Laboratory controls in FDA Form 483 observations

Quality control laboratories: test methods and their validation, laboratory records and equipment, reference standards and chromatography. FDAgent finds 304 such observations in 172 Form 483s issued to 163 facilities, for inspections that ended between October 2008 and September 2026. Each is quoted below from the 483 FDA posted.

A Form 483 lists the investigators' observations at the end of an inspection. It is not FDA's final determination.

Source: Form 483 PDFs in FDA's Electronic Reading Room, last synced Nov 23, 2026, 04:33 UTC. Every quote is from a 483 whose reading passed FDAgent's checks against the scan. FDAgent is independent of FDA.

Observations
304
Each quoted below
Form 483s
172
Posted by FDA
Facilities
163
Distinct FEIs
Most recent
Sep 3, 2026
Inspection end date

Observations by year

These observations by the year the inspection ended. They follow which 483s FDA has posted and FDAgent has read, so they show what is on record, not a trend in FDA's findings. 2026 runs to the latest sync (lighter bar).

Peak: 76 in 2026

Laboratory controls observations in Form 483s by inspection year
YearObservations
20081
20090
20100
20110
20128
20130
20145
20154
201613
201721
201823
201914
20200
20216
202218
202342
202425
202548
202676

How FDAgent sorts observations into this theme

Two fixed rules, applied by code to every observation FDAgent can quote. An observation counts when either holds. No model and no similarity score decides.

It is tied to one of these 21 CFR entries

21 CFR 211.16021 CFR 211.16521 CFR 211.16621 CFR 211.16721 CFR 211.17021 CFR 211.17321 CFR 211.176

The same entries the laboratory controls topic counts. A 483 prints no regulation, so FDAgent ties an observation to a section only when it opens with FDA's own citation wording for that section, word for word.

Or its text contains one of these phrases

“laboratory control*” · “laboratory record*” · “laboratory investigation*” · “laboratory equipment” · “laboratory instrument*” · “laboratory data” · “laboratory test*” · “laboratory result*” · “laboratory analyst*” · “laboratory procedure*” · “laboratory notebook*” · “laboratory error*” · “QC laborator*” · “quality control laborator*” · “test method*” · “analytical method*” · “analytical procedure*” · “method validation” · “HPLC*” · “chromatograph*” · “reference standard*” · “system suitability” · “method transfer”

* matches any ending and … up to three words in between. Case is ignored unless a phrase has capitals.

The 304 observations, newest first

Quoted word for word from each Form 483, with only line breaks joined; … marks where the quote is cut. The highlight shows why the observation is in this theme. Each links to FDAgent's page for the 483.

  1. UCB Farchim SAInspection ended Jul 4, 2023 · FEI 3005023799 · Observation 2

    Appropriate controls are not exercised over computers or related systems to assure that changes in laboratory control records or other records are instituted only by authorized personnel. GMP related computerized systems and equipment have not been validated/qualified to demonstrate the suitability of computer hardware and software to perform assigned tasks. For example, your firm appears to have performed Performance Verification for Empower 3 …

    Contains “laboratory control*”Read the Form 483 →

  2. UCB Farchim SAInspection ended Jul 4, 2023 · FEI 3005023799 · Observation 3

    The accuracy, sensitivity, specificity, and reproducibility of test methods have not been established and documented. Your firm has not performed analytical method transfers for the following active pharmaceutical ingredients: [illegible] Additionally, standard and sample solution stability has not been established by your firm as part of the HPLC method validation reports and testing operations for [illegible].

    Opens with FDA's citation wording for 21 CFR 211.165(e)Read the Form 483 →

  3. Ipca Laboratories LimitedInspection ended Jun 23, 2023 · FEI 3007574780 · Observation 1

    … TP/DR/0691/00 Dissolution (By UV) for (b) (4) Tablets, Effective date 10/25/2019, in line with your analytical method validation report, AMV/R/2017/009/00, Effective date 8/21/2017. Based on this inconclusive hypothesis, you retested the samples for dissolution and based on the passing results, you invalidated the initial OOS results. B. On 12/5/2022, the lab recorded OOS result PIT/OOS/2020/051 for test by HPLC. The OOS result was reported for …

    Contains “analytical method*”Read the Form 483 →

  4. Ipca Laboratories LimitedInspection ended Jun 23, 2023 · FEI 3007574780 · Observation 3

    … However, the sample sequence described in your UV methods (one blank, one standard, one sample) for (b) (4) of the (b) (4) drug products did not establish the system suitability whenever the method was used for dissolution testing. Your firm has been using the UV methods without system suitability established to test the dissolution of the (b) (4) drug products for batch release and stability study since each the Agency (refer to the table …

    Contains “system suitability”Read the Form 483 →

  5. Ipca Laboratories LimitedInspection ended Jun 23, 2023 · FEI 3007574780 · Observation 5

    Laboratory controls do not include the establishment of scientifically sound and appropriate test procedures designed to assure that drug products conform to appropriate standards of identity, strength, quality and purity. Specifically, scientifically sound, and appropriate test procedures are not deployed to establish the identity, strength, quality, and purity of the drug products manufactured at the site. For example: On 6/19/2023, I observed …

    Opens with FDA's citation wording for 21 CFR 211.160(b)Read the Form 483 →

  6. Ipca Laboratories LimitedInspection ended Jun 23, 2023 · FEI 3007574780 · Observation 6

    The responsibilities and procedures applicable to the quality control unit are not fully followed. Specifically, the quality unit failed to control the laboratory test methods created and then transferred in LES Development server as per SOP CSOP/2016/047, "Method Creation on Method Builder Software". Computerized system LES (Laboratory Execution System) is used by the quality control laboratory for documentation and to perform analysis. …

    Contains “laboratory test*”Read the Form 483 →

  7. Qilu Antibiotics Pharmaceutical Co., LTD.Inspection ended May 30, 2023 · FEI 3009831352 · Observation 4

    Laboratory controls do not include the establishment of scientifically sound and appropriate test procedures designed to assure that drug products conform to appropriate standards of identity, strength, quality, and purity. Specifically, a. Sterility testing of sterile injectable products distributed to the U.S. market requires determination of microbial growth in (b) (4) canisters. There is no requirement to document a check for the (b) (4) for …

    Opens with FDA's citation wording for 21 CFR 211.160(b)Read the Form 483 →

  8. Zen Chemicals Pvt. LtdInspection ended Mar 28, 2023 · FEI 3015345351 · Observation 2

    No investigations are conducted to evaluate all potential root causes of multiple sample results. Specifically, in reviewing your firm HPLC and GC computerized analytical data from 2021-2023, you failed to investigate the following: A. Three (3) separate analysis were rerun on HPLC equipment # QC/INST/09. 1. The first set of samples consisted of (b) (4) injections which were analyzed on December 7-8, 2021 for (b) (4) calibration. The injections …

    Contains “HPLC*”Read the Form 483 →

  9. Zen Chemicals Pvt. LtdInspection ended Mar 28, 2023 · FEI 3015345351 · Observation 3

    Appropriate controls governing computer acquired data and/or related systems have not been established to assure that changes and access to computerized software for the laboratory control system is accessed only by authorized personnel. Specifically, on March 23 and 27, 2023, I observed, A. UV which uses software UV Probe 2.70 (equipment # QC/INST/13) does not have audit trail capabilities. This equipment is used for the identification of (b) …

    Contains “laboratory control*”Read the Form 483 →

  10. Lonza AGInspection ended Mar 28, 2023 · FEI 3002930340 · Observation 1

    A) Laboratory records do not include complete data derived from all tests, examinations and assay necessary to assure compliance with established specifications and standards. Specifically, You firm failed to report exact values for TAMC (Total Aerobic Microbial Count) action limit results as indicated in Lonza Test Method, Total Aerobic Microbial Count in (b) (4) Water, Doc. No.: CHVI-15464, Version: 11.0, Effective Date: 13-DEC-2022. For …

    Contains “laboratory record*”Read the Form 483 →

  11. Dynavax GmbHInspection ended Feb 3, 2023 · FEI 3010165220 · Observation 1

    The investigation into the out of specification test result and the corresponding manufacturing investigation were inadequate in that the investigation did not extend into the review of the (b) (4) sanitization of the (b) (4). Specifically, Laboratory investigation #0202-001 and deviation DEV-00031 were initiated on March 16, 2020, for an OOS endotoxin result for the (b) (4). The investigation root cause was assigned to method due to interference …

    Contains “laboratory investigation*”Read the Form 483 →

  12. Octapharma Pharmazeutika Produktionsgesellschaft m.b.H. OPGInspection ended Dec 16, 2022 · FEI 3002809097 · Observation 4

    Deviations from written Quality Control test procedures for holding temperature and time of reference standards are not justified. Specifically, (b)(4) for laboratory testing are not properly stored to ensure the quality is intact. During the Quality walk through, (b)(4) of reference materials used for determination of (b)(4) (b)(4) (b)(4) with recommended storage condition of (b)(4)) and (b)(4) control (lot (b)(4) with recommended storage …

    Contains “reference standard*”Read the Form 483 →

  13. Lupin Limited Unit 1Inspection ended Nov 23, 2022 · FEI 3002807511 · Observation 1

    … In Phase I investigation you noted the test failing for system suitability and through hypothesis test, you concluded instrument malfunction was the root cause for the recorded OOS results. You retested the sample and invalidated the initial OOS results. The Head of Process Development and General Manager of Quality stated that the batch was part of monitoring study, and the batch disposition will be based on overall outcome of the study. …

    Contains “system suitability”Read the Form 483 →

  14. Lupin Limited Unit 1Inspection ended Nov 23, 2022 · FEI 3002807511 · Observation 2

    … Many of your test methods do not contain sequence list, including but not limited to: (b) (4) USP (b) (4) mg/(b) (4) mL Specification No. FPF 1027R (04) Effective date 8/2/2022, and (b) (4) USP (b) (4) mg/(b) (4) mL Specification No. FPF 122R (18) Effective date 10/4/2022. b. On 11/14/2022, during walkthrough of your (b) (4) Room (b) (4) individuals were observed using a common password being used for the (b) (4) ID # (b) (4) 08. Your production …

    Contains “test method*”Read the Form 483 →

  15. Lupin Limited Unit 1Inspection ended Nov 23, 2022 · FEI 3002807511 · Observation 9

    Investigations are inadequate to determine the root cause for the failure of a batch of intermediate or API to meet specifications whether or not the batch has been already distributed. Specifically, On 11/21/2022, your Investigation Manager reviewed OOS/C/20/MDP/FPDS/007 that was initiated to probe OOS result that was reported in related substances test by HPLC during analysis of (b) (4) USP Batch No. (b) (4) such that the observed result of any …

    Contains “HPLC*”Read the Form 483 →

  16. Lupin Limited Unit 1Inspection ended Nov 23, 2022 · FEI 3002807511 · Observation 11

    … Many of your test methods do not contain sequence list, including but not limited to: (b) (4) USP Specification No. FPB 339 (12) Effective date 9/11/2021, and (b) (4) USP Specification No. FPB 101R (14) Effective date 7/7/2021. b. Your General Manager of Quality provided a list of (b) (4) manufacturing equipment in the (b) (4) API manufacturing building. The list documents out of (b) (4) manufacturing equipment, 8 were on SCADA (21 CFR Part 11 …

    Contains “test method*”Read the Form 483 →

  17. QuVa Pharma , Inc.Inspection ended Oct 14, 2022 · FEI 3013931875 · Observation 4

    … C) During walkthrough of the QC laboratory and review of data in the (b) (4) software, we observed that Senior Chemist (b) (6) who routinely conducts analytical testing of finished sterile products, and raw materials, had (b) (4) roles assigned and has final data review and approve privileges in (b) (4) enabled. Prior to (b) (6) role as a Senior Chemist, (b) (6) functioned as a Reviewer; however, Senior Chemist (b) (6) has not functioned in the …

    Contains “QC laborator*”Read the Form 483 →

  18. Catalent Indiana LLCInspection ended Sep 1, 2022 · FEI 3005949964 · Observation 1

    … of QC stated they wanted to retest (b)(4) samples, but the client felt that since (b)(4) was identified, the initial results were contamination from the QC laboratory and therefore, only approved (b)(4) vials to be retested. The retest of the (b)(4) vials resulted in three individual vials with foreign particles ((b)(4) indications of (b)(4) material). There is no evidence to support the particles are isolated to the QC laboratory environment. …

    Contains “QC laborator*”Read the Form 483 →

  19. Catalent Indiana LLCInspection ended Sep 1, 2022 · FEI 3005949964 · Observation 4

    Acceptance criteria for the sampling and testing conducted by the quality control unit is not adequate to assure that batches of drug products meet appropriate statistical quality control criteria as a condition for their approval and release. Specifically, A-SOP-03-04-001, Performance of AQL for Parenteral Inspections, states AQL sampling should be a representation of the entire batch. "During (b)(4) Inspection with (b)(4) (System 2440), if AQL …

    Opens with FDA's citation wording for 21 CFR 211.165(d)Read the Form 483 →

  20. Catalent Indiana LLCInspection ended Sep 1, 2022 · FEI 3005949964 · Observation 9

    In-process samples are not representative. Specifically, There is no scientific justification for not performing particles testing on routinely inspected product. Batches of aseptically filled (b)(4) are inspected by (b)(4). The particulate defects are rejected and the "ejects" are run through the (b)(4) (b)(4) for additional inspection. Instead of performing destructive particles testing on the good vials remaining after the (b)(4) inspection, …

    Opens with FDA's citation wording for 21 CFR 211.160(b)(2)Read the Form 483 →

21 CFR sections these observations open with

The sections FDAgent ties the most of these observations to by FDA's wording, with how many of this theme's observations each covers. Each links to its page of FDA citations, warning letters and the regulation text.

Ask FDAgent about laboratory controls in Form 483s

FDAgent reads FDA's Form 483s, inspections, citations and warning letters and answers in plain English, with links to the records.

Edit it or ask as written. 3 questions free, no card needed.

About this data

  • The observations are the text FDAgent read by OCR from Form 483 PDFs in FDA's Electronic Reading Room, shown only from 483s whose every printed observation passed FDAgent's checks against the scanned text, as on each 483's own page.
  • Which theme an observation belongs to is decided by the fixed rules above; every count on this page is computed from them.
  • A 483 addressed to an individual is listed without the name, and an observation that names a person is left out.
  • FDAgent is independent of FDA.