Inotiv Inc.
2701 Kent Ave, West Lafayette, Indiana, United States
FEI 1000140135
FDA has 8 inspections on file for Inotiv Inc. in West Lafayette, Indiana (FEI 1000140135), from 2011 to 2025. The latest ended on and was classified VAI, voluntary action indicated.
VAI means FDA investigators found objectionable conditions or practices, but FDA was not prepared to take or recommend regulatory action. The firm is expected to correct them voluntarily.
FDA posted the Form 483 from that inspection. FDAgent read its 3 observations from the PDF; they are listed below. One FOIA request has asked FDA for this inspection's 483.
No warning letter, court action or import refusal is filed under this FEI.
Sources:FDA firm profile ↗Form 483 PDF ↗Synced Nov 23, 2026
Ask FDAgent about this site
Email me when FDA posts a new inspection, Form 483 or warning letter for this site
Weekly, on a free account. Pause it any time.
Comparing suppliers? Check this site beside 2 others, free
- Latest inspection
- VAI
- Inspections on file
- 8, 2011–2025
- 0 OAI · 3 VAI · 5 NAI
- Inspections with 483 observations shown
- 2 of 8
- Form 483s posted by FDA
- ↗
- Requested under FOIA
- 1 request, latest Dec 8, 2025
- Warning letters
- None on file
- Import refusals
- None under this FEI
- Import alerts
- Not linked by FEI
- FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗
Inspection history
8 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC
- VAIObjectionable conditions found; FDA took no regulatory actionDrugs · Bioresearch MonitoringPosted 483 · 3 observations
Read this Form 483 on its own page →Form 483 PDF on fda.gov
Read from the FDA-posted Form 483 PDF by OCR. FDAgent keeps an observation only when it matches the scanned text; check the PDF before quoting. The short titles are FDAgent’s.
- Observation 1
Lipemia not evaluated in serum validation
1. During the validation of insulin aspart in human serum (validation report #1000-2403020), the effect of lipemia for selectivity and matrix effect experiments was not evaluated. Specifically, the firm used plasma matrix to study the lipemia effect for selectivity and matrix effect in serum.
- Observation 2
Whole blood stability not assessed
2. Whole blood stability of insulin aspart was not assessed during method validation (validation report #1000-2403020). Specifically, short term stability testing was conducted in plasma.
Show observation 3
- Observation 3
Source records missing whole blood matrix
3. Source records for method validation experiments of long-term stability (validation report #1000-091480-2), dilution integrity, stability in matrix and ion suppression (validation report #1000-091480-1), do not document that lithium heparinized human lysed whole blood was utilized as the matrix. The study samples from study utilized lithium heparinized human lysed whole blood.
- NAINo objectionable conditions that need follow-upDrugs · Bioresearch MonitoringNone
- NAINo objectionable conditions that need follow-upDrugs · Bioresearch MonitoringNone
- NAINo objectionable conditions that need follow-upDrugs · Drug Quality AssuranceNone
- NAINo objectionable conditions that need follow-upDrugs · Drug Quality AssuranceNone
- VAIObjectionable conditions found; FDA took no regulatory actionDrugs · Drug Quality AssuranceCitations not published
- VAIObjectionable conditions found; FDA took no regulatory actionDrugs · Drug Quality Assurance1 citation published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Scientifically sound laboratory controls
Laboratory controls do not include the establishment of scientifically sound and appropriate test procedures designed to assure that components and drug products conform to appropriate standards of identity, strength, quality and purity.
- NAINo objectionable conditions that need follow-upDrugs · Drug Quality AssuranceNone
Requested under FOIA
FDA FOIA logs, last synced Nov 23, 2026, 04:25 UTC
One request to FDA under the Freedom of Information Act named this site, on Dec 8, 2025. It asked for a Form 483.
Inspection ending Oct 24, 2025VAI
1 request. In the inspection history above. FDA has posted its Form 483.
- Dec 8, 2025FDA-FOIA-2025-13838Form 483· 1 of 5 sitesRecords released
A request is listed when it names this site’s FEI number or one of its FDA inspection IDs, or, after review, its exact street address. Requester names are left out.
Regulations FDA cited here
The 1 section of 21 CFR in FDA’s published citations for 1 of this facility’s 8 inspections.
- 21 CFR 211.160General requirements1 of 1
Observations read from posted 483 PDFs carry no CFR section of FDA’s, so they are not counted here.
Go further than this page
FDAgent researches FDA’s records and links every answer to its source.
About this data. Every record here is filed under FEI 1000140135 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
- Every record on this page is filed under FEI 1000140135. Records filed under another FEI or name, including other sites of the same company, are not shown here.
- FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
- Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
- “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
- Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
- FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.