab. G. There are not written procedures that guide the production of (b)(4) or the cleaning/maintenance of the (b)(4) system used for the production of OTC drug products and the cleaning of equipment of the (b)(4) used to …
Contains “supplier*”Read the Form 483 →
Orean Personal Care LtdInspection ended Feb 20, 2026 · FEI 3010999645 · Observation 6
Written procedures are lacking which describe in sufficient detail the receipt, sampling and testing of components of drug products. Specifically, your procedure, OTC Products Sampling and Process WIPROD027 Version 12 (Issued on 11/28/2024) does not contain information about collecting a representative sample. You are in the process of vendor qualification for (b)(4). As of now you have completed the testing of (b)(4) batches of (b)(4) Batch No's …
Contains “vendor qualif*”Read the Form 483 →
Almon Healthcare Private LimitedInspection ended Feb 13, 2026 · FEI 3030509366 · Observation 1
… When the customer requested microbiology testing data, your QC/QA revised the COA to include microbiology specification results, which were not reviewed and approved by QA at the time of release. In the batch release check list QC and QA marked N/A for microbial analysis approved/complies as per specification. d. Your QA did not initiate deviation investigations for production and facility maintenance departments failing to complete work …
Contains “COA*”Read the Form 483 →
Almon Healthcare Private LimitedInspection ended Feb 13, 2026 · FEI 3030509366 · Observation 7
Failure to test the identity of each batch of incoming production material and appropriately qualify suppliers to rely upon their Certificate of Analysis. Your firm failed to perform identity testing on raw materials (b)(4) accepting them based solely on supplier COAs. Your QA Executive confirmed that the supplier for (b)(4) intentionally mislabels and ships (b)(4) [illegible] batches received since 2024. Your firm used this untested [illegible] …
Contains “supplier*”Read the Form 483 →
Almon Healthcare Private LimitedInspection ended Feb 13, 2026 · FEI 3030509366 · Observation 10
… Your firm also sent (b)(4) retain samples to contract testing laboratory for (b)(4) (b)(4) batches (b)(4) analysis.
Contains “contract test*”Read the Form 483 →
Biomylz Private LimitedInspection ended Feb 9, 2026 · FEI 3013697420 · Observation 4
… laboratories that are not registered or inspected by the agency to perform incoming raw material and finished OTC drug product testing, including the Certificate of Analysis (CoA) used to support the release decision. In addition, no Quality Agreements have been established between your firm and the contracted laboratories to define roles, responsibilities, and oversight related to testing, data integrity, and compliance with cGMP requirements. …
Contains “certificate* of analys*”Read the Form 483 →
Yuki Gosei Kogyo, Co., LtdInspection ended Feb 6, 2026 · FEI 3002808534 · Observation 3
Your supplier qualifications are deficient. Specifically, US marketed API, namely, (b)(4) uses the key starting material (b)(4). Your written procedure, SOP# JHA440, Supplier Control Procedure, dated 5/1/2025, section 3.5, requires onsite audit of main raw material supplier's site (b)(4). However, you have not conducted any onsite audit of this supplier after 3/2/2023. Also, no justification is documented to extend the onsite audit time beyond …
Contains “supplier*”Read the Form 483 →
Neeyaan LLCInspection ended Jan 27, 2026 · FEI 3024366553 · Observation 1
Laboratory records are deficient in that they do not include a complete record of all data obtained during testing. Specifically, The original raw data from the firm's own testing and Certificate of Analysis (COA) from the third-party testing laboratory that performed the finished product testing for OTC Drug Product (b) (4), USP ((b) (4) Tablet) Lot # (b) (4) could not be identified or provided during the inspection. Your firm personnel provided …
Contains “certificate* of analys*”Read the Form 483 →
21 CFR sections these observations open with
The sections FDAgent ties the most of these observations to by FDA's wording, with how many of this theme's observations each covers. Each links to its page of FDA citations, warning letters and the regulation text.
- 21 CFR 211.84Testing and approval or rejection of components, drug product containers, and closuresIn this theme's rule15
- 21 CFR 211.22Responsibilities of quality control unit8
- 21 CFR 211.192Production record review6
- 21 CFR 211.160General requirements5
- 21 CFR 211.100Written procedures; deviations4
- 21 CFR 211.113Control of microbiological contamination3
- 21 CFR 211.166Stability testing3
- 21 CFR 211.194Laboratory records3
- 21 CFR 820.75Process validation2
- 21 CFR 1.512What FSVP may I have if I am a very small importer or I am importing certain food from certain small foreign suppliers?1
Ask FDAgent about supplier controls in Form 483s
FDAgent reads FDA's Form 483s, inspections, citations and warning letters and answers in plain English, with links to the records.
Related
About this data
- The observations are the text FDAgent read by OCR from Form 483 PDFs in FDA's Electronic Reading Room, shown only from 483s whose every printed observation passed FDAgent's checks against the scanned text, as on each 483's own page.
- Which theme an observation belongs to is decided by the fixed rules above; every count on this page is computed from them.
- A 483 addressed to an individual is listed without the name, and an observation that names a person is left out.
- FDAgent is independent of FDA.