Skip to content

Laboratory controls in FDA Form 483 observations

Quality control laboratories: test methods and their validation, laboratory records and equipment, reference standards and chromatography. FDAgent finds 304 such observations in 172 Form 483s issued to 163 facilities, for inspections that ended between October 2008 and September 2026. Each is quoted below from the 483 FDA posted.

A Form 483 lists the investigators' observations at the end of an inspection. It is not FDA's final determination.

Source: Form 483 PDFs in FDA's Electronic Reading Room, last synced Nov 23, 2026, 04:33 UTC. Every quote is from a 483 whose reading passed FDAgent's checks against the scan. FDAgent is independent of FDA.

Observations
304
Each quoted below
Form 483s
172
Posted by FDA
Facilities
163
Distinct FEIs
Most recent
Sep 3, 2026
Inspection end date

Observations by year

These observations by the year the inspection ended. They follow which 483s FDA has posted and FDAgent has read, so they show what is on record, not a trend in FDA's findings. 2026 runs to the latest sync (lighter bar).

Peak: 76 in 2026

Laboratory controls observations in Form 483s by inspection year
YearObservations
20081
20090
20100
20110
20128
20130
20145
20154
201613
201721
201823
201914
20200
20216
202218
202342
202425
202548
202676

How FDAgent sorts observations into this theme

Two fixed rules, applied by code to every observation FDAgent can quote. An observation counts when either holds. No model and no similarity score decides.

It is tied to one of these 21 CFR entries

21 CFR 211.16021 CFR 211.16521 CFR 211.16621 CFR 211.16721 CFR 211.17021 CFR 211.17321 CFR 211.176

The same entries the laboratory controls topic counts. A 483 prints no regulation, so FDAgent ties an observation to a section only when it opens with FDA's own citation wording for that section, word for word.

Or its text contains one of these phrases

“laboratory control*” · “laboratory record*” · “laboratory investigation*” · “laboratory equipment” · “laboratory instrument*” · “laboratory data” · “laboratory test*” · “laboratory result*” · “laboratory analyst*” · “laboratory procedure*” · “laboratory notebook*” · “laboratory error*” · “QC laborator*” · “quality control laborator*” · “test method*” · “analytical method*” · “analytical procedure*” · “method validation” · “HPLC*” · “chromatograph*” · “reference standard*” · “system suitability” · “method transfer”

* matches any ending and … up to three words in between. Case is ignored unless a phrase has capitals.

The 304 observations, newest first

Quoted word for word from each Form 483, with only line breaks joined; … marks where the quote is cut. The highlight shows why the observation is in this theme. Each links to FDAgent's page for the 483.

  1. FH Investments Inc. dba Asteria HealthInspection ended Jul 10, 2018 · FEI 3010348724 · Observation 2

    Laboratory controls do not include the establishment of scientifically sound and appropriate test procedures designed to assure that drug products conform to appropriate standards of identity, strength, quality, and purity. Specifically, sterility testing of hormone pellets during (b) (4) only provides information for the exterior surfaces of the implantable pellets. There is no evaluation of the entire pellet.

    Opens with FDA's citation wording for 21 CFR 211.160(b)Read the Form 483 →

  2. Biotika ASInspection ended Jun 1, 2018 · FEI 3003370212 · Observation 2

    Failure to document changes to computerized systems and provide adequate control to prevent omission or alteration of data. Specifically, a. There is no record of the review audit trails associated with your chromatography data systems used to analyze and to ensure data has not been altered or deleted. SOP 10.36/2017 Ver. 03 effective May 25, 2018 requires the review of audit trails. This procedure lacks details to describe how audit trail review …

    Contains “chromatograph*”Read the Form 483 →

  3. Glenmark Pharmaceuticals Ltd.Inspection ended May 24, 2018 · FEI 3008565058 · Observation 1

    The accuracy, sensitivity, specificity and reproducibility of test methods have not been established and documented. Specifically, Your firm's QC laboratory has not validated analytical methods used in the determination of active residue from equipment's used for processing of the following products: HP® Capsules USP? mg &P® mg, 2)? Tablets USP me.§ mg {§ mg &[H me. 3)@@ Tablets, 4)0@ Tablets, 5)0@ Capsules USP™® mg, and 6)P@ Tablets USPP® mg & …

    Opens with FDA's citation wording for 21 CFR 211.165(e)Read the Form 483 →

  4. Glenmark Pharmaceuticals Ltd.Inspection ended May 24, 2018 · FEI 3008565058 · Observation 2

    Laboratory controls do not include the establishment of scientifically sound and appropriate sampling plans designed to assure that conform to appropriate standards of identity, strength, quality and purity. Specifically, Your SOP CM/QA034 "Sampling Methodology for [7 is deficient in section 5.5 ¢ sample of ®® |, subsection 5.5.2 indicates "The ®@ sample shall be collected as gram unless and otherwise mentioned in the sampling [ / process …

    Opens with FDA's citation wording for 21 CFR 211.160(b)Read the Form 483 →

  5. Glenmark Pharmaceuticals Ltd.Inspection ended May 24, 2018 · FEI 3008565058 · Observation 3

    There is a failure to thoroughly review any unexplained discrepancy whether or not the batch has been already distributed. Specifically, e Your firm's QC unit did not investigate in to the issues of API peak seen in the blank injections during method validation for related substances by HPLC method. Additionally, system suitability (precision) standard injections (| ppm) peaks were not integrated A The reintegration of these peaks changed the …

    Contains “method validation”Read the Form 483 →

  6. Glenmark Pharmaceuticals Ltd.Inspection ended May 24, 2018 · FEI 3008565058 · Observation 5

    The responsibilities and procedures applicable to the quality control unit are not fully followed. Specifically, Your firm's Quality Unit has no adequate oversight on the raw data pertaining to production and QC laboratory. For example, 1) On 05/15/2018, we observed torn balance printouts, shredded documents, torn draft investigation reports, etc. Your firm's interim investigations pertaining to balance printouts confirmed that production …

    Contains “QC laborator*”Read the Form 483 →

  7. Eagle Pharmacy Inc.Inspection ended May 18, 2018 · FEI 3011688532 · Observation 1

    Laboratory controls do not include the specifications and test procedures designed standards of identity, strength, quality, and purity. Specifically, finished product release testing does include preservative content for injectable drug products. All drug products are sold and include preservatives, such as 5 4 ti) (4)

    Contains “laboratory control*”Read the Form 483 →

  8. Galaxy Surfactants Ltd.Inspection ended Mar 30, 2018 · FEI 3004879807 · Observation 2

    2. Your firm failed to (adequately) validate the analytical method for stability studies of APL®® to support the intended commercialization period and expiry. Specifically, a. GC method, GQM 157 used for the test is neither a USP method nor an internal validated method. b. Your firm is currently using GC method, GQM 157 for the stability testing of APL®® but you have not completed forced degradation studies with this method. c. You have conducted …

    Contains “analytical method*”Read the Form 483 →

  9. Galaxy Surfactants Ltd.Inspection ended Mar 23, 2018 · FEI 3010166890 · Observation 2

    2. Your firm failed to validate the analytical methods required for the trace analysis of API residue as part of the cleaning validation program. Specifically, there is not validation report that confirms the sensitivity of <USP621> analytical method to detect the established acceptable level of the residue or contaminants. The method's attainable recovery level is also not established.

    Contains “analytical method*”Read the Form 483 →

  10. Aurobindo Pharma Ltd.Inspection ended Feb 20, 2018 · FEI 3008461619 · Observation 7

    The statistical quality control criteria fail to include appropriate acceptance levels and rejection levels. Specifically, your procedure for receiving and inspection of the components and packaging materials (procedure # FU4-QC-GEN-026, Sampling of packing materials) follows an unspecified level of AQL of ANSI Z1.4-2008 with no clear criteria for acceptance and rejection level. This procedure is deficient to provide clear reject and acceptance …

    Opens with FDA's citation wording for 21 CFR 211.165(d)Read the Form 483 →

  11. Aurobindo Pharma Ltd.Inspection ended Feb 20, 2018 · FEI 3008461619 · Observation 8

    Established laboratory control mechanisms are not followed and documented at the time of performance. Specifically, laboratory control is deficient in that USP reference standards, code #USP- | with lot #% , used to determine the identity, strength, quality and purity of ® | injection which were received in laboratory on January 11, 2018: a) were not received in laboratory inventory management system (LIMS) as required by your procedure, however …

    Opens with FDA's citation wording for 21 CFR 211.160(a)Read the Form 483 →

  12. Suven Life Sciences Ltd.Inspection ended Feb 15, 2018 · FEI 1000623515 · Observation 1

    Appropriate controls are not exercised over computer or related systems to assure that changes in master production and control records or other records are instituted only by authorized personnel. Specifically, A. Your firm has a total of (b) (4) HPLCs and (b) (4) UPLCs used to analyze raw material, finished product, and related substances for products marketed and distributed to the US. The HPLCs and UPLCs use either Empower 3 or EZ CHROM …

    Contains “HPLC*”Read the Form 483 →

  13. Shilpa Medicare LimitedInspection ended Nov 30, 2017 · FEI 3009876430 · Observation 4

    … Your firm invalidated initial out-of-specification (OOS) laboratory results in favor of passing results without an adequate investigation.

    Contains “laboratory result*”Read the Form 483 →

  14. Shilpa Medicare LimitedInspection ended Nov 30, 2017 · FEI 3009876430 · Observation 6

    The written stability program for drug products does not include reliable, meaningful, and specific test methods. The methods used during the stability program for [illegible] Tablets [illegible] mg & [illegible] mg; Injection, ®® Injection, § mg/ml and ®® & Injection, ® mg/vial have not been proven to be stability indicating. A. [illegible] Tablets [illegible] mg: The stress studies performed during the validation of [illegible] tablets resulted …

    Opens with FDA's citation wording for 21 CFR 211.166(a)(3)Read the Form 483 →

  15. Shilpa Medicare LimitedInspection ended Nov 30, 2017 · FEI 3009876430 · Observation 7

    The following issues were noted during the review of the analytical methods used in the finished product release of product for the U.S. market. Specifically, the software used to conduct high performance liquid chromatography (HPLC) analysis of finished product for unknown impurities is configured to permit extensive use of multiple processing methods without scientific justification. For example, during the analysis of impurities, an analyst …

    Contains “analytical method*”Read the Form 483 →

  16. Shilpa Medicare LimitedInspection ended Nov 30, 2017 · FEI 3009876430 · Observation 8

    Acceptance criteria for the sampling and testing conducted by the quality control unit is not adequate to assure that batches of drug products meet each appropriate specification as a condition for their approval and release. Specifically there is no evidence that each vial produced meets specification. A. One vial of ®® mg/vial Injection contains [illegible] ppm. The assay method, degradation products method, and [illegible] method requires …

    Opens with FDA's citation wording for 21 CFR 211.165(d)Read the Form 483 →

  17. Divi's Laboratories Ltd. Unit IIInspection ended Sep 19, 2017 · FEI 3004149463 · Observation 2

    OBSERVATION 2 Laboratory failures are not thoroughly investigated. Specifically, Out of Specification (OOS) investigations where laboratory error was not clearly identified were invalidated without adequate investigation into manufacturing, process, maintenance, and engineering activities. For example: (b) (4) (b) (4) a. Investigation No./ 2-PLV/OOS/IP-IM/W/012 for lot number [illegible] any other individual impurity out of specification …

    Contains “laboratory error*”Read the Form 483 →

  18. Divi's Laboratories Ltd. Unit IIInspection ended Sep 19, 2017 · FEI 3004149463 · Observation 3

    … and/or compared to specification to ensure OOS results were not obtained and sample result screening is not occurring prior to invalidating the run for system suitability or chromatograph issues. For instance: a. On January 12, 2017, Incident 2-AT/FP[illegible]/W/002 was initiated after testing of 6 lots for theoretical plate count failure of peak in standard solution. Sample results were invalidated without evaluation of results to ensure OOS …

    Contains “system suitability”Read the Form 483 →

  19. Dr. Reddy's Laboratories Unit 1Inspection ended Jun 16, 2017 · FEI 3009193040 · Observation 1

    OBSERVATION 1 Testing and release of drug product for distribution do not include appropriate laboratory determination of satisfactory conformance to the final specifications prior to release. ... (b)(4) Specifically, in testing results of (b)(4) Powder (b)(4) mg packed in sachets for exhibit batches/lots we observed: 1. On 16 December 2015, the initial release testing results was generated using manual integration for related substance testing …

    Opens with FDA's citation wording for 21 CFR 211.165(a)Read the Form 483 →

  20. Strides Shasun LimitedInspection ended May 26, 2017 · FEI 3004554612 · Observation 2

    Samples taken of in-process materials for determination of conformance to specifications are not representative. Specifically, there is no scientific rationale or data to support the sampling plan your firm uses for testing the ®® capsules in ®® on ®®. Furthermore, your manufacturing batch record directions are adequate in that they only specify that operators are to take samples from ®® and that your operators collect only a total of capsules …

    Opens with FDA's citation wording for 21 CFR 211.160(b)(2)Read the Form 483 →

21 CFR sections these observations open with

The sections FDAgent ties the most of these observations to by FDA's wording, with how many of this theme's observations each covers. Each links to its page of FDA citations, warning letters and the regulation text.

Ask FDAgent about laboratory controls in Form 483s

FDAgent reads FDA's Form 483s, inspections, citations and warning letters and answers in plain English, with links to the records.

Edit it or ask as written. 3 questions free, no card needed.

About this data

  • The observations are the text FDAgent read by OCR from Form 483 PDFs in FDA's Electronic Reading Room, shown only from 483s whose every printed observation passed FDAgent's checks against the scanned text, as on each 483's own page.
  • Which theme an observation belongs to is decided by the fixed rules above; every count on this page is computed from them.
  • A 483 addressed to an individual is listed without the name, and an observation that names a person is left out.
  • FDAgent is independent of FDA.