Laboratory controls in FDA Form 483 observations
Quality control laboratories: test methods and their validation, laboratory records and equipment, reference standards and chromatography. FDAgent finds 304 such observations in 172 Form 483s issued to 163 facilities, for inspections that ended between October 2008 and September 2026. Each is quoted below from the 483 FDA posted.
A Form 483 lists the investigators' observations at the end of an inspection. It is not FDA's final determination.
Source: Form 483 PDFs in FDA's Electronic Reading Room, last synced Nov 23, 2026, 04:33 UTC. Every quote is from a 483 whose reading passed FDAgent's checks against the scan. FDAgent is independent of FDA.
- Observations
- 304
- Each quoted below
- Form 483s
- 172
- Posted by FDA
- Facilities
- 163
- Distinct FEIs
- Most recent
- Sep 3, 2026
- Inspection end date
Observations by year
These observations by the year the inspection ended. They follow which 483s FDA has posted and FDAgent has read, so they show what is on record, not a trend in FDA's findings. 2026 runs to the latest sync (lighter bar).
Peak: 76 in 2026
| Year | Observations |
|---|---|
| 2008 | 1 |
| 2009 | 0 |
| 2010 | 0 |
| 2011 | 0 |
| 2012 | 8 |
| 2013 | 0 |
| 2014 | 5 |
| 2015 | 4 |
| 2016 | 13 |
| 2017 | 21 |
| 2018 | 23 |
| 2019 | 14 |
| 2020 | 0 |
| 2021 | 6 |
| 2022 | 18 |
| 2023 | 42 |
| 2024 | 25 |
| 2025 | 48 |
| 2026 | 76 |
How FDAgent sorts observations into this theme
Two fixed rules, applied by code to every observation FDAgent can quote. An observation counts when either holds. No model and no similarity score decides.
It is tied to one of these 21 CFR entries
21 CFR 211.16021 CFR 211.16521 CFR 211.16621 CFR 211.16721 CFR 211.17021 CFR 211.17321 CFR 211.176
The same entries the laboratory controls topic counts. A 483 prints no regulation, so FDAgent ties an observation to a section only when it opens with FDA's own citation wording for that section, word for word.
Or its text contains one of these phrases
“laboratory control*” · “laboratory record*” · “laboratory investigation*” · “laboratory equipment” · “laboratory instrument*” · “laboratory data” · “laboratory test*” · “laboratory result*” · “laboratory analyst*” · “laboratory procedure*” · “laboratory notebook*” · “laboratory error*” · “QC laborator*” · “quality control laborator*” · “test method*” · “analytical method*” · “analytical procedure*” · “method validation” · “HPLC*” · “chromatograph*” · “reference standard*” · “system suitability” · “method transfer”
* matches any ending and … up to three words in between. Case is ignored unless a phrase has capitals.
The 304 observations, newest first
Quoted word for word from each Form 483, with only line breaks joined; … marks where the quote is cut. The highlight shows why the observation is in this theme. Each links to FDAgent's page for the 483.
Cipla Ltd. Virgonagar SiteInspection ended Jul 19, 2019 · FEI 3002806691 · Observation 1
… Review of the overlays of the chromatographic data for 0 - 36 months stability testing for Lots: (6) 4) and (0) (4) shows that the main (0)(4) peak shifts from around (b)' minutes to around (8) © minutes for both lots during the 0 - 36 month period. However, you do not evaluate why the main (©)(4) peak shifts significantly between stability pulls, or compare it to the USP reference standard. Note that these lots of (b) (4) USP are validation …
Contains “chromatograph*”Read the Form 483 →
Sun Pharmaceutical Industries Ltd.Inspection ended Jun 11, 2019 · FEI 3002809586 · Observation 2
OBSERVATION 2 Established laboratory control mechanisms are not documented at the time of performance. The assurance of the data from the all analytical testing data is not provided. Specifically, Environmental monitoring of plates on June 5, 2019 included review of more than [illegible] sample sets. In my review of the data in support of the EM review, it was noted that in selected examples, the primary reviewer performed the plate analysis …
Opens with FDA's citation wording for 21 CFR 211.160(a)Read the Form 483 →
Torrent Pharmaceuticals LimitedInspection ended Apr 16, 2019 · FEI 3005029956 · Observation 1
… The investigation could not conclusively identify if there is any laboratory error or manufacturing error and no CAPA was initiated. The (b)(4) was released using re-tested reserve sample set results. (b)(4) was released to U.S. market with an expiry date of (b)(4). In addition, the documents provided to the investigators in support of OOS Investigation Number OOS/IN/F/FP/17/233 included an unsigned OOS Amendment Report. This document had a print …
Contains “laboratory error*”Read the Form 483 →
Torrent Pharmaceuticals LimitedInspection ended Apr 16, 2019 · FEI 3005029956 · Observation 4
… Furthermore, since the method validation for solution stability was deficiently established only for (b)(4), a second set of samples was analyzed in duplicate to invalidate the original failing results. Passing results from initial sequence for other batches were also discarded, and re-tested results were reported without a valid justification. No consideration was given to report the average of all passing results when a definitive root cause is …
Contains “method validation”Read the Form 483 →
Front Door Pharmacy, LLC dba Pure PharmacyInspection ended Mar 18, 2019 · FEI 3012937475 · Observation 8
Laboratory test equipment used in the preparation and testing of collected samples are inadequate. Specifically, your firm uses a [redacted model] for the incubation of collected media-fills. The unit fails to have adequate temperature and related humidity controls to perform and control these environmental conditions.
Contains “laboratory test*”Read the Form 483 →
Lupin LimitedInspection ended Feb 8, 2019 · FEI 3004819820 · Observation 1
… CQA-004-17 “Handling of out of Specification test results” Section 5.1.12 (a) states — ‘Retesting shall be done by BE analysis following the approved laboratory procedure preferably by original analyst from the portion of sample available in the laboratory’. The procedure lacks statistical justification to release the batch based on retest data. a) OOS/C/18/GA/FP/042 was initiated on April 09, 2018 to probe the failure in assay for © @ in [BNE …
Contains “laboratory procedure*”Read the Form 483 →
Daiichi Sankyo Chemical Pharma Co., Ltd.Inspection ended Jan 25, 2019 · FEI 3003279188 · Observation 1
Appropriate documentation of testing performed to fully establish the identity and purity of the in-house primary reference standard was not retained. Specifically, Your firm does not have complete documentation to verify the GMP status of the Sankyo Investigational Medicinal Product GMP laboratory equipment used to conduct the full characterization of the primary reference standard for ^(b) (4) lot ^(b) (4) ^(b) (4) in-house reference standard …
Contains “reference standard*”Read the Form 483 →
Daiichi Sankyo Chemical Pharma Co., Ltd.Inspection ended Jan 25, 2019 · FEI 3003279188 · Observation 2
Suitability of each batch of in-house secondary reference standard is not determined by comparison against a primary reference standard. Specifically, Since 2004, ^(b) (4) in-house reference standard lots have not been fully characterized or compared against a fully characterized primary reference standard. ^(b) (4) in-house reference standard is used to conduct finished product release testing of ^(b) (4) drug substance.
Contains “reference standard*”Read the Form 483 →
Daiichi Sankyo Chemical Pharma Co., Ltd.Inspection ended Jan 25, 2019 · FEI 3003279188 · Observation 4
… The investigation did not include analytical methods to identify or verify the identity of the material as glass and the investigation did not evaluate possible sources of contamination other than the production equipment. The investigation was closed without identifying a root cause and it was determined there was no impact to product quality. 2) Deviation Number COMD2018-007 - The ^(b) (4) (W26) in the F-10 Freezer, was not appropriately …
Contains “analytical method*”Read the Form 483 →
Dishman Carbogen Amcis Ltd.Inspection ended Oct 26, 2018 · FEI 3004161218 · Observation 2
LABORATORY CONTROL SYSTEM 2.) Laboratory investigations for OOS intermediate batches are not conducted. Specifically, according to SOP BDQC-311 for OOS investigations (effective 07/16/2018), OOS is to be conducted on "...all quantitative and qualitative tests of...intermediates..." An investigation of an OOS impurity result in the [illegible] intermediate batch or [illegible] was not conducted.
Contains “laboratory control*”Read the Form 483 →
Lupin Limited Unit IIIInspection ended Oct 18, 2018 · FEI 3009107538 · Observation 3
Appropriate controls are not exercised over computerized systems, master production and control records or other records. Specifically, a) control over the computerized system in [illegible] permits modification to the sequences in HPLC and GC. The reviewer does not review the modification to the sequence. Your QC reviewer even stated that they do not investigate modifications to the sequence during normal review and accepts the audit trail entry …
Contains “HPLC*”Read the Form 483 →
Sun Pharmaceutical Industries LtdInspection ended Aug 31, 2018 · FEI 3002809586 · Observation 1
Laboratory controls do not include the establishment of scientifically sound and appropriate test procedures designed to assure that drug products conform to appropriate standards of identity, strength, quality and purity. For example, electronic data for mg exhibit batch for 0-month stability shows: dissolution at buffer stage L1 was conducted on 9/26/15 with results for the sampling time point at 5 hours, to have a minimum of] Blo release, a …
Opens with FDA's citation wording for 21 CFR 211.160(b)Read the Form 483 →
Sun Pharmaceutical Industries LtdInspection ended Aug 31, 2018 · FEI 3002809586 · Observation 3
The written stability program for drug products does not include sample size based on statistical criteria for each attribute examined to assure valid estimates of stability. Specifically, on 8/28/18, we observed unused stability samples stored inside a styrofoam cooler, on top of a stool, located inside stability chamber QCC-380, 25 degrees 60% relative humidity. Your Senior General Manager, Quality Non-Sterile Manufacturing, stated these were …
Opens with FDA's citation wording for 21 CFR 211.166(a)(1)Read the Form 483 →
Auro PharmaciesInspection ended Aug 2, 2018 · FEI 3011893599 · Observation 8
OBSERVATION 8 The written stability testing program is not followed. Specifically, Your firm has not conducted any stability studies to establish Beyond Use Dates (BUD) for your non-sterile drug products. For example, we observed: "TETRACAINE HCL TOPICAL 2% SOLUTION" of lot 180212@21, produced on 02/12/2018 where a BUD of 05/13/2018 was applied. You have no assurance that this BUD is appropriate. This is a repeat observation from the 2015 FDA …
Opens with FDA's citation wording for 21 CFR 211.166(a)Read the Form 483 →
Unichem Laboratories Ltd.Inspection ended Jul 27, 2018 · FEI 3006456888 · Observation 1
… did not completely describe the equipment to be cleaned, procedures, materials, acceptable cleaning levels, parameters to be monitored and controlled, and analytical methods. In addition, stability of the worst case condition defined by the completion of a full campaign of [illegible] batches was not defined, even though finished API has been distributed into the market for years.
Contains “analytical method*”Read the Form 483 →
Unichem Laboratories Ltd.Inspection ended Jul 27, 2018 · FEI 3006456888 · Observation 2
… procedure PPQC141/00, confirmed the company has a policy permitting any response attributed to a blank in HPLC analysis to be [illegible] without confirming the identity of the compound eluted, investigation, and notification of QA when a possible OOS occurs.
Contains “HPLC*”Read the Form 483 →
Unichem Laboratories Ltd.Inspection ended Jul 27, 2018 · FEI 3006456888 · Observation 3
3. Laboratory controls were not followed and documented at the time of performance. Specifically, Review of Unichem Laboratories Ltd. procedures for Media Preparation and Microbial Limits testing do not provide for an assurance of GMP compliance in the following examples: a. Media are not placed in a secure, quarantine state pending growth promotion before being released for use in Microbial Limits testing of raw materials, finished goods or …
Contains “laboratory control*”Read the Form 483 →
Celltrion Inc.Inspection ended Jul 17, 2018 · FEI 3005241015 · Observation 4
Procedures describing the calibration of instruments, apparatus, gauges, and recording devices are deficiently written or followed. Specifically, Your firm does not adequately monitor the incubator rooms used to incubate environmental, media fill, and sterility samples. Your most recent temperature mapping study determined hot and cold spots for the incubators; however your routine temperature monitoring gauge is not located in these areas.
Opens with FDA's citation wording for 21 CFR 211.160(b)(4)Read the Form 483 →
Celltrion Inc.Inspection ended Jul 17, 2018 · FEI 3005241015 · Observation 8
During the review of QC equipment logsheets, we observed your firm failed to document the time and temperature of release, stability, and reference standard samples during transfer between freezers. For example: Samples from freezer LEQ-94297 ([illegible]) located in Suite C2-372, were transferred to LEQ-94201 for Maintenance of Validation (MOV) on June 6, 2018. Stability samples from freezer LEQ-94255 ([illegible]) located in Suite W2-307, were …
Contains “reference standard*”Read the Form 483 →
FH Investments Inc. dba Asteria HealthInspection ended Jul 10, 2018 · FEI 3010348724 · Observation 1
Procedures designed to prevent microbiological contamination of drug products purporting to be sterile do not include adequate validation of the sterilization process. Specifically, during (b) (4) sterilization validation, a. The bioburden and sterility method validation and suitability testing used for dose verification testing were not performed using testosterone and estradiol pellets made by your firm. b. The sterility and bioburden testing …
Contains “method validation”Read the Form 483 →
21 CFR sections these observations open with
The sections FDAgent ties the most of these observations to by FDA's wording, with how many of this theme's observations each covers. Each links to its page of FDA citations, warning letters and the regulation text.
- 21 CFR 211.160General requirementsIn this theme's rule40
- 21 CFR 211.165Testing and release for distributionIn this theme's rule24
- 21 CFR 211.166Stability testingIn this theme's rule21
- 21 CFR 211.194Laboratory records17
- 21 CFR 211.192Production record review15
- 21 CFR 211.22Responsibilities of quality control unit12
- 21 CFR 211.167Special testing requirementsIn this theme's rule6
- 21 CFR 211.170Reserve samplesIn this theme's rule6
- 21 CFR 211.68Automatic, mechanical, and electronic equipment5
- 21 CFR 211.110Sampling and testing of in-process materials and drug products3
Ask FDAgent about laboratory controls in Form 483s
FDAgent reads FDA's Form 483s, inspections, citations and warning letters and answers in plain English, with links to the records.
Related
- Laboratory controls: FDA's Form 483 citations of the same 21 CFR entries
- Aseptic processing: FDA's Form 483 citations of overlapping 21 CFR entries
- Stability: FDA's Form 483 citations of overlapping 21 CFR entries
- Every Form 483 theme, with its counts
- What a Form 483 is, and the latest 483s
- Every 21 CFR section in FDA's Form 483 citations
About this data
- The observations are the text FDAgent read by OCR from Form 483 PDFs in FDA's Electronic Reading Room, shown only from 483s whose every printed observation passed FDAgent's checks against the scanned text, as on each 483's own page.
- Which theme an observation belongs to is decided by the fixed rules above; every count on this page is computed from them.
- A 483 addressed to an individual is listed without the name, and an observation that names a person is left out.
- FDAgent is independent of FDA.