Laboratory controls in FDA Form 483 observations

Quality control laboratories: test methods and their validation, laboratory records and equipment, reference standards and chromatography. FDAgent finds 304 such observations in 172 Form 483s issued to 163 facilities, for inspections that ended between October 2008 and September 2026. Each is quoted below from the 483 FDA posted.

A Form 483 lists the investigators' observations at the end of an inspection. It is not FDA's final determination.

Source: Form 483 PDFs in FDA's Electronic Reading Room, last synced Nov 23, 2026, 04:33 UTC. Every quote is from a 483 whose reading passed FDAgent's checks against the scan. FDAgent is independent of FDA.

Observations
304
Each quoted below
Form 483s
172
Posted by FDA
Facilities
163
Distinct FEIs
Most recent
Sep 3, 2026
Inspection end date

Observations by year

These observations by the year the inspection ended. They follow which 483s FDA has posted and FDAgent has read, so they show what is on record, not a trend in FDA's findings. 2026 runs to the latest sync (lighter bar).

Peak: 76 in 2026

Laboratory controls observations in Form 483s by inspection year
YearObservations
20081
20090
20100
20110
20128
20130
20145
20154
201613
201721
201823
201914
20200
20216
202218
202342
202425
202548
202676

How FDAgent sorts observations into this theme

Two fixed rules, applied by code to every observation FDAgent can quote. An observation counts when either holds. No model and no similarity score decides.

It is tied to one of these 21 CFR entries

21 CFR 211.16021 CFR 211.16521 CFR 211.16621 CFR 211.16721 CFR 211.17021 CFR 211.17321 CFR 211.176

The same entries the laboratory controls topic counts. A 483 prints no regulation, so FDAgent ties an observation to a section only when it opens with FDA's own citation wording for that section, word for word.

Or its text contains one of these phrases

“laboratory control*” · “laboratory record*” · “laboratory investigation*” · “laboratory equipment” · “laboratory instrument*” · “laboratory data” · “laboratory test*” · “laboratory result*” · “laboratory analyst*” · “laboratory procedure*” · “laboratory notebook*” · “laboratory error*” · “QC laborator*” · “quality control laborator*” · “test method*” · “analytical method*” · “analytical procedure*” · “method validation” · “HPLC*” · “chromatograph*” · “reference standard*” · “system suitability” · “method transfer”

* matches any ending and … up to three words in between. Case is ignored unless a phrase has capitals.

The 304 observations, newest first

Quoted word for word from each Form 483, with only line breaks joined; … marks where the quote is cut. The highlight shows why the observation is in this theme. Each links to FDAgent's page for the 483.

  1. Nubratori, IncInspection ended Jul 26, 2024 · FEI 3010166491 · Observation 4

    Acceptance criteria for the sampling and testing conducted by the quality control unit is not adequate to assure that batches of drug products meet each appropriate specification and appropriate statistical quality control criteria as a condition for their approval and release. Specifically, Your firm's visual inspection program is inadequate. a. You do not document the complete results [illegible] Quality control after the 100% visual inspection …

    Opens with FDA's citation wording for 21 CFR 211.165(d)Read the Form 483 →

  2. Serum Institute of India Pvt. Limited Manjari SiteInspection ended Jul 2, 2024 · FEI 3020223359 · Observation 4

    There is a lack of (b) (4) data in the qualification report for (b) (4) reference standard lot # (b) (4). The (b) (4) is used to (b) (4) of this reference standard and serves as supportive information for the (b) (4) for the release of (b) (4) and drug product. Moreover, no investigation was included in the qualification report regarding the failure of producing valid test results for (b) (4) for the working (b) (4) reference standard.

    Contains “reference standard*”Read the Form 483 →

  3. Hikma Pharmaceuticals USA Inc.Inspection ended Jun 12, 2024 · FEI 2220525 · Observation 2

    Laboratory controls do not include the establishment of scientifically sound and appropriate test procedures designed to assure that drug products conform to appropriate standards of identity, strength, quality and purity. Specifically, your firm's co-validated method transfer, Report CHH15181 Method Transfer Report for SOP CHH15817 (TMD227) Determination of Fentanyl in Fentanyl Citrate in 0.9% Sodium Chloride Injection IV Bags omits established …

    Opens with FDA's citation wording for 21 CFR 211.160(b)Read the Form 483 →

  4. PCAS Finland OyInspection ended Jun 7, 2024 · FEI 3003469991 · Observation 1

    … per SOP Vaatimuksesta poikkeavien (OOS/OOT) laboratoriotulosten kasittely ja raportointi, DOC-00002873 (English: Processing and reporting of non-conforming laboratory results (OOS/OOT)). In addition, one major known (b)(4) can only be seen using the in-house method. This is expected but has not been investigated as to whether the amount is an acceptable output of the manufacturing process. The following most recent API batches have been released …

    Contains “laboratory result*”Read the Form 483 →

  5. Swati Spentose Private LimitedInspection ended May 31, 2024 · FEI 3009765202 · Observation 1

    Employees are not given role specific training to perform their function. Specifically, As per SOP # QCD/0117/000, "Procedure for Review of Analytical Documents and Reviewer Qualification", the QC laboratory reviewer who reviewed the testing data needs to be trained and qualified to perform the review of the analyzed test however your firm failed to ensure that all Quality Control (QC) laboratory reviewers are adequately trained to review …

    Contains “QC laborator*”Read the Form 483 →

  6. Apotex Research Private LimitedInspection ended Mar 15, 2024 · FEI 3006076314 · Observation 3

    … (b) (4) CAPA has not been implemented for the following: LIR Batch Sub lot Mean %SD Anomalous result (b) (4) B) The Quality Unit has not ensured that all system suitability injections were performed in chromatographic sample set sequence for analyzing the degradation products for (b) (4) mg prior to injecting the samples. During the review of the degradation results of Batch (b) (4) (Stability Sample, RT (b) (4)) you failed to inject the Solution …

    Contains “system suitability”Read the Form 483 →

  7. Fukuzyu Pharmaceutical Co., Ltd.Inspection ended Jan 19, 2024 · FEI 3002807025 · Observation 3

    Validation of analytical methods is incomplete. Specifically, your firm did not provide records during this inspection that documented analysts in your laboratory performed method validation, verification, or transfer for the in-house gas chromatography (GC) residual solvents method used to analyze samples for release testing of (b)(4). Your firm used results from this method to release all (b)(4) API, including batch (b)(4) lot (b)(4) that was …

    Contains “analytical method*”Read the Form 483 →

  8. Fukuzyu Pharmaceutical Co., Ltd.Inspection ended Jan 19, 2024 · FEI 3002807025 · Observation 4

    Appropriate procedures are not established and followed for control of computerized systems and electronic records in the QC laboratory. Specifically, A) The Vice Manager of QA is the sole administrator for Empower 3 software used to collect and process data from multiple chromatography instruments in your laboratory. On 01/18/2023, I observed the Vice Manager of QA login to the software as the administrator and then access system policies, audit …

    Contains “QC laborator*”Read the Form 483 →

  9. Jiangsu Hengrui Pharmceuticals Co., LtdInspection ended Jan 16, 2024 · FEI 3007373503 · Observation 4

    … review of electronic data by the Quality Assurance Unit for standalone and network systems, to ensure completeness, consistency, and accuracy of all chromatographic and non-chromatographic electronic data generated by the Quality Control Laboratory.

    Contains “chromatograph*”Read the Form 483 →

  10. Laurus Synthesis Private LimitedInspection ended Dec 12, 2023 · FEI 3011524794 · Observation 1

    … Non-Conformance Investigation NCM/LSV1/003/22 was initiated for non-conforming assay and purity results by HPLC during (b) (4) hold-time study of key starting material (b) (4). One out of (b) (4) batches on hold time study (2 of (b) (4) samples) analyzed failed for assay and impurities. The investigation concluded that it was an isolated case and no further actions were documented. C. OOS investigation OOS/LSV1/001/22 identified improper …

    Contains “HPLC*”Read the Form 483 →

  11. Alkem Laboratories LimitedInspection ended Dec 1, 2023 · FEI 3008450172 · Observation 1

    Validated methods have not been established for testing of API's. Specifically, your firm does not have a validated method for (b) (4) API's to which the site is used as an alternative testing facility. The site is documented to test for identification of (b) (4) API's by (IR) and water content by (KF). However, during the inspection, when reviewing the method validation and verification documentations, no method verification / validation is …

    Contains “method validation”Read the Form 483 →

  12. Antaria Pty Ltd.Inspection ended Nov 17, 2023 · FEI 3025978151 · Observation 3

    Laboratory records are deficient in that they do not include a complete record of all data obtained during testing and do not always include documentation of pertinent information related to the preparation and testing of a sample. For example, a) There is no documentation of the preparation of standard stock solutions, working standards, reference standard solutions, sample solutions, diluents, mobile phases, reagents, and other preparations …

    Contains “laboratory record*”Read the Form 483 →

  13. Antaria Pty Ltd.Inspection ended Nov 17, 2023 · FEI 3025978151 · Observation 4

    Your firm does not assign an expiration date for prepared working standards, mobile phases, diluents and other preparations used for performing testing of drug products, including (b) (4) API. Test methods include testing for assay, elemental impurities, and other required USP tests. For example, the bottle of working standard (b) (4) ug/mL (b) (4) standard solution) that is currently being used for assay by (b) (4) chromatography is marked with …

    Contains “test method*”Read the Form 483 →

  14. Antaria Pty Ltd.Inspection ended Nov 17, 2023 · FEI 3025978151 · Observation 5

    There is no written testing program designed to assess the stability characteristics of drug products. Specifically, your firm has not placed at least one lot of (b) (4) API (b) (4) on stability annually. For example, you currently have (b) (4) lots on long term stability; (b) (4) lot placed on stability in 2022 and (b) (4) in 2023. No lots have been placed on stability for product manufactured at the Brisbane site in 2019, 2020, or 2021. In …

    Opens with FDA's citation wording for 21 CFR 211.166(a)Read the Form 483 →

  15. Antaria Pty Ltd.Inspection ended Nov 17, 2023 · FEI 3025978151 · Observation 7

    Your firm is not always following the approved test method. For example, a) SOP206.15 USP (b) (4) Monograph Test Methods, rev 8 issued 09AUG2023, requires the use of a (b) (4) (b) (4) for the performance of the test for (b) (4) and colour of solution. This is consistent with the USP test method. Your firm is using a (b) (4) b) SOP206.15 USP (b) (4) Monograph Test Methods, rev 8 issued 09AUG2023, requires drying of the (b) (4) and sample at (b) …

    Contains “test method*”Read the Form 483 →

  16. Antaria Pty Ltd.Inspection ended Nov 17, 2023 · FEI 3025978151 · Observation 8

    Your firm has no documentation to show the test methods used for stability testing of (b) (4) USP (API) and other drug products are stability indicating.

    Contains “test method*”Read the Form 483 →

  17. Jiangsu Intco Medical Products Co., LtdInspection ended Nov 17, 2023 · FEI 3005260401 · Observation 2

    There is no written testing program designed to assess the stability characteristics of drug products. Specifically, Your firm’s stability program is deficient. For example: (a) You did not maintain batch production [illegible] to support your stability testing used to establish the expiration date for your OTC liquid [illegible] and OTC [illegible]. The following batches were used for your accelerated and long-term stability study, however your …

    Opens with FDA's citation wording for 21 CFR 211.166(a)Read the Form 483 →

  18. Catalent Indiana, LLCInspection ended Nov 15, 2023 · FEI 3005949964 · Observation 4

    … Justification provided for why this is allowable (document A-VPPQ-00316-S-VL-016AUG23, b) (4 Visual Inspection Process [1)] Vials) Test Method Validation (TMV) Summary Report, VL-016AUG23) is inadequate in that the justification appears to attempt to validate a [(6f (4)] process and the document's qualification methodology was not designed to mimic the conditions of the individual inspector qualifications (§3.2).

    Contains “test method*”Read the Form 483 →

  19. Catalent Indiana, LLCInspection ended Nov 15, 2023 · FEI 3005949964 · Observation 5

    Laboratory controls do not include the establishment of scientifically sound and appropriate test procedures designed to assure that drug products conform to appropriate standards of identity, strength, quality and purity. Specifically, There is a failure to validate the endotoxin sample storage of drug products under project code: (6)(4) I a((o) (4) I up tq (6) (4) I prior to the testing. A [(o)(4) I hold study or low endotoxin recovery data …

    Opens with FDA's citation wording for 21 CFR 211.160(b)Read the Form 483 →

  20. Sumitomo Osaka Cement Co., LtdInspection ended Oct 27, 2023 · FEI 3003943727 · Observation 2

    Out of Specification investigation (OOS) is deficient and further no CAPA is initiated to prevent recurrence. Specifically, 1. OOS investigation OOS21001 was initiated on 3/3/2021 to probe the failure of (b)(4) Assay (b)(4) to (b)(4)% in (b)(4) API batch (b)(4) % using Ion chromatograph. The root cause of failure was attributed to error in the first standard preparation without any scientific justification or comparison with historical area …

    Contains “chromatograph*”Read the Form 483 →

21 CFR sections these observations open with

The sections FDAgent ties the most of these observations to by FDA's wording, with how many of this theme's observations each covers. Each links to its page of FDA citations, warning letters and the regulation text.

Ask FDAgent about laboratory controls in Form 483s

FDAgent reads FDA's Form 483s, inspections, citations and warning letters and answers in plain English, with links to the records.

Edit it or ask as written. 3 questions free, no card needed.

About this data

  • The observations are the text FDAgent read by OCR from Form 483 PDFs in FDA's Electronic Reading Room, shown only from 483s whose every printed observation passed FDAgent's checks against the scanned text, as on each 483's own page.
  • Which theme an observation belongs to is decided by the fixed rules above; every count on this page is computed from them.
  • A 483 addressed to an individual is listed without the name, and an observation that names a person is left out.
  • FDAgent is independent of FDA.