Skip to content

Aurorium LLC

500 N 9th St, Midlothian, Texas, United States

FEI 1628757DUNS 167166230

FDA has 5 inspections on file for Aurorium LLC in Midlothian, Texas (FEI 1628757), from 2010 to 2026. The latest ended on and was classified VAI, voluntary action indicated.

VAI means FDA investigators found objectionable conditions or practices, but FDA was not prepared to take or recommend regulatory action. The firm is expected to correct them voluntarily.

FDA posted the Form 483 from that inspection. FDAgent read its 5 observations from the PDF; they are listed below. One FOIA request has asked FDA for this inspection's 483.

No warning letter, court action or import refusal is filed under this FEI.

As of Nov 21, 2026, the site is registered with FDA as a drug establishment for API manufacturing, through Dec 31, 2026. Its registrant is Aurorium LLC (D-U-N-S 167166230).

Sources:FDA firm profile ↗Form 483 PDF ↗FDA drug establishment registrations ↗Synced Nov 23, 2026

Ask FDAgent about this site

Free account: 3 questions, no card needed. Each answer links to its FDA sources.

Email me when FDA posts a new inspection, Form 483 or warning letter for this site

Weekly, on a free account. Pause it any time.

Watch this facility

Comparing suppliers? Check this site beside 2 others, free

Latest inspection
VAI
Inspections on file
5, 2010–2026
0 OAI · 3 VAI · 2 NAI
Inspections with 483 observations shown
1 of 5
Form 483s posted by FDA
↗
Requested under FOIA
1 request, latest May 18, 2026
Warning letters
None on file
Import refusals
None under this FEI
Import alerts
Not linked by FEI
FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗
Drug establishment registration
Through

Inspection history

5 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC

  1. VAIObjectionable conditions found; FDA took no regulatory actionDrugs · Drug Quality AssurancePosted 483 · 5 observations

    Read this Form 483 on its own page →Form 483 PDF on fda.gov

    Read from the FDA-posted Form 483 PDF by OCR. FDAgent keeps an observation only when it matches the scanned text; check the PDF before quoting. The short titles are FDAgent’s.

    1. Observation 1

      Poor facility maintenance and corrosion

      OBSERVATION 1 Failure to properly maintain buildings, facilities and equipment used in the manufacture of active pharmaceutical ingredients (API). Specifically, a) During our walk-through of the manufacturing facility on March 2 and 3, 2026, we noted there were areas of the building that had excessive corrosion, water leaking from equipment, and deterioration of walls and floors. Examples include the following. + We observed water leaking from a pipe that went from the (b) (4) to the (b) (4) ((b) (4) (b) (4) « The platform where the (b) (4) is located is deteriorating and has excessive corrosion. * The wall between the (b) (4) room and production has water damage and is warped and the panels are separating. On the side in the (b) (4) room there is exposed insulation from a large hole in the corner of the room near the (b) (4) and (b) (4) tanks. * We observed a leak on the (b) (4) unit and the metal support for the skid had excessive corrosion. « There were remnants of an old platform on the (b) (4) platform area (b) (4) where the concrete had broken into pieces. « There is water leaking from the exhaust for the dryer. b) On March 3, 2026, we observed the presence of a clear tape placed around a phalange connection on the (b) (4). The system is under (b) (4); (b) (4) can enter the system. The (b) (4) system (b) (4) (b) (4). c) Tank P* used to store (b) (4) from the (b) (4) that will be (b) (4) tanks is not completely covered and is open to the environment.

    2. Observation 2

      Failure to investigate OOS endotoxin

      OBSERVATION 2 Failure to investigate all out of specification test results. Specifically, a review of (b) (4) test results from March 2024 to February 2026 revealed four (4) instances where your firm received an out of specification test result for endotoxins. Your firm has no documentation to show that these out of specification test results were investigated nor did you perform any product impact assessment. Per MIDL-SOP-307 (b) (4), effective January 1, 2024, the specification for endotoxin is (b) (4) EU/mL. + Sample from (b) (4) system collected 05/29/2024 - test result was 0.430 EU/mL « Sample from (b) (4) system collected 09/25/2024 - test result was 1.508 EU/mL « Sample from (b) (4) system — QC Laboratory collected 09/25/2024 - test result was 0.439 EU/mL « Sample from (b) (4) system — QC Laboratory collected 03/31/2025 - test result was 17.589 EU/mL « Sample from (b) (4) system collected 05/27/2025 - test result was 0.418 EU/mL In addition, several of the lab reports for the testing of the (b) (4) indicate that the sample for endotoxin testing was received outside of the recommended temperature range. The laboratory report does not state the temperature the samples were received at nor does it state what the recommended temperature range is. There is no evaluation of this issue to determine if it has any impact on testing of the (b) (4) to receive accurate results.

    Show observations 3 to 5
    1. Observation 3

      Quality Unit oversight failures

      OBSERVATION 3 Failure of the Quality Unit to fulfill its oversight responsibilities. Specifically, a) CAPAs are not fully documented in accordance with your established procedure entitled Corrective and Preventative Actions (MIDL-SOP-124) effective January 1, 2024; they also are not closed out in a timely manner. For example: + CP25-014 (initiated 8/15/2025), CP25-010 (initiated 7/01/2025), CP25-019 (initiated 11/17/2025), CP24-12 (initiated 10/11/2024), and CP24-13 (initiated 9/24/204), either lack or have insufficient documentation of an effectiveness check. + CP24-12, CP24-13 and CP25-019 have not been finalized and signed off as completed or reviewed by quality. b) Your firm does not have written procedures for the issuance, use and reconciliation of loose-leaf forms used for documentation of manufacturing (e.g. batch record) and laboratory testing. Examples of these forms include: + All forms and documents that are part of the batch record « Form 239A Reagent/Solution Preparation Log * Form 120A Label Control Form « Form 121A Label Issuance + Form 112B Pre-Shipment Review Log + Form 133D Total Unacceptable (b) (4) Count c) There is a failure to review and approve all quality-related documents. The analytical testing lab books (data collection form) used for testing of (b) (4), USP (API) are not controlled documents. The documents are not assigned numbers or revisions nor have they been reviewed and approved by the Quality Unit. These forms include the following: * USP/FCC/JP Chemistry Testing of (b) (4) * USP/FCC/EP/JP Loss on Drying of (b) (4) + USP/FCC/EP/JP Equivalent (b) (4) Test for (b) (4) * (b) (4) and Morphology d) Your firm has no written SOP governing the circumstances that would warrant the creation of a work order to ensure that all maintenance of equipment is reported. For example, on March 3, 2026, we observed the presence of a clear tape placed around a phalange connection on the (b) (4). No work order has been initiated to fix the issue and your firm has no documentation to show how long the tape has been on the equipment and who made the decision to do so. e) There is a failure of the Quality Unit to appropriately review all (b) (4) test results. « There is no documented Quality Unit review of the results from August 2025 to present. « From March 2024 to February 2026 there were four (4) instances of out of specification results for endotoxin that were not investigated however, the Quality Unit documented that the results were reviewed.

    2. Observation 4

      Missing annual product quality review

      OBSERVATION 4 Your firm failed to conduct an annual product quality review of your API. During the current inspection of your facility we observed that you had not conducted your 2024 annual product quality review for your API (b) (4), USP. On August 6, 2025, your firm initiated a deviation report however, no content was provided in the report explaining why the 2024 review had not been completed or when it would be finalized. The deviation is still open.

    3. Observation 5

      Lack of required material testing

      OBSERVATION 5 Failure to perform testing of the (b) (4) used in production and in the QC Laboratory. Specifically, your firm could not provide test results for the (b) (4) for the (b) (4) of (b) (4) and (b) (4). MIDL-SOP-307 (b) (4), effective January 1, 2024 requires (b) (4) testing of the (b) (4) for (b) (4), endotoxin, and microbial (standard plate count and pseudomonas). In addition, your firm has no documentation and/or procedure to show exactly where the samples are being taken by your 3rd party contractor. The laboratory results simply state (b) (4), "Lab" and "Production."

  2. VAIObjectionable conditions found; FDA took no regulatory actionDrugs · Drug Quality AssuranceCitations not published
  3. VAIObjectionable conditions found; FDA took no regulatory actionDrugs · Drug Quality AssuranceCitations not published
  4. NAINo objectionable conditions that need follow-upDrugs · Drug Quality AssuranceNone
  5. NAINo objectionable conditions that need follow-upDrugs · Drug Quality AssuranceNone

Requested under FOIA

FDA FOIA logs, last synced Nov 23, 2026, 04:25 UTC

One request to FDA under the Freedom of Information Act named this site, on May 18, 2026. It asked for a Form 483.

  • Inspection ending Mar 6, 2026VAI

    1 request. In the inspection history above. FDA has posted its Form 483.

    1. May 18, 2026FDA-FOIA-2026-5049Form 483· 1 of 5 sitesRecords released

A request is listed when it names this site’s FEI number or one of its FDA inspection IDs, or, after review, its exact street address. Requester names are left out.

FDA drug establishment registration

FDA registration search
Registered operations
API manufacturing
Registration runs through
Dec 31, 2026
Registrant
Aurorium LLCD-U-N-S 167166230

From FDA’s Drug Establishments Current Registration Site, as of Nov 21, 2026. A registration says what a site told FDA it does. It is not an inspection result. FDA’s registration search looks firms up by name.

Other FEIs recorded as Aurorium LLC

Matched on the exact firm name FDA recorded. Sites recorded under other names are not listed, and a shared name does not show common ownership.

  • Aurorium LLCFEI 1824624Zeeland, Michigan, United States

    Last inspected 2025-07-31VAI

  • Aurorium LLCFEI 3002906664Greensboro, North Carolina, United States

    Last inspected 2012-11-21NAI

Go further than this page

FDAgent researches FDA’s records and links every answer to its source.

Free account: 3 questions, no card needed.

About this data. Every record here is filed under FEI 1628757 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
  • Every record on this page is filed under FEI 1628757. Records filed under another FEI or name, including other sites of the same company, are not shown here.
  • FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
  • Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
  • “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
  • Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
  • FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.