Dr. Reddy's Laboratories Unit 1 Form 483, June 2017
FDA issued this Form 483 to Dr. Reddy's Laboratories Unit 1 in Srikakulam, India (FEI 3009193040) at the end of a drug inspection that ended June 16, 2017. It lists 1 observation. FDA classified the inspection Voluntary Action Indicated (VAI).
A Form 483 lists the investigators' observations at the end of an inspection. It is not FDA's final determination.
Source: FDA's Electronic Reading Room, posted Jun 29, 2017. Form 483 PDF on fda.gov. FDAgent is independent of FDA.
- Inspection ended
- Jun 16, 2017
- Drugs · Drug Quality Assurance
- FDA's outcome
- VAI
- Voluntary Action Indicated
- Posted by FDA
- Jun 29, 2017
- 13 days after the inspection
The 1 observation on this Form 483
Read from the FDA-posted Form 483 PDF by OCR. FDAgent keeps an observation only when it matches the scanned text; check the PDF before quoting. The short titles are FDAgent's.
- 1
Inadequate lab processing for release testing
OBSERVATION 1 Testing and release of drug product for distribution do not include appropriate laboratory determination of satisfactory conformance to the final specifications prior to release. ... (b)(4) Specifically, in testing results of (b)(4) Powder (b)(4) mg packed in sachets for exhibit batches/lots we observed: 1. On 16 December 2015, the initial release testing results was generated using manual integration for related substance testing for all three exhibit batches. 2. Different processing methods were used to process system suitability solution, standard, and sample injections for the following release testing and stability samples: (b)(4) A. Powder release testing (b)(4) and accelerated samples: o 3 month, lots o 6 month, lots o 5 month, lots In addition, related substance for assay testing for [illegible], which is a component in (b)(4) Powder (b)(4) mg, was also processed using separate processing methods for system suitability solution, standard, and sample injections during the 3 and 6 month long term and accelerated stability samples, and 5 month accelerated stability sample of exhibit lots. C. Tablets (b)(4) [illegible] mg tablets, lot [illegible] packed in [illegible] and [illegible] count bottles, for nine month long term stability sample. [illegible] 09 mg tablets, lot packed in [illegible] and [illegible] count bottles, for nine month long term stability sample. [illegible] mg tablets, lots [illegible] and for initial release testing sample. [illegible] mg tablets, lot for initial release testing sample.
Other Form 483s FDA posted for this facility
Every Form 483 FDA posts under FEI 3009193040, newest first.
- Form 483, December 2025FDA record date Dec 12, 2025 · Posted Jan 5, 2026
- Form 483, January 2019Inspection ended Jan 15, 2019 · Posted Feb 26, 2019VAI
Ask FDAgent about this Form 483
FDAgent reads the 483, the facility's inspections, warning letters and FDA's citations and answers in plain English, with links to the records.
Related
About this data
- The PDF and its dates come from FDA's Electronic Reading Room, last synced Nov 23, 2026, 04:33 UTC; the inspection, its outcome and FDA's citations from FDA's Data Dashboard.
- The observations are the text FDAgent read from this PDF by OCR, shown only when every printed observation passed FDAgent’s checks against the scanned text.
- FDAgent leaves out the names of the people the Form 483 is addressed to and of the investigators.