Environmental monitoring in FDA Form 483 observations
Monitoring cleanrooms and the people in them: viable and nonviable air sampling, settle and contact plates, glove and fingertip samples, and the results that should trigger action. FDAgent finds 179 such observations in 116 Form 483s issued to 109 facilities, for inspections that ended between October 2008 and September 2026. Each is quoted below from the 483 FDA posted.
A Form 483 lists the investigators' observations at the end of an inspection. It is not FDA's final determination.
Source: Form 483 PDFs in FDA's Electronic Reading Room, last synced Nov 23, 2026, 04:33 UTC. Every quote is from a 483 whose reading passed FDAgent's checks against the scan. FDAgent is independent of FDA.
- Observations
- 179
- Each quoted below
- Form 483s
- 116
- Posted by FDA
- Facilities
- 109
- Distinct FEIs
- Most recent
- Sep 28, 2026
- Inspection end date
Observations by year
These observations by the year the inspection ended. They follow which 483s FDA has posted and FDAgent has read, so they show what is on record, not a trend in FDA's findings. 2026 runs to the latest sync (lighter bar).
Peak: 64 in 2026
| Year | Observations |
|---|---|
| 2008 | 2 |
| 2009 | 0 |
| 2010 | 0 |
| 2011 | 0 |
| 2012 | 2 |
| 2013 | 0 |
| 2014 | 2 |
| 2015 | 2 |
| 2016 | 11 |
| 2017 | 9 |
| 2018 | 6 |
| 2019 | 5 |
| 2020 | 1 |
| 2021 | 14 |
| 2022 | 14 |
| 2023 | 13 |
| 2024 | 11 |
| 2025 | 23 |
| 2026 | 64 |
How FDAgent sorts observations into this theme
Two fixed rules, applied by code to every observation FDAgent can quote. An observation counts when either holds. No model and no similarity score decides.
It is tied to one of these 21 CFR entries
A 483 prints no regulation, so FDAgent ties an observation to a section only when it opens with FDA's own citation wording for that section, word for word.
Or its text contains one of these phrases
“environmental monitoring” · “viable” · “nonviable” · “non viable” · “settle plate*” · “contact plate*” · “personnel monitoring” · “fingertip*” · “glove print*” · “glove sampl*” · “particle count*” · “air sampl*” · “surface sampl*” · “EM program*” · “EM sampl*” · “EM data” · “EM excursion*” · “EM result*” · “EM plate*”
* matches any ending and … up to three words in between. Case is ignored unless a phrase has capitals.
The 179 observations, newest first
Quoted word for word from each Form 483, with only line breaks joined; … marks where the quote is cut. The highlight shows why the observation is in this theme. Each links to FDAgent's page for the 483.
Strive Pharmacy LLCInspection ended Sep 28, 2026 · FEI 3021042485 · Observation 1
Microbial contamination was present in the ISO 5 area and areas adjacent to production areas. Specifically, Periodic Air and Surface Environmental Monitoring A. There was an action level (b)(4) active air sample in ISO 5 LFH (Hood) on 11/26/25. Sample ID #BHT112625NHDH1, result = 15 cfu, action limit for ISO 5 air samples is [illegible] cfu. Microorganisms were identified as Micrococcus species and coag-negative Staphylococcus species. There was …
Contains “environmental monitoring”Read the Form 483 →
Strive Pharmacy LLCInspection ended Sep 28, 2026 · FEI 3021042485 · Observation 9
Inadequate routine environmental monitoring in the classified areas. Specifically, stools positioned at all (b)(4) ISO 5 LFH are not included in the EM surface sample program. I observed technicians in the Lane ISO 7 buffer room and the Lane ISO 7 buffer room directly contact a stool with their gloved hands during filling operations for sterile injectable drugs. Technicians sit on the stools while filling. Technicians sit on a bench in the ISO 8 …
Contains “environmental monitoring”Read the Form 483 →
Victory Medical Center PharmacyInspection ended Sep 10, 2026 · FEI 3010485868 · Observation 4
… This vent has not been monitored for unidirectional airflow and is not included in your firm's (b) (4) environmental monitoring sampling program. On August 31, 2026, I observed open and exposed vials of ELECTRO B COMPLEX (MDV) MAGNESIUM CHLORIDE/CALCIUM CHLORIDE/B COMPLEX/SODIUM BICARBONATE (105MG/28MG/B-COMPLEX,12MG/ML), Lot 2055939, waiting to be stoppered, and exposed rubber stoppers directly in front of the air intake vent. I observed that …
Contains “environmental monitoring”Read the Form 483 →
Victory Medical Center PharmacyInspection ended Sep 10, 2026 · FEI 3010485868 · Observation 8
Lack of routine environmental monitoring in the ISO 5 area and classified areas. Specifically, the following deficiencies were observed: 1) Absence of Historical Surface Sampling Data - Your firm was unable to provide surface sampling environmental monitoring data for your ISO 5 Laminar Airflow Hood (LAFH) and classified cleanrooms prior to July 29, 2026. Despite this absence of surface sampling data, between June 5, 2026 and July 28, 2026, your …
Contains “environmental monitoring”Read the Form 483 →
Greenwich RxInspection ended Sep 9, 2026 · FEI 3043120213 · Observation 3
Microbial contamination was present in the ISO 5 area and areas adjacent to production areas. Specifically, during my review of your firm's environmental monitoring (EM) records, I found the following deficiencies: A. During a review of your firm's 2025 (b)(4) EM monitor sampling of the ISO Classified Cleanroom Suite I found two (2) different occurrences where 1 CFU was found during ISO 5 LAFC surface sampling (Acceptance Limits ISO Class 5: …
Contains “environmental monitoring”Read the Form 483 →
STAQ Pharma IncInspection ended Sep 8, 2026 · FEI 3015144909 · Observation 2
Procedures designed to prevent microbiological contamination of drug products purporting to be sterile are not. Specifically, on September 1, 2026, Sterile Filling Technician [illegible] was observed in ISO-7 [illegible] Production Room 26102671A, at an ISO-5 (b)(4) Laminar Air Flow Workstation. While exchanging an environmental monitoring (EM) active air sample plate, [illegible] was observed extending her arms fully into the hood, up to her …
Contains “environmental monitoring”Read the Form 483 →
STAQ Pharma IncInspection ended Sep 8, 2026 · FEI 3015144909 · Observation 3
… In Room [illegible], environmental monitoring equipment used during the production of drug products is placed (b)(4) of (b)(4). Cleaning of (b)(4) of (b)(4) occurs (b)(4) however, your firm has not assessed the discolorations for potential risk to cleanroom environment and product.
Contains “environmental monitoring”Read the Form 483 →
RC Outsourcing LLCInspection ended Sep 3, 2026 · FEI 3011967886 · Observation 1
… Absence of Active Air Sampling During AFV Studies. In all three August 2025 video records covering both LFH and LFH, no active air sampling was performed concurrent with the AFV studies. Active air sampling during AFV studies is necessary to assess and document the microbiological quality of the environment under simulated or actual operational conditions. The AFV studies for ISO 5 LFH and LFH conducted in both August 2025 and February 2026, were …
Contains “air sampl*”Read the Form 483 →
RC Outsourcing LLCInspection ended Sep 3, 2026 · FEI 3011967886 · Observation 5
… For example, your firm operates incubators (b)(4); and (b)(4) refrigerators (b)(4), all of which are used for product storage and environmental monitoring (EM) sample incubation. A review of the initial qualification records for these units revealed that temperature mapping was performed exclusively under empty-state conditions. When asked whether temperature mapping under operationally loaded conditions had been performed to reflect the units' …
Contains “environmental monitoring”Read the Form 483 →
RC Outsourcing LLCInspection ended Sep 3, 2026 · FEI 3011967886 · Observation 7
… Your firm has not established training or qualification requirements to ensure that personnel responsible for reading and interpreting EM plate results are able to accurately enumerate and differentiate microbial growth and surface colonization.
Contains “EM plate*”Read the Form 483 →
OPS International Inc dba Olympia PharmaceuticalsInspection ended Sep 3, 2026 · FEI 3009724085 · Observation 1
Aseptic processing areas are deficient regarding the system for monitoring environmental conditions. Specifically, (A) The firm's environmental monitoring procedure does not require viable environmental monitoring during the initial aseptic setup of the (b)(4) filling line within the (b)(4) or during the initial aseptic setup of the (b)(4) automated filling machine positioned under [illegible] HEPA-filtered unidirectional airflow enclosure. Per …
Opens with FDA's citation wording for 21 CFR 211.42(c)(10)(iv)Read the Form 483 →
Houston Rx LLCInspection ended Aug 28, 2026 · FEI 3045022201 · Observation 5
Microbial contamination was present in the ISO 5 area. On August 3, 2026, your firm recovered 1 CFU in the ISO 5 Laminar Flow Hood (b)(4) (viable surface sample) after the filling/stoppering of lot #TZN-080326-20/2-2 of Tirzepatide/Niacinamide 20/2 mg/mL (2mL vial). No investigation was performed as your firm's action level for ISO 5 areas is (b)(4) CFU, which is not appropriate. The lot was released and has been partially distributed.
Contains “viable”Read the Form 483 →
TAILSTORM HEALTH INCInspection ended Aug 28, 2026 · FEI 3015929581 · Observation 1
… For example, ISO 5 Filling Line was not fully assembled or in-operation, and there was one cleanroom technician outside the (b) (4) during dynamic nonviable particulate (NVP) monitoring. We observed up to (b) (4) technicians inside the ISO 7 buffer room for the ISO 5 (b) (4) during filling of Lidocaine 2% Injection, 10 mL vial, batch 2608002. Records reviewed included: Project AZ-MVDT261801C-1, (b) (4) Cleanroom Testing and Certification, dated …
Contains “nonviable”Read the Form 483 →
TAILSTORM HEALTH INCInspection ended Aug 28, 2026 · FEI 3015929581 · Observation 2
… Personnel Monitoring G. Personnel monitoring specifications for (b) (4) hands, forearms, chest, and forehead are not scientifically justified. Your firm uses the ISO 7 microbial monitoring limits ((b) (4) CFU/plate alert and CFU/plate action) to evaluate samples from operators that use (b) (4) to perform closed-door interventions in the ISO 5 Filling Line. Samples collected from operators after open-door interventions are the only personnel …
Contains “personnel monitoring”Read the Form 483 →
TAILSTORM HEALTH INCInspection ended Aug 28, 2026 · FEI 3015929581 · Observation 4
… The (b) (4) Environmental Monitoring Trend Report for the Raw Material (b) (4) storage area (May 1-31, 2026) documents the temperature readings from sensors (b) (4) and (b) (4) exceeded the High Warning Limit of (b) (4) on multiple occasions throughout the month of May 2026. Maximum temperature readings on both sensors approached or reached the High Alarm Limit of (b) (4) during the same period. Quality management stated they notified the …
Contains “environmental monitoring”Read the Form 483 →
Carolina Infusion LLCInspection ended Aug 18, 2026 · FEI 3015826782 · Observation 8
Lack of and Inadequate routine environmental monitoring in the ISO 5 area and classified areas. Your staff stated they perform (b)(4) surface sampling within the ISO 5 and ISO 7 area after cleaning and disinfection. Sampling after cleaning does not accurately represent your production environment. Additionally, your staff also stated and demonstrated that when (b)(4) plates are unavailable, the firm's procedure is to use (b)(4) (b)(4), (b)(4) on …
Contains “environmental monitoring”Read the Form 483 →
IntegraDose Compounding Services LLCInspection ended Aug 12, 2026 · FEI 3014483112 · Observation 2
… (b)(4)) for viable air sampling in your cleanrooms. However, your firm only has SOP-0064 (b)(4) Microbiological (Viable) Air Sampler - Calibration, Operation, Use, and Maintenance, for viable air sampling equipment. There is no SOP outlining the use of the (b)(4) Microbial Sampler.
Contains “viable”Read the Form 483 →
IntegraDose Compounding Services LLCInspection ended Aug 12, 2026 · FEI 3014483112 · Observation 4
Aseptic processing areas are deficient regarding the system for monitoring environmental conditions. Specifically, On August 5, 2026, video recordings of the production of Hydromorphone HCL 20mg/100mL CADDS lot #20260720HYD-1, EXP November 17, 2026, completed (b)(4), were reviewed. During this video review the following was observed: 1. Your production operators were observed placing the settle plates in the corners of the LAFHs before the start …
Opens with FDA's citation wording for 21 CFR 211.42(c)(10)(iv)Read the Form 483 →
Brookfield Medical Surgical Supplies IncInspection ended Jul 10, 2026 · FEI 3011430551 · Observation 3
OBSERVATION 3 Procedures designed to prevent microbiological contamination are not followed. Specifically, The environmental monitoring program of your firm allows for a CFU limit for gloved fingertips in an ISO-5 area. For a COMP compliant 503B facility, the action level must be any detected organism. Furthermore, your firm failed to investigate and identify the microbial growth recovered from the ISO-5 environment at the facility. All the …
Contains “environmental monitoring”Read the Form 483 →
Glenmark Pharmaceuticals LimitedInspection ended Jun 30, 2026 · FEI 3004672766 · Observation 1
… aerobic Microbial count: NMT (b)(4) cfu/plate" fails to demonstrate disinfectant efficacy for two reasons: The acceptance criteria is identical to routine environmental monitoring limits for Grade D areas and therefore does not demonstrate disinfectant efficacy. The acceptance criteria measures surface cleanliness, meaning there is no scientific basis to distinguish between effective disinfection and low baseline contamination. There is no …
Contains “environmental monitoring”Read the Form 483 →
21 CFR sections these observations open with
The sections FDAgent ties the most of these observations to by FDA's wording, with how many of this theme's observations each covers. Each links to its page of FDA citations, warning letters and the regulation text.
- 21 CFR 211.42Design and construction featuresIn this theme's rule30
- 21 CFR 211.113Control of microbiological contamination27
- 21 CFR 211.160General requirements6
- 21 CFR 211.192Production record review5
- 21 CFR 211.100Written procedures; deviations3
- 21 CFR 211.22Responsibilities of quality control unit2
- 21 CFR 211.63Equipment design, size, and location2
- 21 CFR 211.67Equipment cleaning and maintenance2
- 21 CFR 211.68Automatic, mechanical, and electronic equipment2
- 21 CFR 211.194Laboratory records2
Ask FDAgent about environmental monitoring in Form 483s
FDAgent reads FDA's Form 483s, inspections, citations and warning letters and answers in plain English, with links to the records.
Related
About this data
- The observations are the text FDAgent read by OCR from Form 483 PDFs in FDA's Electronic Reading Room, shown only from 483s whose every printed observation passed FDAgent's checks against the scanned text, as on each 483's own page.
- Which theme an observation belongs to is decided by the fixed rules above; every count on this page is computed from them.
- A 483 addressed to an individual is listed without the name, and an observation that names a person is left out.