Data integrity in FDA Form 483 observations

Records that are incomplete, unprotected or not made at the time: audit trails, deleted or overwritten data, shared logins, uncontrolled forms and backdating. FDAgent finds 120 such observations in 85 Form 483s issued to 84 facilities, for inspections that ended between November 2012 and September 2026. Each is quoted below from the 483 FDA posted.

A Form 483 lists the investigators' observations at the end of an inspection. It is not FDA's final determination.

Source: Form 483 PDFs in FDA's Electronic Reading Room, last synced Nov 23, 2026, 04:33 UTC. Every quote is from a 483 whose reading passed FDAgent's checks against the scan. FDAgent is independent of FDA.

Observations
120
Each quoted below
Form 483s
85
Posted by FDA
Facilities
84
Distinct FEIs
Most recent
Sep 3, 2026
Inspection end date

Observations by year

These observations by the year the inspection ended. They follow which 483s FDA has posted and FDAgent has read, so they show what is on record, not a trend in FDA's findings. 2026 runs to the latest sync (lighter bar).

Peak: 49 in 2026

Data integrity observations in Form 483s by inspection year
YearObservations
20121
20130
20140
20154
20168
20175
20189
20194
20200
20211
20227
202312
20248
202512
202649

How FDAgent sorts observations into this theme

Two fixed rules, applied by code to every observation FDAgent can quote. An observation counts when either holds. No model and no similarity score decides.

It is tied to one of these 21 CFR entries

21 CFR 211.68(b)21 CFR 211.18821 CFR 211.194(a)21 CFR 211.186(a)21 CFR 211.18221 CFR 212.60(g)21 CFR 212.110(b)

The same entries the data integrity topic counts. A 483 prints no regulation, so FDAgent ties an observation to a section only when it opens with FDA's own citation wording for that section, word for word.

Or its text contains one of these phrases

“data integrity” · “audit trail*” · “backdat*” · “trial injection*” · “test* into compliance” · “contemporaneous*” · “uncontrolled form*” · “uncontrolled blank form*” · “uncontrolled worksheet*” · “uncontrolled record*” · “uncontrolled document*” · “electronic data” · “electronic raw data” · “shared user*” · “shared account*” · “shared login*” · “shared password*” · “user access” · “access privilege*” · “access level*” · “administrator privilege*” · “administrator right*” · “falsif*” · “delet* … data” · “delet* … result*” · “delet* … chromatogram*” · “overwr* … data”

* matches any ending and … up to three words in between. Case is ignored unless a phrase has capitals.

The 120 observations, newest first

Quoted word for word from each Form 483, with only line breaks joined; … marks where the quote is cut. The highlight shows why the observation is in this theme. Each links to FDAgent's page for the 483.

  1. Aurobindo Pharma Ltd.Inspection ended Feb 20, 2018 · FEI 3008461619 · Observation 9

    Appropriate controls are not exercised over computers or related systems to assure that changes in master production and control records or other records are instituted only by authorized personnel. Specifically, master batch records that are submitted in“ for | injection, have different batch codes (document number) than the batch records that are currently in use for commercial production. There are no traceability for the changes made to the …

    Opens with FDA's citation wording for 21 CFR 211.68(b)Read the Form 483 →

  2. Suven Life Sciences Ltd.Inspection ended Feb 15, 2018 · FEI 1000623515 · Observation 1

    … The HPLCs and UPLCs use either Empower 3 or EZ CHROM ELITE, which have audit trail capabilities. However, your firm could not provide documentation these audit trails have been electronically reviewed by quality personnel. B. Your firm's Manager of Quality Assurance and Deputy Manager of QC verified your firm did not conduct electronic data review for any analysis performed in the QC Formulations Laboratory prior to August 2017.

    Contains “audit trail*”Read the Form 483 →

  3. Divi's Laboratories Ltd. Unit IIInspection ended Sep 19, 2017 · FEI 3004149463 · Observation 6

    OBSERVATION 6 Batch records are not always recorded contemporaneously at the completion of an operation. For example: On 9/12/2017 during the production of [illegible] batch the in-charge production employee had signed the "CHECKED BY" column at step [illegible] in the manufacturing record to increase the batch [illegible] to [illegible]. However the "DONE BY" column was not signed by the chemist and the operation was not completed. On 9/12/2017 …

    Contains “contemporaneous*”Read the Form 483 →

  4. Cadila Healthcare LimitedInspection ended Mar 1, 2017 · FEI 3005430968 · Observation 3

    Batch production and control records do not include complete information relating to the production and control of each batch. A. The batch manufacturing records for (b) (4) batches (b) (4) manufacturing records for (b) (4) Capsules USP (b) (4) mg and (b) (4) mg exhibit respectively) and intended batch Capsules USP (b) (4) mg BMR/ZB/0586-00 and (b) (4) (b) (4) Capsules USP (b) (4) mg BMR/ZB/0587-00 do not include specific instructions describing …

    Opens with FDA's citation wording for 21 CFR 211.188Read the Form 483 →

  5. Bayer AG PharmaInspection ended Jan 20, 2017 · FEI 3002806462 · Observation 9

    … Review of audit trail entries for 19 January 2017 observed two non-conforming data points that were not reported and were not further evaluated.

    Contains “audit trail*”Read the Form 483 →

  6. Bayer AG PharmaInspection ended Jan 20, 2017 · FEI 3002806462 · Observation 10

    Test procedures for analytical testing have not been established. 1. No procedures have been established to review audit trails for the laboratory software systems. Examples include, but are not limited to, Empower (Chromatography), Opus (IR Spectroscopy), and Disso Net (Dissolution by UV). 2. Manual integration of chromatograms is permitted. Procedures describing when manual integration can be performed, how it is to be performed, and how it …

    Contains “audit trail*”Read the Form 483 →

  7. Bayer AG PharmaInspection ended Jan 20, 2017 · FEI 3002806462 · Observation 11

    Data is not recorded contemporaneously. 1. Production and laboratory personnel do not document contemporaneously. The practice of signing for multiple steps all at once is permitted. 2. Supervisors reviewing logs signed one page rather than each individual page requiring a signature. 3. Microbiologists enter results into the electronic (b) (4) system for multiple samples all at one time during plate reading. 4. Microbiologists document in the (b) …

    Contains “contemporaneous*”Read the Form 483 →

  8. Zhejiang Huahai Pharmaceutical Co. Ltd.Inspection ended Nov 18, 2016 · FEI 3003999190 · Observation 4

    OBSERVATION #4 Data is not recorded contemporaneously. . ®@ © 1. The “checked by” entries for batch of Tablets were not documented at the time of batch production. 2. The QC analyst responsible for environmental monitoring does not document sampling at the time it occurs. It ‘was reported the analyst remembers the times and documents all at once after finishing.

    Contains “contemporaneous*”Read the Form 483 →

  9. Sun Pharmaceuticals Industries, Inc.Inspection ended Nov 16, 2016 · FEI 3002807979 · Observation 2

    Laboratory records do not include complete data derived from all tests, examinations and assay necessary to assure compliance with established specifications and standards. Specifically, There is a failure to document repeat testing in the laboratory equipment logbook when samples are reanalyzed from the same sample preparation vial. On October 29, 2016, raw material batch (BX was tested twice for “and © “Bement via Atomic Absorption Spectrometry …

    Opens with FDA's citation wording for 21 CFR 211.194(a)Read the Form 483 →

  10. Sun Pharmaceuticals Industries, Inc.Inspection ended Nov 16, 2016 · FEI 3002807979 · Observation 3

    Appropriate controls are not exercised over computers or related systems to assure that changes in master production and control records or other records are instituted only by authorized personnel. Specifically, There is no password protection for the Sartorius GP3202 precision balance instrument ID MQABAL04 that is used in the quality control testing laboratory to weigh for analysis. Laboratory analysts have the ability to change instrument …

    Opens with FDA's citation wording for 21 CFR 211.68(b)Read the Form 483 →

  11. Sun Pharmaceuticals Industries, Inc.Inspection ended Nov 16, 2016 · FEI 3002807979 · Observation 5

    The quality control unit lacks authority to review production records to assure that no errors have occurred. Specifically, SOP OPO13152 "Good Documentation Practices (GDP)" effective date July 28, 2016 explains that Quality Assurance is to periodically review electronic data as a part of self-inspection. However, periodic review of electronic data is only being performed for the quality control testing laboratory and is not being is not …

    Contains “electronic data”Read the Form 483 →

  12. Sekisui MedicalInspection ended Jun 17, 2016 · FEI 3002806840 · Observation 1

    … Review of electronic files from the HPLC and GC computers found the presence of hundreds of deleted or unreported chromatograms. Deleted files were also observed on the IR-3 system. Examples of deleted or unused chromatograms included: a. The computer for GC-4 and GC-6 had deleted individual chromatogram files and deleted folders containing sequences with associated chromatograms in the "Recycle Bin". Deleted data for testing of lot" as part of …

    Contains “delet* … chromatogram*”Read the Form 483 →

  13. Sekisui MedicalInspection ended Jun 17, 2016 · FEI 3002806840 · Observation 2

    … The data back-up process for the standalone HPLC and GC systems does not include back-up of audit trails associated with the analysis. During the inspection the stand-alone computer associated with HPLC-19 was not functional. Audit trail information could not be retrieved from the back-up. 3. Electronic chromatography data is stored on stand-alone systems for seven of the) | HPLCs and one of the PT GCs. There are no controls to prevent deletion …

    Contains “audit trail*”Read the Form 483 →

  14. Cosma SpAInspection ended Jan 15, 2016 · FEI 3002806754 · Observation 1

    Observation 1 Failure to maintain complete data derived from all testing and to ensure compliance with established specifications and standards pertaining to data retention and management. Specifically, I. 1. Data can be overwritten and/or deleted on 4 out of the HPLCs and 2 out of @ GCs used in the testing of APIs manufactured for the U.S. market. These systems do not have audit trails or password protection for individual operators. [illegible] …

    Contains “audit trail*”Read the Form 483 →

  15. Cosma SpAInspection ended Jan 15, 2016 · FEI 3002806754 · Observation 9

    … Documentation is not performed contemporaneously during the execution of batch records. Production operators document time requirements to the nearest 5 minutes.

    Contains “contemporaneous*”Read the Form 483 →

  16. Zhejiang Medicine Co.Inspection ended Jun 18, 2015 · FEI 3003631275 · Observation 2

    OBSERVATION 2 QC Laboratory records are not documented contemporaneously. Specifically, A) During our inspection of your finished API Quality Control Laboratory on June 15, 2015, we observed that a QC analyst pasted balance weighing slips, which were generated on June 11, 2015, onto the “Testing Original Records of®@ DS” worksheet (Lot #: 9 and signed her name and back dated as June 11, 2015. B) During our inspection of your finished API Quality …

    Contains “contemporaneous*”Read the Form 483 →

  17. Zhejiang Medicine Co.Inspection ended Jun 18, 2015 · FEI 3003631275 · Observation 4

    … written document describing the User Requirement Specifications (URS) B) There are a total of three common login IDs (Admin, Manager, Operator) with various access levels. There are no unique usernames/passwords established. There is no written procedure to define the access level (e.g. ability to alter recipe) for each of the three login IDs. C) The validation package was written, performed, and reviewed by a third party contractor. There is no …

    Contains “access level*”Read the Form 483 →

  18. Zhejiang Medicine Co.Inspection ended Jun 18, 2015 · FEI 3003631275 · Observation 5

    OBSERVATION 5 Electronic records are not adequately maintained, archived and retrieved. Specifically, . : (0) (4) A) The electronic raw data for your ©®@p3001 and D3002, controlled using the and 5h software, respectively, is transferred onto a USB drive and then to another desktop for archival. After transferring to this second desktop, the original electronic data on the first desktop and USB drive are deleted. There are no written procedures …

    Contains “electronic raw data”Read the Form 483 →

  19. Zhejiang Medicine Co.Inspection ended Jun 18, 2015 · FEI 3003631275 · Observation 7

    OBSERVATION 7 Laboratory electronic raw data is not controlled and protected. Specifically, during our review of the electronic potency raw data for collected using the ZY-3001V Multifunction Microbiology Analyzer, we noted the following deficiencies regarding data security: A) There is no provision in place to create raw data records for each analysis performed. Test results are required to be saved manually. B) There are no individual user …

    Contains “electronic raw data”Read the Form 483 →

  20. Boston Analytical, Inc.Inspection ended Nov 9, 2012 · FEI 1221426 · Observation 1

    … b.) SOP P-1050, Version 16, Laboratory Investigation Policy, is not being followed which states testing into compliance without regards for scientific justification is not acceptable. For analysis of Levofloxacin (Project Number (b)(4), Sample #(b)(4)), Assay and Impurities test reported OOS results on 10/06/11, the sample was tested three times and the third analytical findings were reported on C of A. c.) SOP P-510, Version Number: 12 Log-In of …

    Contains “test* into compliance”Read the Form 483 →

21 CFR sections these observations open with

The sections FDAgent ties the most of these observations to by FDA's wording, with how many of this theme's observations each covers. Each links to its page of FDA citations, warning letters and the regulation text.

Ask FDAgent about data integrity in Form 483s

FDAgent reads FDA's Form 483s, inspections, citations and warning letters and answers in plain English, with links to the records.

Edit it or ask as written. 3 questions free, no card needed.