Laboratory controls in FDA Form 483 observations

Quality control laboratories: test methods and their validation, laboratory records and equipment, reference standards and chromatography. FDAgent finds 304 such observations in 172 Form 483s issued to 163 facilities, for inspections that ended between October 2008 and September 2026. Each is quoted below from the 483 FDA posted.

A Form 483 lists the investigators' observations at the end of an inspection. It is not FDA's final determination.

Source: Form 483 PDFs in FDA's Electronic Reading Room, last synced Nov 23, 2026, 04:33 UTC. Every quote is from a 483 whose reading passed FDAgent's checks against the scan. FDAgent is independent of FDA.

Observations
304
Each quoted below
Form 483s
172
Posted by FDA
Facilities
163
Distinct FEIs
Most recent
Sep 3, 2026
Inspection end date

Observations by year

These observations by the year the inspection ended. They follow which 483s FDA has posted and FDAgent has read, so they show what is on record, not a trend in FDA's findings. 2026 runs to the latest sync (lighter bar).

Peak: 76 in 2026

Laboratory controls observations in Form 483s by inspection year
YearObservations
20081
20090
20100
20110
20128
20130
20145
20154
201613
201721
201823
201914
20200
20216
202218
202342
202425
202548
202676

How FDAgent sorts observations into this theme

Two fixed rules, applied by code to every observation FDAgent can quote. An observation counts when either holds. No model and no similarity score decides.

It is tied to one of these 21 CFR entries

21 CFR 211.16021 CFR 211.16521 CFR 211.16621 CFR 211.16721 CFR 211.17021 CFR 211.17321 CFR 211.176

The same entries the laboratory controls topic counts. A 483 prints no regulation, so FDAgent ties an observation to a section only when it opens with FDA's own citation wording for that section, word for word.

Or its text contains one of these phrases

“laboratory control*” · “laboratory record*” · “laboratory investigation*” · “laboratory equipment” · “laboratory instrument*” · “laboratory data” · “laboratory test*” · “laboratory result*” · “laboratory analyst*” · “laboratory procedure*” · “laboratory notebook*” · “laboratory error*” · “QC laborator*” · “quality control laborator*” · “test method*” · “analytical method*” · “analytical procedure*” · “method validation” · “HPLC*” · “chromatograph*” · “reference standard*” · “system suitability” · “method transfer”

* matches any ending and … up to three words in between. Case is ignored unless a phrase has capitals.

The 304 observations, newest first

Quoted word for word from each Form 483, with only line breaks joined; … marks where the quote is cut. The highlight shows why the observation is in this theme. Each links to FDAgent's page for the 483.

  1. Macau-Union Pharmaceutical LimitedInspection ended Dec 17, 2025 · FEI 3011219786 · Observation 6

    Written procedures are not established for the cleaning and maintenance of equipment, including utensils, used in the manufacture, processing, packing or holding of a drug product. Specifically, Your firm has not validated your cleaning procedures and UV-Vis test method to demonstrate that cleaning of shared equipment is effective and reproducible at removing drug product residue and other contaminants that could adversely affect the quality of …

    Contains “test method*”Read the Form 483 →

  2. Macau-Union Pharmaceutical LimitedInspection ended Dec 17, 2025 · FEI 3011219786 · Observation 7

    Reserve samples from representative sample lots or batches of drug products selected by acceptable statistical procedures are not examined visually at least once a year for evidence of deterioration. Specifically, Your firm does not have procedures for performing an annual visual examination of reserve samples of over-the-counter human finished drug products, including drug products intended for the U.S. market, (b)(4) such as: (b)(4) Your firm …

    Opens with FDA's citation wording for 21 CFR 211.170(b)Read the Form 483 →

  3. Dr. Reddy’s Laboratories LimitedInspection ended Dec 12, 2025 · FEI 3009193040 · Observation 1

    OBSERVATION 1 There is a failure to thoroughly review any unexplained discrepancy and the failure of a batch or any of its components to meet any of its specifications whether or not the batch has been already distributed. Specifically, (b)(4)(b)(4) A. On 4/28/2025, you recorded non-conformances 310028189 and 310028200 for (b)(4)(b)(4)(b)(4) for Dissolution by HPLC. The OOS results were recorded (b)(4)(b)(4)(b)(4) for Batch No’s. (OOS No. …

    Contains “HPLC*”Read the Form 483 →

  4. Dr. Reddy’s Laboratories LimitedInspection ended Dec 12, 2025 · FEI 3009193040 · Observation 4

    OBSERVATION 4 Testing and release of drug product for distribution do not include appropriate laboratory determination of satisfactory conformance to the final specifications prior to release. Specifically, A. Section 5.10.6 of your procedure, SOP-GLOB-QC-0085 Version 13.0 Good Analytical Practices (b)(4) (Effective date 7/25/2025) requires measurements from each preparation for assay test methods by HPLC. However, you have validated the assay …

    Opens with FDA's citation wording for 21 CFR 211.165(a)Read the Form 483 →

  5. Wizcure Pharmaa Private LimitedInspection ended Dec 10, 2025 · FEI 3030304532 · Observation 1

    Laboratory records do not include complete data derived from all tests, examinations and assay necessary to assure compliance with established specifications and standards Specifically, Microbiologists responsible for performing sterility testing and collecting environmental monitoring, personnel monitoring, and (b)(4) samples confirmed they do not collect, test, and incubate all samples. A. During inspection of the (b)(4) Biological incubators …

    Contains “laboratory record*”Read the Form 483 →

  6. Wizcure Pharmaa Private LimitedInspection ended Dec 10, 2025 · FEI 3030304532 · Observation 7

    … You performed a change to your method validation for your (b)(4) and did not document this revalidation via any change control. B. At the initiation of this inspection on December 3, 2025, we found GMP documents that were disposed inside a bag which was inside a box collected from QC Documentation Room (b)(4). The evaluation of the contents of the bag revealed presence of (b)(4) Analysis sheets from (b)(4) validation data summary sheets, (b)(4) …

    Contains “method validation”Read the Form 483 →

  7. Huons Co., LtdInspection ended Nov 21, 2025 · FEI 3007163376 · Observation 1

    Laboratory records do not include complete data derived from all tests, examinations, and assay necessary to assure compliance with established specifications and standards. Specifically, A. During review of the bioburden and endotoxin tests performed on 11/03/2025 for the (b)(4) Line media fill (b)(4) samples, we found the following. a) The (b)(4) Tank and Tank Transfer Line samples endotoxin out of specification (OOS) results obtained on …

    Contains “laboratory record*”Read the Form 483 →

  8. Huons Co., LtdInspection ended Nov 21, 2025 · FEI 3007163376 · Observation 3

    … Your Quality Control (QC) team failed to thoroughly document laboratory investigations such as, but not limited to Column Failure and Device Malfunction. Additionally, there is no documentation available to demonstrate that your Quality Assurance (QA) team reviewed and evaluated the laboratory investigations at the form utilized to document the investigation, at the batch release review and release stage, and the annual product quality review.

    Contains “laboratory investigation*”Read the Form 483 →

  9. Huons Co., LtdInspection ended Nov 21, 2025 · FEI 3007163376 · Observation 5

    … The audit trail log printout attached to each final product release record is incomplete where only the beginning of the laboratory testing is printed. There are no audit trail log printouts dedicated to stability testing. Additionally, the written procedure for audit trail reviews prior to July 2023 only require random equipment and time length of audit trail logs reviewed. There are a number of batches intended for the U.S. Market and within …

    Contains “laboratory test*”Read the Form 483 →

  10. HEC Pharm Co. Ltd.Inspection ended Nov 21, 2025 · FEI 3006259537 · Observation 1

    Analytical method used for determination of impurities in API is inadequately validated. Specifically, the impurity method used for determination or quantification of impurities in (b)(4) Active Pharmaceutical Ingredient (API) did not detect the presence of an unknown impurity (b)(4) was initiated for assay OOT for (b)(4) in the test sample solution. On 02/20/2025, OOT (b)(4) During the investigation, an unknown peak was detected in (b)(4) batch …

    Contains “analytical method*”Read the Form 483 →

  11. HEC Pharm Co. Ltd.Inspection ended Nov 21, 2025 · FEI 3006259537 · Observation 2

    Laboratory instruments are inadequately qualified for their intended use. Specifically, the Ultra High Performance Liquid Chromatography-Mass Spectrometer (UHPLC-MS) used for determination (b)(4) impurities in (b)(4) API has not been fully qualified for its intended use. Although, the installation, operation and performance qualification of the instrument was performed in July 2025 and the instrument has been used for method validation and …

    Contains “laboratory instrument*”Read the Form 483 →

  12. Eugia Pharma Specialities LimitedInspection ended Nov 14, 2025 · FEI 3009883410 · Observation 6

    Laboratory controls do not include the establishment of scientifically sound and appropriate test procedures designed to assure that drug products conform to appropriate standards of identity, strength, quality and purity. There is no assurance that your laboratory test methods have been validated or demonstrated to be suitable for their intended use to consistently produce accurate and reliable results. For example, repeated Out-of-Specification …

    Opens with FDA's citation wording for 21 CFR 211.160(b)Read the Form 483 →

  13. Eugia Pharma Specialities LimitedInspection ended Nov 14, 2025 · FEI 3009883410 · Observation 8

    … Operators are trained by reading procedures without hands-on experience using (b) (4) Additionally, your firm does not maintain (b) (4) with reference standards to train and qualify operators' ability to (b) (4) This deficiency compromises the consistency and reliability of (b) (4) detection during manufacturing operations. B. Standard Operating Procedure EP2-PFB-SOP-038-02 "In-Process Checks During Batch Operation" fails to establish adequate …

    Contains “reference standard*”Read the Form 483 →

  14. Asahi Kasei Finechem Co., Ltd.Inspection ended Nov 14, 2025 · FEI 3002806404 · Observation 1

    Methods used in testing must meet proper standards of accuracy and reliability as applied to the product tested. Specifically, Your firm has not performed method validation studies on in-house developed test method, nor method verification studies on USP methods used to perform finished product release testing on (b)(4) drug products manufactured for the US market. A) The following are examples of unvalidated in-house analytical methods used for …

    Contains “method validation”Read the Form 483 →

  15. ICON plcInspection ended Oct 24, 2025 · FEI 3038011239 · Observation 1

    The following observation pertaining to method validation activities for (b) (4) and (b) (4) (method validations # JJP130EL-151303-B, JJP1953A-1953AX-A and JJP20016-20016X-B): The site did not evaluate the method's specificity for (b) (4) and (b) (4) in the presence of all co-administered medications during bioanalytical method validation. Specifically, the method's specificity for (b) (4) and (b) (4) was only evaluated in the presence of …

    Contains “method validation”Read the Form 483 →

  16. Inotiv Inc.Inspection ended Oct 24, 2025 · FEI 1000140135 · Observation 2

    2. Whole blood stability of insulin aspart was not assessed during method validation (validation report #1000-2403020). Specifically, short term stability testing was conducted in plasma.

    Contains “method validation”Read the Form 483 →

  17. Inotiv Inc.Inspection ended Oct 24, 2025 · FEI 1000140135 · Observation 3

    3. Source records for method validation experiments of long-term stability (validation report #1000-091480-2), dilution integrity, stability in matrix and ion suppression (validation report #1000-091480-1), do not document that lithium heparinized human lysed whole blood was utilized as the matrix. The study samples from study utilized lithium heparinized human lysed whole blood.

    Contains “method validation”Read the Form 483 →

  18. Farmakeio Outsourcing LLCInspection ended Sep 5, 2025 · FEI 3014982757 · Observation 2

    Laboratory controls do not include the establishment of scientifically sound and appropriate specifications and test procedures designed to assure that drug products conform to appropriate standards of identity, strength, quality and purity. Specifically, A. During a review of your firm's batch record for Product ID#364, Testosterone/Triamcinolone Acetonide Pellet (200mg/40mcg), Lot# 3131, including laboratory test chromatograms, …

    Opens with FDA's citation wording for 21 CFR 211.160(b)Read the Form 483 →

  19. Farmakeio Outsourcing LLCInspection ended Sep 5, 2025 · FEI 3014982757 · Observation 3

    Each batch of controlled-release dosage form drug product is not laboratory tested to determine conformance to the specifications for the rate of release for each active ingredient. Specifically, your firm failed to require or perform dissolution testing on each batch of controlled-release dosage form drug product. The drug product is not laboratory tested to determine conformance to the specifications at a rate of release for each active …

    Opens with FDA's citation wording for 21 CFR 211.167(c)Read the Form 483 →

  20. Aptuit LtdInspection ended Aug 22, 2025 · FEI 3003444315 · Observation 3

    Procedures designed to prevent microbiological contamination of products purporting to be non-sterile are not established. Specifically, a. The quality unit fails to ensure that the contract laboratory that performs the testing for the [illegible] system used for the manufacture of APIs (i.e. [illegible] and excipient, [illegible]) has adequate and validated test methods. b. The quality control unit failed to ensure that the [illegible] system …

    Contains “test method*”Read the Form 483 →

21 CFR sections these observations open with

The sections FDAgent ties the most of these observations to by FDA's wording, with how many of this theme's observations each covers. Each links to its page of FDA citations, warning letters and the regulation text.

Ask FDAgent about laboratory controls in Form 483s

FDAgent reads FDA's Form 483s, inspections, citations and warning letters and answers in plain English, with links to the records.

Edit it or ask as written. 3 questions free, no card needed.

About this data

  • The observations are the text FDAgent read by OCR from Form 483 PDFs in FDA's Electronic Reading Room, shown only from 483s whose every printed observation passed FDAgent's checks against the scanned text, as on each 483's own page.
  • Which theme an observation belongs to is decided by the fixed rules above; every count on this page is computed from them.
  • A 483 addressed to an individual is listed without the name, and an observation that names a person is left out.