Supplier controls and component testing in FDA Form 483 observations
Qualifying suppliers and contract labs, testing incoming components, and relying on a supplier's certificate of analysis. FDAgent finds 128 such observations in 89 Form 483s issued to 86 facilities, for inspections that ended between March 2010 and September 2026. Each is quoted below from the 483 FDA posted.
A Form 483 lists the investigators' observations at the end of an inspection. It is not FDA's final determination.
Source: Form 483 PDFs in FDA's Electronic Reading Room, last synced Nov 23, 2026, 04:33 UTC. Every quote is from a 483 whose reading passed FDAgent's checks against the scan. FDAgent is independent of FDA.
- Observations
- 128
- Each quoted below
- Form 483s
- 89
- Posted by FDA
- Facilities
- 86
- Distinct FEIs
- Most recent
- Sep 9, 2026
- Inspection end date
Observations by year
These observations by the year the inspection ended. They follow which 483s FDA has posted and FDAgent has read, so they show what is on record, not a trend in FDA's findings. 2026 runs to the latest sync (lighter bar).
Peak: 48 in 2026
| Year | Observations |
|---|---|
| 2010 | 1 |
| 2011 | 0 |
| 2012 | 1 |
| 2013 | 2 |
| 2014 | 2 |
| 2015 | 3 |
| 2016 | 8 |
| 2017 | 1 |
| 2018 | 3 |
| 2019 | 7 |
| 2020 | 1 |
| 2021 | 4 |
| 2022 | 7 |
| 2023 | 14 |
| 2024 | 3 |
| 2025 | 23 |
| 2026 | 48 |
How FDAgent sorts observations into this theme
Two fixed rules, applied by code to every observation FDAgent can quote. An observation counts when either holds. No model and no similarity score decides.
It is tied to one of these 21 CFR entries
The same entries the supplier controls topic counts. A 483 prints no regulation, so FDAgent ties an observation to a section only when it opens with FDA's own citation wording for that section, word for word.
Or its text contains one of these phrases
“supplier*” · “vendor qualif*” · “vendor audit*” · “vendor approv*” · “approved vendor*” · “certificate* of analys*” · “CoA*” · “COA*” · “purchasing control*” · “contract manufactur*” · “contract laborator*” · “contract test*” · “identity test*” · “component* … not … tested”
* matches any ending and … up to three words in between. Case is ignored unless a phrase has capitals.
The 128 observations, newest first
Quoted word for word from each Form 483, with only line breaks joined; … marks where the quote is cut. The highlight shows why the observation is in this theme. Each links to FDAgent's page for the 483.
Neeyaan LLCInspection ended Jan 27, 2026 · FEI 3024366553 · Observation 13
Specific identification tests are not conducted on components that have been accepted based on the suppliers report of analysis. Specifically, No comprehensive testing was conducted to establish the reliability of the API (b) (4), USP vendor which is not in accordance with section 5.5 of SOP-20 "Vendor Qualification". The SOP requires that "in addition to the questionnaire, a sample of the material supplied by the vendor in question will be …
Contains “supplier*”Read the Form 483 →
Neeyaan LLCInspection ended Jan 27, 2026 · FEI 3024366553 · Observation 15
… According to the Interim Production Supervisor, the labels had been packed for shipment and return to the supplier; however, there was no documentation in the Label Room Log to trace the movement of labels. B) Batch production records for OTC drug products, including but not limited to (b) (4) tablet Lot #(b) (4), (b) (4) tablet Lot #(b) (4) and (b) (4) tablet Lot #(b) (4), lacked the following documentation: Packaging and label reconciliation to …
Contains “supplier*”Read the Form 483 →
Neeyaan LLCInspection ended Jan 27, 2026 · FEI 3024366553 · Observation 19
You did not conduct appropriate tests or examinations or rely on a certificate of analysis to determine whether components met established specifications. Specifically, You manufactured (b) (4) Tablet, Batch No #: (b) (4), which was distributed on 4/30/25. The following was observed: You provided a certificate of analysis using Neeyaan letterhead dated 04/05/2025 indicating (b) (4) Batch No: (b) (4) was used to manufacture (b) (4) Tablet, Batch …
Contains “certificate* of analys*”Read the Form 483 →
Neeyaan LLCInspection ended Jan 27, 2026 · FEI 3024366553 · Observation 20
Your quality control operations did not include determining whether each finished batch conforms to established product specifications. Specifically, You manufactured (b) (4) Tablet, Batch No #: (b) (4), which was distributed on 4/30/25. You provided a certificate of analysis using Neeyaan letterhead dated 05-22-2025. You do not conduct testing in-house and use a third-party lab for testing. You do not have testing records from the lab to …
Contains “certificate* of analys*”Read the Form 483 →
Neeyaan LLCInspection ended Jan 27, 2026 · FEI 3024366553 · Observation 21
… However, your certificate of analysis for (b) (4) Tablet, Batch No #: (b) (4) was signed as approved on 05/22/2025, which is after the distribution date of 4/30/25. In addition, your certificate of analysis for (b) (4) Capsule, Batch No #: (b) (4) was signed as approved on 05/19/2025, which is after the distribution date of 04/28/25.
Contains “certificate* of analys*”Read the Form 483 →
Shimoga ChemicalsInspection ended Jan 23, 2026 · FEI 3033881432 · Observation 9
The responsibilities and procedures applicable to the quality control unit are not in writing and fully followed. Specifically, A. Your firm failed to follow its vendor qualification procedure. Vendor qualification questionnaires were available for only (b)(4) raw materials, and your firm did not document a rationale for limiting vendor qualification to these materials. B. Your firm failed to establish quality agreements with your broker and …
Contains “vendor qualif*”Read the Form 483 →
US Specialty Formulations LLCInspection ended Jan 14, 2026 · FEI 3010680515 · Observation 1
… (D) Your Quality Unit failed to ensure that identity testing Sarracenia Purpurea Bulk Distillate production, which g/mL for Injection. An assay method is utilized (E) Following discovery of an incorrect expiration date b)(4 vials were relabeled for Sarracenia Purpurea 2/28/2026). However, there is no QA approval for the relabeling activity, no documented relabeling procedure, no reconciliation of removed labels, and no reconciliation with the …
Contains “identity test*”Read the Form 483 →
US Specialty Formulations LLCInspection ended Jan 14, 2026 · FEI 3010680515 · Observation 6
Certificates of testing of containers and closures are of the supplier's test results through appropriate validation Specifically, your firm accepts supplier certificates demonstrating the reliability of the supplier's testing. Although the not performed periodic verification of supplier results thro agreement with the stopper supplier, and has not conducted confidence in the supplier's analytical data. As are supplier test results for these …
Contains “supplier*”Read the Form 483 →
Smithfield Biosciencs, Inc.Inspection ended Dec 23, 2025 · FEI 1527761 · Observation 2
Supplier qualification do not include evaluations that provide adequate evidence that the manufacturer can consistently provide material meeting specifications Specifically, A.) Your manufacturer of (b) (4) has stated that "This product is unsuitable for use in the production of APIs as (b) (4) do not follow required GMP practices and do not sell pharmaceutical grade products". You proceeded to qualify this manufacturer of (b) (4) without …
Contains “supplier*”Read the Form 483 →
Macau-Union Pharmaceutical LimitedInspection ended Dec 17, 2025 · FEI 3011219786 · Observation 2
The identity of each component of a drug product is not verified by conducting at least one test to verify the identity, using specific identity tests if they exist. Specifically, Your firm does not perform identity testing for components used to manufacture drug products intended for the U.S. market. Components without identity testing performed include but are not limited to: (b)(4) (b)(4) These components are the active pharmaceutical …
Opens with FDA's citation wording for 21 CFR 211.84(d)(1)Read the Form 483 →
Macau-Union Pharmaceutical LimitedInspection ended Dec 17, 2025 · FEI 3011219786 · Observation 3
Each component is not tested for conformity with all appropriate written specifications for purity, strength, and quality. Specifically, Your firm does not test components used to manufacture drug products intended for the U.S. market for conformance to all appropriate specifications for quality. For example, your firm does not test the active pharmaceutical ingredients for (b)(4) (b)(4) to demonstrate conformance to all appropriate …
Opens with FDA's citation wording for 21 CFR 211.84(d)(2)Read the Form 483 →
Wizcure Pharmaa Private LimitedInspection ended Dec 10, 2025 · FEI 3030304532 · Observation 1
… Furthermore, your contract testing laboratory (b)(4) did not perform growth promotion testing after media fill unit testing performed at their laboratory. 5. EM plate data appears erroneous such that periodically clustered data is recorded for EM logs. Specifically, approximately all Grade B - Grade C spaces in your (b)(4) CFUs. Grade C data is as follows for the recoveries in Grade B plates are (b)(4) period of (b)(4) (b)(4). The CFU data …
Contains “contract test*”Read the Form 483 →
Huons Co., LtdInspection ended Nov 21, 2025 · FEI 3007163376 · Observation 1
… The passing results were reported in the final certificate of analysis (COA). b) The media plates that showed alert level colony forming units (CFU) at the completion of incubation on 11/10/2025 were discarded. No worksheet was generated to document the positive bioburden data. On 11/12/2025 when inquired by the FDA investigator, the QC Microbiology Team Lead changed camera time and date to take photos of the plates with microbial growth. She …
Contains “certificate* of analys*”Read the Form 483 →
Eugia Pharma Specialities LimitedInspection ended Nov 14, 2025 · FEI 3009883410 · Observation 7
Establishment of the reliability of the component supplier's report of analyses is deficient in that the test results are not appropriately validated at appropriate intervals. Specifically, your firm relies exclusively on the vendor's Certificate of Analysis (COA) to ensure the sterility of (b) (4) used on filling line (b) (4) for (b) (4) Following an initial analysis conducted in March 2024, your firm has purchased over (b) (4) bottles of …
Opens with FDA's citation wording for 21 CFR 211.84(d)(2)Read the Form 483 →
Lonza Guangzhou Pharmaceutical Ltd.Inspection ended Oct 31, 2025 · FEI 3006532186 · Observation 3
OBSERVATION 3 The quality system that approves contract laboratories to perform testing of [illegible] impurities is deficient. Specifically, The quality unit fails to ensure that the contract laboratory used to perform testing of [illegible] content in API (codename [illegible]) has the capabilities, equipment, and the knowledge to perform the required testing in an accurate, precise, and reliable way.
Contains “contract laborator*”Read the Form 483 →
Farmakeio Outsourcing LLCInspection ended Sep 5, 2025 · FEI 3014982757 · Observation 2
… Your firm's management reported they are aware of the unknown peaks and at the time of this inspection is working with a contract testing laboratory to identify and determine the pharmacological activity of the foreign peaks for the Testosterone/Triamcinolone Acetonide pellets. Your firm released and distributed this lot and continued producing and distributing more lots of this pellet product without identifying the unknowns. B. …
Contains “contract test*”Read the Form 483 →
Aptuit LtdInspection ended Aug 22, 2025 · FEI 3003444315 · Observation 3
Procedures designed to prevent microbiological contamination of products purporting to be non-sterile are not established. Specifically, a. The quality unit fails to ensure that the contract laboratory that performs the testing for the [illegible] system used for the manufacture of APIs (i.e. [illegible] and excipient, [illegible]) has adequate and validated test methods. b. The quality control unit failed to ensure that the [illegible] system …
Contains “contract laborator*”Read the Form 483 →
Aptuit LtdInspection ended Aug 22, 2025 · FEI 3003444315 · Observation 4
The quality system to approve contract laboratories to perform microbiological testing of APIs, and excipients is deficient. Specifically, 1. The quality unit fails to ensure that the approved contract laboratory used to perform testing of APIs and excipient has the capabilities, equipment, test methods, knowledge, and reliability needed to perform the testing in an accurate and reliable way. 2. On December 11, 2024, the quality unit learned that …
Contains “contract laborator*”Read the Form 483 →
Aptuit LtdInspection ended Aug 22, 2025 · FEI 3003444315 · Observation 6
The identity of each raw material of drug substance is not verified by conducting at least one test to verify the identity, using specific identity test. Specifically, The quality control unit fails to establish a specific identification test method for a raw material used for the manufacture of API, or [illegible].
Contains “identity test*”Read the Form 483 →
Aptuit LtdInspection ended Aug 22, 2025 · FEI 3003444315 · Observation 7
The amount of material taken from each container is not based upon appropriate criteria. Specifically, The quality unit fails to ensure that the amount of raw material taken for testing is based upon appropriate criteria such as criticality of the material, material variability, past history of the supplier, and the quantity needed for analysis.
Opens with FDA's citation wording for 21 CFR 211.84(b)Read the Form 483 →
21 CFR sections these observations open with
The sections FDAgent ties the most of these observations to by FDA's wording, with how many of this theme's observations each covers. Each links to its page of FDA citations, warning letters and the regulation text.
- 21 CFR 211.84Testing and approval or rejection of components, drug product containers, and closuresIn this theme's rule15
- 21 CFR 211.22Responsibilities of quality control unit8
- 21 CFR 211.192Production record review6
- 21 CFR 211.160General requirements5
- 21 CFR 211.100Written procedures; deviations4
- 21 CFR 211.113Control of microbiological contamination3
- 21 CFR 211.166Stability testing3
- 21 CFR 211.194Laboratory records3
- 21 CFR 820.75Process validation2
- 21 CFR 1.512What FSVP may I have if I am a very small importer or I am importing certain food from certain small foreign suppliers?1
Ask FDAgent about supplier controls in Form 483s
FDAgent reads FDA's Form 483s, inspections, citations and warning letters and answers in plain English, with links to the records.
Related
About this data
- The observations are the text FDAgent read by OCR from Form 483 PDFs in FDA's Electronic Reading Room, shown only from 483s whose every printed observation passed FDAgent's checks against the scanned text, as on each 483's own page.
- Which theme an observation belongs to is decided by the fixed rules above; every count on this page is computed from them.
- A 483 addressed to an individual is listed without the name, and an observation that names a person is left out.
- FDAgent is independent of FDA.