Equipment cleaning and cleaning validation in FDA Form 483 observations
Cleaning procedures for manufacturing equipment, the studies that show they work, and the residues, swabs and logs that prove each cleaning. FDAgent finds 197 such observations in 144 Form 483s issued to 138 facilities, for inspections that ended between October 2008 and September 2026. Each is quoted below from the 483 FDA posted.
A Form 483 lists the investigators' observations at the end of an inspection. It is not FDA's final determination.
Source: Form 483 PDFs in FDA's Electronic Reading Room, last synced Nov 23, 2026, 04:33 UTC. Every quote is from a 483 whose reading passed FDAgent's checks against the scan. FDAgent is independent of FDA.
- Observations
- 197
- Each quoted below
- Form 483s
- 144
- Posted by FDA
- Facilities
- 138
- Distinct FEIs
- Most recent
- Sep 28, 2026
- Inspection end date
Observations by year
These observations by the year the inspection ended. They follow which 483s FDA has posted and FDAgent has read, so they show what is on record, not a trend in FDA's findings. 2026 runs to the latest sync (lighter bar).
Peak: 63 in 2026
| Year | Observations |
|---|---|
| 2008 | 1 |
| 2009 | 0 |
| 2010 | 0 |
| 2011 | 0 |
| 2012 | 0 |
| 2013 | 0 |
| 2014 | 1 |
| 2015 | 1 |
| 2016 | 10 |
| 2017 | 15 |
| 2018 | 20 |
| 2019 | 5 |
| 2020 | 5 |
| 2021 | 10 |
| 2022 | 7 |
| 2023 | 22 |
| 2024 | 13 |
| 2025 | 24 |
| 2026 | 63 |
How FDAgent sorts observations into this theme
Two fixed rules, applied by code to every observation FDAgent can quote. An observation counts when either holds. No model and no similarity score decides.
It is tied to one of these 21 CFR entries
The same entries the equipment cleaning topic counts. A 483 prints no regulation, so FDAgent ties an observation to a section only when it opens with FDA's own citation wording for that section, word for word.
Or its text contains one of these phrases
“cleaning validat*” · “validat* … cleaning” · “cleaning procedure*” · “cleaning process*” · “cleaning record*” · “cleaning log*” · “cleaning agent*” · “cleaning verification” · “cleaning stud*” · “cleaning method*” · “cleaning effectiveness” · “not cleaned” · “not adequately cleaned” · “not properly cleaned” · “residue*” · “swab*” · “carry over” · “carryover”
* matches any ending and … up to three words in between. Case is ignored unless a phrase has capitals.
The 197 observations, newest first
Quoted word for word from each Form 483, with only line breaks joined; … marks where the quote is cut. The highlight shows why the observation is in this theme. Each links to FDAgent's page for the 483.
Unilab Chemicals and Pharmaceuticals Private Ltd.Inspection ended Feb 27, 2026 · FEI 3004985767 · Observation 1
… This includes the lack of verification for Identity test results, (b) (4) Limit test results, (b) (4) Limit test results, Residue on Evaporation test results, pH test results, Solvent Acidity test results, Denaturant Level test results, and/or Impurity test results. In addition, the non-USP in-house test method used for Gas Chromatography (GC) purity testing has not been validated. According to the QA Manager, the test parameters for the in-house …
Contains “residue*”Read the Form 483 →
Unilab Chemicals and Pharmaceuticals Private Ltd.Inspection ended Feb 27, 2026 · FEI 3004985767 · Observation 3
… The Microbiology Department's Cleaning Record of Equipment Log No. F/MIC/008 was pre-dated with equipment cleaning (Frequency: (b) (4)) already completed on (b) (4), which is (b) (4) since the last equipment cleaning documented for completion on (b) (4). The logbook entry pre-dated for (b) (4) also included Done By, Checked By, and Reviewed By signatures that were already filled in as of (b) (4) C. According to the QA Manager, the Done By …
Contains “cleaning record*”Read the Form 483 →
Unilab Chemicals and Pharmaceuticals Private Ltd.Inspection ended Feb 27, 2026 · FEI 3004985767 · Observation 5
There is a failure to clean equipment and utensils to prevent contamination or carry-over of a material that would alter the quality of the API beyond the official or other established specifications. Specifically, There are no standard operating procedures or established instructions to perform and document a visual inspection of equipment subsequent to cleaning operations. For example: During my walkthrough of the Bulk Drug (API) Manufacturing …
Contains “carry over”Read the Form 483 →
RayzeBio, IncInspection ended Feb 24, 2026 · FEI 3030530655 · Observation 2
… Your firm lacks post-batch cleaning procedures to prevent contamination. SOP-00098 ("Cleaning Procedure: (b)(4) ISO 5 (b)(4)," v2, effective 04/11/2025) requires only (b)(4) cleaning of the (b)(4) (ISO 5) and production room (ISO 7). Batch records confirm no post-batch cleaning is performed. No risk assessment or justification supports this practice. D. On 02/02/2026, during the set-up and aseptic fill for (b)(4), lot (b)(4), in Suite (b)(4) …
Contains “cleaning procedure*”Read the Form 483 →
RayzeBio, IncInspection ended Feb 24, 2026 · FEI 3030530655 · Observation 5
Equipment are not cleaned, maintained, sanitized or sterilized at appropriate intervals to prevent malfunctions or contamination that would alter the safety, identity, strength, quality, or purity of the drug product. Specifically, Your firm uses manual cleaning for (b)(4) ISO 5 Isolators in Suite (b)(4) (Room 1012) and Suite (b)(4) (Room 1013) but has not established or validated a decontamination method (e.g., (b)(4)). The firm's Validation …
Contains “not cleaned”Read the Form 483 →
Orean Personal Care LtdInspection ended Feb 20, 2026 · FEI 3010999645 · Observation 1
Written procedures are not established for the cleaning and maintenance of equipment, including utensils, used in the manufacture, processing, packing or holding of a drug product. Specifically, you manufacture Over the Counter (OTC) drug products for the US market on non-dedicated manufacturing equipment. You designed a cleaning matrix that identified most difficult product to clean, (b)(4) FDA product Code No. (b)(4) also called (b)(4). You did …
Opens with FDA's citation wording for 21 CFR 211.67(b)Read the Form 483 →
Orean Personal Care LtdInspection ended Feb 20, 2026 · FEI 3010999645 · Observation 2
… The Cleaning Validation Protocols and Reports were not reviewed and signed by quality assurance. c. The Process Validation Protocol and Reports were not reviewed and signed by quality assurance.
Contains “cleaning validat*”Read the Form 483 →
Dreyer's Grand Ice Cream, Inc.Inspection ended Feb 20, 2026 · FEI 2939502 · Observation 1
… Year Total environmental swabs taken Total confirmed (+) for Listeria spp. Total confirmed (+) for L. monocytogenes 1/1/26-2/5/26 2025 2024 2023 2022 2021 (b)(4) [illegible]. You did not implement your Corrective Action procedure indicated in your "Pathogen Monitoring Program Procedure" which states "If an area is having repeat positives, a team must be formed to investigate the issue. The team will consist of (b)(4) [illegible] should be used in …
Contains “swab*”Read the Form 483 →
MicroBioPharm Japan Co., LtdInspection ended Feb 13, 2026 · FEI 3003992082 · Observation 1
Your firm failed to establish adequate written procedures for cleaning and maintenance of equipment used in the manufacture, processing, or holding of drug substances, and failed to validate cleaning procedures to ensure that equipment is routinely cleaned to prevent contamination. Specifically, Your firm has not validated cleaning procedures for major manufacturing and processing equipment used in the production of active pharmaceutical …
Contains “validat* … cleaning”Read the Form 483 →
Almon Healthcare Private LimitedInspection ended Feb 13, 2026 · FEI 3030509366 · Observation 5
… Cleaning logs showed (b)(4) cleaning on February 10, 2026. ii. (b)(4) stains were visible on walls and ceiling. An incomplete construction area in the corner next to (b)(4)-001 (b)(4) contained apparent dirt like material. iii. (b)(4) clean room: apparent stains were present on walls. An incomplete construction area containing apparent dirt-like material.
Contains “cleaning log*”Read the Form 483 →
Almon Healthcare Private LimitedInspection ended Feb 13, 2026 · FEI 3030509366 · Observation 6
Failure to clean equipment and utensils to prevent contamination or carry-over of a material that would alter the quality of the intermediates and API. a. Your firm uses a (b)(4) (b)(4) to [illegible]. No usage or cleaning logbook exists for this [illegible]. This is used in the manufacturing of different APIs including (b)(4) (b)(4) (b)(4) for the US market. b. On February 9, 2026, a (b)(4) (b)(4) (002) on them [illegible] floor showed apparent …
Contains “carry over”Read the Form 483 →
Almon Healthcare Private LimitedInspection ended Feb 13, 2026 · FEI 3030509366 · Observation 9
Failure to adequately validate written procedures for the cleaning and maintenance of equipment. Your firm has not performed cleaning validation studies for non-dedicated manufacturing equipment (b)(4) including (b)(4). Your firm manufactures APIs including (b)(4) (b)(4) (b)(4) for the US market. Your Executive QA stated that the cleaning verification was carried out with rinse samples and swab testing of the selected equipment, however cleaning …
Contains “cleaning validat*”Read the Form 483 →
Biomylz Private LimitedInspection ended Feb 9, 2026 · FEI 3013697420 · Observation 5
Written procedures are not established for the cleaning and maintenance of equipment, including utensils, used in the manufacture, processing, packing or holding of a drug product. Specifically, your firm failed to develop and approve cleaning procedures describing in sufficient detail the methods, equipment and materials to be used for cleaning the production equipment (i.e., (b)(4) tank, holding tank and filling line) and utensils or auxiliary …
Opens with FDA's citation wording for 21 CFR 211.67(b)Read the Form 483 →
Biomylz Private LimitedInspection ended Feb 9, 2026 · FEI 3013697420 · Observation 6
… In addition, (b)(4) is used as a cleaning agent for equipment and utensils (b)(4) that come into direct contact with the OTC finished product during manufacturing operations.
Contains “cleaning agent*”Read the Form 483 →
SNJ Labs Private Ltd.Inspection ended Feb 6, 2026 · FEI 3026587575 · Observation 4
… The (b) (4) cleaning record showed that the room was last cleaned on (b) (4) and there was no documentation on the logbook for (b) (4) cleaning of ceiling. b. Clean room (b) (4) Plant (b) used for (b) (4) API (b) (4) was not in clean condition on February 2, 2026. Floors showed (b) (4) liquid like substance leaking from (b) (4) (b) (4) of (b) (4) materials scattered on the floor next to (b) (4) and (b) (4) liquid stains on floors and walls. The …
Contains “cleaning record*”Read the Form 483 →
Centrient Pharmaceuticals India Private LimitedInspection ended Jan 27, 2026 · FEI 3004497364 · Observation 1
Failure to clean equipment and utensils to prevent contamination or carry-over of a material that would alter the quality of the intermediates and Active Pharmaceutical Ingredient beyond the official or other established specifications. Specifically, The (b)(4) used at the (b)(4) stages of APIs manufacturing have not been non-dedicated (b)(4) appropriately cleaned since their installation 27+ years ago. For example, A. On 19 and 20 January 2026, …
Contains “carry over”Read the Form 483 →
Neeyaan LLCInspection ended Jan 27, 2026 · FEI 3024366553 · Observation 10
Written procedures are not established and followed for the cleaning and maintenance of equipment, including utensils, used in the manufacture, processing, packing or holding of a drug product. Specifically, A) Your firm has not performed cleaning validation studies and established clean hold times and dirty hold times for equipment, including, but not limited to, the Mixing Machine (b) (4), used to manufacture both dietary supplements and drug …
Opens with FDA's citation wording for 21 CFR 211.67(b)Read the Form 483 →
Neeyaan LLCInspection ended Jan 27, 2026 · FEI 3024366553 · Observation 17
… I reviewed the logbook titled "EQUIPMENT USAGE & CLEANING LOGBOOK - Mixing Machine (b) (4)," and there is no entry of (b) (4) Tablet, Batch No #: (b) (4) using this equipment. Step No. (b) (4) was initialed in the batch manufacturing record as done and checked by QA. Step No. (b) (4) of the "Compression" section states Room (b) (4) Tablet Deduster (b) ID: (b) (4), Metal Detector (b) (4) ID: (b) (4), Tablet Press ID: (b) (4), Tablet Deduster (b) …
Contains “cleaning log*”Read the Form 483 →
Shimoga ChemicalsInspection ended Jan 23, 2026 · FEI 3033881432 · Observation 2
… use (including (b)(4) 003) was routinely moved to non-GMP areas for earlier production stages and returned, with no validation of impact on product quality, cleaning procedures, or verification of clean status. D. Your firm has not performed revalidation for (b)(4) USP API and key starting materials, despite process validation reports for (b)(4) stating, "Revalidation has to be performed by (b)(4)." In addition, the process validation report for …
Contains “cleaning procedure*”Read the Form 483 →
Shimoga ChemicalsInspection ended Jan 23, 2026 · FEI 3033881432 · Observation 3
Batch production and control records do not include complete information relating to the production and control of each batch. Specifically, A. Your firm fails to maintain accurate and consistent cleaning logs and batch records, demonstrating significant data integrity deficiencies in the documentation of (b)(4) 001 cleaning and (b)(4) of (b)(4) USP API. Review of the following batch (b)(4) batches for (b)(4) reveal discrepancies in cleaning …
Contains “cleaning log*”Read the Form 483 →
21 CFR sections these observations open with
The sections FDAgent ties the most of these observations to by FDA's wording, with how many of this theme's observations each covers. Each links to its page of FDA citations, warning letters and the regulation text.
- 21 CFR 211.67Equipment cleaning and maintenanceIn this theme's rule31
- 21 CFR 211.113Control of microbiological contamination10
- 21 CFR 211.42Design and construction features6
- 21 CFR 211.22Responsibilities of quality control unit3
- 21 CFR 211.100Written procedures; deviations2
- 21 CFR 211.192Production record review2
- 21 CFR 211.28Personnel responsibilities1
- 21 CFR 211.56Sanitation1
- 21 CFR 211.58Maintenance1
- 21 CFR 211.94Drug product containers and closures1
Ask FDAgent about equipment cleaning in Form 483s
FDAgent reads FDA's Form 483s, inspections, citations and warning letters and answers in plain English, with links to the records.
Related
About this data
- The observations are the text FDAgent read by OCR from Form 483 PDFs in FDA's Electronic Reading Room, shown only from 483s whose every printed observation passed FDAgent's checks against the scanned text, as on each 483's own page.
- Which theme an observation belongs to is decided by the fixed rules above; every count on this page is computed from them.
- A 483 addressed to an individual is listed without the name, and an observation that names a person is left out.
- FDAgent is independent of FDA.