Skip to content

Laboratory controls in FDA Form 483 observations

Quality control laboratories: test methods and their validation, laboratory records and equipment, reference standards and chromatography. FDAgent finds 304 such observations in 172 Form 483s issued to 163 facilities, for inspections that ended between October 2008 and September 2026. Each is quoted below from the 483 FDA posted.

A Form 483 lists the investigators' observations at the end of an inspection. It is not FDA's final determination.

Source: Form 483 PDFs in FDA's Electronic Reading Room, last synced Nov 23, 2026, 04:33 UTC. Every quote is from a 483 whose reading passed FDAgent's checks against the scan. FDAgent is independent of FDA.

Observations
304
Each quoted below
Form 483s
172
Posted by FDA
Facilities
163
Distinct FEIs
Most recent
Sep 3, 2026
Inspection end date

Observations by year

These observations by the year the inspection ended. They follow which 483s FDA has posted and FDAgent has read, so they show what is on record, not a trend in FDA's findings. 2026 runs to the latest sync (lighter bar).

Peak: 76 in 2026

Laboratory controls observations in Form 483s by inspection year
YearObservations
20081
20090
20100
20110
20128
20130
20145
20154
201613
201721
201823
201914
20200
20216
202218
202342
202425
202548
202676

How FDAgent sorts observations into this theme

Two fixed rules, applied by code to every observation FDAgent can quote. An observation counts when either holds. No model and no similarity score decides.

It is tied to one of these 21 CFR entries

21 CFR 211.16021 CFR 211.16521 CFR 211.16621 CFR 211.16721 CFR 211.17021 CFR 211.17321 CFR 211.176

The same entries the laboratory controls topic counts. A 483 prints no regulation, so FDAgent ties an observation to a section only when it opens with FDA's own citation wording for that section, word for word.

Or its text contains one of these phrases

“laboratory control*” · “laboratory record*” · “laboratory investigation*” · “laboratory equipment” · “laboratory instrument*” · “laboratory data” · “laboratory test*” · “laboratory result*” · “laboratory analyst*” · “laboratory procedure*” · “laboratory notebook*” · “laboratory error*” · “QC laborator*” · “quality control laborator*” · “test method*” · “analytical method*” · “analytical procedure*” · “method validation” · “HPLC*” · “chromatograph*” · “reference standard*” · “system suitability” · “method transfer”

* matches any ending and … up to three words in between. Case is ignored unless a phrase has capitals.

The 304 observations, newest first

Quoted word for word from each Form 483, with only line breaks joined; … marks where the quote is cut. The highlight shows why the observation is in this theme. Each links to FDAgent's page for the 483.

  1. An individual (name left out)Inspection ended Mar 13, 2026 · FEI 3042366712 · Observation 1

    … Laboratory results were documented as reviewed by you on November 19, 2024. A separate electronic signature dated November 11, 2024, indicates you reviewed and agreed with all inclusion/exclusion criteria prior to reviewing the laboratory results necessary for eligibility determination. (c) Critical Data Gap: At the time of your November 11, 2024 electronic signature, laboratory results required for eligibility determination—including (B) …

    Contains “laboratory result*”Read the Form 483 →

  2. Hospira, Inc.Inspection ended Mar 10, 2026 · FEI 1925262 · Observation 6

    Observation 6: Your firm failed to establish adequate controls to ensure documentation is complete, accurate, appropriately reviewed and traceable. Specifically, On 6 March 2026, a review of laboratory notebooks managed in the chemistry laboratory revealed deficiencies in the issuance, reconciliation and archiving process. According to SOP 95718, "Chemistry Laboratory Operation Procedure,” v71, Effective date 16 November 2025, ag, audit is …

    Contains “laboratory notebook*”Read the Form 483 →

  3. Aurorium LLCInspection ended Mar 6, 2026 · FEI 1628757 · Observation 3

    … batch record) and laboratory testing. Examples of these forms include: + All forms and documents that are part of the batch record « Form 239A Reagent/Solution Preparation Log * Form 120A Label Control Form « Form 121A Label Issuance + Form 112B Pre-Shipment Review Log + Form 133D Total Unacceptable (b) (4) Count c) There is a failure to review and approve all quality-related documents. The analytical testing lab books (data collection form) used …

    Contains “laboratory test*”Read the Form 483 →

  4. Aurorium LLCInspection ended Mar 6, 2026 · FEI 1628757 · Observation 5

    … The laboratory results simply state (b) (4), "Lab" and "Production."

    Contains “laboratory result*”Read the Form 483 →

  5. Accura-Pac, Inc.Inspection ended Mar 5, 2026 · FEI 1818911 · Observation 1

    … (b) (4) is the software system installed in HPLC/GC for the testing of raw materials and finished drug products. During the walkthrough of the laboratory on 3/4/2026, after the QC Laboratory Supervisor logged into (b) (4), the QC Laboratory Supervisor was able to change the date and time on the computer. b. The written procedures do not clearly delineate data backup responsibilities. Two procedures address data backup but contain conflicting …

    Contains “HPLC*”Read the Form 483 →

  6. Unilab Chemicals and Pharmaceuticals Private Ltd.Inspection ended Feb 27, 2026 · FEI 3004985767 · Observation 1

    … In addition, the non-USP in-house test method used for Gas Chromatography (GC) purity testing has not been validated. According to the QA Manager, the test parameters for the in-house methods were solely determined based on the firm's laboratory capability. C) The firm's designated contract testing laboratory for (b) (4) water TOC testing has not been qualified. In addition, there is no quality agreement that defines GMP expectations and …

    Contains “test method*”Read the Form 483 →

  7. Unilab Chemicals and Pharmaceuticals Private Ltd.Inspection ended Feb 27, 2026 · FEI 3004985767 · Observation 4

    … There is also a lack GMP and SOP training for the Production Worker responsible for routinely cleaning microbiology and QC laboratory equipment.

    Contains “QC laborator*”Read the Form 483 →

  8. Eugia Pharma Specialities LimitedInspection ended Feb 27, 2026 · FEI 3011960448 · Observation 3

    OBSERVATION 3 Laboratory controls do not include the establishment of scientifically sound and appropriate test procedures designed to assure that drug products conform to appropriate standards of identity, strength, quality and purity. Specifically, Employees engaged in visual inspection of sterile finished pharmaceutical products work under (b)(14); however, the qualification of visual inspection the first time the employee is qualified is not …

    Opens with FDA's citation wording for 21 CFR 211.160(b)Read the Form 483 →

  9. RayzeBio, IncInspection ended Feb 24, 2026 · FEI 3030530655 · Observation 7

    Routine calibration of automatic and electronic equipment is not performed according to a written program designed to assure proper performance. Specifically, On 02/18/2026, I performed a review of your (b)(4) system used for the workflows and activities associated with preventive maintenance (PM) and calibration of equipment in the production and QC laboratory. The review revealed a total of (b)(4) pipettes ((b)(4) pipettes owned by Production …

    Contains “QC laborator*”Read the Form 483 →

  10. RayzeBio, IncInspection ended Feb 24, 2026 · FEI 3030530655 · Observation 9

    Laboratory records do not include complete records of any testing and standardization of laboratory reference standards. Specifically, Your firm did not perform reference standard characterizations of the (b)(4) reference standard used in the final release testing of (b)(4). Your firm uses the reference standard (b)(4) ((b)(4)) lot# (b)(4) to conduct the radiochemical purity, identity, and peptide concentration determination by HPLC of (b)(4).

    Contains “laboratory record*”Read the Form 483 →

  11. Orean Personal Care LtdInspection ended Feb 20, 2026 · FEI 3010999645 · Observation 3

    Laboratory controls do not include the establishment of scientifically sound and appropriate specifications, standards, sampling plans, test procedures designed to assure that components of drug product, in-process materials, and drug products conform to appropriate standards of identity, strength, quality and purity. Specifically, A. Your firm failed to monitor the (b)(4) system to ensure that it consistently produces (b)(4) that, at a minimum, …

    Contains “laboratory control*”Read the Form 483 →

  12. Almon Healthcare Private LimitedInspection ended Feb 13, 2026 · FEI 3030509366 · Observation 2

    … The stability indicating aspects of in-house methods (b)(4) used for USP (b)(4) and the compendial method for (b)(4) were not evaluated through method validation and forced degradation studies to ensure no chromatographic co-elution of impurity peaks with the active peak. The following batches were on stability studies: Batch No. Manufacturing date Stability study start date (b)(4) [illegible].

    Contains “method validation”Read the Form 483 →

  13. Almon Healthcare Private LimitedInspection ended Feb 13, 2026 · FEI 3030509366 · Observation 3

    Failure to ensure that all specifications are scientifically sound and appropriate to ensure that your raw materials and API conform to established standards of quality and purity. a. Your firm failed to conduct method validation for the in-house method and verification for compendial methods for the following KSM and APIs. Test Description STP number KSM (b)(4) [illegible]. b. Analytical method verification carried out by your firm for (b)(4) …

    Contains “method validation”Read the Form 483 →

  14. Almon Healthcare Private LimitedInspection ended Feb 13, 2026 · FEI 3030509366 · Observation 4

    Your firm failed to exercise sufficient controls over computerized systems to prevent unauthorized access or changes to data and failure to have adequate controls to prevent omission of data. a. HPLC and GC data from instruments GC HS QC/001, HPLC QC/002, QC/024 and QC/0025 covering January 2024 to June 2025 saved on standalone computer were backed up only once in June 2025 to an external hard drive without assurance of correct data was saved and …

    Contains “HPLC*”Read the Form 483 →

  15. Biomylz Private LimitedInspection ended Feb 9, 2026 · FEI 3013697420 · Observation 1

    … Plant Manager due to repeated and ongoing data integrity violations, including, but not limited to, destruction of batch production records, destruction of laboratory-controlled records, and documentation of false production and laboratory data. According to your Quality Head, employees working in the production and laboratory areas were instructed by the former Plant Manager to destroy, alter, or manipulate records and felt compelled to comply …

    Contains “laboratory control*”Read the Form 483 →

  16. Biomylz Private LimitedInspection ended Feb 9, 2026 · FEI 3013697420 · Observation 3

    Laboratory records do not include complete data derived from all tests, examinations and assay necessary to assure compliance with established specifications and standards. Your firm failed to maintain complete laboratory records, including raw data, for laboratory analyses conducted on commercial drug products. Specifically, raw data supporting laboratory test results used to release commercial batches, to validate the commercial manufacturing …

    Contains “laboratory record*”Read the Form 483 →

  17. Biomylz Private LimitedInspection ended Feb 9, 2026 · FEI 3013697420 · Observation 4

    The responsibilities and procedures applicable to the quality control unit are not fully followed. Specifically, A. There is no assurance that laboratory testing used to determine raw material suitability and finished drug product quality is performed in accordance with applicable cGMP requirements. For example, your firm uses contracted laboratories that are not registered or inspected by the agency to perform incoming raw material and finished …

    Contains “laboratory test*”Read the Form 483 →

  18. Biomylz Private LimitedInspection ended Feb 9, 2026 · FEI 3013697420 · Observation 7

    Reserve samples from representative sample lots or batches of drug products selected by acceptable statistical procedures are not examined visually at least once a year for evidence of deterioration. Specifically, according to your procedure titled "Procedure for Collection and Storage of Control Sample", SOP Number: QCD/009, Version: R04, Effective Date: April 07, 2025, Section 5.1.1 states, "The control sample shall be collected from each batch …

    Opens with FDA's citation wording for 21 CFR 211.170(b)Read the Form 483 →

  19. SNJ Labs Private Ltd.Inspection ended Feb 6, 2026 · FEI 3026587575 · Observation 1

    … Your firm did not thoroughly investigate the OOS (OOS/24/005) results for (b) (4) assay by HPLC in (b) (4) (b) (4) API intended for the US market. The assay results were (b) (4) % (first replicate) and (b) (4) % (second replicate); specification limit: (b) (4) % to (b) (4) % w/w. The identified root cause was due to a (b) (4) failure at (b) (4) in the HPLC system; however, the chromatographic data acquisition was documented at (b) (4) The …

    Contains “HPLC*”Read the Form 483 →

  20. SNJ Labs Private Ltd.Inspection ended Feb 6, 2026 · FEI 3026587575 · Observation 2

    Your firm failed to use validated and stability-indicating test procedures. Your firm failed to use validated, stability-indicating methods to test active pharmaceutical ingredients (API) (b) (4) (b) (4) intended for the US market when assigning retest dates of (b) (4) The analytical method used for testing stability samples is not stability indicating and cannot identify degradation products. Furthermore, your firm started stress studies for (b) …

    Contains “analytical method*”Read the Form 483 →

21 CFR sections these observations open with

The sections FDAgent ties the most of these observations to by FDA's wording, with how many of this theme's observations each covers. Each links to its page of FDA citations, warning letters and the regulation text.

Ask FDAgent about laboratory controls in Form 483s

FDAgent reads FDA's Form 483s, inspections, citations and warning letters and answers in plain English, with links to the records.

Edit it or ask as written. 3 questions free, no card needed.

About this data

  • The observations are the text FDAgent read by OCR from Form 483 PDFs in FDA's Electronic Reading Room, shown only from 483s whose every printed observation passed FDAgent's checks against the scanned text, as on each 483's own page.
  • Which theme an observation belongs to is decided by the fixed rules above; every count on this page is computed from them.
  • A 483 addressed to an individual is listed without the name, and an observation that names a person is left out.
  • FDAgent is independent of FDA.