Laboratory controls in FDA Form 483 observations
Quality control laboratories: test methods and their validation, laboratory records and equipment, reference standards and chromatography. FDAgent finds 304 such observations in 172 Form 483s issued to 163 facilities, for inspections that ended between October 2008 and September 2026. Each is quoted below from the 483 FDA posted.
A Form 483 lists the investigators' observations at the end of an inspection. It is not FDA's final determination.
Source: Form 483 PDFs in FDA's Electronic Reading Room, last synced Nov 23, 2026, 04:33 UTC. Every quote is from a 483 whose reading passed FDAgent's checks against the scan. FDAgent is independent of FDA.
- Observations
- 304
- Each quoted below
- Form 483s
- 172
- Posted by FDA
- Facilities
- 163
- Distinct FEIs
- Most recent
- Sep 3, 2026
- Inspection end date
Observations by year
These observations by the year the inspection ended. They follow which 483s FDA has posted and FDAgent has read, so they show what is on record, not a trend in FDA's findings. 2026 runs to the latest sync (lighter bar).
Peak: 76 in 2026
| Year | Observations |
|---|---|
| 2008 | 1 |
| 2009 | 0 |
| 2010 | 0 |
| 2011 | 0 |
| 2012 | 8 |
| 2013 | 0 |
| 2014 | 5 |
| 2015 | 4 |
| 2016 | 13 |
| 2017 | 21 |
| 2018 | 23 |
| 2019 | 14 |
| 2020 | 0 |
| 2021 | 6 |
| 2022 | 18 |
| 2023 | 42 |
| 2024 | 25 |
| 2025 | 48 |
| 2026 | 76 |
How FDAgent sorts observations into this theme
Two fixed rules, applied by code to every observation FDAgent can quote. An observation counts when either holds. No model and no similarity score decides.
It is tied to one of these 21 CFR entries
21 CFR 211.16021 CFR 211.16521 CFR 211.16621 CFR 211.16721 CFR 211.17021 CFR 211.17321 CFR 211.176
The same entries the laboratory controls topic counts. A 483 prints no regulation, so FDAgent ties an observation to a section only when it opens with FDA's own citation wording for that section, word for word.
Or its text contains one of these phrases
“laboratory control*” · “laboratory record*” · “laboratory investigation*” · “laboratory equipment” · “laboratory instrument*” · “laboratory data” · “laboratory test*” · “laboratory result*” · “laboratory analyst*” · “laboratory procedure*” · “laboratory notebook*” · “laboratory error*” · “QC laborator*” · “quality control laborator*” · “test method*” · “analytical method*” · “analytical procedure*” · “method validation” · “HPLC*” · “chromatograph*” · “reference standard*” · “system suitability” · “method transfer”
* matches any ending and … up to three words in between. Case is ignored unless a phrase has capitals.
The 304 observations, newest first
Quoted word for word from each Form 483, with only line breaks joined; … marks where the quote is cut. The highlight shows why the observation is in this theme. Each links to FDAgent's page for the 483.
Catalent Indiana LLCInspection ended Sep 1, 2022 · FEI 3005949964 · Observation 12
Your firm failed to establish laboratory controls that include scientifically sound and appropriate specifications, standards, sampling plans, and test procedures designed to assure that components, drug product containers, closures, in-process materials, labeling, and drug products conform to appropriate standards of identity, strength, quality, and purity. A. The verification of method AM-(b)(4) 100-018 Supplemental Destructive Visual …
Contains “laboratory control*”Read the Form 483 →
AnazaoHealth CorporationInspection ended Jul 29, 2022 · FEI 3011152407 · Observation 4
The accuracy, sensitivity, specificity, and reproducibility of test methods have not been established and documented. Specifically, on May 19, 2022, your contract testing laboratory detected visual turbidity in the sterility testing (b) (4) for your (b) (4) Nutrient Cocktail lot (b) (4), and confirmed that this turbidity was caused by microbial contamination with Staphylococcus epidermidis. However, your contract testing laboratory failed to …
Opens with FDA's citation wording for 21 CFR 211.165(e)Read the Form 483 →
Medlmmune UK Limited (AstraZeneca/Supply Biologics)Inspection ended Jun 14, 2022 · FEI 3004066112 · Observation 2
The bacterial endotoxin procedure is deficient to ensure adequate detection of endotoxin in the samples. A. Analytical method used to detect bacterial endotoxin in LAIV Monovalent Bulk (MVB), and LAIV Finished Product are inadequate. Specifically, SOP-0067568, entitled "The Endotoxin Testing of Product by [illegible]", v40.0 is deficient in that the mixing and subsequent preparations are not described. Without sample requirement for the qualified …
Contains “analytical method*”Read the Form 483 →
Serum Institute of India Pvt. Ltd.Inspection ended Apr 19, 2022 · FEI 3020223359 · Observation 3
There is no data to demonstrate that the potency of the (b)(4) reference standard Lot # [illegible] used for potency determination of [illegible] and DP has no significant change (b)(4) reference between the initial qualification performed in September 2021 and the re-qualification of the standard performed in January 2022.
Contains “reference standard*”Read the Form 483 →
Cadila Healthcare LimitedInspection ended Mar 10, 2022 · FEI 3013712903 · Observation 1
OBSERVATION 1 The written stability testing program is not followed. Specifically, the stability program is not followed to complete the testing within predefined time frame. The Quality Unit failed to test about 250 stability samples within [illegible] days as required under, "SOP for Stability Management (Document# 0317-SOP-QC-00132, Effective Date: 7/23/2021)." Examples of stability samples that were delayed testing include: Product/Material …
Opens with FDA's citation wording for 21 CFR 211.166(a)Read the Form 483 →
Davis City Pharmacy Inc.Inspection ended Feb 18, 2022 · FEI 3016774626 · Observation 3
Testing and release of drug product for distribution do not include appropriate laboratory determination of satisfactory conformance to the final specifications prior to release. Specifically, Your firm failed to conduct appropriate tests to determine the identity and strength of each active ingredient in the following non-sterile drug products prior to release and distribution: o RX[(B)(B): Benzocaine 20%, Lidocaine 10%, Tetracaine 10% Lot …
Opens with FDA's citation wording for 21 CFR 211.165(a)Read the Form 483 →
Davis City Pharmacy Inc.Inspection ended Feb 18, 2022 · FEI 3016774626 · Observation 4
Laboratory controls do not include the establishment of scientifically sound and appropriate specifications designed to assure that drug products conform to appropriate standards of identity, strength, quality and purity. Specifically, Your firm failed to establish microbial specifications for your finished non-sterile compounded drug products to assure that it meets the appropriate standards for microbial quality. The firm failed to perform …
Opens with FDA's citation wording for 21 CFR 211.160(b)Read the Form 483 →
Bioiberica, S.A.UInspection ended Feb 4, 2022 · FEI 1000418405 · Observation 3
The quality control unit lacks responsibility to approve all procedures which assure strength, quality, and purity of drug substances. Specifically, A. During my review of electronic HPLC data for assay testing, [illegible] encountered during analysis of drug substance, samples are re-tested until desirable results are achieved. Original test results were not reported, and no laboratory investigation procedure. Specifically during review of your …
Contains “HPLC*”Read the Form 483 →
Bioiberica, S.A.UInspection ended Feb 4, 2022 · FEI 1000418405 · Observation 4
There is no written testing program designed to assess the stability characteristics [illegible]. Specifically, the assay test method for active ingredients [illegible] substance is not stability indicative (b)(4) [illegible] it can detect the change with [illegible]. Microbiological properties of the (b)(4) USP drug substance are specific so that the content and degradation products can be accurately measured without interference. As a result, …
Contains “test method*”Read the Form 483 →
Dr. Reddy's Laboratories Ltd.Inspection ended Oct 29, 2021 · FEI 3006549835 · Observation 1
… The reports w The SAP system does not include 2) The ma=i~.i~~ ~£.i~b.;rmacopoe 11 1 for the SP test method of pharmacopoeia updates procedure pharmacopeia/ supplements, not implemented after shall be routed through current version of USP test methods the USP monograph. The - was updated in December testing and release of drug There is no incident [illegible] monograph.
Contains “test method*”Read the Form 483 →
Dr. Reddy's Laboratories Ltd.Inspection ended Oct 29, 2021 · FEI 3006549835 · Observation 2
Laboratory controls do not include the establishment specifications, sampling plans and test procedures designed to assure that components, in-process materials and drug products conform to appropriate standards. Specifically, the drug substance and drug product plans and test methods. The following examples are for the Drug product, B. The viscosity specification for equipment not being capable of producing precise in-process viscosity failures …
Contains “laboratory control*”Read the Form 483 →
Fujifilm Diosynth Biotechnologies USAInspection ended Apr 21, 2021 · FEI 3000718852 · Observation 5
… At least ten (10) Endotoxin testing results did not meet the Endotoxin System Suitability Test. You failed to conduct adequate investigations to identify the root causes. ii. Under Protocol Event for (b) (4) three (3) Endotoxin Out of Specification results for in-process hold time studies were reported (PR206361, PR206365, and PR307208); however, no definitive root causes were identified. The firm concluded that the OOS results were "aberrant …
Contains “system suitability”Read the Form 483 →
Wuxi BiologicsInspection ended Apr 2, 2021 · FEI 3010606982 · Observation 2
… For example: a) On 03/22/2021, Form WX-FORM-000970-09, Registration Form of Samples for Analyses, was observed in the Quality Control laboratory. A review of your documentation controls revealed this controlled record is issued to Quality Control by your Quality Assurance department, however, there is not an adequate reconciliation plan in place. b) On 03/22/2021, Form WX-FORM-002218-02, Incoming Material Checklist, was observed in the Warehouse …
Contains “quality control laborator*”Read the Form 483 →
Wuxi BiologicsInspection ended Apr 2, 2021 · FEI 3010606982 · Observation 4
The specifications established for components used in the production of drug substance are deficient in that they do not include a description of the sampling plan. Specifically, raw material specifications do not provide sufficient instructions to ensure samples collected are representative and reproducible. For example, the raw material specifications for (b) (4) chromatography media (Document WX-SPEC-000233), (b) (4) (Document WX-SPEC-000234), …
Contains “chromatograph*”Read the Form 483 →
Imprimis NJOF, -LLCInspection ended Jan 29, 2021 · FEI 3013024146 · Observation 7
(REPEAT OBSERVATION) Laboratory controls do not include the establishment of scientifically sound and appropriate sampling plans designed to assure that drug products conform to appropriate standards of identity, strength, quality and purity. Specifically, the number of units analyzed for endotoxin testing is not representative of the entire commercial batch size. Four examples of compounded sterile drug products are listed below: a) Moxifloxacin …
Contains “laboratory control*”Read the Form 483 →
SCA Pharmaceuticals, Inc.Inspection ended Nov 7, 2019 · FEI 3010683157 · Observation 1
Each batch of drug product required to be free of objectionable microorganisms is not tested through appropriate laboratory testing. Specifically, your method allows for [(o) (4) with the (o) (4) for sterility testing which is not equivalent to USP <71>.
Opens with FDA's citation wording for 21 CFR 211.165(b)Read the Form 483 →
Aurobindo Pharma Ltd. Unit 5Inspection ended Oct 28, 2019 · FEI 3004051616 · Observation 1
Laboratory controls do not always include the establishment of scientifically sound and appropriate procedures designed to assure that drug substances conform to appropriate standards of identity, strength, quality and purity. Specifically, when there is no scientific basis found for invalidating initial out-of-specification (OOS) API test results and no assignable cause for the OOS is determined, procedures (pZQA005) allow re-testing (following …
Contains “laboratory control*”Read the Form 483 →
Rapid Equine Solutions, LLCInspection ended Aug 9, 2019 · FEI 3006708880 · Observation 1
Testing and release of drug product for distribution do not include appropriate laboratory determination of satisfactory conformance to the identity and strength of each active ingredient prior to release. Specifically, your firm did not perform testing to confirm the identify and strength of drug products. For example, the firm produced Toltrazuril Pyrimethamine Paste, Lot#20190509-11, BUD 11/05/2019 on 05/09/2019, of which (b) (4) of (b) (4) …
Opens with FDA's citation wording for 21 CFR 211.165(a)Read the Form 483 →
Rapid Equine Solutions, LLCInspection ended Aug 9, 2019 · FEI 3006708880 · Observation 8
Procedures describing the calibration of instruments, apparatus, gauges and recording devices are not written or followed. Specifically, review of the Certificate of Analysis for (b) (4) Cartridges lot# (b) (4) revealed that the firm is not conducting cartridge calibration per SOP P-604 Testing Via the (b) (4) System, v0 effective 2/25/2015, section 8.3.1.2. The following drug products Folic Acid 10 mg/mL (b) (4) Lot#20190722-1, BUD: 01/18/2020, …
Opens with FDA's citation wording for 21 CFR 211.160(b)(4)Read the Form 483 →
Southern California Compounding Pharmacy, LLCInspection ended Aug 1, 2019 · FEI 3010006900 · Observation 5
Laboratory controls do not include the establishment of scientifically sound and appropriate specifications, sampling plans, and test procedures designed to assure drug products conform to appropriate standards of identity, strength, quality and purity. Specifically, finished non-sterile drug products are not tested for the presence of microorganisms, for example: Ketoprofen 20%/Lidocaine 10%, lot no. 07419K20L10A, submitted for [illegible] …
Contains “laboratory control*”Read the Form 483 →
21 CFR sections these observations open with
The sections FDAgent ties the most of these observations to by FDA's wording, with how many of this theme's observations each covers. Each links to its page of FDA citations, warning letters and the regulation text.
- 21 CFR 211.160General requirementsIn this theme's rule40
- 21 CFR 211.165Testing and release for distributionIn this theme's rule24
- 21 CFR 211.166Stability testingIn this theme's rule21
- 21 CFR 211.194Laboratory records17
- 21 CFR 211.192Production record review15
- 21 CFR 211.22Responsibilities of quality control unit12
- 21 CFR 211.167Special testing requirementsIn this theme's rule6
- 21 CFR 211.170Reserve samplesIn this theme's rule6
- 21 CFR 211.68Automatic, mechanical, and electronic equipment5
- 21 CFR 211.110Sampling and testing of in-process materials and drug products3
Ask FDAgent about laboratory controls in Form 483s
FDAgent reads FDA's Form 483s, inspections, citations and warning letters and answers in plain English, with links to the records.
Related
- Laboratory controls: FDA's Form 483 citations of the same 21 CFR entries
- Aseptic processing: FDA's Form 483 citations of overlapping 21 CFR entries
- Stability: FDA's Form 483 citations of overlapping 21 CFR entries
- Every Form 483 theme, with its counts
- What a Form 483 is, and the latest 483s
- Every 21 CFR section in FDA's Form 483 citations
About this data
- The observations are the text FDAgent read by OCR from Form 483 PDFs in FDA's Electronic Reading Room, shown only from 483s whose every printed observation passed FDAgent's checks against the scanned text, as on each 483's own page.
- Which theme an observation belongs to is decided by the fixed rules above; every count on this page is computed from them.
- A 483 addressed to an individual is listed without the name, and an observation that names a person is left out.
- FDAgent is independent of FDA.