Laboratory controls in FDA Form 483 observations
Quality control laboratories: test methods and their validation, laboratory records and equipment, reference standards and chromatography. FDAgent finds 304 such observations in 172 Form 483s issued to 163 facilities, for inspections that ended between October 2008 and September 2026. Each is quoted below from the 483 FDA posted.
A Form 483 lists the investigators' observations at the end of an inspection. It is not FDA's final determination.
Source: Form 483 PDFs in FDA's Electronic Reading Room, last synced Nov 23, 2026, 04:33 UTC. Every quote is from a 483 whose reading passed FDAgent's checks against the scan. FDAgent is independent of FDA.
- Observations
- 304
- Each quoted below
- Form 483s
- 172
- Posted by FDA
- Facilities
- 163
- Distinct FEIs
- Most recent
- Sep 3, 2026
- Inspection end date
Observations by year
These observations by the year the inspection ended. They follow which 483s FDA has posted and FDAgent has read, so they show what is on record, not a trend in FDA's findings. 2026 runs to the latest sync (lighter bar).
Peak: 76 in 2026
| Year | Observations |
|---|---|
| 2008 | 1 |
| 2009 | 0 |
| 2010 | 0 |
| 2011 | 0 |
| 2012 | 8 |
| 2013 | 0 |
| 2014 | 5 |
| 2015 | 4 |
| 2016 | 13 |
| 2017 | 21 |
| 2018 | 23 |
| 2019 | 14 |
| 2020 | 0 |
| 2021 | 6 |
| 2022 | 18 |
| 2023 | 42 |
| 2024 | 25 |
| 2025 | 48 |
| 2026 | 76 |
How FDAgent sorts observations into this theme
Two fixed rules, applied by code to every observation FDAgent can quote. An observation counts when either holds. No model and no similarity score decides.
It is tied to one of these 21 CFR entries
21 CFR 211.16021 CFR 211.16521 CFR 211.16621 CFR 211.16721 CFR 211.17021 CFR 211.17321 CFR 211.176
The same entries the laboratory controls topic counts. A 483 prints no regulation, so FDAgent ties an observation to a section only when it opens with FDA's own citation wording for that section, word for word.
Or its text contains one of these phrases
“laboratory control*” · “laboratory record*” · “laboratory investigation*” · “laboratory equipment” · “laboratory instrument*” · “laboratory data” · “laboratory test*” · “laboratory result*” · “laboratory analyst*” · “laboratory procedure*” · “laboratory notebook*” · “laboratory error*” · “QC laborator*” · “quality control laborator*” · “test method*” · “analytical method*” · “analytical procedure*” · “method validation” · “HPLC*” · “chromatograph*” · “reference standard*” · “system suitability” · “method transfer”
* matches any ending and … up to three words in between. Case is ignored unless a phrase has capitals.
The 304 observations, newest first
Quoted word for word from each Form 483, with only line breaks joined; … marks where the quote is cut. The highlight shows why the observation is in this theme. Each links to FDAgent's page for the 483.
PQ Pharmacy LLCInspection ended Jun 1, 2026 · FEI 3017374013 · Observation 7
Laboratory controls do not include the establishment of scientifically sound and appropriate specifications and test procedures designed to assure that drug product containers conform to appropriate standards of identity, strength, quality and purity. Specifically, A) Your finished product testing used to release finished drug products for commercial distribution and perform stability testing does not include an evaluation of impurities and …
Contains “laboratory control*”Read the Form 483 →
Nippon Kayaku Co., LtdInspection ended May 29, 2026 · FEI 3002807722 · Observation 1
… Your firm does not review 100% of the audit trails and electronic data in your Lab Solutions chromatography software. Your firm reviews the audit trails related to batch release testing, but your firm does not review all of the data generated between batch release testing. You are unable to confirm the presence or the absence of trial injections or other illegitimate testing due to the lack of 100% review. Your firm uses Lab Solutions for …
Contains “chromatograph*”Read the Form 483 →
Shilpa Medicare LimitedInspection ended May 29, 2026 · FEI 3018694237 · Observation 2
… On 10/22/2025, you recorded a calibration failure in test parameter wavelength accuracy during the calibration of Agilent HPLC ID #QC-E080 where peak was not eluted in the standard (b) (4) solution. You notified your service provider on 10/28/2025 and visited your firm on 12/26/2025 for repair. You attributed the root cause to deterioration of DAD lamp and source of lens after replacing the DAD lamp and recalibrated the equipment. You then stated …
Contains “HPLC*”Read the Form 483 →
PETNET Solutions Inc.Inspection ended May 22, 2026 · FEI 3036193770 · Observation 1
… to introducing the precursor (b) (4) into the manufacture of (b) (4) process validation (PV) batches of (b) (4) Injection (Batch #s (b) (4) In addition, the HPLC identity test performed to receive the precursor raw material at central facility is not specific to confirm the identity of precursor that yields as an active ingredient. B. Your firm accepted transfer tubing — a material used during product transfer in the manufacture of (b) (4) …
Contains “HPLC*”Read the Form 483 →
Shanghai Henlius Biopharmaceutical Co., Ltd.Inspection ended May 20, 2026 · FEI 3016529861 · Observation 6
Your quality control laboratory procedures and controls for test and sample management are inadequate or not followed. Specifically, a. Permission assigned to analysts in Empower chromatography software includes the privilege to view quantitation peak fields in review. This allows analysts to view area accounts and results at the time of integration. QC analysts were observed changing the integration parameters without saving interim results. …
Contains “quality control laborator*”Read the Form 483 →
Shanghai Henlius Biopharmaceutical Co., Ltd.Inspection ended May 20, 2026 · FEI 3016529861 · Observation 7
Your firm failed to establish adequate controls to prevent unauthorized deletion of electronic laboratory data during data archiving process. Specifically, the deletion protection setting for (b)(4) analyzer (b)(4) was removed when IT was archiving electronic data from (b)(4) to (b)(4). During audit on May 12, 2026, an operator demonstrated unauthorized deletion of data files (i.e., (b)(4)) from (b)(4) analyzer (b)(4) during data archiving.
Contains “laboratory data”Read the Form 483 →
LGM Pharma Solutions, LLCInspection ended May 18, 2026 · FEI 3017140477 · Observation 2
Laboratory controls do not include the establishment of scientifically sound and appropriate specifications and test procedures designed to assure that components, in-process materials and drug products conform to appropriate standards of identity, strength, quality and purity. Specifically, 1. Your specifications for assay of (b) (4) tablet, a (b) (4) drug, does not appear to be sufficiently (b) (4) as the specifications are set at (b) (4) 2. …
Opens with FDA's citation wording for 21 CFR 211.160(b)Read the Form 483 →
LGM Pharma Solutions, LLCInspection ended May 18, 2026 · FEI 3017140477 · Observation 4
Reserve samples from representative sample lots or batches of drug products selected by acceptable statistical procedures are not examined visually at least once a year for evidence of deterioration. Specifically, Your firm's annual product review process includes visual examination of a subset lots of reserve samples during the first year, according to SOP titled "Annual Product Review" (DOC-001071, Rev 10, Effective Date 14 March 2023). Your …
Opens with FDA's citation wording for 21 CFR 211.170(b)Read the Form 483 →
Fresenius Kabi USA, LLCInspection ended May 14, 2026 · FEI 3001833549 · Observation 3
Verification of the suitability of the testing methods is deficient in that they are not performed under actual conditions of use. Specifically, your firm does not perform comparative testing for all in-house analytical method transfer studies that compare analytical results between the transferring lab and your receiving lab to determine a successful method transfer. For example, you did not compare the results of the transferring lab with the …
Contains “analytical method*”Read the Form 483 →
USV Private LimitedInspection ended Apr 17, 2026 · FEI 3004086192 · Observation 4
… Absence of Training Materials: No defect kit or reference standards have been established to train operators on identifying unacceptable product attributes during visual inspection. 3. No Written Procedure: There is no documented procedure describing how to properly prepare or transfer the (b) (4) solution from the (b) (4) container to the test tube used for visual inspection, creating risk of contamination or sample integrity issues. (b) (4) All …
Contains “reference standard*”Read the Form 483 →
USV Private LimitedInspection ended Apr 17, 2026 · FEI 3004086192 · Observation 9
Laboratory records do not include complete data derived from all tests, examinations, and assay necessary to assure compliance with established specifications and standards. Specifically, Your analysts have been given permission within the Empower software to view numerical results in the review window, allowing them to see area counts of peaks while changing the processing methods, without requiring the chromatograms to be saved. Using this …
Contains “laboratory record*”Read the Form 483 →
Novo NordiskInspection ended Apr 17, 2026 · FEI 3003037054 · Observation 1
Not all method validation activities were documented in source records from study. For instance, stability of (b)(4) in whole blood was evaluated after 1-hour, 2-hour, and 4-hour incubation periods at room temperature. However, source records have no time stamps to support the incubation time points. A similar finding was observed in the shipping and short-term stability experiments.
Contains “method validation”Read the Form 483 →
Novo NordiskInspection ended Apr 17, 2026 · FEI 3003037054 · Observation 2
Existing standard operating procedures (SOPs) were not always followed during the conduct of method validation experiments. For instance, 2a. SOP Q053698, Section 9.8.2, states that only failed stability QCs should be included in subsequent confirmatory runs. Yet, confirmatory run 8 included a new full set of stability QCs instead of including only failing stability QCs from run 7. 2b. SOP Q053698, Section 5.1, states that an average percent …
Contains “method validation”Read the Form 483 →
Cliantha Research LimitedInspection ended Apr 17, 2026 · FEI 3017682080 · Observation 5
5. Your Quality unit failed to perform thorough and/or adequate laboratory investigations (identified by firm as event or incident) as required by the control procedure 100AB014, Version 4, Laboratory Event Management, Effective date: 23Jun2025; to ensure the appropriate root cause, propose adequate corrective / preventive actions, scientific justification to invalidate original result and management of events for any bioanalytical phase of …
Contains “laboratory investigation*”Read the Form 483 →
Jiangsu Hengrui PharmaceuticalsInspection ended Apr 15, 2026 · FEI 3006689263 · Observation 8
The TLC test method for determining (b)(4) degradation impurities used during stability has not been proven to be stability indicating.
Contains “test method*”Read the Form 483 →
ChemPacific CorporationInspection ended Apr 7, 2026 · FEI 3003299802 · Observation 1
The firm failed to maintain adequate quality control systems, including data integrity controls and procedural oversight, Specifically, a. Your firm failed to ensure that laboratory records include complete data derived from all tests conducted on GMP materials and failed to maintain adequate documentation of laboratory testing activities. Review of HPLC sample injection testing data from instruments (b)(4) and (b)(4) (software: (b)(4)) revealed …
Contains “laboratory record*”Read the Form 483 →
Dendreon Pharmaceuticals LLCInspection ended Mar 25, 2026 · FEI 3008821247 · Observation 1
… f) Laboratory Investigation LIR-2023-0144 dated Nov 17, 2023, examined a Day (b) (4) BacT alarm for an (b) (4) at (b) (4) for batch (b) (4) final product. Preliminary (b) (4) (b) (4). The investigation used a (b) (4) BacT (b) (4) that had been (b) (4) for (b) (4), violating work instruction WI-000496.1, entitled "Work Instruction for the Completion of Sterility Test Laboratory Investigations" effective January 31, 2023, which requires (b) (4) per …
Contains “laboratory investigation*”Read the Form 483 →
Dendreon Pharmaceuticals LLCInspection ended Mar 25, 2026 · FEI 3008821247 · Observation 2
Laboratory records do not include the initials or signature of a second person showing that the original records have been reviewed for accuracy. Your firm's test method, (b) (4) entitled (b) (4) effective June 29, 2022, for Sipuleucel-T release lacks adequate independent verification of microbial contamination results. The procedure requires (b) (4) (b) (4) entitled "(b) (4)" effective March 10, 2026, no microscopic examination of (b) (4) by QC …
Contains “laboratory record*”Read the Form 483 →
Cerovene Inc.Inspection ended Mar 23, 2026 · FEI 3005890568 · Observation 3
… testing of commercial drug products, in this cabinet: Product Type of Preparation | Expiry Date | Expiry after according to Brgparation (b) (4) Tmpuity Ju 25, Method Validation Tablets (b) (4), — od MVR (6) (4) Related (b) (4) § Substances 2018 95 Stock standard | January 30, (®)@) for assay 2026 Method Validation Tablets (©) @) Working standard for assay January 30, 2018 2026 10) (EE _Assay_ Stock standard January 16, dissolution 2026 (b) 4) …
Contains “method validation”Read the Form 483 →
SGS North America Inc.Inspection ended Mar 20, 2026 · FEI 1418028 · Observation 1
Laboratory controls do not include the establishment of scientifically sound and appropriate specifications, standards, sampling plans and test procedures designed to assure that components and drug products conform to appropriate standards of identity, strength, quality and purity. Specifically, Your firm does not perform minimum verification testing to qualify the reference standards you use for testing of various drug products and raw …
Opens with FDA's citation wording for 21 CFR 211.160(b)Read the Form 483 →
21 CFR sections these observations open with
The sections FDAgent ties the most of these observations to by FDA's wording, with how many of this theme's observations each covers. Each links to its page of FDA citations, warning letters and the regulation text.
- 21 CFR 211.160General requirementsIn this theme's rule40
- 21 CFR 211.165Testing and release for distributionIn this theme's rule24
- 21 CFR 211.166Stability testingIn this theme's rule21
- 21 CFR 211.194Laboratory records17
- 21 CFR 211.192Production record review15
- 21 CFR 211.22Responsibilities of quality control unit12
- 21 CFR 211.167Special testing requirementsIn this theme's rule6
- 21 CFR 211.170Reserve samplesIn this theme's rule6
- 21 CFR 211.68Automatic, mechanical, and electronic equipment5
- 21 CFR 211.110Sampling and testing of in-process materials and drug products3
Ask FDAgent about laboratory controls in Form 483s
FDAgent reads FDA's Form 483s, inspections, citations and warning letters and answers in plain English, with links to the records.
Related
- Laboratory controls: FDA's Form 483 citations of the same 21 CFR entries
- Aseptic processing: FDA's Form 483 citations of overlapping 21 CFR entries
- Stability: FDA's Form 483 citations of overlapping 21 CFR entries
- Every Form 483 theme, with its counts
- What a Form 483 is, and the latest 483s
- Every 21 CFR section in FDA's Form 483 citations
About this data
- The observations are the text FDAgent read by OCR from Form 483 PDFs in FDA's Electronic Reading Room, shown only from 483s whose every printed observation passed FDAgent's checks against the scanned text, as on each 483's own page.
- Which theme an observation belongs to is decided by the fixed rules above; every count on this page is computed from them.
- A 483 addressed to an individual is listed without the name, and an observation that names a person is left out.
- FDAgent is independent of FDA.