Quality unit responsibilities in FDA Form 483 observations
The quality control unit's authority and duties: approving or rejecting materials and batches, reviewing records and procedures, and overseeing investigations. FDAgent finds 141 such observations in 103 Form 483s issued to 100 facilities, for inspections that ended between March 2012 and August 2026. Each is quoted below from the 483 FDA posted.
A Form 483 lists the investigators' observations at the end of an inspection. It is not FDA's final determination.
Source: Form 483 PDFs in FDA's Electronic Reading Room, last synced Nov 23, 2026, 04:33 UTC. Every quote is from a 483 whose reading passed FDAgent's checks against the scan. FDAgent is independent of FDA.
- Observations
- 141
- Each quoted below
- Form 483s
- 103
- Posted by FDA
- Facilities
- 100
- Distinct FEIs
- Most recent
- Aug 12, 2026
- Inspection end date
Observations by year
These observations by the year the inspection ended. They follow which 483s FDA has posted and FDAgent has read, so they show what is on record, not a trend in FDA's findings. 2026 runs to the latest sync (lighter bar).
Peak: 50 in 2026
| Year | Observations |
|---|---|
| 2012 | 1 |
| 2013 | 0 |
| 2014 | 0 |
| 2015 | 1 |
| 2016 | 5 |
| 2017 | 5 |
| 2018 | 4 |
| 2019 | 2 |
| 2020 | 0 |
| 2021 | 3 |
| 2022 | 9 |
| 2023 | 17 |
| 2024 | 12 |
| 2025 | 32 |
| 2026 | 50 |
How FDAgent sorts observations into this theme
Two fixed rules, applied by code to every observation FDAgent can quote. An observation counts when either holds. No model and no similarity score decides.
It is tied to one of these 21 CFR entries
A 483 prints no regulation, so FDAgent ties an observation to a section only when it opens with FDA's own citation wording for that section, word for word.
Or its text contains one of these phrases
“quality unit*” · “quality control unit*” · “QCU” · “QU” · “quality assurance unit*” · “quality assurance department*” · “QA unit*”
* matches any ending and … up to three words in between. Case is ignored unless a phrase has capitals.
The 141 observations, newest first
Quoted word for word from each Form 483, with only line breaks joined; … marks where the quote is cut. The highlight shows why the observation is in this theme. Each links to FDAgent's page for the 483.
Farmakeio Outsourcing LLCInspection ended Sep 5, 2025 · FEI 3014982757 · Observation 1
… Additionally, your firm's metal detection system for finished pellets was not operational until May 19, 2025, metal detection performance qualification (PQ) quality unit approval date. Before this date, no finished pellets underwent metal detection as part of its manufacturing process. The [(D)(4)] intermediate product, which is (b)(4) and then processed into pellets, was visually inspected, packaged, [(6)(4)] sterilized, and distributed without …
Contains “quality unit*”Read the Form 483 →
Aptuit LtdInspection ended Aug 22, 2025 · FEI 3003444315 · Observation 1
The responsibilities applicable to the quality unit are not fully followed. Specifically, Since June 2023, approximately [illegible] batches of [illegible] were released to the market without assurance that they had safety levels of [illegible]. Additionally, the quality unit failed to develop a Risk Assessment of the Presence of [illegible].
Contains “quality unit*”Read the Form 483 →
Aptuit LtdInspection ended Aug 22, 2025 · FEI 3003444315 · Observation 2
There is a failure to thoroughly review any unexplained discrepancy and the failure to [illegible] of a batch or any of its components to meet any of its specifications whether or not the batch has been already distributed. Specifically, The quality unit failed to perform a thorough customer complaint investigation regarding [illegible] batches of [illegible] API that obtained [illegible] contents above the specifications limits. The quality unit …
Contains “quality unit*”Read the Form 483 →
Aptuit LtdInspection ended Aug 22, 2025 · FEI 3003444315 · Observation 3
Procedures designed to prevent microbiological contamination of products purporting to be non-sterile are not established. Specifically, a. The quality unit fails to ensure that the contract laboratory that performs the testing for the [illegible] system used for the manufacture of APIs (i.e. [illegible] and excipient, [illegible]) has adequate and validated test methods. b. The quality control unit failed to ensure that the [illegible] system …
Contains “quality unit*”Read the Form 483 →
Aptuit LtdInspection ended Aug 22, 2025 · FEI 3003444315 · Observation 4
The quality system to approve contract laboratories to perform microbiological testing of APIs, and excipients is deficient. Specifically, 1. The quality unit fails to ensure that the approved contract laboratory used to perform testing of APIs and excipient has the capabilities, equipment, test methods, knowledge, and reliability needed to perform the testing in an accurate and reliable way. 2. On December 11, 2024, the quality unit learned that …
Contains “quality unit*”Read the Form 483 →
Aptuit LtdInspection ended Aug 22, 2025 · FEI 3003444315 · Observation 5
The accuracy, sensitivity, specificity, reproducibility, and reliability of microbiological test methods has not been established. Specifically, The quality unit fails to ensure that the microbiological test methods (e.g., TVC, pathogens identification, and endotoxins) utilized for the release of APIs [e.g., [illegible] (intended for drug product), [illegible] (intended for sterile injectable) and excipient [illegible]] are validated.
Contains “quality unit*”Read the Form 483 →
Aptuit LtdInspection ended Aug 22, 2025 · FEI 3003444315 · Observation 6
The identity of each raw material of drug substance is not verified by conducting at least one test to verify the identity, using specific identity test. Specifically, The quality control unit fails to establish a specific identification test method for a raw material used for the manufacture of API, or [illegible].
Contains “quality control unit*”Read the Form 483 →
Aptuit LtdInspection ended Aug 22, 2025 · FEI 3003444315 · Observation 7
The amount of material taken from each container is not based upon appropriate criteria. Specifically, The quality unit fails to ensure that the amount of raw material taken for testing is based upon appropriate criteria such as criticality of the material, material variability, past history of the supplier, and the quantity needed for analysis.
Contains “quality unit*”Read the Form 483 →
Tripak Pharmaceuticals Pty Ltd.Inspection ended Jul 25, 2025 · FEI 3008549263 · Observation 1
The quality control unit lacks the responsibility and authority to approve, reject all drug products. Specifically, Your firm manufactured several products using inactive ingredients that failed testing. Following are the examples I observed during my review, but are not limited to: a) On 09/24/2024, your firm opened an OOS investigation #122 for one of the inactive ingredients used in the manufacturing of (b) (4) mL), Lot #(b) (4) (Bulk Lot #(b) …
Contains “quality control unit*”Read the Form 483 →
Veganic SKN LimitedInspection ended Jul 21, 2025 · FEI 3024609429 · Observation 1
OBSERVATION 1 The responsibilities and procedures applicable to the quality control unit are not in writing. Specifically, Your firm has not established written procedures defining the quality control unit's responsibilities and authority. Standard operating procedures describing the quality control unit's role and scope of authority were not available during the inspection. As a result, personnel were unclear about their specific duties and …
Opens with FDA's citation wording for 21 CFR 211.22(d)Read the Form 483 →
Veganic SKN LimitedInspection ended Jul 21, 2025 · FEI 3024609429 · Observation 2
OBSERVATION 2 The quality control unit lacks the responsibility and authority to approve, reject all drug products. Specifically, a) Manufacturing operations are currently proceeding without qualified Quality Assurance (QA) oversight present during critical process steps. This deficiency includes a lack of real-time review and approval of critical process parameters and deviations by qualified QA personnel. Additionally, there is insufficient QA …
Contains “quality control unit*”Read the Form 483 →
Guangxi Yulin Pharmaceutical Group Co. Ltd.Inspection ended Mar 21, 2025 · FEI 1000589993 · Observation 7
… (B) Your quality unit failed to implement appropriate change control procedures following changes to your supplier qualification process. For example, you changed your procedures for conducting routine supplier audits. However, these changes were not included in a change control process, and you do not have justification for why these changes were not incorporated into a change control.
Contains “quality unit*”Read the Form 483 →
Hangzhou Minsheng Binjiang Pharmaceutical Co., Ltd.Inspection ended Mar 7, 2025 · FEI 3007498452 · Observation 3
The responsibilities and procedures applicable to the quality control unit are not fully followed. Specifically, during the inspection, we observed torn and discarded GMP documents in the trash bin located in the QC laboratory. Your firm was unaware that these GMP documents were altered and discarded. GMP documents are not managed according to your procedures, document # 02-QA-2-001 and 02-QA-2-004.
Opens with FDA's citation wording for 21 CFR 211.22(d)Read the Form 483 →
Anhui Poly Pharm Co., LtdInspection ended Feb 27, 2025 · FEI 3020871572 · Observation 1
In-process materials are not tested for quality by the quality control unit after storage for long periods. Specifically, In the bulk hold time studies that your firm performed for the following API products, you did not include a study study samples and then placement of the finished API on stability: that assessed the bulk intermediate AP(b)(4) (b)(4) (b)(4) The allowable bulk hold time for (b)(4) API intermediates (b)(4) is (b)(4)
Contains “quality control unit*”Read the Form 483 →
TrueCare BiomedixInspection ended Feb 21, 2025 · FEI 3010166683 · Observation 1
The responsibilities and procedures applicable to the quality control unit are not fully followed. Specifically, written procedures do not include sufficient details to ensure consistent implementation and written procedures are not always followed in that: i) your procedure for good documentation practices, SOP No: QA-013-01, titled Procedure for Maintaining Good Documentation Practice, Effective January 1, 2024 states, "All documents should …
Opens with FDA's citation wording for 21 CFR 211.22(d)Read the Form 483 →
Shouguang Fukang Pharmaceutical Co., Ltd.Inspection ended Feb 21, 2025 · FEI 3003941038 · Observation 1
The quality control unit lacks responsibility to approve final testing results for product release. Specifically, a review of completed analytical testing records for the API [illegible] regarding HPLC L-002 and GC L-0[illegible] testing results noted that QC performs the review/approval without review/approval by the QA department. For example, the chemist uses Excel Spreadsheets to manual input testing result data from both the HPLC and GC. The …
Contains “quality control unit*”Read the Form 483 →
Piramal, Piramal Pharma LimitedInspection ended Feb 17, 2025 · FEI 3008763768 · Observation 1
OBSERVATION 1 The responsibilities and procedures applicable to the quality control unit are not in writing and fully followed. Specifically, A. On February 11th, 2025, I conducted a walkthrough of Room [illegible] used to [illegible]. Approximately 1:10 pm, I observed liquid present throughout the floor of the Room and visually leaking from the adjacent (b)(4) vessel. (b)(4) was used to store the (b)(4) material. (b)(4) was also used for …
Opens with FDA's citation wording for 21 CFR 211.22(d)Read the Form 483 →
Piramal, Piramal Pharma LimitedInspection ended Feb 17, 2025 · FEI 3008763768 · Observation 3
… According to SOP CIC/QA/11 Handling of Customer Complaints, Version 10, Effective Date: February 22, 2022, Section 5.5, the Quality Control unit must examine the retain sample of the complaint batch and check for physical parameters. In addition, compare the retain sample with the sample, if received along with the complaint.
Contains “quality control unit*”Read the Form 483 →
Glenmark Pharmaceuticals Ltd.Inspection ended Feb 14, 2025 · FEI 3008565058 · Observation 5
The quality control unit lacks responsibility to approve all procedures or specifications impacting on the quality of drug products. Specifically, Since 01 Jan 2023, approximately [illegible] US commercial samples [illegible] completed in a timely manner. These testing delays at [illegible] stations as shown below: This delay is potentially causing several commercial batches to be in the market without having concurrently [illegible] supporting …
Opens with FDA's citation wording for 21 CFR 211.22(c)Read the Form 483 →
Laurus Labs LimitedInspection ended Jan 31, 2025 · FEI 3014255847 · Observation 1
Written procedures, including changes thereof, are not drafted, reviewed and approved by appropriate organization units, and approved by the quality control unit. Specifically, You failed to ensure that equipment qualification protocols are reviewed and approved by the quality unit. During the inspection, we observed multiple equipment used for the manufacturing of active pharmaceutical ingredients (i.e., [illegible]) qualified without an …
Contains “quality control unit*”Read the Form 483 →
21 CFR sections these observations open with
The sections FDAgent ties the most of these observations to by FDA's wording, with how many of this theme's observations each covers. Each links to its page of FDA citations, warning letters and the regulation text.
- 21 CFR 211.22Responsibilities of quality control unitIn this theme's rule46
- 21 CFR 211.192Production record review6
- 21 CFR 211.113Control of microbiological contamination5
- 21 CFR 211.165Testing and release for distribution5
- 21 CFR 211.63Equipment design, size, and location2
- 21 CFR 211.100Written procedures; deviations2
- 21 CFR 211.160General requirements2
- 21 CFR 211.180General requirements2
- 21 CFR 211.25Personnel qualifications1
- 21 CFR 211.42Design and construction features1
Ask FDAgent about quality unit in Form 483s
FDAgent reads FDA's Form 483s, inspections, citations and warning letters and answers in plain English, with links to the records.
Related
About this data
- The observations are the text FDAgent read by OCR from Form 483 PDFs in FDA's Electronic Reading Room, shown only from 483s whose every printed observation passed FDAgent's checks against the scanned text, as on each 483's own page.
- Which theme an observation belongs to is decided by the fixed rules above; every count on this page is computed from them.
- A 483 addressed to an individual is listed without the name, and an observation that names a person is left out.
- FDAgent is independent of FDA.