Laboratory controls in FDA Form 483 observations

Quality control laboratories: test methods and their validation, laboratory records and equipment, reference standards and chromatography. FDAgent finds 304 such observations in 172 Form 483s issued to 163 facilities, for inspections that ended between October 2008 and September 2026. Each is quoted below from the 483 FDA posted.

A Form 483 lists the investigators' observations at the end of an inspection. It is not FDA's final determination.

Source: Form 483 PDFs in FDA's Electronic Reading Room, last synced Nov 23, 2026, 04:33 UTC. Every quote is from a 483 whose reading passed FDAgent's checks against the scan. FDAgent is independent of FDA.

Observations
304
Each quoted below
Form 483s
172
Posted by FDA
Facilities
163
Distinct FEIs
Most recent
Sep 3, 2026
Inspection end date

Observations by year

These observations by the year the inspection ended. They follow which 483s FDA has posted and FDAgent has read, so they show what is on record, not a trend in FDA's findings. 2026 runs to the latest sync (lighter bar).

Peak: 76 in 2026

Laboratory controls observations in Form 483s by inspection year
YearObservations
20081
20090
20100
20110
20128
20130
20145
20154
201613
201721
201823
201914
20200
20216
202218
202342
202425
202548
202676

How FDAgent sorts observations into this theme

Two fixed rules, applied by code to every observation FDAgent can quote. An observation counts when either holds. No model and no similarity score decides.

It is tied to one of these 21 CFR entries

21 CFR 211.16021 CFR 211.16521 CFR 211.16621 CFR 211.16721 CFR 211.17021 CFR 211.17321 CFR 211.176

The same entries the laboratory controls topic counts. A 483 prints no regulation, so FDAgent ties an observation to a section only when it opens with FDA's own citation wording for that section, word for word.

Or its text contains one of these phrases

“laboratory control*” · “laboratory record*” · “laboratory investigation*” · “laboratory equipment” · “laboratory instrument*” · “laboratory data” · “laboratory test*” · “laboratory result*” · “laboratory analyst*” · “laboratory procedure*” · “laboratory notebook*” · “laboratory error*” · “QC laborator*” · “quality control laborator*” · “test method*” · “analytical method*” · “analytical procedure*” · “method validation” · “HPLC*” · “chromatograph*” · “reference standard*” · “system suitability” · “method transfer”

* matches any ending and … up to three words in between. Case is ignored unless a phrase has capitals.

The 304 observations, newest first

Quoted word for word from each Form 483, with only line breaks joined; … marks where the quote is cut. The highlight shows why the observation is in this theme. Each links to FDAgent's page for the 483.

  1. Sumitomo Osaka Cement Co., LtdInspection ended Oct 27, 2023 · FEI 3003943727 · Observation 4

    … The analytical method for the raw material (b)(4) used in the manufacturing of (b)(4) API is not verified. The firm does not perform testing of assay and instead confirm assay with the COA of supplier to ensure if the results meet the firms' specification. 6. The operating (b)(4) of the (b)(4) Id # APK-002 utilized in (b)(4) of raw material to manufacture (b)(4) (b)(4) and (b)(4) API is (b)(4) to (b)(4). However, the calibration of the old (b)(4) …

    Contains “analytical method*”Read the Form 483 →

  2. Sanofi PasteurInspection ended Sep 29, 2023 · FEI 3002972083 · Observation 4

    Laboratory controls do not include the establishment of scientifically sound and appropriate standards designed to assure that drug products conform to appropriate standards of identity, strength, quality, and purity. Specifically, reference standards are not included in analytical sequences to verify (b)(4) control for the following critical test assays. A. SOP: MLE 1-005714, titled, [illegible] is used for the Purified Vi Polysaccharide bulk …

    Opens with FDA's citation wording for 21 CFR 211.160(b)Read the Form 483 →

  3. Linhai Tianyu Pharmaceutical Co., Ltd.Inspection ended Sep 15, 2023 · FEI 3013345425 · Observation 3

    … During our review of your Empower 3 chromatography software, interrupted sequences were observed, which generated "Data Incomplete" chromatographic data. Your firm documented these interrupted injections due to a power outage via investigation no. DR-3-22003R, dated 02/27/2022, showing that no chromatograms had been generated for (b) (4) drug substance batch numbers (b) (4). However, during the investigation, the interrupted sample injections …

    Contains “chromatograph*”Read the Form 483 →

  4. Linhai Tianyu Pharmaceutical Co., Ltd.Inspection ended Sep 15, 2023 · FEI 3013345425 · Observation 4

    Laboratory records are not completed contemporaneously and do not include complete data derived from all tests to assure compliance with established specifications and standards. Specifically, A. Electronic and physical laboratory records data review are not completed contemporaneously by your QC reviewers. During our walkthrough of your QC laboratory data review room on 09/11/23, I observed your QC analyst/reviewer checking off boxes on a form …

    Contains “laboratory record*”Read the Form 483 →

  5. Linhai Tianyu Pharmaceutical Co., Ltd.Inspection ended Sep 15, 2023 · FEI 3013345425 · Observation 5

    There is no written testing program designed to assess the stability characteristics of drug substances and intermediates. Specifically, the assay and related substance test methods for intermediate (b) (4) used for stability testing are not stability indicating, where they can detect the change with time in chemical, physical or microbiological properties of the intermediate drug substance so that the content of active ingredients and …

    Contains “test method*”Read the Form 483 →

  6. Linhai Tianyu Pharmaceutical Co., Ltd.Inspection ended Sep 15, 2023 · FEI 3013345425 · Observation 6

    … For Example, during our review of your Empower 3 chromatography software, interrupted sequences were observed, which generated "Data Incomplete" chromatographic data. Your firm has not demonstrated full understanding of the different types of communication errors and circumstances which may lead to a "Data incomplete" or Bad Checksum/Data Missing chromatography considering the relationship between the Empower application and database server, …

    Contains “chromatograph*”Read the Form 483 →

  7. Linhai Tianyu Pharmaceutical Co., Ltd.Inspection ended Sep 15, 2023 · FEI 3013345425 · Observation 7

    … Quality Control (QC) personnel perform review electronic test data for all analytical batch records using an approved review procedure and checklist for HPLC/GC network equipment. This review process does not appear to be adequate, in that all potential data generated during that specific review period for that product are not reviewed. For example, your firm utilizes up to five different software's for testing of intermediate and API products on …

    Contains “HPLC*”Read the Form 483 →

  8. Central Glass Co., Ltd.Inspection ended Sep 15, 2023 · FEI 3002806649 · Observation 1

    Laboratory Controls are inadequate. Specifically, a) The working standards used during release testing of (b) (4) for assay and volatile analogues not always elude peaks for the corresponding volatile analogues. During qualification of your working standards one of the requirements is to verify the presence of all peaks. However, injection of standard lot (b) (4) during analysis of commercial (b) (4) API lots (b) (4) does not contain peaks for …

    Contains “laboratory control*”Read the Form 483 →

  9. Central Glass Co., Ltd.Inspection ended Sep 15, 2023 · FEI 3002806649 · Observation 3

    … e) Your test method used from 2015-2020 to evaluate cleaning validation samples had an LOQ of (b) (4) ppm. This test was not suitable for the evaluation of cleaning validation samples (b) (4) specifications for residue were (b) (4) ppm during this time.

    Contains “test method*”Read the Form 483 →

  10. Central Glass Co., Ltd.Inspection ended Sep 15, 2023 · FEI 3002806649 · Observation 4

    … On 8/6/23 UQSE20-19 test method used to evaluate rinse samples from cleaning validation was changed to obtain a lower LOD and LOQ the residue limit on equipment was also changed from (b) (4) ppm to (b) (4) ppm. The reason for not performing this change using the change control program was not documented. b) Change control CCQ22-17 enacted to address change in storage of (b) (4) medium used by QC from refrigeration to room temperature states that …

    Contains “test method*”Read the Form 483 →

  11. Wickliffe LLCInspection ended Sep 15, 2023 · FEI 3002992930 · Observation 4

    Acceptance criteria for the sampling and testing conducted by the quality control unit is not adequate to assure that batches of drug products meet each appropriate specification as a condition for their approval and release. Specifically, Your firm only tests potency of a product the first time it is compounded. Once the first batch is successfully produced for potency then all subsequent batches are not tested for potency. On 5/24/2023, your …

    Opens with FDA's citation wording for 21 CFR 211.165(d)Read the Form 483 →

  12. Wickliffe LLCInspection ended Sep 15, 2023 · FEI 3002992930 · Observation 6

    There is no written testing program designed to assess the stability characteristics of drug products. Specifically, Batches are not put on stability that have been prescribed for office stock. Products prescribed for office stock include Doxycycline Hyclate in Oil Suspension 500mg/mL and Methimazole Suspension 2.5mg/mL. Potency over time or (b) (4) is tested for new sterile products, however, other drug quality attributes are not evaluated under …

    Opens with FDA's citation wording for 21 CFR 211.166(a)Read the Form 483 →

  13. Samsung Biologics Co. Ltd.Inspection ended Sep 1, 2023 · FEI 3010479596 · Observation 1

    … You received observations and recommendations from a client audit 06 July 2022 and 01 September 2022 specific to the MSAT Laboratory data management practices, with Samsung senior executive tour of the MSAT Laboratory 14 September 2022 identifying sticky notes with system password written on PCs, use of shared administration account, no label on chemical solutions, uncontrolled spreadsheets used to track samples, use of uncontrolled sheets to …

    Contains “laboratory data”Read the Form 483 →

  14. Samsung Biologics Co. Ltd.Inspection ended Sep 1, 2023 · FEI 3010479596 · Observation 4

    … -HPLC ID 150-HPLC-001: Was due for revalidation (b) (4) from the last revalidation which was performed on 20-Aug-2023. However, this equipment was observed being used past its revalidation due date. -HPLC 150-HPLC-006: Was due for revalidation (b) (4) from the last revalidation which was performed on 16-Aug-2023. However, this equipment was observed being used past its revalidation due date. -ELISA Binding ID 150-FACS-003: Was due for …

    Contains “HPLC*”Read the Form 483 →

  15. Samsung Biologics Co. Ltd.Inspection ended Sep 1, 2023 · FEI 3010479596 · Observation 5

    … In addition, laboratory investigations for SST failures did not identify abnormalities using PEMME (people, equipment, method, material, environment) tool and investigations concluded that invalidation rate will be monitored continuously since root cause of invalid runs was not identified. This is a repeat observation as high invalid rates were previously observed during FDA inspection of the facility in 2022 and corrective actions implemented by …

    Contains “laboratory investigation*”Read the Form 483 →

  16. Jiangsu Hengrui Pharmaceuticals CO. LTD.Inspection ended Aug 29, 2023 · FEI 3003404148 · Observation 1

    Samples taken of in-process materials for determination of conformance to specifications are not representative. Specifically, samples are taken for testing bioburden, appearance, and assay after (b) (4) compounding of (b) (4) Injection, a sterile (b) (4) product distributed to the U.S. market. The sample obtained is not the (b) (4) bulk solution. The compounded solution is transported from Workshop (b) (4) room (b) (4) to the Workshop (b) (4) …

    Opens with FDA's citation wording for 21 CFR 211.160(b)(2)Read the Form 483 →

  17. Jiangsu Hengrui Pharmaceuticals CO. LTD.Inspection ended Aug 29, 2023 · FEI 3003404148 · Observation 2

    Laboratory controls do not include the establishment of scientifically sound and appropriate test procedures designed to assure that drug products conform to appropriate standards of identity, strength, quality and purity. Specifically, a) SOP QC-868-009, The Determination of Microbiological Experiment Result, does not assure that the (b) (4) will not be disrupted during the (b) (4) reads of the sterility test samples. The procedure requires, …

    Opens with FDA's citation wording for 21 CFR 211.160(b)Read the Form 483 →

  18. Jiangsu Hengrui Pharmaceuticals CO. LTD.Inspection ended Aug 29, 2023 · FEI 3003404148 · Observation 7

    Established test procedures are not documented at the time of performance. Specifically, after incubation is completed of environmental monitoring samples and bioburden test samples, the first microbiologist performs the reads of all (b) (4) plates without documenting the results. Only after the second microbiologist performs review of the (b) (4) plates and an agreement is reach on the read result, is the results documented by the first …

    Opens with FDA's citation wording for 21 CFR 211.160(a)Read the Form 483 →

  19. Novavax CZInspection ended Jul 18, 2023 · FEI 3017847976 · Observation 2

    There is a lack of data to support the use of (b) (4) reference standard (b) (4) for the potency determination of (b) (4) batches for the (b) (4) and (b) (4) by the (b) (4) method.

    Contains “reference standard*”Read the Form 483 →

  20. UCB Farchim SAInspection ended Jul 4, 2023 · FEI 3005023799 · Observation 1

    Quality unit oversight over quality control operations is deficient, including laboratory and productions electronic systems and data review. A. During my review of your Empower 3 chromatography software, interrupted sequences were observed, which generated "Data Incomplete" and "Missing Data" chromatographic data. As a result, the interrupted sample injections were not adequately documented to have occurred within your quality control system, …

    Contains “chromatograph*”Read the Form 483 →

21 CFR sections these observations open with

The sections FDAgent ties the most of these observations to by FDA's wording, with how many of this theme's observations each covers. Each links to its page of FDA citations, warning letters and the regulation text.

Ask FDAgent about laboratory controls in Form 483s

FDAgent reads FDA's Form 483s, inspections, citations and warning letters and answers in plain English, with links to the records.

Edit it or ask as written. 3 questions free, no card needed.

About this data

  • The observations are the text FDAgent read by OCR from Form 483 PDFs in FDA's Electronic Reading Room, shown only from 483s whose every printed observation passed FDAgent's checks against the scanned text, as on each 483's own page.
  • Which theme an observation belongs to is decided by the fixed rules above; every count on this page is computed from them.
  • A 483 addressed to an individual is listed without the name, and an observation that names a person is left out.
  • FDAgent is independent of FDA.