Quality unit responsibilities in FDA Form 483 observations
The quality control unit's authority and duties: approving or rejecting materials and batches, reviewing records and procedures, and overseeing investigations. FDAgent finds 141 such observations in 103 Form 483s issued to 100 facilities, for inspections that ended between March 2012 and August 2026. Each is quoted below from the 483 FDA posted.
A Form 483 lists the investigators' observations at the end of an inspection. It is not FDA's final determination.
Source: Form 483 PDFs in FDA's Electronic Reading Room, last synced Nov 23, 2026, 04:33 UTC. Every quote is from a 483 whose reading passed FDAgent's checks against the scan. FDAgent is independent of FDA.
- Observations
- 141
- Each quoted below
- Form 483s
- 103
- Posted by FDA
- Facilities
- 100
- Distinct FEIs
- Most recent
- Aug 12, 2026
- Inspection end date
Observations by year
These observations by the year the inspection ended. They follow which 483s FDA has posted and FDAgent has read, so they show what is on record, not a trend in FDA's findings. 2026 runs to the latest sync (lighter bar).
Peak: 50 in 2026
| Year | Observations |
|---|---|
| 2012 | 1 |
| 2013 | 0 |
| 2014 | 0 |
| 2015 | 1 |
| 2016 | 5 |
| 2017 | 5 |
| 2018 | 4 |
| 2019 | 2 |
| 2020 | 0 |
| 2021 | 3 |
| 2022 | 9 |
| 2023 | 17 |
| 2024 | 12 |
| 2025 | 32 |
| 2026 | 50 |
How FDAgent sorts observations into this theme
Two fixed rules, applied by code to every observation FDAgent can quote. An observation counts when either holds. No model and no similarity score decides.
It is tied to one of these 21 CFR entries
A 483 prints no regulation, so FDAgent ties an observation to a section only when it opens with FDA's own citation wording for that section, word for word.
Or its text contains one of these phrases
“quality unit*” · “quality control unit*” · “QCU” · “QU” · “quality assurance unit*” · “quality assurance department*” · “QA unit*”
* matches any ending and … up to three words in between. Case is ignored unless a phrase has capitals.
The 141 observations, newest first
Quoted word for word from each Form 483, with only line breaks joined; … marks where the quote is cut. The highlight shows why the observation is in this theme. Each links to FDAgent's page for the 483.
RayzeBio, IncInspection ended Feb 24, 2026 · FEI 3030530655 · Observation 4
… The Quality Unit failed to adequately investigate and implement corrective actions after identifying the CFUs, determine microorganism sources or trace them to production operations, and evaluate potential quality impact on sterile drug product manufacturing. B. A review of the firm's environmental monitoring excursions list (EME (b)(4)) for the period from 05/20/2025 to 01/20/2026 revealed that approximately 55 of 113 EME investigations remain …
Contains “quality unit*”Read the Form 483 →
Orean Personal Care LtdInspection ended Feb 20, 2026 · FEI 3010999645 · Observation 2
The responsibilities and procedures applicable to the quality control unit are not in writing and/or fully followed. Specifically, A. You released the batches of finished products before completing all the test to ensure the OTC drug products meet all the specification. Your procedure, WIPROD027 OTC Products Sampling and Processes Version 12 (Issued on 11/28/2024) require a review of all the test to ensure the product meets specification for the …
Opens with FDA's citation wording for 21 CFR 211.22(d)Read the Form 483 →
Almon Healthcare Private LimitedInspection ended Feb 13, 2026 · FEI 3030509366 · Observation 1
Failure of your quality unit to exercise its responsibility to ensure the active pharmaceutical ingredient (API) and intermediates manufactured at your facility are in compliance with CGMP. a. Your Quality unit lacked adequate control over the shredder used in the QA document control room. On February 9, 2026, a shredder was found three-fourths full of shredded quality related documents. One partially recognizable fragment contained a batch …
Contains “quality unit*”Read the Form 483 →
Biomylz Private LimitedInspection ended Feb 9, 2026 · FEI 3013697420 · Observation 4
The responsibilities and procedures applicable to the quality control unit are not fully followed. Specifically, A. There is no assurance that laboratory testing used to determine raw material suitability and finished drug product quality is performed in accordance with applicable cGMP requirements. For example, your firm uses contracted laboratories that are not registered or inspected by the agency to perform incoming raw material and finished …
Opens with FDA's citation wording for 21 CFR 211.22(d)Read the Form 483 →
SNJ Labs Private Ltd.Inspection ended Feb 6, 2026 · FEI 3026587575 · Observation 1
Failure of your quality unit to ensure that critical deviations are investigated and resolved. a. Your firm had four OOS results for the (b) (4) content in the finished active pharmaceutical ingredient (API) (b) (4) from April-September 2024. Root causes identified included loose wire connections to the (b) (4) instrument, black coating on the (b) (4) poor (b) (4) performance, and instrument malfunction. However, no preventive maintenance was …
Contains “quality unit*”Read the Form 483 →
Neeyaan LLCInspection ended Jan 27, 2026 · FEI 3024366553 · Observation 4
The responsibilities and procedures applicable to the quality control unit are not in writing and fully followed. Specifically, A) A deviation was not initiated, in accordance with SOP-50 "Deviations", on 01/14/2026 when compression operations for (b) (4) Lot #(b) (4) tablets on the (b) (4) Tooling Tablet Press ((b) (4)) (Compression Room #(4) were halted, per the Interim Production Supervisor, supposedly due to tablet sticking and breaking …
Opens with FDA's citation wording for 21 CFR 211.22(d)Read the Form 483 →
Shimoga ChemicalsInspection ended Jan 23, 2026 · FEI 3033881432 · Observation 9
The responsibilities and procedures applicable to the quality control unit are not in writing and fully followed. Specifically, A. Your firm failed to follow its vendor qualification procedure. Vendor qualification questionnaires were available for only (b)(4) raw materials, and your firm did not document a rationale for limiting vendor qualification to these materials. B. Your firm failed to establish quality agreements with your broker and …
Opens with FDA's citation wording for 21 CFR 211.22(d)Read the Form 483 →
Genzyme Ireland LimitedInspection ended Jan 20, 2026 · FEI 3003809840 · Observation 2
The responsibilities and procedures applicable to the quality control unit are not in writing and fully followed. Specifically, your procedure requires a change control to be issued for implementing changes. This requirement is delineated in SOP (b) (4)-0000514 titled "Change Control Management". However, several examples were found where a Work Order was used for conducting investigations instead of following the established change control …
Opens with FDA's citation wording for 21 CFR 211.22(d)Read the Form 483 →
US Specialty Formulations LLCInspection ended Jan 14, 2026 · FEI 3010680515 · Observation 1
… market without assay and two finished products lots (Lot#02RS1502A was not performed for the lots produced pr (B) Your quality unit failed to ensure that potency and i Sarracenia Purpurea Bulk Distillate were performed filling of Sarracenia Purpurea 0.17 g/mL for released lots of Sarracenia Purpurea Bulk Distillate were Purpurea 0.17 g/mL for Injection, totaling approximately[( (C) Your quality unit failed to ensure that sufficient data support …
Contains “quality unit*”Read the Form 483 →
US Specialty Formulations LLCInspection ended Jan 14, 2026 · FEI 3010680515 · Observation 2
… The testing methodology invol Quality Unit did not confirm that appropriate - was completed by your third party lab.
Contains “quality unit*”Read the Form 483 →
Smithfield Biosciencs, Inc.Inspection ended Dec 23, 2025 · FEI 1527761 · Observation 3
Batch records and laboratory records of critical steps are not reviewed and approved by the quality unit appropriately. Specifically, A.) Your analytical laboratory paper records do not include signature and date of the second person reviewing (b) (4) (b) (4) impurity data on HPLC-1 for batches (b) (4) These batches were later released by QA without investigating whether they were reviewed or not. In addition, your firm does not include an …
Contains “quality unit*”Read the Form 483 →
Smithfield Biosciencs, Inc.Inspection ended Dec 23, 2025 · FEI 1527761 · Observation 4
… Your quality unit failed to get a positive identification on the "black specks". (b) (4) (b) (4) and released on 12/16/2025. batches were reprocessed into (b) (4) was rejected. C.) In October 2024, your firm recorded two out-of-specification (OOS) results for Limit of (b) (4) in batches (b) (4) (b) (4) in Batch (b) (4) was retested twice, and the OOS result was confirmed on both occasions. Batch (b) (4) was retested six times, with the OOS result …
Contains “quality unit*”Read the Form 483 →
Wizcure Pharmaa Private LimitedInspection ended Dec 10, 2025 · FEI 3030304532 · Observation 1
… Your analysts harbor access and permissions to generate and print data sheets without Quality Unit oversight. F. You lack control over certificate of analysis creation such as version control or page numbering. Specifically, we observed uncontrolled access to your PC used to generate and print CoAs. This PC lacks usage logs and lacks an audit trail. Furthermore, we observed numerous CoAs discovered in your trash on December 3rd, 2025.
Contains “quality unit*”Read the Form 483 →
Wizcure Pharmaa Private LimitedInspection ended Dec 10, 2025 · FEI 3030304532 · Observation 7
The responsibilities and procedures applicable to the quality control unit are not in writing or fully followed. A. You do not document quality system events such as deviation, OOSs and changes. 1. No deviations or OOS were generated for the EM/PM deficiencies nor for any discarded QC records identified during the inspection. 2. per Microbiology (b)(4) testing forms changes were made approximately (b)(4) (b)(6) samples. No Manager, subsequently …
Opens with FDA's citation wording for 21 CFR 211.22(d)Read the Form 483 →
Huons Co., LtdInspection ended Nov 21, 2025 · FEI 3007163376 · Observation 4
… No smoke study was completed for Ampoule Filling Line (b)(4) (Grade A area), and there was no risk assessment completed by your Quality Unit to scientifically justify the lack of a smoke study. Additionally, a (b)(4) was applied to one of the bolts within the Grade A area to secure it; however, there was no documentation available to confirm. There is no written procedure to require documentation of the (b)(4) being (b)(4) for the Ampoule Filling …
Contains “quality unit*”Read the Form 483 →
Eugia Pharma Specialities LimitedInspection ended Nov 14, 2025 · FEI 3009883410 · Observation 1
The responsibilities and procedures applicable to the Quality Unit are not fully followed. Aseptic processing operations are not adequately designed to provide sufficient visual monitoring of critical interventions and aseptic manipulations to ensure compliance with current good manufacturing practice (CGMP) requirements. For example, during inspection of aseptic filling lines (b) (4) it was observed that multiple corrective interventions (b) (4) …
Contains “quality unit*”Read the Form 483 →
Lonza Guangzhou Pharmaceutical Ltd.Inspection ended Oct 31, 2025 · FEI 3006532186 · Observation 1
OBSERVATION 1 The responsibilities applicable to the quality unit are not fully followed. Specifically, a. Since the year 2023, approximately [illegible] batches of API [illegible] (codename [illegible]) were released to the U.S. market without assurance that they had safety levels of [illegible] impurities (e.g., [illegible]). b. Since the year 2023, approximately [illegible] batches of [illegible] API (codename [illegible]) were released to the …
Contains “quality unit*”Read the Form 483 →
Lonza Guangzhou Pharmaceutical Ltd.Inspection ended Oct 31, 2025 · FEI 3006532186 · Observation 2
OBSERVATION 2 There is a failure to thoroughly review any unexplained discrepancy and the failure to [illegible] of a batch or any of its components to meet any of its specifications whether or not the batch has been already distributed. Specifically, The quality unit failed to provide adequate complaint investigations. For example, a. The complaint investigation associated to the presence of black particles during the use of API, batch …
Contains “quality unit*”Read the Form 483 →
Lonza Guangzhou Pharmaceutical Ltd.Inspection ended Oct 31, 2025 · FEI 3006532186 · Observation 3
OBSERVATION 3 The quality system that approves contract laboratories to perform testing of [illegible] impurities is deficient. Specifically, The quality unit fails to ensure that the contract laboratory used to perform testing of [illegible] content in API (codename [illegible]) has the capabilities, equipment, and the knowledge to perform the required testing in an accurate, precise, and reliable way.
Contains “quality unit*”Read the Form 483 →
UBI Pharma Inc.Inspection ended Oct 14, 2025 · FEI 3004713402 · Observation 4
The responsibilities and procedures applicable to the quality control unit are not in writing and/or fully followed. Specifically, A. You manufacture (b)(4) injection for your US customer. You do not have a Quality Agreement with your (b)(4). The initial QA was signed on 5/20/2019 and was effective for five years until 5/19/2024. You did not review or renew the QA with [illegible]. B. Your initial QA did not have a provision [illegible]. Section …
Opens with FDA's citation wording for 21 CFR 211.22(d)Read the Form 483 →
21 CFR sections these observations open with
The sections FDAgent ties the most of these observations to by FDA's wording, with how many of this theme's observations each covers. Each links to its page of FDA citations, warning letters and the regulation text.
- 21 CFR 211.22Responsibilities of quality control unitIn this theme's rule46
- 21 CFR 211.192Production record review6
- 21 CFR 211.113Control of microbiological contamination5
- 21 CFR 211.165Testing and release for distribution5
- 21 CFR 211.63Equipment design, size, and location2
- 21 CFR 211.100Written procedures; deviations2
- 21 CFR 211.160General requirements2
- 21 CFR 211.180General requirements2
- 21 CFR 211.25Personnel qualifications1
- 21 CFR 211.42Design and construction features1
Ask FDAgent about quality unit in Form 483s
FDAgent reads FDA's Form 483s, inspections, citations and warning letters and answers in plain English, with links to the records.
Related
About this data
- The observations are the text FDAgent read by OCR from Form 483 PDFs in FDA's Electronic Reading Room, shown only from 483s whose every printed observation passed FDAgent's checks against the scanned text, as on each 483's own page.
- Which theme an observation belongs to is decided by the fixed rules above; every count on this page is computed from them.
- A 483 addressed to an individual is listed without the name, and an observation that names a person is left out.
- FDAgent is independent of FDA.