Lights Medical Manufacture Co Ltd
19 Quanda Road, Wuqing District, China
FEI 3005310494DUNS 529128649
FDA has 4 inspections on file for Lights Medical Manufacture Co Ltd in Wuqing District, China (FEI 3005310494), from 2013 to 2024. The latest ended on and was classified VAI, voluntary action indicated.
VAI means FDA investigators found objectionable conditions or practices, but FDA was not prepared to take or recommend regulatory action. The firm is expected to correct them voluntarily.
FDA posted the Form 483 from that inspection. FDAgent read its 2 observations from the PDF; they are listed below.
1 import refusal is filed under this FEI, on . No warning letter or court action is filed under this FEI.
As of Nov 21, 2026, the site is registered with FDA as a drug establishment for manufacturing, through Dec 31, 2026. Its registrant is Lights Medical Manufacture Co., Ltd. (D-U-N-S 529128649).
Sources:FDA firm profile ↗Form 483 PDF ↗FDA drug establishment registrations ↗Synced Nov 23, 2026
Ask FDAgent about this site
Email me when FDA posts a new inspection, Form 483 or warning letter for this site
Weekly, on a free account. Pause it any time.
Comparing suppliers? Check this site beside 2 others, free
- Latest inspection
- VAI
- Inspections on file
- 4, 2013–2024
- 0 OAI · 4 VAI · 0 NAI
- Inspections with 483 observations shown
- 1 of 4
- Form 483s posted by FDA
- ↗
- Warning letters
- None on file
- Import refusals
- 1 refusal
- Import alerts
- Not linked by FEI
- FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗
- Drug establishment registration
- Through
Inspection history
4 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC
- VAIObjectionable conditions found; FDA took no regulatory actionDrugs · Drug Quality AssurancePosted 483 · 2 observations
Read this Form 483 on its own page →Form 483 PDF on fda.gov
Read from the FDA-posted Form 483 PDF by OCR. FDAgent keeps an observation only when it matches the scanned text; check the PDF before quoting. The short titles are FDAgent’s.
- Observation 1
Change control not completed for equipment
Procedures applicable to the quality unit are not fully followed Specifically, The firm's SOP LIG/PCD-ZL-09-E/4 "Change Control Procedure", effective date: 10/16/2024, requires change control to be completed for all equipment used directly in production and quality control testing that impacts product quality. However, your firm did not have a change control documentation for the Potentiometric Titrator (ID: HY-121) and pH Meter (ID: HY-120), which are used to test the assay and pH of finished over-the-counter drug (OTC) drug products, including [illegible].
- Observation 2
No equipment use logs maintained
Written records of major equipment use are not included in individual equipment logs. Specifically, Your firm does not maintain equipment use logs for equipment used in the manufacturing process of over-the-counter drug products. This includes, but is not limited to, the following equipment: Name of Manufacturing Equipment Equipment ID(s) [illegible]
- VAIObjectionable conditions found; FDA took no regulatory actionDrugs · Drug Quality AssuranceCitations not published
- VAIObjectionable conditions found; FDA took no regulatory actionDrugs · Drug Quality AssuranceCitations not published
- VAIObjectionable conditions found; FDA took no regulatory actionDrugs · Drug Quality AssuranceCitations not published
Import refusals
FDA Data Dashboard, last synced Nov 23, 2026, 04:17 UTC. Shipments FDA refused entry into the US, filed under FEI 3005310494, one row per refused product line.
1 refusal on 2019-04-01 · charged 708NODFL on 1 · FDA sample analysis on 1
FDA firm profile- 2019-04-01
ISOPROPYL ALCOHOL (ANTI-INFECTIVE, TOPICAL)62FAY23
708NODFL 502(c), 801(a)(3); MISBRANDING
FDA sample analysis
Division of Southwest Imports9M3-1057089-3/21/1
FDA drug establishment registration
FDA registration search- Registered operations
- Manufacturing
- Registration runs through
- Dec 31, 2026
- Registrant
- Lights Medical Manufacture Co., Ltd.D-U-N-S 529128649
From FDA’s Drug Establishments Current Registration Site, as of Nov 21, 2026. A registration says what a site told FDA it does. It is not an inspection result. FDA’s registration search looks firms up by name.
Go further than this page
FDAgent researches FDA’s records and links every answer to its source.
About this data. Every record here is filed under FEI 3005310494 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
- Every record on this page is filed under FEI 3005310494. Records filed under another FEI or name, including other sites of the same company, are not shown here.
- FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
- Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
- “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
- Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
- FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.