Skip to content

NOVAVAX CZ a.s.

Libalova 2348/1, Prague, Czech Republic

FEI 3017847976

FDA has one inspection on file for NOVAVAX CZ a.s. in Prague, Czech Republic (FEI 3017847976). It ended on and was classified VAI, voluntary action indicated.

VAI means FDA investigators found objectionable conditions or practices, but FDA was not prepared to take or recommend regulatory action. The firm is expected to correct them voluntarily.

FDA posted the Form 483 from that inspection. FDAgent read its 6 observations from the PDF; they are listed below. One FOIA request has asked FDA for this inspection's 483.

No warning letter, court action or import refusal is filed under this FEI.

Sources:FDA firm profile ↗Form 483 PDF ↗Synced Nov 23, 2026

Ask FDAgent about this site

Free account: 3 questions, no card needed. Each answer links to its FDA sources.

Email me when FDA posts a new inspection, Form 483 or warning letter for this site

Weekly, on a free account. Pause it any time.

Watch this facility

Comparing suppliers? Check this site beside 2 others, free

Latest inspection
VAI
Inspections on file
1, in 2023
Inspections with 483 observations shown
1 of 1
Form 483s posted by FDA
↗
Requested under FOIA
1 request, latest Apr 24, 2025
Warning letters
None on file
Import refusals
None under this FEI
Import alerts
Not linked by FEI
FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗

Inspection history

1 inspection · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC

  1. VAIObjectionable conditions found; FDA took no regulatory actionBiologics · Vaccines and Allergenic ProductsPosted 483 · 6 observations

    Read this Form 483 on its own page →Form 483 PDF on fda.gov

    Read from the FDA-posted Form 483 PDF by OCR. FDAgent keeps an observation only when it matches the scanned text; check the PDF before quoting. The short titles are FDAgent’s.

    1. Observation 1

      Inadequate validation for maximum reuse cycles

      There is inadequate data in the (b) (4) and (b) (4) validation report approved on April 24, 2023 to support the maximum proposed number of (b) (4) uses for the (b) (4) and (b) (4) uses for the (b) (4) used for the (b) (4) of (b) (4) The deficiencies are as follows: a. The (b) (4) for the (b) (4) was evaluated only up to (b) (4) uses of (b) (4) b. The critical quality attributes such as (b) (4), (b) (4) and (b) (4) were not evaluated as In-Process Controls for each batch after (b) (4) and only (b) (4) was evaluated for each batch after (b) (4) (b) (4) c. The (b) (4) and (b) (4) were not evaluated during the (b) (4) of the (b) (4) and (b) (4)

    2. Observation 2

      Lack of data supporting reference standard use

      There is a lack of data to support the use of (b) (4) reference standard (b) (4) for the potency determination of (b) (4) batches for the (b) (4) and (b) (4) by the (b) (4) method.

    Show observations 3 to 6
    1. Observation 3

      Unvalidated maximum holding times for intermediates

      The proposed maximum holding times for the process intermediates including (b) (4) (b) (4) (b) (4) (b) (4) (b) (4) (b) (4) (b) (4) (b) (4) and (b) (4) have not been validated for the (b) (4) manufacturing process at production scale.

    2. Observation 4

      Disinfectant studies lacked facility environmental isolates

      Vendor performed Disinfectant efficacy studies failed to include environmental isolates from the facility. Firm provided a disinfectant effective study protocol (b) (4) that will include facility isolates for the study.

    3. Observation 5

      Untimely CAPA for repeated integrity failures

      Corrective action was not always taken in a timely manner to prevent the repeated recurrence for (b) (4) integrity failures, for example: (b) (4) used in production process and equipment are currently visually inspected (b) (4); however, current visual inspection procedure had failed to prevent multiple (b) (4) integrity failures (A total of 25 deviations related to (b) (4) integrity test failures were reported from August 2021 to April 2023), and CAPA to introduce pre-use (b) (4) integrity test has not been introduced in a timely manner to prevent the re-occurrence of integrity failures.

    4. Observation 6

      Deficient procedures lacking time limits and validation

      Current procedures for (b) (4) and (b) (4) of (b) (4) are deficient. For example: No (b) (4) time limit is specified in the procedures for time between (b) (4) product removed from (b) (4) and (b) (4) or (b) (4) In addition, (b) (4) validation to (b) (4) is pending for concurrent validation.

Requested under FOIA

FDA FOIA logs, last synced Nov 23, 2026, 04:25 UTC

One request to FDA under the Freedom of Information Act named this site, on Apr 24, 2025. It asked for a Form 483.

  • Inspection ending Jul 18, 2023VAI

    1 request. In the inspection history above. FDA has posted its Form 483.

    1. Apr 24, 2025FDA-FOIA-2025-3689Form 483· 1 of 3 sitesRecords released

A request is listed when it names this site’s FEI number or one of its FDA inspection IDs, or, after review, its exact street address. Requester names are left out.

Go further than this page

FDAgent researches FDA’s records and links every answer to its source.

Free account: 3 questions, no card needed.

About this data. Every record here is filed under FEI 3017847976 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
  • Every record on this page is filed under FEI 3017847976. Records filed under another FEI or name, including other sites of the same company, are not shown here.
  • FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
  • Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
  • “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
  • Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
  • FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.