Laboratory controls in FDA Form 483 observations
Quality control laboratories: test methods and their validation, laboratory records and equipment, reference standards and chromatography. FDAgent finds 304 such observations in 172 Form 483s issued to 163 facilities, for inspections that ended between October 2008 and September 2026. Each is quoted below from the 483 FDA posted.
A Form 483 lists the investigators' observations at the end of an inspection. It is not FDA's final determination.
Source: Form 483 PDFs in FDA's Electronic Reading Room, last synced Nov 23, 2026, 04:33 UTC. Every quote is from a 483 whose reading passed FDAgent's checks against the scan. FDAgent is independent of FDA.
- Observations
- 304
- Each quoted below
- Form 483s
- 172
- Posted by FDA
- Facilities
- 163
- Distinct FEIs
- Most recent
- Sep 3, 2026
- Inspection end date
Observations by year
These observations by the year the inspection ended. They follow which 483s FDA has posted and FDAgent has read, so they show what is on record, not a trend in FDA's findings. 2026 runs to the latest sync (lighter bar).
Peak: 76 in 2026
| Year | Observations |
|---|---|
| 2008 | 1 |
| 2009 | 0 |
| 2010 | 0 |
| 2011 | 0 |
| 2012 | 8 |
| 2013 | 0 |
| 2014 | 5 |
| 2015 | 4 |
| 2016 | 13 |
| 2017 | 21 |
| 2018 | 23 |
| 2019 | 14 |
| 2020 | 0 |
| 2021 | 6 |
| 2022 | 18 |
| 2023 | 42 |
| 2024 | 25 |
| 2025 | 48 |
| 2026 | 76 |
How FDAgent sorts observations into this theme
Two fixed rules, applied by code to every observation FDAgent can quote. An observation counts when either holds. No model and no similarity score decides.
It is tied to one of these 21 CFR entries
21 CFR 211.16021 CFR 211.16521 CFR 211.16621 CFR 211.16721 CFR 211.17021 CFR 211.17321 CFR 211.176
The same entries the laboratory controls topic counts. A 483 prints no regulation, so FDAgent ties an observation to a section only when it opens with FDA's own citation wording for that section, word for word.
Or its text contains one of these phrases
“laboratory control*” · “laboratory record*” · “laboratory investigation*” · “laboratory equipment” · “laboratory instrument*” · “laboratory data” · “laboratory test*” · “laboratory result*” · “laboratory analyst*” · “laboratory procedure*” · “laboratory notebook*” · “laboratory error*” · “QC laborator*” · “quality control laborator*” · “test method*” · “analytical method*” · “analytical procedure*” · “method validation” · “HPLC*” · “chromatograph*” · “reference standard*” · “system suitability” · “method transfer”
* matches any ending and … up to three words in between. Case is ignored unless a phrase has capitals.
The 304 observations, newest first
Quoted word for word from each Form 483, with only line breaks joined; … marks where the quote is cut. The highlight shows why the observation is in this theme. Each links to FDAgent's page for the 483.
Aptuit LtdInspection ended Aug 22, 2025 · FEI 3003444315 · Observation 4
The quality system to approve contract laboratories to perform microbiological testing of APIs, and excipients is deficient. Specifically, 1. The quality unit fails to ensure that the approved contract laboratory used to perform testing of APIs and excipient has the capabilities, equipment, test methods, knowledge, and reliability needed to perform the testing in an accurate and reliable way. 2. On December 11, 2024, the quality unit learned that …
Contains “test method*”Read the Form 483 →
Aptuit LtdInspection ended Aug 22, 2025 · FEI 3003444315 · Observation 5
The accuracy, sensitivity, specificity, reproducibility, and reliability of microbiological test methods has not been established. Specifically, The quality unit fails to ensure that the microbiological test methods (e.g., TVC, pathogens identification, and endotoxins) utilized for the release of APIs [e.g., [illegible] (intended for drug product), [illegible] (intended for sterile injectable) and excipient [illegible]] are validated.
Contains “test method*”Read the Form 483 →
Aptuit LtdInspection ended Aug 22, 2025 · FEI 3003444315 · Observation 6
The identity of each raw material of drug substance is not verified by conducting at least one test to verify the identity, using specific identity test. Specifically, The quality control unit fails to establish a specific identification test method for a raw material used for the manufacture of API, or [illegible].
Contains “test method*”Read the Form 483 →
Tripak Pharmaceuticals Pty Ltd.Inspection ended Jul 25, 2025 · FEI 3008549263 · Observation 5
There is no written testing program designed to assess the stability characteristics of drug products. Specifically, Your firm failed to establish a written testing program designed to assess the stability characteristics of (b) (4) Your firm currently proposes a (b) (4) expiration period without adequate supporting stability data. The proposed expiration date is based solely on 6-month accelerated condition (ACC) data for bulk, which is …
Opens with FDA's citation wording for 21 CFR 211.166(a)Read the Form 483 →
Tripak Pharmaceuticals Pty Ltd.Inspection ended Jul 25, 2025 · FEI 3008549263 · Observation 7
Laboratory controls shall include the establishment of scientifically sound and appropriate specifications, standards, sampling plans, and test procedures designed to assure that components, drug product containers, drug product closures, in-process materials, labeling, and drug products conform to appropriate standards of identity, strength, quality, and purity. Specifically, From 05/04/2022 to 05/19/2025 your firm released approximately (b) (4) …
Contains “laboratory control*”Read the Form 483 →
Raptim Research Private LimitedInspection ended Jul 25, 2025 · FEI 3007267856 · Observation 4
The firm did not consistently follow good documentation practices in their raw data management during conduct of method validation of (b) (4) (RRL-ANA-MVR1409). For example, in raw data forms (17242, 7215, 0597) several corrections were made to address QC findings. However, the changes were made without documenting when the change was made and who made the change.
Contains “method validation”Read the Form 483 →
Veganic SKN LimitedInspection ended Jul 21, 2025 · FEI 3024609429 · Observation 3
OBSERVATION 3 Laboratory controls shall include the establishment of scientifically sound and appropriate specifications, standards, sampling plans, and test procedures designed to assure that components, drug product containers, drug product closures, in-process materials, labeling, and drug products conform to appropriate standards of identity, strength, quality, and purity. Specifically, From 06/02/2024 to 05/05/2025, your firm released ma …
Contains “laboratory control*”Read the Form 483 →
Medi-Fare Drug Pharmaceutical Compounding LLCInspection ended Jun 6, 2025 · FEI 3009925820 · Observation 7
An adequate number of batches of each drug product are not tested to determine an appropriate expiration date. Specifically, You discontinued your stability program in approximately 2018 per your site leadership personnel. You have continued to produce various drug products to include but not limited to Fentanyl [illegible], lot [illegible] in bags, Succinylcholine [illegible], lot [illegible] m syringes. Succinylcholine is produced in lots of …
Opens with FDA's citation wording for 21 CFR 211.166(b)Read the Form 483 →
AGC Biologics, Inc.Inspection ended Apr 18, 2025 · FEI 3000209996 · Observation 1
Review of laboratory data for drug substance in-process and release testing is not adequate: [illegible] does not match a) We observed raw data from the analytical packet for lot #[illegible] used to document the [illegible] final results. This data is generated as part of Final Bulk Drug Substance release testing for residual [illegible] The analysis packet was reviewed by a QC reviewer on 19 NOV 2024 and did not catch the error. b) On 15 APR …
Contains “laboratory data”Read the Form 483 →
Guangxi Yulin Pharmaceutical Group Co. Ltd.Inspection ended Mar 21, 2025 · FEI 1000589993 · Observation 1
Testing and release of drug product for distribution do not include appropriate laboratory determination of satisfactory conformance to the identity and strength of each active ingredient prior to release. Specifically, a) Your firm does not perform analytical testing to determine the concentrations of (b) (4) active ingredients in the finished OTC drug product. (b) (4) batches of this drug product were distributed to the US in July 2023. b) …
Opens with FDA's citation wording for 21 CFR 211.165(a)Read the Form 483 →
Guangxi Yulin Pharmaceutical Group Co. Ltd.Inspection ended Mar 21, 2025 · FEI 1000589993 · Observation 2
Laboratory controls do not include the establishment of scientifically sound and appropriate specifications, standards, and test procedures designed to assure that components and drug products conform to appropriate standards of identity, strength, quality and purity. Specifically, a) Your firm has not validated or verified all the component intermediate/in-process and finished product analytical test methods performed on (b) (4) OTC drug …
Opens with FDA's citation wording for 21 CFR 211.160(b)Read the Form 483 →
Guangxi Yulin Pharmaceutical Group Co. Ltd.Inspection ended Mar 21, 2025 · FEI 1000589993 · Observation 3
There is no written testing program designed to assess the stability characteristics of drug products. Specifically, (A) You have failed to establish how your stability testing methods are stability indicating. You do not have data to show your stability test methods meet the appropriate identity, strength, and quality attributes to support the (b) (4) shelf life of your OTC (b) (4) products, (b) (4) (B) Microbial content and identification …
Opens with FDA's citation wording for 21 CFR 211.166(a)Read the Form 483 →
Nanjing Cuccess Pharmaceutical Co., LtdInspection ended Mar 21, 2025 · FEI 3027129265 · Observation 3
The written stability testing program is not followed. Specifically, your firm assigned ay shelf-life to the three yy process validation lots manufactured in January 2024. There was no accelerated stability study conducted at 40° Celsius and 949% relative humidity for six months to support the y 4 expiration date. Your firm released py process validation lot bw | expiration date gy ong and distributed the drug on about
Opens with FDA's citation wording for 21 CFR 211.166(a)Read the Form 483 →
2seventy bio, Inc.Inspection ended Mar 20, 2025 · FEI 3013427685 · Observation 1
Laboratory records do not include complete records of any testing and standardization of laboratory reference standards. Specifically, your firm has not adequately qualified the DNA plasmid reference standard (b) (4) used in the test method TM-0059 Determination of Lentiviral Vector Titer by (b) (4) (b) (4) in Quality Control. This test method relates to the viral vector Anti-BCMA02 CAR aLVV used in manufacturing drug product ABECMA. Three …
Contains “laboratory record*”Read the Form 483 →
Hubei JXBio Pharmaceutical Co., Ltd.Inspection ended Mar 14, 2025 · FEI 3016784508 · Observation 1
… discrepancy, explain the process used to determine the root cause of contaminated or document the (b) (4) change control for update of the specification and test method. In addition, the investigation did not (b) (4) extend to other batches that may have been affected by the contaminated D. An investigation for out-of-specification OOS-C005A-24002 was initiated for a related substance impurity (b) (4) (b) (4) OOS for Batch During the …
Contains “test method*”Read the Form 483 →
Lin, James P MD Inc. d.b.a. Reproductive Feritlity CenterInspection ended Mar 12, 2025 · FEI 3005405736 · Observation 1
HCT/P donors were not determined to be eligible based on the results of donor screening and testing. Specifically, a) On (b) (6) directed oocyte donor (b) (6) tested (b) (6) (b) (7)(C) as documented on the laboratory test report dated (b) (6) (b) (7)(C) This directed oocyte donor (b) (6) was determined eligible and donated (b) (6) oocytes on that were cryopreserved, thawed on (b) (6) fertilized, and (b) (6) resulting tissues cryopreserved. b) On …
Contains “laboratory test*”Read the Form 483 →
Hangzhou Minsheng Binjiang Pharmaceutical Co., Ltd.Inspection ended Mar 7, 2025 · FEI 3007498452 · Observation 3
The responsibilities and procedures applicable to the quality control unit are not fully followed. Specifically, during the inspection, we observed torn and discarded GMP documents in the trash bin located in the QC laboratory. Your firm was unaware that these GMP documents were altered and discarded. GMP documents are not managed according to your procedures, document # 02-QA-2-001 and 02-QA-2-004.
Contains “QC laborator*”Read the Form 483 →
Zhejiang Guobang Pharmaceutical Co., LtdInspection ended Mar 3, 2025 · FEI 3008617115 · Observation 1
Laboratory controls lack scientifically justified and appropriate specifications and test procedures necessary to ensure that components meet established standards for identity, quality, and purity. Specifically, (b)(4) water is utilized in the critical cleaning of equipment in classified rooms during the water was also used as a component in the manufacturing processes of your drug substances. (b)(4) Batch manufacturing of drug substances …
Contains “laboratory control*”Read the Form 483 →
Shouguang Fukang Pharmaceutical Co., Ltd.Inspection ended Feb 21, 2025 · FEI 3003941038 · Observation 1
The quality control unit lacks responsibility to approve final testing results for product release. Specifically, a review of completed analytical testing records for the API [illegible] regarding HPLC L-002 and GC L-0[illegible] testing results noted that QC performs the review/approval without review/approval by the QA department. For example, the chemist uses Excel Spreadsheets to manual input testing result data from both the HPLC and GC. The …
Contains “HPLC*”Read the Form 483 →
Shouguang Fukang Pharmaceutical Co., Ltd.Inspection ended Feb 21, 2025 · FEI 3003941038 · Observation 5
Analytical methods are not suitable for their intended use and/or have not been established. Specifically, A. The validation report, "Validation Report of Analytical Procedure for Related Substances in (b)(1) (USP/EP)," does not reference the batch numbers associated with the validation. There is no assurance that any (b)(1) batches were associated with this method validation. B. The validation report, "Validation Report of Analytical Procedure …
Contains “analytical method*”Read the Form 483 →
21 CFR sections these observations open with
The sections FDAgent ties the most of these observations to by FDA's wording, with how many of this theme's observations each covers. Each links to its page of FDA citations, warning letters and the regulation text.
- 21 CFR 211.160General requirementsIn this theme's rule40
- 21 CFR 211.165Testing and release for distributionIn this theme's rule24
- 21 CFR 211.166Stability testingIn this theme's rule21
- 21 CFR 211.194Laboratory records17
- 21 CFR 211.192Production record review15
- 21 CFR 211.22Responsibilities of quality control unit12
- 21 CFR 211.167Special testing requirementsIn this theme's rule6
- 21 CFR 211.170Reserve samplesIn this theme's rule6
- 21 CFR 211.68Automatic, mechanical, and electronic equipment5
- 21 CFR 211.110Sampling and testing of in-process materials and drug products3
Ask FDAgent about laboratory controls in Form 483s
FDAgent reads FDA's Form 483s, inspections, citations and warning letters and answers in plain English, with links to the records.
Related
- Laboratory controls: FDA's Form 483 citations of the same 21 CFR entries
- Aseptic processing: FDA's Form 483 citations of overlapping 21 CFR entries
- Stability: FDA's Form 483 citations of overlapping 21 CFR entries
- Every Form 483 theme, with its counts
- What a Form 483 is, and the latest 483s
- Every 21 CFR section in FDA's Form 483 citations
About this data
- The observations are the text FDAgent read by OCR from Form 483 PDFs in FDA's Electronic Reading Room, shown only from 483s whose every printed observation passed FDAgent's checks against the scanned text, as on each 483's own page.
- Which theme an observation belongs to is decided by the fixed rules above; every count on this page is computed from them.
- A 483 addressed to an individual is listed without the name, and an observation that names a person is left out.
- FDAgent is independent of FDA.