Skip to content

Data integrity in FDA Form 483 observations

Records that are incomplete, unprotected or not made at the time: audit trails, deleted or overwritten data, shared logins, uncontrolled forms and backdating. FDAgent finds 120 such observations in 85 Form 483s issued to 84 facilities, for inspections that ended between November 2012 and September 2026. Each is quoted below from the 483 FDA posted.

A Form 483 lists the investigators' observations at the end of an inspection. It is not FDA's final determination.

Source: Form 483 PDFs in FDA's Electronic Reading Room, last synced Nov 23, 2026, 04:33 UTC. Every quote is from a 483 whose reading passed FDAgent's checks against the scan. FDAgent is independent of FDA.

Observations
120
Each quoted below
Form 483s
85
Posted by FDA
Facilities
84
Distinct FEIs
Most recent
Sep 3, 2026
Inspection end date

Observations by year

These observations by the year the inspection ended. They follow which 483s FDA has posted and FDAgent has read, so they show what is on record, not a trend in FDA's findings. 2026 runs to the latest sync (lighter bar).

Peak: 49 in 2026

Data integrity observations in Form 483s by inspection year
YearObservations
20121
20130
20140
20154
20168
20175
20189
20194
20200
20211
20227
202312
20248
202512
202649

How FDAgent sorts observations into this theme

Two fixed rules, applied by code to every observation FDAgent can quote. An observation counts when either holds. No model and no similarity score decides.

It is tied to one of these 21 CFR entries

21 CFR 211.68(b)21 CFR 211.18821 CFR 211.194(a)21 CFR 211.186(a)21 CFR 211.18221 CFR 212.60(g)21 CFR 212.110(b)

The same entries the data integrity topic counts. A 483 prints no regulation, so FDAgent ties an observation to a section only when it opens with FDA's own citation wording for that section, word for word.

Or its text contains one of these phrases

“data integrity” · “audit trail*” · “backdat*” · “trial injection*” · “test* into compliance” · “contemporaneous*” · “uncontrolled form*” · “uncontrolled blank form*” · “uncontrolled worksheet*” · “uncontrolled record*” · “uncontrolled document*” · “electronic data” · “electronic raw data” · “shared user*” · “shared account*” · “shared login*” · “shared password*” · “user access” · “access privilege*” · “access level*” · “administrator privilege*” · “administrator right*” · “falsif*” · “delet* … data” · “delet* … result*” · “delet* … chromatogram*” · “overwr* … data”

* matches any ending and … up to three words in between. Case is ignored unless a phrase has capitals.

The 120 observations, newest first

Quoted word for word from each Form 483, with only line breaks joined; … marks where the quote is cut. The highlight shows why the observation is in this theme. Each links to FDAgent's page for the 483.

  1. Zhejiang Huahai Pharmaceutical Co., Ltd.Inspection ended Jan 24, 2025 · FEI 3003999190 · Observation 7

    OBSERVATION #7 The quality control unit lacks authority to review production records to assure that no errors have occurred. 1. Procedures for audit trail reviews for production equipment have not been established: a. Production operators can make changes to compression settings. Quality personnel do not review audit trails that document parameters and changes. b. Production personnel can make changes to recipes for production and (6)l.J n the …

    Contains “audit trail*”Read the Form 483 →

  2. Hankook Cosmetics Manufacturing Co., LtdInspection ended Nov 21, 2024 · FEI 3012656014 · Observation 2

    Laboratory records do not include complete data derived from all tests, examinations and assay necessary to assure compliance with established specifications and standards. Specifically, Records related to chemical and microbiological testing do not include all the information related to the execution of the test. For example, (b)(4) is tested for (b)(4) and microbial limits. The laboratory records do not include complete information related to …

    Opens with FDA's citation wording for 21 CFR 211.194(a)Read the Form 483 →

  3. Lights Medical Manufacture Co., LtdInspection ended Nov 8, 2024 · FEI 3005310494 · Observation 2

    Written records of major equipment use are not included in individual equipment logs. Specifically, Your firm does not maintain equipment use logs for equipment used in the manufacturing process of over-the-counter drug products. This includes, but is not limited to, the following equipment: Name of Manufacturing Equipment Equipment ID(s) [illegible]

    Opens with FDA's citation wording for 21 CFR 211.182Read the Form 483 →

  4. Huangshi Hygenic Material Medicine Co.Inspection ended Aug 30, 2024 · FEI 3005543486 · Observation 4

    Laboratory records are deficient in that they do not include a statement of the results of tests and how they compare to the established specifications. Specifically, Your firm did not perform growth promotion studies on each batch of media used for microbial testing of your finished drug products (b) (4)

    Opens with FDA's citation wording for 21 CFR 211.194(a)(6)Read the Form 483 →

  5. Huangshi Hygenic Material Medicine Co.Inspection ended Aug 30, 2024 · FEI 3005543486 · Observation 5

    The responsibilities and procedures applicable to the quality control unit are not in writing and fully followed. Specifically, A) Your quality unit does not review electronic raw data including but not limited to identify testing of a finished product and APIs by HPLC and GC. B) Your quality unit does not review audit trails on test results including for HPLC (Waters Chromatography Data System Empower 3 and Lab Solutions) and GC (Chromleon and …

    Contains “electronic raw data”Read the Form 483 →

  6. Huangshi Hygenic Material Medicine Co.Inspection ended Aug 30, 2024 · FEI 3005543486 · Observation 8

    Laboratory records do not include the initials or signature of a second person showing that the original records have been reviewed for accuracy, completeness and compliance with established standards. Specifically, The reviewer of microbiological test results does not verify the original raw data for accuracy and completeness.

    Opens with FDA's citation wording for 21 CFR 211.194(a)(8)Read the Form 483 →

  7. Fukuzyu Pharmaceutical Co., Ltd.Inspection ended Jan 19, 2024 · FEI 3002807025 · Observation 4

    … On 01/18/2023, I observed the Vice Manager of QA login to the software as the administrator and then access system policies, audit trails, and stored data for (b)(4). The list of permissions for the Administrator user type indicated the user has unlimited ability to change or delete electronic records. The Vice QA Manager is responsible for oversight of API release. Your firm used Empower 3 software to process chromatographic data used to release …

    Contains “audit trail*”Read the Form 483 →

  8. Jiangsu Hengrui Pharmceuticals Co., LtdInspection ended Jan 16, 2024 · FEI 3007373503 · Observation 4

    … E) There is no adequate data integrity program in place to include statistically sound comprehensive review of electronic data by the Quality Assurance Unit for standalone and network systems, to ensure completeness, consistency, and accuracy of all chromatographic and non-chromatographic electronic data generated by the Quality Control Laboratory.

    Contains “data integrity”Read the Form 483 →

  9. Jiangsu Hengrui Pharmceuticals Co., LtdInspection ended Jan 16, 2024 · FEI 3007373503 · Observation 6

    Appropriate controls are not exercised over computers or related systems to assure that changes in master production and control records or other records are instituted only by authorized personnel. Specifically, Your firm's GMP related computerized systems and equipment spread across manufacturing workshops are not 21 CFR part 11 compliant. For example: A) Your firm operates portable non-viable particle monitoring equipment used to perform and …

    Opens with FDA's citation wording for 21 CFR 211.68(b)Read the Form 483 →

  10. Antaria Pty Ltd.Inspection ended Nov 17, 2023 · FEI 3025978151 · Observation 3

    Laboratory records are deficient in that they do not include a complete record of all data obtained during testing and do not always include documentation of pertinent information related to the preparation and testing of a sample. For example, a) There is no documentation of the preparation of standard stock solutions, working standards, reference standard solutions, sample solutions, diluents, mobile phases, reagents, and other preparations …

    Opens with FDA's citation wording for 21 CFR 211.194(a)(4)Read the Form 483 →

  11. Sumitomo Osaka Cement Co., LtdInspection ended Oct 27, 2023 · FEI 3003943727 · Observation 5

    Review of audit trail for analytical data generated by your firm is inadequate. Specifically, Review of the electronic audit trails from the Specific surface area measuring machine (BET) - Macsorb HM-model 1201 (ID-RK062 for API samples and ID-RK131 for raw material samples) indicated repeated events of Measurement error, Measurement interruption, Measurement warning and File delete. For example, during the test runs, the following incidents were …

    Contains “audit trail*”Read the Form 483 →

  12. Linhai Tianyu Pharmaceutical Co., Ltd.Inspection ended Sep 15, 2023 · FEI 3013345425 · Observation 4

    Laboratory records are not completed contemporaneously and do not include complete data derived from all tests to assure compliance with established specifications and standards. Specifically, A. Electronic and physical laboratory records data review are not completed contemporaneously by your QC reviewers. During our walkthrough of your QC laboratory data review room on 09/11/23, I observed your QC analyst/reviewer checking off boxes on a form …

    Contains “contemporaneous*”Read the Form 483 →

  13. Linhai Tianyu Pharmaceutical Co., Ltd.Inspection ended Sep 15, 2023 · FEI 3013345425 · Observation 6

    … Your firm appears to have performed Performance Verification and data integrity assessments for Empower 3 software, however the verification and assessments do not evaluate the consistent performance of the software/equipment over a specified period and operating environment. For Example, during our review of your Empower 3 chromatography software, interrupted sequences were observed, which generated "Data Incomplete" chromatographic data. Your …

    Contains “data integrity”Read the Form 483 →

  14. Samsung Biologics Co. Ltd.Inspection ended Sep 1, 2023 · FEI 3010479596 · Observation 1

    The Manufacturing Scientific Analytical Technology (MSAT) laboratory used in support of application submission testing data had inadequate controls over data integrity. Specifically, A) The MSAT laboratory was used in support of application submission data for (b) (4) and (b) (4) with the laboratory a development laboratory having inadequate controls in assurance of data integrity since 2013. Although you conducted an internal data integrity …

    Contains “data integrity”Read the Form 483 →

  15. Jiangsu Hengrui Pharmaceuticals CO. LTD.Inspection ended Aug 29, 2023 · FEI 3003404148 · Observation 3

    Batch production and control records for each batch of drug product produced do not include an accurate reproduction of the appropriate master production or control record which was checked for accuracy, dated and signed. Specifically, logbook for (b) (4) Tube Replacement Records in Workshop (b) (4) were replaced, with the information in logbook rewritten in the new logbook. a) The logbook used to document the number of (b) (4) tubing uses …

    Opens with FDA's citation wording for 21 CFR 211.188(a)Read the Form 483 →

  16. Jiangsu Hengrui Pharmaceuticals CO. LTD.Inspection ended Aug 29, 2023 · FEI 3003404148 · Observation 9

    The batch production and control records are deficient in that they do not include documentation of the accomplishment of each significant step in manufacturing, processing, and holding. Specifically, a) Workshop (b) (4) tanks used for holding in-process sterile drug solutions (b) (4) have (b) (4) used for (b) (4) into the tanks. The (b) (4) had labels that documented number of uses, (b) (4) installation, (b) (4) model, batch number, replacement …

    Opens with FDA's citation wording for 21 CFR 211.188(b)Read the Form 483 →

  17. UCB Farchim SAInspection ended Jul 4, 2023 · FEI 3005023799 · Observation 1

    … This discrepancy in your firm's ability to review, document, and investigate all electronic data is a gap in your firm's Data Integrity Program. B. There is no adequate data integrity program in place to include a statistically sound comprehensive review of all electronic data by the Quality Assurance Unit for standalone and network systems, to ensure completeness, consistency, and accuracy of all chromatographic and non-chromatographic …

    Contains “electronic data”Read the Form 483 →

  18. Ipca Laboratories LimitedInspection ended Jun 23, 2023 · FEI 3007574780 · Observation 3

    The suitability of all testing methods is not verified under actual conditions of use. Specifically, Your firm uses the UV method to quantitate the release of drug product in dissolution testing of (b) (4) drug products. However, the sample sequence described in your UV methods (one blank, one standard, one sample) for (b) (4) of the (b) (4) drug products did not establish the system suitability whenever the method was used for dissolution …

    Opens with FDA's citation wording for 21 CFR 211.194(a)(2)Read the Form 483 →

  19. Zen Chemicals Pvt. LtdInspection ended Mar 28, 2023 · FEI 3015345351 · Observation 3

    Appropriate controls governing computer acquired data and/or related systems have not been established to assure that changes and access to computerized software for the laboratory control system is accessed only by authorized personnel. Specifically, on March 23 and 27, 2023, I observed, A. UV which uses software UV Probe 2.70 (equipment # QC/INST/13) does not have audit trail capabilities. This equipment is used for the identification of (b) …

    Contains “audit trail*”Read the Form 483 →

  20. FRUEH VERPACKUNGSTECHNIK AGInspection ended Mar 24, 2023 · FEI 3010363503 · Observation 3

    The batch production and control records are deficient in that they do not include documentation of the accomplishment of each significant step in manufacturing, processing, and packing. Specifically, during packaging operations of drug product, we observed (b)(4) This practice of (b)(4) is not detailed in your batch record, nor is it described in a procedure.

    Opens with FDA's citation wording for 21 CFR 211.188(b)Read the Form 483 →

21 CFR sections these observations open with

The sections FDAgent ties the most of these observations to by FDA's wording, with how many of this theme's observations each covers. Each links to its page of FDA citations, warning letters and the regulation text.

Ask FDAgent about data integrity in Form 483s

FDAgent reads FDA's Form 483s, inspections, citations and warning letters and answers in plain English, with links to the records.

Edit it or ask as written. 3 questions free, no card needed.

About this data

  • The observations are the text FDAgent read by OCR from Form 483 PDFs in FDA's Electronic Reading Room, shown only from 483s whose every printed observation passed FDAgent's checks against the scanned text, as on each 483's own page.
  • Which theme an observation belongs to is decided by the fixed rules above; every count on this page is computed from them.
  • A 483 addressed to an individual is listed without the name, and an observation that names a person is left out.
  • FDAgent is independent of FDA.