Novavax CZ Form 483, July 2023
FDA issued this Form 483 to Novavax CZ in Prague, Czech Republic (FEI 3017847976) at the end of a biologics inspection that ended July 18, 2023. It lists 6 observations. FDA classified the inspection Voluntary Action Indicated (VAI).
A Form 483 lists the investigators' observations at the end of an inspection. It is not FDA's final determination.
Source: FDA's Electronic Reading Room, posted Nov 17, 2026. Form 483 PDF on fda.gov. FDAgent is independent of FDA.
- Inspection ended
- Jul 18, 2023
- Biologics · Vaccines and Allergenic Products
- FDA's outcome
- VAI
- Voluntary Action Indicated
- Posted by FDA
- Nov 17, 2026
- 1,218 days after the inspection
The 6 observations on this Form 483
Read from the FDA-posted Form 483 PDF by OCR. FDAgent keeps an observation only when it matches the scanned text; check the PDF before quoting. The short titles are FDAgent's.
- 1
Inadequate validation for maximum reuse cycles
There is inadequate data in the (b) (4) and (b) (4) validation report approved on April 24, 2023 to support the maximum proposed number of (b) (4) uses for the (b) (4) and (b) (4) uses for the (b) (4) used for the (b) (4) of (b) (4) The deficiencies are as follows: a. The (b) (4) for the (b) (4) was evaluated only up to (b) (4) uses of (b) (4) b. The critical quality attributes such as (b) (4), (b) (4) and (b) (4) were not evaluated as In-Process Controls for each batch after (b) (4) and only (b) (4) was evaluated for each batch after (b) (4) (b) (4) c. The (b) (4) and (b) (4) were not evaluated during the (b) (4) of the (b) (4) and (b) (4)
- 2
Lack of data supporting reference standard use
There is a lack of data to support the use of (b) (4) reference standard (b) (4) for the potency determination of (b) (4) batches for the (b) (4) and (b) (4) by the (b) (4) method.
- 3
Unvalidated maximum holding times for intermediates
The proposed maximum holding times for the process intermediates including (b) (4) (b) (4) (b) (4) (b) (4) (b) (4) (b) (4) (b) (4) (b) (4) and (b) (4) have not been validated for the (b) (4) manufacturing process at production scale.
- 4
Disinfectant studies lacked facility environmental isolates
Vendor performed Disinfectant efficacy studies failed to include environmental isolates from the facility. Firm provided a disinfectant effective study protocol (b) (4) that will include facility isolates for the study.
- 5
Untimely CAPA for repeated integrity failures
Corrective action was not always taken in a timely manner to prevent the repeated recurrence for (b) (4) integrity failures, for example: (b) (4) used in production process and equipment are currently visually inspected (b) (4); however, current visual inspection procedure had failed to prevent multiple (b) (4) integrity failures (A total of 25 deviations related to (b) (4) integrity test failures were reported from August 2021 to April 2023), and CAPA to introduce pre-use (b) (4) integrity test has not been introduced in a timely manner to prevent the re-occurrence of integrity failures.
- 6
Deficient procedures lacking time limits and validation
Current procedures for (b) (4) and (b) (4) of (b) (4) are deficient. For example: No (b) (4) time limit is specified in the procedures for time between (b) (4) product removed from (b) (4) and (b) (4) or (b) (4) In addition, (b) (4) validation to (b) (4) is pending for concurrent validation.
Ask FDAgent about this Form 483
FDAgent reads the 483, the facility's inspections, warning letters and FDA's citations and answers in plain English, with links to the records.
Related
About this data
- The PDF and its dates come from FDA's Electronic Reading Room, last synced Nov 23, 2026, 04:33 UTC; the inspection, its outcome and FDA's citations from FDA's Data Dashboard.
- The observations are the text FDAgent read from this PDF by OCR, shown only when every printed observation passed FDAgent’s checks against the scanned text.
- FDAgent leaves out the names of the people the Form 483 is addressed to and of the investigators.
- FDAgent is independent of FDA.