FDA issued this Form 483 to Inotiv Inc. in West Lafayette, Indiana (FEI 1000140135) at the end of a drug inspection that ran from October 20 to October 24, 2025. It lists 3 observations. FDA classified the inspection Voluntary Action Indicated (VAI).

A Form 483 lists the investigators' observations at the end of an inspection. It is not FDA's final determination.

Source: FDA's Electronic Reading Room, posted Jan 16, 2026. Form 483 PDF on fda.gov. FDAgent is independent of FDA.

Inspection
Oct 20–24, 2025
Drugs · Bioresearch Monitoring
FDA's outcome
VAI
Voluntary Action Indicated
Posted by FDA
Jan 16, 2026
84 days after the inspection
Facility
FEI 1000140135
West Lafayette, Indiana

The 3 observations on this Form 483

Read from the FDA-posted Form 483 PDF by OCR. FDAgent keeps an observation only when it matches the scanned text; check the PDF before quoting. The short titles are FDAgent's.

  1. 1

    Lipemia not evaluated in serum validation

    1. During the validation of insulin aspart in human serum (validation report #1000-2403020), the effect of lipemia for selectivity and matrix effect experiments was not evaluated. Specifically, the firm used plasma matrix to study the lipemia effect for selectivity and matrix effect in serum.

  2. 2

    Whole blood stability not assessed

    2. Whole blood stability of insulin aspart was not assessed during method validation (validation report #1000-2403020). Specifically, short term stability testing was conducted in plasma.

  3. 3

    Source records missing whole blood matrix

    3. Source records for method validation experiments of long-term stability (validation report #1000-091480-2), dilution integrity, stability in matrix and ion suppression (validation report #1000-091480-1), do not document that lithium heparinized human lysed whole blood was utilized as the matrix. The study samples from study utilized lithium heparinized human lysed whole blood.

Ask FDAgent about this Form 483

FDAgent reads the 483, the facility's inspections, warning letters and FDA's citations and answers in plain English, with links to the records.

Edit it or ask as written. 3 questions free, no card needed.

About this data

  • The PDF and its dates come from FDA's Electronic Reading Room, last synced Nov 23, 2026, 04:33 UTC; the inspection, its outcome and FDA's citations from FDA's Data Dashboard.
  • The observations are the text FDAgent read from this PDF by OCR, shown only when every printed observation passed FDAgent’s checks against the scanned text.
  • FDAgent leaves out the names of the people the Form 483 is addressed to and of the investigators.
  • FDAgent is independent of FDA.