Corrective and preventive action (CAPA) in FDA Form 483 observations
Corrective and preventive actions: whether firms open them, carry them out on time and check that they worked. FDAgent finds 141 such observations in 111 Form 483s issued to 107 facilities, for inspections that ended between November 2008 and September 2026. Each is quoted below from the 483 FDA posted.
A Form 483 lists the investigators' observations at the end of an inspection. It is not FDA's final determination.
Source: Form 483 PDFs in FDA's Electronic Reading Room, last synced Nov 23, 2026, 04:33 UTC. Every quote is from a 483 whose reading passed FDAgent's checks against the scan. FDAgent is independent of FDA.
- Observations
- 141
- Each quoted below
- Form 483s
- 111
- Posted by FDA
- Facilities
- 107
- Distinct FEIs
- Most recent
- Sep 28, 2026
- Inspection end date
Observations by year
These observations by the year the inspection ended. They follow which 483s FDA has posted and FDAgent has read, so they show what is on record, not a trend in FDA's findings. 2026 runs to the latest sync (lighter bar).
Peak: 51 in 2026
| Year | Observations |
|---|---|
| 2008 | 1 |
| 2009 | 0 |
| 2010 | 1 |
| 2011 | 0 |
| 2012 | 1 |
| 2013 | 1 |
| 2014 | 0 |
| 2015 | 2 |
| 2016 | 2 |
| 2017 | 5 |
| 2018 | 9 |
| 2019 | 4 |
| 2020 | 1 |
| 2021 | 7 |
| 2022 | 8 |
| 2023 | 16 |
| 2024 | 7 |
| 2025 | 25 |
| 2026 | 51 |
How FDAgent sorts observations into this theme
Two fixed rules, applied by code to every observation FDAgent can quote. An observation counts when either holds. No model and no similarity score decides.
It is tied to one of these 21 CFR entries
A 483 prints no regulation, so FDAgent ties an observation to a section only when it opens with FDA's own citation wording for that section, word for word.
Or its text contains one of these phrases
“CAPA*” · “corrective and preventive action*” · “corrective and preventative action*” · “corrective action*” · “preventive action*” · “preventative action*” · “effectiveness check*” · “effectiveness verification*”
* matches any ending and … up to three words in between. Case is ignored unless a phrase has capitals.
The 141 observations, newest first
Quoted word for word from each Form 483, with only line breaks joined; … marks where the quote is cut. The highlight shows why the observation is in this theme. Each links to FDAgent's page for the 483.
Auro PharmaciesInspection ended Aug 2, 2018 · FEI 3011893599 · Observation 2
… Your firm continued to produce sterile drugs without implementing a corrective and preventive action. A. On 01/08/2018, your firm recovered 1 CFU/cubic meter from viable air samples collected from ISO 5 Hood #3 during production of "TESTOSTERONE CYPIONATE/ENANTHATE OIL INJECTION 126/54 MG/ML INJECTABLE", lot 180108@25, RX (b) (6). B. On 02/28/2018, your third-party contractor recovered 1 CFU/cubic meter from viable air particle monitoring during …
Contains “corrective and preventive action*”Read the Form 483 →
Auro PharmaciesInspection ended Aug 2, 2018 · FEI 3011893599 · Observation 3
… After this media fill failure, your firm continued to produce sterile drugs without implementing corrective and preventive actions during a six-month time frame. 2. On 03/08/2018, a media fill batch was performed using (b) (4) mL vials in ISO 5 classified Hood #5. Of the (b) (4) vials filled, one vial resulted in microbial growth. After this media fill failure, your firm continued to produce sterile drug products without implementing corrective …
Contains “corrective and preventive action*”Read the Form 483 →
Plasma Industries Belgium SCRLInspection ended Mar 16, 2018 · FEI 3003208230 · Observation 2
… Event 94220 which was opened due to the detection of particles during the changing of (b)(4) has 2 overdue CAPAs. CAPA 103334 was due 31-Dec-2017 and CAPA 103335 was due 31-Jan-2018. The due dates for these CAPAs have not been extended by the Quality Unit.
Contains “CAPA*”Read the Form 483 →
Plasma Industries Belgium SCRLInspection ended Mar 16, 2018 · FEI 3003208230 · Observation 3
The current procedure for CIP of the (b)(4) fails to require the initiation of a deviation when an atypical event is detected during the (b)(4) of the (b)(4) Specifically, CAPA 81235 was initiated as part of Event Record 79659 to update SOP (b)(4) 1879 version 3.0 to include a clean and (b)(4) clean or a (b)(4) clean. This procedure does not require the operator who performs the (b)(4) or a (b)(4) clean to initiate a deviation when a (b)(4) …
Contains “CAPA*”Read the Form 483 →
One Way Drug LLC dba Partell Specialty PharmacyInspection ended Mar 8, 2018 · FEI 3011769075 · Observation 8
Environmental monitoring in your aseptic processing areas is not adequate. Specifically, a. Review of the last (b)(4) Environmental Monitoring report, dated 11/30/17, revealed that Micrococcus species and Aspergillus was detected when viable air sampling was conducted in the ISO-5 Laminar Air Flow Hood. As a corrective action, your firm performed cleaning (b)(4) times in the controlled environment, and air sampling was repeated. An investigation …
Contains “corrective action*”Read the Form 483 →
One Way Drug LLC dba Partell Specialty PharmacyInspection ended Mar 8, 2018 · FEI 3011769075 · Observation 10
… The firm's corrective action includes cleaning and disinfecting the controlled environment (b)(4) times. Demonstration of the effectiveness of the cleaning operations was not assessed because environmental and personnel monitoring was not performed afterward.
Contains “corrective action*”Read the Form 483 →
Suven Life Sciences Ltd.Inspection ended Feb 15, 2018 · FEI 1000623515 · Observation 7
Your firm's Product Quality Reviews (PQRs) are deficient. Specifically, A. 2017 PQR for (b) (4) Capsules (review period January 2016 - December 2017): 1. Does not address corrective and preventative actions implemented for out of specification (OOS16-005) for (b) (4) content for (b) (4) Capsules (b) (4) mg, Batch #(b) (4) 2. Is missing change control (EN/CCP/17/5036) for the installation of new equipment in the packing line ((b) (4) Inserting …
Contains “corrective and preventative action*”Read the Form 483 →
Empower PharmacyInspection ended Jan 24, 2018 · FEI 3011887629 · Observation 1
… 2) Your firm did not implement any corrective actions or preventative measures to assure that future endotoxin failures will not occur when manufacturing this product. 3) Your firm did not evaluate whether other associated lots were impacted. Review of your firm's potency failure, occurring on 10/16/2017 for Lot No. 35979; HCG Lyophilized 3000 IU/vial, revealed that the investigation was incomplete in that: 1) Your firm failed to document a …
Contains “corrective action*”Read the Form 483 →
Strides Shasun LimitedInspection ended May 26, 2017 · FEI 3004554612 · Observation 1
… your firm conducted investigations into the manufacturing process, your investigations did not arrive at the actual root cause to take the appropriate corrective actions. For example, the following complaints were related to the sticking together of your capsules within your firm’s product bottles. Complaint Number | Product Name | Nature of Complaint ODF/MC/2016/106 | Capsules USP®® mg | Capsules sticking together ODF/MC/2015/103 | Capsules mcg …
Contains “corrective action*”Read the Form 483 →
Meridian Medical Technologies, Inc. a Pfizer CompanyInspection ended Mar 24, 2017 · FEI 1950222 · Observation 1
… No new CAPAs were identified for the Lot 6M1454 investigation and your investigation concluded the filler was operating within its capabilities.
Contains “CAPA*”Read the Form 483 →
Meridian Medical Technologies, Inc. a Pfizer CompanyInspection ended Mar 24, 2017 · FEI 1950222 · Observation 13
… version 10.0, Management of Changes for Equipment, Facilities and Manufacturing Processes, Effective Date: 23 MAR 2016" fails to require documentation of effectiveness checks for completed change controls. Process validations are completed (b) (4) but are not evaluated or trended to determine impact to the overall manufacturing process; additionally, complete process validations for autoinjector manufacturing processes such as EpiPen and …
Contains “effectiveness check*”Read the Form 483 →
Alkem Laboratories LimitedInspection ended Mar 10, 2017 · FEI 3006370533 · Observation 1
The written testing stability program is not followed. The required stability testing on numerous drug products (over approximately 1,000 individual samples) was not completed within the time frames established in procedure Stability Management QC/063/07, current at the time the samples were due for testing. Deviation B/DQ/15/014 was opened November 13, 2015 and CAPA B/CAPA/15/122 was opened December 15, 2015 to address missing the analysis time …
Contains “CAPA*”Read the Form 483 →
Bayer AG PharmaInspection ended Jan 20, 2017 · FEI 3002806462 · Observation 7
Failure to perform a thorough investigation into a discrepancy. 1. During environmental monitoring of the (b) (4) filling area, numerous alert and action level findings were not thoroughly investigated. From April-June 2015 there were 5 alert and 3 action level investigations opened for monitoring of the (b) (4). A thorough investigation with root cause evaluation and preventive actions was not performed. Trending analysis conclusions recommended …
Contains “preventive action*”Read the Form 483 →
Indoco Remedies LimitedInspection ended Sep 4, 2016 · FEI 3005124189 · Observation 3
… The procedure for Corrective and Preventive Action, SOP QS/023, section 5.3.1.3.2, requires preventive action to be taken to prevent reoccurrence of an incident.
Contains “corrective and preventive action*”Read the Form 483 →
Cipla LimitedInspection ended Aug 30, 2016 · FEI 3004081307 · Observation 1
… Although®® line ®® | is believed to be an appropriate corrective action to eliminate the defect, prevalence and severity of the defect has not been thoroughly investigated, and inspection for the container-closure defect has not yet been assessed throughout an entire filling run. The product has not been filled since January, 2016, but there are approximately 11 lots on the market within expiry, in addition to lot 9 for which the integrity of the …
Contains “corrective action*”Read the Form 483 →
Alphatec Spine, Inc.Inspection ended Mar 13, 2015 · FEI 2027467 · Observation 5
Corrective and preventive action activities and/or results have not been adequately documented. Specifically, the firm acquired the (b) (4) (spinal implant device) from another medical device company in (b) (4). Previously, on (b) (4), the (b) (4) device was recalled, however, the firm has no records of the recall/corrections and removals on file to include receipt and final disposition of the returned units under recall nor any records …
Opens with FDA's citation wording for 21 CFR 820.100(b)Read the Form 483 →
Alphatec Spine, Inc.Inspection ended Mar 13, 2015 · FEI 2027467 · Observation 7
Procedures for corrective and preventive action have not been adequately established. Specifically, a) Procedure No.: SOP-025 - Corrective and Preventive Action (CAPA) System, Revision L is not defined in that the procedure does not provide instructions or a linkage to an action or alert level for the (b) (4) complaint trending data, which is reviewed at the (b) (4) trend review meetings. For example, the (b) (4) complaint rate by product line is …
Opens with FDA's citation wording for 21 CFR 820.100(a)Read the Form 483 →
Village Fertility Pharmacy, Inc.Inspection ended Mar 13, 2013 · FEI 3008335877 · Observation 2
The firm failed to adequately evaluate product integrity incident reports, received from patients and healthcare practitioners, for injectable Progesterone produced in various solutions (e.g. sesame oil and ethyl oleate). Neither root-cause analysis nor corrective/preventive actions were identified or documented by the firm to avoid future occurrences of the incidents. For example: A. Two different incident reports were received for Progesterone …
Contains “preventive action*”Read the Form 483 →
Cangene CorporationInspection ended Jun 21, 2012 · FEI 3003153579 · Observation 1
… The corrective actions identified in CAPA 219683 as a result of investigations described in deviation 208223 and deviation 211915 failed to prevent (b)(4) fouling in VIGIV bulk lot (b)(4) Specifically, exclusion of specific (b)(4) lots of plasma did not preclude (b)(4) fouling.
Contains “corrective action*”Read the Form 483 →
Regeneron Pharmaceuticals, Inc.Inspection ended Mar 31, 2010 · FEI 1000514603 · Observation 3
Review of the firm's manual log for SCARs (Supplier Corrective Action Requests) which was stated to be the original GMP document for tracking these events found numerous entries for the past two years which were not closed out. This log is used for tracking of potential quality issues related to raw materials received from vendors.
Contains “corrective action*”Read the Form 483 →
21 CFR sections these observations open with
The sections FDAgent ties the most of these observations to by FDA's wording, with how many of this theme's observations each covers. Each links to its page of FDA citations, warning letters and the regulation text.
- 21 CFR 211.192Production record review24
- 21 CFR 211.22Responsibilities of quality control unit11
- 21 CFR 820.100Corrective and preventive actionIn this theme's rule6
- 21 CFR 211.42Design and construction features5
- 21 CFR 211.113Control of microbiological contamination5
- 21 CFR 211.63Equipment design, size, and location2
- 21 CFR 211.67Equipment cleaning and maintenance2
- 21 CFR 211.198Complaint files2
- 21 CFR 211.100Written procedures; deviations1
- 21 CFR 211.110Sampling and testing of in-process materials and drug products1
Ask FDAgent about CAPA in Form 483s
FDAgent reads FDA's Form 483s, inspections, citations and warning letters and answers in plain English, with links to the records.
Related
About this data
- The observations are the text FDAgent read by OCR from Form 483 PDFs in FDA's Electronic Reading Room, shown only from 483s whose every printed observation passed FDAgent's checks against the scanned text, as on each 483's own page.
- Which theme an observation belongs to is decided by the fixed rules above; every count on this page is computed from them.
- A 483 addressed to an individual is listed without the name, and an observation that names a person is left out.
- FDAgent is independent of FDA.