June 2026 · 2026-06-01 to 2026-06-30
FDA enforcement, June 2026
In June 2026 (2026-06-01 to 2026-06-30), FDA posted 69 warning letters from its drug, biologics and device offices, and listed 1 Form 483 in its Electronic Reading Room for the first time. FDA's Enforcement Report added 366 drug and device recalls, 71 of them Class I. FDAgent recorded 5 new Official Action Indicated (OAI) classifications in FDA's inspection data. FDA published 22 firm entries on its import alert lists, 18 of them on a Red List.
Counts are FDA's records dated in June as of each source's last sync, which each section states; a late FDA update can still change them.
- Warning letters posted
- 69
- On 6 posting dates · 26 in May
- Form 483s listed
- 1
- From 1 facility · 14 in May
- Inspections newly classified
- 293
- 5 OAI · 86 VAI · 202 NAI
- New OAI classifications
- 5
- Newly listed by FDA · 11 in May
- Recalled products
- 366
- 116 recall events · 71 Class I
- Import alert entries
- 22
- 18 Red List · 4 Green List
Ask FDAgent about this month
FDAgent reads the warning letters FDA posted this month and answers in plain English, with links to the letters.
Compared with May 2026 and June 2025
Each measure is counted the same way in every month. A year-over-year figure is shown only where FDAgent holds FDA's own dates for June 2025.
Warning letters posted69
May 2026: 26 · +43 (+165%)Jun 2025: 27 · +42 (+156%)
Form 483s listed in the reading room1
May 2026: 14 · −13 (−93%)Jun 2025: 21 · −20 (−95%)
Inspections newly classified293
May 2026: 275 · +18 (+7%)Jun 2025: —¹
New OAI classifications5
May 2026: 11 · −6 (−55%)Jun 2025: —¹
Recalled products (drug and device)366
May 2026: 326 · +40 (+12%)Jun 2025: 207 · +159 (+77%)
Recall events116
May 2026: 104 · +12 (+12%)Jun 2025: 99 · +17 (+17%)
Class I recall events17
May 2026: 15 · +2 (+13%)Jun 2025: 11 · +6 (+55%)
Import alert firm entries22
May 2026: 27 · −5 (−19%)Jun 2025: —²
¹ FDAgent has recorded when an inspection first appears in FDA's data only since late January 2026, so months before February 2026 have no count. ² FDA deletes an import alert entry when it removes a firm, and FDAgent keeps removed entries only from November 2026, so earlier months count only the entries FDA still lists, and the year before is not compared. Sources: FDA warning letters page, last synced Nov 22, 2026, 04:45 UTC; Electronic Reading Room, last synced Nov 22, 2026, 15:42 UTC; FDA Data Dashboard inspections, last synced Nov 22, 2026, 04:55 UTC; openFDA Enforcement Report, FDA last updated Nov 11, 2026; FDA import alert pages, last synced Nov 20, 2026, 03:55 UTC.
Inspections FDA newly classified
293 inspections FDAgent first found in FDA's inspection classification data in June, by FDA's classification. FDA gives no date for a classification, so each counts in the month FDAgent first recorded it. When FDA later removes an inspection from its Data Dashboard, it stops counting, so a finished month's count can fall. They ended a median of 46 days before FDAgent first recorded them, and FDA's Data Dashboard lists Form 483 citations for 77 of them. Inspection classifications last synced Nov 22, 2026, 04:55 UTC.
| Product type | OAI | VAI | NAI | Total |
|---|---|---|---|---|
| Drugs | 4 | 44 | 53 | 101 |
| Devices | 1 | 29 | 64 | 94 |
| Biologics | 0 | 13 | 85 | 98 |
| All | 5 | 86 | 202 | 293 |
The 5 OAI inspections are listed under New OAI classifications below.
Regulations cited most
The 21 CFR sections behind the Form 483 citations FDA's Data Dashboard lists for the inspections it newly classified in June (138 citations on 53 inspections), and the sections June's warning letters cite in their numbered violations (71 violations in 25 of the 69 letters). Each section links to its page. Inspection citations last synced Nov 22, 2026, 04:55 UTC.
| On inspections | Citations |
|---|---|
| 21 CFR 211.67Equipment cleaning and maintenance | 13 |
| 21 CFR 211.160General requirements | 9 |
| 21 CFR 211.192Production record review | 9 |
| 21 CFR 211.84Testing and approval or rejection of components, drug product containers, and closures | 8 |
| 21 CFR 211.22Responsibilities of quality control unit | 7 |
| 21 CFR 211.68Automatic, mechanical, and electronic equipment | 7 |
| 21 CFR 211.166Stability testing | 7 |
| 21 CFR 801.20Label to bear a unique device identifier | 6 |
| In warning letters | Letters |
|---|---|
| 21 CFR 211.22Responsibilities of quality control unit | 13 |
| 21 CFR 211.84Testing and approval or rejection of components, drug product containers, and closures | 10 |
| 21 CFR 211.192Production record review | 7 |
| 21 CFR 211.113Control of microbiological contamination | 6 |
| 21 CFR 211.42Design and construction features | 5 |
| 21 CFR 211.160General requirements | 3 |
| 21 CFR 211.194Laboratory records | 3 |
| 21 CFR 806.10Reports of corrections and removals | 3 |
Sections with the same count are ordered by section number. A section without its own page is shown without a link. 106 citations name no 21 CFR section and are not counted; device citations can name ISO 13485:2016 clauses instead. FDA can list an inspection's citations weeks after its classification, so these counts can grow.
Form 483s listed in the reading room
1 Form 483 FDA listed for the first time in its Electronic Reading Room in June, dated by the publish date FDA gives each, from 1 facility. FDA sometimes lists a 483 days after that date, so a finished month can still gain 483s. FDA posts only some 483s, often long after the inspection: it listed these a median of 75 days after their record date. Each week's page lists every 483 with its PDF. Reading room last synced Nov 22, 2026, 15:42 UTC.
- 20261
- 20250
- 20240
- 20230
- 2022 and earlier0
Warning letters
69 letters FDA posted on its warning letters page in June, from its drug, biologics and device offices. 56 were issued in June and 13 earlier; FDA posted them a median of 8 days after issue. Warning letters last synced Nov 22, 2026, 04:45 UTC.
- CDER · Drugs
- 62
- CDRH · Devices
- 5
- CBER · Biologics
- 2
3B Medical, Inc. dba Reach Health, Inc.
CGMP/QSR/Medical Devices/Adulterated
CDRH · issued 2026-05-20 · posted 2026-06-30
Investigational Device Exemptions (IDE)/Premarket Approval Application (PMA) Adulterated Device
CDRH · issued 2026-06-04 · posted 2026-06-30
CGMP/QSR/Medical Devices/PMA/Adulterated/Misbranded
CDRH · issued 2026-05-20 · posted 2026-06-30
CGMP/Finished Pharmaceuticals/Adulterated
CDER · issued 2026-06-25 · posted 2026-06-30
CGMP/Deviations/Biologics License Application (BLA)
CBER · issued 2026-06-22 · posted 2026-06-30
Wizcure Pharmaa Private Limited
CGMP/Finished Pharmaceuticals/Adulterated
CDER · issued 2026-06-24 · posted 2026-06-30
Yangzhou Hongshengding Chemical Co., Ltd.
Refusal to Provide Access to and Copying of Records
CDER · issued 2026-06-15 · posted 2026-06-30
Great India Overseas Pvt. Ltd. dba Silkrute
Unapproved New Drugs
CDER · issued 2026-06-17 · posted 2026-06-23
Herbal Dermals LLC FZ dba Gotucream
Unapproved New Drugs/Misbranded
CDER · issued 2026-06-17 · posted 2026-06-23
CGMP/Finished Pharmaceuticals/Adulterated
CDER · issued 2026-06-15 · posted 2026-06-23
Jubilant HollisterStier General Partnership
CGMP/Finished Pharmaceuticals/Adulterated
CDER · issued 2026-05-28 · posted 2026-06-23
Telos Trading Limited dba Alwinshop
Unapproved New Drugs/Misbranded
CDER · issued 2026-06-17 · posted 2026-06-23
Show 49 more
Altru Telehealth, LLC dba Altru Telehealth
False & Misleading Claims/Misbranded (Telehealth)
CDER · issued 2026-06-08 · posted 2026-06-16
False & Misleading Claims/Misbranded (Telehealth)
CDER · issued 2026-06-08 · posted 2026-06-16
Clover Meds LLC dba Clover Meds
False & Misleading Claims/Misbranded (Telehealth)
CDER · issued 2026-06-08 · posted 2026-06-16
False & Misleading Claims/Misbranded (Telehealth)
CDER · issued 2026-06-08 · posted 2026-06-16
Eden Health International Inc. dba Eden
False & Misleading Claims/Misbranded (Telehealth)
CDER · issued 2026-06-08 · posted 2026-06-16
False & Misleading Claims/Misbranded (Telehealth)
CDER · issued 2026-06-08 · posted 2026-06-16
False & Misleading Claims/Misbranded (Telehealth)
CDER · issued 2026-06-08 · posted 2026-06-16
Glow Medispa, LLC dba Mint Med
False & Misleading Claims/Misbranded (Telehealth)
CDER · issued 2026-06-08 · posted 2026-06-16
False & Misleading Claims/Misbranded (Telehealth)
CDER · issued 2026-06-08 · posted 2026-06-16
Electronic Drug Registration and Listing System (eDRLS)/Violations
CDER · issued 2026-06-03 · posted 2026-06-16
Joi and Blokes dba Joi + Blokes
False & Misleading Claims/Misbranded (Telehealth)
CDER · issued 2026-06-08 · posted 2026-06-16
Maximus Health, Inc. dba Maximus
False & Misleading Claims/Misbranded (Telehealth)
CDER · issued 2026-06-08 · posted 2026-06-16
False & Misleading Claims/Misbranded (Telehealth)
CDER · issued 2026-06-08 · posted 2026-06-16
Momentum Health 360 dba Momentum Health
False & Misleading Claims/Misbranded (Telehealth)
CDER · issued 2026-06-08 · posted 2026-06-16
False & Misleading Claims/Misbranded (Telehealth)
CDER · issued 2026-06-08 · posted 2026-06-16
False & Misleading Claims/Misbranded (Telehealth)
CDER · issued 2026-06-08 · posted 2026-06-16
Nexus Health Solutions LLC dba Harper Meds
False & Misleading Claims/Misbranded (Telehealth)
CDER · issued 2026-06-08 · posted 2026-06-16
False & Misleading Claims/Misbranded (Telehealth)
CDER · issued 2026-06-08 · posted 2026-06-16
OrderlyMeds LLC dba OrderlyMeds
False & Misleading Claims/Misbranded (Telehealth)
CDER · issued 2026-06-08 · posted 2026-06-16
CGMP/Active Pharmaceutical Ingredient (API)/Adulterated
CDER · issued 2026-05-27 · posted 2026-06-16
Public Health Solution LLC dba DrMedHealth
False & Misleading Claims/Misbranded (Telehealth)
CDER · issued 2026-06-08 · posted 2026-06-16
False & Misleading Claims/Misbranded (Telehealth)
CDER · issued 2026-06-08 · posted 2026-06-16
False & Misleading Claims/Misbranded (Telehealth)
CDER · issued 2026-06-08 · posted 2026-06-16
CGMP/Finished Pharmaceuticals/Adulterated
CDER · issued 2026-06-05 · posted 2026-06-16
False & Misleading Claims/Misbranded (Telehealth)
CDER · issued 2026-06-08 · posted 2026-06-16
Trinity HealthCare Supply, LLC dba altRx
False & Misleading Claims/Misbranded (Telehealth)
CDER · issued 2026-06-08 · posted 2026-06-16
False & Misleading Claims/Misbranded (Telehealth)
CDER · issued 2026-06-08 · posted 2026-06-16
CGMP/QSR/Medical Devices/Adulterated
CDRH · issued 2026-04-30 · posted 2026-06-16
Investigational Device Exemptions (IDE)/Premarket Approval Application (PMA) Adulterated Device
CDRH · issued 2026-06-15 · posted 2026-06-15
Asanuma Corporation - Sagamihara Factory
CGMP/OTC Finished Pharmaceuticals/Adulterated
CDER · issued 2026-06-02 · posted 2026-06-09
Erkul Kozmetik Sanayi ve Ticaret A.S.
CGMP/OTC Finished Pharmaceuticals/Adulterated
CDER · issued 2026-06-02 · posted 2026-06-09
CGMP/OTC Finished Pharmaceuticals/Adulterated
CDER · issued 2026-06-02 · posted 2026-06-09
Hubei Gedian Humanwell Pharmaceutical Co., Ltd.
CGMP/Active Pharmaceutical Ingredient (API)/Adulterated
CDER · issued 2026-06-03 · posted 2026-06-09
CGMP/OTC Finished Pharmaceuticals/Adulterated
CDER · issued 2026-06-02 · posted 2026-06-09
CGMP/OTC Finished Pharmaceuticals/Adulterated
CDER · issued 2026-06-02 · posted 2026-06-09
Shantou Qiwei Industry Co., LTD
CGMP/OTC Finished Pharmaceuticals/Adulterated
CDER · issued 2026-06-02 · posted 2026-06-09
Umendra Life Sciences Private Limited
CGMP/OTC Finished Pharmaceuticals/Adulterated
CDER · issued 2026-06-02 · posted 2026-06-09
Request for Records/CGMP/Finished Pharmaceuticals/Adulterated
CDER · issued 2026-06-02 · posted 2026-06-09
Aeroflex Industria de Aerosol Ltda.
Refusal to Provide Access to and Copying of Records
CDER · issued 2026-05-22 · posted 2026-06-02
Blue Horizon International, LLC
Unapproved New Drugs/Unlicensed Biological Product Violations
CBER · issued 2026-05-26 · posted 2026-06-02
HealthPartners Neuroscience Center Research and Innovation
Good Laboratory Practice (GLP) for Nonclinical Laboratory Studies
CDER · issued 2026-05-19 · posted 2026-06-02
Laboratorios Jaloma S.A. de C.V.
CGMP/OTC/Finished Pharmaceuticals/Adulterated
CDER · issued 2026-05-22 · posted 2026-06-02
An FEI links to its facility page only when the letter names exactly one FEI and FDAgent holds FDA records for it.
Recalls by class
Drug and device recalls in FDA's Enforcement Report dated in June, from openFDA, which FDA last updated Nov 11, 2026. Each openFDA record is one recalled product; a recall event can cover many products, so events are counted too.
| Class | Drug products | Device products | EventsRecall events |
|---|---|---|---|
| Class I | 3 | 68 | 17 |
| Class II | 65 | 224 | 93 |
| Class III | 6 | 0 | 6 |
| All | 74 | 292 | 116 |
The largest Class I recall event, from ARROW INTERNATIONAL, LLC, covers 35 of the 71 Class I products. The Class I products are listed below. Class II and III recalls are listed on each week's page.
Class I71 recallsShow all
- Z-2360-2026openFDA record
Abiomed, Inc.
Automated Impella Controller (AIC), for use with left heart support blood pump. Labeled as the following with corresponding Product Codes: 1. Impella Controller, Packaged, AU; Product Code: 0042-0000-AU. 2. Impella Controller, Packaged, CA; Product Code: 0042-0000-CA. 3. Impella…
Reason: Potential software error in the Automated Impella Controller ("AIC" ) when used in conjunction with left ventricular Impella devices.
Device · initiated 2026-05-14 · Enforcement Report 2026-06-24 · Ongoing
- Z-2457-2026openFDA record
Abiomed, Inc.
Impella CP Set with SmartAssist. Product Codes: 1000080, 1000851, 0048-0014, 0048-0024-JP.
Reason: Exposure to the low purge pressure occurrence may result in persistent low purge pressure alarms and, in some cases, interruption or loss of mechanical circulatory support.
Device · initiated 2026-05-18 · Enforcement Report 2026-06-24 · Ongoing
- Z-2363-2026openFDA record
ARROW INTERNATIONAL, LLC
PARACENTESIS KIT, REF: AK-00376, ASK-00376-VCU
Reason: Lidocaine, Bupivacaine, and 0.9% sodium chloride, the subject of a supplier recall, were included into kits and sets. The supplier was found to have quality issues and limited reports of drug ineffectiveness, and if sterility of a water-based injectable product is compromised th…
Device · initiated 2026-04-21 · Enforcement Report 2026-06-24 · Ongoing
- Z-2364-2026openFDA record
ARROW INTERNATIONAL, LLC
Description/REF: ACCESS TRAY/ASK-04001-SUH-S; ACCESS TRAY/ASK-04001-VM1; STIMUCATH SFTY KIT: 19G CTH, 17Gx4CM NDL/AB-19604-KS; STIMUCATH SFTY KIT: 19G CTH, 17Gx8CM NDL/AB-19608-KS; STIMUCATH SFTY KIT: 20G CTH, 18Gx8CM NDL/AB-20608-KS; STIMUCATH KIT: 19G CATH, 17G x 9CM NDL/AB-05…
Reason: Lidocaine, Bupivacaine, and 0.9% sodium chloride, the subject of a supplier recall, were included into kits and sets. The supplier was found to have quality issues and limited reports of drug ineffectiveness, and if sterility of a water-based injectable product is compromised th…
Device · initiated 2026-04-21 · Enforcement Report 2026-06-24 · Ongoing
- Z-2365-2026openFDA record
ARROW INTERNATIONAL, LLC
Description/REF: BIOPSY SYSTEM COMPREHENSIVE KIT/ASK-09458-AH, ASK-09458-COC, ASK-09458-EUH, ASK-09458-SD, ASK-09458-VU
Reason: Lidocaine, Bupivacaine, and 0.9% sodium chloride, the subject of a supplier recall, were included into kits and sets. The supplier was found to have quality issues and limited reports of drug ineffectiveness, and if sterility of a water-based injectable product is compromised th…
Device · initiated 2026-04-21 · Enforcement Report 2026-06-24 · Ongoing
- Z-2366-2026openFDA record
ARROW INTERNATIONAL, LLC
Description/REF: PACING PSI KIT: 6 FR/AI-06210-KS, PACING PSI KIT: 5 FR 2-L/AI-07155-KS, PACING PSI KIT: 5 FR 2L/ASK-07155-UPM
Reason: Lidocaine, Bupivacaine, and 0.9% sodium chloride, the subject of a supplier recall, were included into kits and sets. The supplier was found to have quality issues and limited reports of drug ineffectiveness, and if sterility of a water-based injectable product is compromised th…
Device · initiated 2026-04-21 · Enforcement Report 2026-06-24 · Ongoing
- Z-2367-2026openFDA record
ARROW INTERNATIONAL, LLC
Description/REF: AGBA PICC G4 Stylet: 3L 6FR x 55CM BP/DLX-45563-VPSB; AGBA PICC G4 Stylet: 3L 6FR x 55CM TCG/DLX-45563-VPSC; ARROW(R) VPS(R) ACCESS KIT/ASK-04001-UCHD1; ARROW(R) VPS(R) ACCESS KIT/ASK-04001-UCHS1; ARROW(R) VPS(R) ACCESS KIT/ASK-04001-DU13; ARROW(R) VPS(R) ACCESS…
Reason: Lidocaine, Bupivacaine, and 0.9% sodium chloride, the subject of a supplier recall, were included into kits and sets. The supplier was found to have quality issues and limited reports of drug ineffectiveness, and if sterility of a water-based injectable product is compromised th…
Device · initiated 2026-04-21 · Enforcement Report 2026-06-24 · Ongoing
- Z-2368-2026openFDA record
ARROW INTERNATIONAL, LLC
Description/REF: RA CATH KIT: 20 GA X 1 3/4 IN/TJ-04020-1; RA CATH KIT: 20 GA X 1 1/2IN/NA-04220-S1A; ARTERIAL LINE KIT/AK-04510-S; ARTERIAL CATH KIT: 18 GA X 6IN (16 CM)/ASK-04018-HUMC; ARTERIAL LINE KIT: 18 GA X 6 IN (16 CM)/ASK-04018-OH1; RA CATH KIT: 20 GA X 1 3/4IN (4.45 CM…
Reason: Lidocaine, Bupivacaine, and 0.9% sodium chloride, the subject of a supplier recall, were included into kits and sets. The supplier was found to have quality issues and limited reports of drug ineffectiveness, and if sterility of a water-based injectable product is compromised th…
Device · initiated 2026-04-21 · Enforcement Report 2026-06-24 · Ongoing
- Z-2369-2026openFDA record
ARROW INTERNATIONAL, LLC
Description/REF: PSI KIT: 7 FR X 4IN ANTIMICROBIAL/ASK-09701-UMH; PSI KIT: 8.5 FR X 4 1/8 IN (10 CM)/ASK-09803-OSU4; PSI/CVC KIT: 9 FR/7 FR X 6IN (16CM)/ASK-09903-AMC1; PSI KIT: 9 FR X 4 1/8 IN (10 CM)/ASK-09903-CMC1; PSI KIT: 9 FR X 4 1/8 IN (10 CM)/ASK-09903-OSU4; PSI KIT: 9 F…
Reason: Lidocaine, Bupivacaine, and 0.9% sodium chloride, the subject of a supplier recall, were included into kits and sets. The supplier was found to have quality issues and limited reports of drug ineffectiveness, and if sterility of a water-based injectable product is compromised th…
Device · initiated 2026-04-21 · Enforcement Report 2026-06-24 · Ongoing
- Z-2370-2026openFDA record
ARROW INTERNATIONAL, LLC
Description/REF: PI AGBA PICC KIT: 3L 6 FR X 40 CM W/VPS/ASK-44063-NS AGBA PICC/DELTA: 2L 5.5 FR X 50 CM/ASK-45052-NS AGBA PICC/DELTA KIT: 1L 4.5 FR X 55 CM/ASK-45541-MHBV PICC KIT: 2L 5.5 FR X 55 CM CG+ VPS/ASK-45552-RH AGBA PICC NaviCurve: 2L 5.5FR x 55CM BP/DLX-45552-CURVB AG…
Reason: Lidocaine, Bupivacaine, and 0.9% sodium chloride, the subject of a supplier recall, were included into kits and sets. The supplier was found to have quality issues and limited reports of drug ineffectiveness, and if sterility of a water-based injectable product is compromised th…
Device · initiated 2026-04-21 · Enforcement Report 2026-06-24 · Ongoing
- Z-2371-2026openFDA record
ARROW INTERNATIONAL, LLC
Description/REF: EPIDURAL CATHETER KIT: 19 GA X 90 CM/ASK-05502-VM1; EPIDURAL CATHETERIZATION KIT: 19 GA/AK-05501; EPIDURAL CATHETERIZATION KIT/AK-05502; EPIDURAL CATHETERIZATION KIT/AK-05502D; EPIDURAL CATHETERIZATION KIT/AK-05503; EPIDURAL CATHETERIZATION KIT/AK-05503-L; SPINA…
Reason: Lidocaine, Bupivacaine, and 0.9% sodium chloride, the subject of a supplier recall, were included into kits and sets. The supplier was found to have quality issues and limited reports of drug ineffectiveness, and if sterility of a water-based injectable product is compromised th…
Device · initiated 2026-04-21 · Enforcement Report 2026-06-24 · Ongoing
- Z-2372-2026openFDA record
ARROW INTERNATIONAL, LLC
Description/REF: HEMODIALYSIS KIT: 2L 14 FR X 20 CM AGB CDC-25142-X1A
Reason: Lidocaine, Bupivacaine, and 0.9% sodium chloride, the subject of a supplier recall, were included into kits and sets. The supplier was found to have quality issues and limited reports of drug ineffectiveness, and if sterility of a water-based injectable product is compromised th…
Device · initiated 2026-04-21 · Enforcement Report 2026-06-24 · Ongoing
- Z-2373-2026openFDA record
ARROW INTERNATIONAL, LLC
Description/REF: HEMODIALYSIS KIT: 2-LUMEN 8 FR X 11 CM/AK-11802-F; HEMODIALYSIS KIT: 2-LUMEN 12 FR X 16 CM/AK-22122; HEMODIALYSIS KIT 2L: 14 FR X 15CM AGB/AK-22142-CDC; HEMODIALYSIS KIT 2L: 14 FR X 20CM AGB/AK-25142-CDC; HEMODIALYSIS KIT: 2L 14 FR X 15 CM/ASK-22142-UMD; HEMODIA…
Reason: Lidocaine, Bupivacaine, and 0.9% sodium chloride, the subject of a supplier recall, were included into kits and sets. The supplier was found to have quality issues and limited reports of drug ineffectiveness, and if sterility of a water-based injectable product is compromised th…
Device · initiated 2026-04-21 · Enforcement Report 2026-06-24 · Ongoing
- Z-2374-2026openFDA record
ARROW INTERNATIONAL, LLC
Description/REF: AHDC KIT: 3-L 12 FR X 16 CM/AK-12123-PCDC; AHDC KIT: 3-L 12 FR X 20 CM/AK-15123-PCDC; AHDC KIT: 3-L 12 FR X 20 CM/CDC-15123-XP1A; AHDC KIT: 3-L 12 FR X 25 CM/CDC-16123-XP1A; AHDC KIT: 3-L 12 FR X 25 CM/CDC-16123-XPN1A; AHDC AGB U-BEND KIT: 2-L 12 FR X 20 CM/CDC-…
Reason: Lidocaine, Bupivacaine, and 0.9% sodium chloride, the subject of a supplier recall, were included into kits and sets. The supplier was found to have quality issues and limited reports of drug ineffectiveness, and if sterility of a water-based injectable product is compromised th…
Device · initiated 2026-04-21 · Enforcement Report 2026-06-24 · Ongoing
- Z-2375-2026openFDA record
ARROW INTERNATIONAL, LLC
Description/REF: PI CVC KIT: 3L 7 FR X 16 CM AGB+/ASK-42703-PNHS2; PI CVC KIT: 3L 7 FR X 20 CM AGB+/ASK-45703-PNHS2; PI CVC KIT: 4L 8.5 FR X 20 CM AGB+/ASK-45854-PNHS2
Reason: Lidocaine, Bupivacaine, and 0.9% sodium chloride, the subject of a supplier recall, were included into kits and sets. The supplier was found to have quality issues and limited reports of drug ineffectiveness, and if sterility of a water-based injectable product is compromised th…
Device · initiated 2026-04-21 · Enforcement Report 2026-06-24 · Ongoing
- Z-2376-2026openFDA record
ARROW INTERNATIONAL, LLC
Description/REF: CVC/PSI KIT: 2 LUMEN 8 FR X 6 IN/ASK-42802-PLH; CVC KIT: 3-LUMEN 12 FR X 16 CM/AK-12123-F; CVC KIT: 3-LUMEN 12 FR X 20 CM/AK-15123-F; 3L CVC KIT: 7 FR X 16 CM/ASK-12703-PMAH; 3L PI CVC KIT: 7 FR X 16 CM/ASK-12703-PNM; 3L CVC KIT: 7 FR X 16 CM/ASK-12703-PUMH; MUL…
Reason: Lidocaine, Bupivacaine, and 0.9% sodium chloride, the subject of a supplier recall, were included into kits and sets. The supplier was found to have quality issues and limited reports of drug ineffectiveness, and if sterility of a water-based injectable product is compromised th…
Device · initiated 2026-04-21 · Enforcement Report 2026-06-24 · Ongoing
- Z-2377-2026openFDA record
ARROW INTERNATIONAL, LLC
Description/REF: ARTERIAL LINE KIT: SL 20 GA X 5 IN/ASK-04510-MM6 RA CATH KIT: 20 GA X 1-3/4IN/NA-04020-X1ASP ACCESS TRAY/ASK-04001-MS RA CATH KIT: 20 GA X 8 CM/ASK-00820-UM1 VESSEL CATH KIT: 20GA X 5IN/AK-04550-S CVC KIT: 1-LUMEN 5 FR X 20 CM/AK-04301-SP CVC KIT: 16 GA X 6IN (1…
Reason: Lidocaine, Bupivacaine, and 0.9% sodium chloride, the subject of a supplier recall, were included into kits and sets. The supplier was found to have quality issues and limited reports of drug ineffectiveness, and if sterility of a water-based injectable product is compromised th…
Device · initiated 2026-04-21 · Enforcement Report 2026-06-24 · Ongoing
- Z-2378-2026openFDA record
ARROW INTERNATIONAL, LLC
Description/REF: PI PICC: 3: 6FR x 55CM w CHG/ DLX-35563-HPKC; AGBA PICC: 2L 5.5 FR X 45 CM/ASK-44552-ABKB; AGBA PICC/DELTA KIT: 5.5 FR X 55 CM/ASK-45552-MHBV; AGBA PICC: 1L 4.5FR x 55CM w Biopatch/DLX-45541-HPKB; PICC/DELTA KIT: 2L 5.5 FR X 50 CM/ASK-45052-MC; AGBA PI PICC KIT:…
Reason: Lidocaine, Bupivacaine, and 0.9% sodium chloride, the subject of a supplier recall, were included into kits and sets. The supplier was found to have quality issues and limited reports of drug ineffectiveness, and if sterility of a water-based injectable product is compromised th…
Device · initiated 2026-04-21 · Enforcement Report 2026-06-24 · Ongoing
- Z-2379-2026openFDA record
ARROW INTERNATIONAL, LLC
Description/REF: PI JACC KIT: 2L 5.5 FR X 15 CM CG+/ASK-41552-LTAC; PI JACC KIT: 3L 6 FR X 20 CM CG+/ASK-42063-LTAC; AGBA PI JACC KIT: 1-L 4.5 FR X 15 CM/CDC-41541-JX1A; AGBA PI JACC KIT: 2-L 5.5 FR X 15 CM/CDC-41552-JX1A; AGBA PI JACC KIT: 3-L 6 FR X 15 CM/CDC-41563-JX1A; AGBA…
Reason: Lidocaine, Bupivacaine, and 0.9% sodium chloride, the subject of a supplier recall, were included into kits and sets. The supplier was found to have quality issues and limited reports of drug ineffectiveness, and if sterility of a water-based injectable product is compromised th…
Device · initiated 2026-04-21 · Enforcement Report 2026-06-24 · Ongoing
- Z-2380-2026openFDA record
ARROW INTERNATIONAL, LLC
Description/REF: PERITONEAL LAVAGE CATH KIT:8FR/AK-09000
Reason: Lidocaine, Bupivacaine, and 0.9% sodium chloride, the subject of a supplier recall, were included into kits and sets. The supplier was found to have quality issues and limited reports of drug ineffectiveness, and if sterility of a water-based injectable product is compromised th…
Device · initiated 2026-04-21 · Enforcement Report 2026-06-24 · Ongoing
- Z-2381-2026openFDA record
ARROW INTERNATIONAL, LLC
Description/REF: HEMODIALYSIS KIT: 2-LUMEN 12 FR X 16 CM/AK-12122-F
Reason: Lidocaine, Bupivacaine, and 0.9% sodium chloride, the subject of a supplier recall, were included into kits and sets. The supplier was found to have quality issues and limited reports of drug ineffectiveness, and if sterility of a water-based injectable product is compromised th…
Device · initiated 2026-04-21 · Enforcement Report 2026-06-24 · Ongoing
- Z-2382-2026openFDA record
ARROW INTERNATIONAL, LLC
Description/REF: DRAINAGE KIT/AK-01600, DRAINAGE KIT/AK-01601, DRAINAGE KIT/ASK-01601-MM, THORACENTESIS KIT/AK-01000-T, THORACENTESIS KIT/AK-01000, THORACENTESIS KIT/ASK-01000-CHP, DRAINAGE KIT/ASK-01000-NW, THORACENTESIS KIT/ASK-01000-UPMS, PNEUMOTHORAX KIT/AK-01500, PNEUMOTHOR…
Reason: Lidocaine, Bupivacaine, and 0.9% sodium chloride, the subject of a supplier recall, were included into kits and sets. The supplier was found to have quality issues and limited reports of drug ineffectiveness, and if sterility of a water-based injectable product is compromised th…
Device · initiated 2026-04-21 · Enforcement Report 2026-06-24 · Ongoing
- Z-2383-2026openFDA record
ARROW INTERNATIONAL, LLC
Description/REF: PACING/PSI KIT: 5 FR 2L/ASK-07155-MGU
Reason: Lidocaine, Bupivacaine, and 0.9% sodium chloride, the subject of a supplier recall, were included into kits and sets. The supplier was found to have quality issues and limited reports of drug ineffectiveness, and if sterility of a water-based injectable product is compromised th…
Device · initiated 2026-04-21 · Enforcement Report 2026-06-24 · Ongoing
- Z-2384-2026openFDA record
ARROW INTERNATIONAL, LLC
Description/REF: ACCESS TRAY/ASK-04001-JHH; ACCESS TRAY/ASK-04001-LMDTG; PI CVC KIT: 3 LUMEN 7 FR X 6 IN (16 CM)/ASK-12703-PSPH1; 3L CVC KIT: 7 FR X 16 CM/ASK-12703-WMC2; MULTILUMEN PI CVC KIT: 3L 7 FR X 20CM/ASK-15703-LMDT1; MULTILUMEN PI CVC KIT: 3L 7 FR X 20CM/ASK-15703-LMDTG…
Reason: Lidocaine, Bupivacaine, and 0.9% sodium chloride, the subject of a supplier recall, were included into kits and sets. The supplier was found to have quality issues and limited reports of drug ineffectiveness, and if sterility of a water-based injectable product is compromised th…
Device · initiated 2026-04-21 · Enforcement Report 2026-06-24 · Ongoing
- Z-2385-2026openFDA record
ARROW INTERNATIONAL, LLC
Description/REF: ACCESS TRAY/GU-04020; PSI KIT: 8.5 FR/AK-09803-CDC; PSI KIT: 9 FR/AK-09903-CDC; PSI KIT: 8.5 FR X 4 IN (10 CM)/ASK-09810-PCMH1; PSI KIT: 9 FR X 10 CM/ASK-09903-KH; PSI KIT: 9 FR X 10 CM/ASK-09903-PHP1; PSI KIT: 9 FR X 4 IN (10 CM)/ASK-09903-UPA; PSI KIT: 9 FR X…
Reason: Lidocaine, Bupivacaine, and 0.9% sodium chloride, the subject of a supplier recall, were included into kits and sets. The supplier was found to have quality issues and limited reports of drug ineffectiveness, and if sterility of a water-based injectable product is compromised th…
Device · initiated 2026-04-21 · Enforcement Report 2026-06-24 · Ongoing
- Z-2386-2026openFDA record
ARROW INTERNATIONAL, LLC
Description/REF: PARACENTESIS KIT/ASK-00376-JHH
Reason: Lidocaine, Bupivacaine, and 0.9% sodium chloride, the subject of a supplier recall, were included into kits and sets. The supplier was found to have quality issues and limited reports of drug ineffectiveness, and if sterility of a water-based injectable product is compromised th…
Device · initiated 2026-04-21 · Enforcement Report 2026-06-24 · Ongoing
- Z-2387-2026openFDA record
ARROW INTERNATIONAL, LLC
Description/REF: PI MIDLINE 2L: 5FR X 20CM/CDC-32052-MPKC
Reason: Lidocaine, Bupivacaine, and 0.9% sodium chloride, the subject of a supplier recall, were included into kits and sets. The supplier was found to have quality issues and limited reports of drug ineffectiveness, and if sterility of a water-based injectable product is compromised th…
Device · initiated 2026-04-21 · Enforcement Report 2026-06-24 · Ongoing
- Z-2388-2026openFDA record
ARROW INTERNATIONAL, LLC
Description/REF: HEMODIALYSIS KIT: 2-LUMEN 12 FR X 20 CM/AK-15122-F, AHDC KIT: 3L 12 FR X 16 CM/ASK-12123-UOC1, AHDC KIT: 3-L 12 FR X 16 CM/ASK-12123-UPM, AHDC KIT: 3L 12 FR X 20 CM/ASK-15123-UOC1, AHDC KIT: 3-L 12 FR X 20 CM/ASK-15123-UPM, AHDC KIT: 3-L 12 FR X 16 CM/CDC-12123-…
Reason: Lidocaine, Bupivacaine, and 0.9% sodium chloride, the subject of a supplier recall, were included into kits and sets. The supplier was found to have quality issues and limited reports of drug ineffectiveness, and if sterility of a water-based injectable product is compromised th…
Device · initiated 2026-04-21 · Enforcement Report 2026-06-24 · Ongoing
- Z-2389-2026openFDA record
ARROW INTERNATIONAL, LLC
Description/REF: SPINAL ANESTHESIA KIT/ASA-24090-SB; SPINAL ANESTHESIA KIT/ASA-24103-SB; SPINAL ANESTHESIA KIT/ASA-25090-SB; SPINAL ANESTHESIA KIT: 24 GA X 9 CM/ASK-24090-SA; SPINAL ANESTHESIA KIT: 25 GA X 9 CM/ASK-24090-SPH; SPINAL ANESTHESIA KIT/ASK-25090-BIDMC; SPINAL ANESTHE…
Reason: Lidocaine, Bupivacaine, and 0.9% sodium chloride, the subject of a supplier recall, were included into kits and sets. The supplier was found to have quality issues and limited reports of drug ineffectiveness, and if sterility of a water-based injectable product is compromised th…
Device · initiated 2026-04-21 · Enforcement Report 2026-06-24 · Ongoing
- Z-2390-2026openFDA record
ARROW INTERNATIONAL, LLC
Description/REF: SAC KIT: 18 GA X 12CM W BIO/SAC-01218-B; SAC KIT: 18 GA X 12CM W CHG/SAC-01218-C; SAC KIT: 18 GA X 16CM W BIO/SAC-01618-B
Reason: Lidocaine, Bupivacaine, and 0.9% sodium chloride, the subject of a supplier recall, were included into kits and sets. The supplier was found to have quality issues and limited reports of drug ineffectiveness, and if sterility of a water-based injectable product is compromised th…
Device · initiated 2026-04-21 · Enforcement Report 2026-06-24 · Ongoing
- Z-2391-2026openFDA record
ARROW INTERNATIONAL, LLC
Description/REF: RA CATH KIT: 20 GA X 1-3/4IN/AK-04020-SPC; RA CATH KIT: 20 GA X 5 IN/ASK-04020-MW; ARTERIAL LINE KIT: SL 20 GA X 5 IN/ASK-04510-MM5; CATH SL 20 GA X 5 CM HYDROPHILIC/ASK-00520-JH; CATH S-L 20 GA X 5 CM HYDROPHILIC/ASK-00520-UHC; ARTERIAL LINE KIT: 20 GA x 5IN/AS…
Reason: Lidocaine, Bupivacaine, and 0.9% sodium chloride, the subject of a supplier recall, were included into kits and sets. The supplier was found to have quality issues and limited reports of drug ineffectiveness, and if sterility of a water-based injectable product is compromised th…
Device · initiated 2026-04-21 · Enforcement Report 2026-06-24 · Ongoing
- Z-2392-2026openFDA record
ARROW INTERNATIONAL, LLC
Description/REF: ACCESS TRAY/ASK-04001-MC3
Reason: Lidocaine, Bupivacaine, and 0.9% sodium chloride, the subject of a supplier recall, were included into kits and sets. The supplier was found to have quality issues and limited reports of drug ineffectiveness, and if sterility of a water-based injectable product is compromised th…
Device · initiated 2026-04-21 · Enforcement Report 2026-06-24 · Ongoing
- Z-2393-2026openFDA record
ARROW INTERNATIONAL, LLC
Description/REF: PSI KIT: 7 FR X 4 IN/ASK-09701-UP; PSI KIT: 8.5 FR X 4 1/8 IN (10 CM)/ASK-09803-OSU5; PSI KIT: 8.5 FR/ASK-09807-UC; PSI KIT: 9 FR X 4 1/8IN (10 CM)/ASK-09903-HH2; PSI KIT: 9 FR X 10 CM/ASK-09903-LMDTG; PSI KIT: 9 FR X 4 1/8 IN (10 CM)/ASK-09903-OSU5; PSI KIT: 9…
Reason: Lidocaine, Bupivacaine, and 0.9% sodium chloride, the subject of a supplier recall, were included into kits and sets. The supplier was found to have quality issues and limited reports of drug ineffectiveness, and if sterility of a water-based injectable product is compromised th…
Device · initiated 2026-04-21 · Enforcement Report 2026-06-24 · Ongoing
- Z-2394-2026openFDA record
ARROW INTERNATIONAL, LLC
Description/REF: MIDLINE CATHETER KIT: 3 FR X 20 CM/CDC-02031-MK1A; AM/AT PI MIDLINE 1L: 4.5FR X 15CM/CDC-41541-MPK1A; AM/AT PI MIDLINE 2L: 5.5FR X 15CM/CDC-41552-MPK1A; AGBA PI MIDLINE 1L: 4.5FR X 15CM/ASK-41541-VFHM; AGBA PI MIDLINE 1-L: 4.5FR X 15CM/CDC-41541-MPK; AGBA PI MID…
Reason: Lidocaine, Bupivacaine, and 0.9% sodium chloride, the subject of a supplier recall, were included into kits and sets. The supplier was found to have quality issues and limited reports of drug ineffectiveness, and if sterility of a water-based injectable product is compromised th…
Device · initiated 2026-04-21 · Enforcement Report 2026-06-24 · Ongoing
- Z-2395-2026openFDA record
ARROW INTERNATIONAL, LLC
Description/REF: PICC ACCESS TRAY/ASK-04001-CG1
Reason: Lidocaine, Bupivacaine, and 0.9% sodium chloride, the subject of a supplier recall, were included into kits and sets. The supplier was found to have quality issues and limited reports of drug ineffectiveness, and if sterility of a water-based injectable product is compromised th…
Device · initiated 2026-04-21 · Enforcement Report 2026-06-24 · Ongoing
- Z-2396-2026openFDA record
ARROW INTERNATIONAL, LLC
Description/REF: CVC KIT: 16 GA X 8IN/AK-04210, CVC KIT: 1-L 16 GA X 8IN (20 CM)/AK-04300, CVC KIT: 14 GA X 6 IN/CK-04711
Reason: Lidocaine, Bupivacaine, and 0.9% sodium chloride, the subject of a supplier recall, were included into kits and sets. The supplier was found to have quality issues and limited reports of drug ineffectiveness, and if sterility of a water-based injectable product is compromised th…
Device · initiated 2026-04-21 · Enforcement Report 2026-06-24 · Ongoing
- Z-2397-2026openFDA record
ARROW INTERNATIONAL, LLC
Description/REF: RA CATH KIT: 20 GA X 1-1/2IN/NA-04220-1A; RA CATH KIT: 20 GA X 1-1/2IN/NA-04220-X1A; ARTERIAL CATHETER KIT: 18 GA X 16 CM/ASK-04018-HCA; ARTERIAL CATHETER KIT: 18 GA X 16 CM/ASK-04018-UPM; RA CATH KIT: 20 GA X 1-3/4IN/ASK-04020-HC; ARTERIAL LINE KIT: 20 GA X 5IN…
Reason: Lidocaine, Bupivacaine, and 0.9% sodium chloride, the subject of a supplier recall, were included into kits and sets. The supplier was found to have quality issues and limited reports of drug ineffectiveness, and if sterility of a water-based injectable product is compromised th…
Device · initiated 2026-04-21 · Enforcement Report 2026-06-24 · Ongoing
- Z-2357-2026openFDA record
Datex-Ohmeda Inc
Giraffe Stand-alone Infant Resuscitation System; Powered emergency ventilator (resuscitator) Product identifiers: 2082663-001, 2082663-002, M1241424 Product impacted only if service blender M1091607-R was installed during the last Preventative Maintenance
Reason: Blender knob shaft can loosen, which can affect the delivered oxygen concentration. The system can deliver an oxygen concentration that does not match the intended concentration set by the user, potentially resulting in hypoxia or hyperoxia in the patient.
Device · initiated 2026-05-08 · Enforcement Report 2026-06-24 · Ongoing
- Z-2358-2026openFDA record
Datex-Ohmeda Inc
Giraffe Warmer with Integrated Resuscitation System (iRes), Panda Warmer with Integrated Resuscitation System (iRes) Product Identifiers: M1118179, M1112198 Impacted only if service blender M1091607-R was installed during the last Preventative Maintenance.
Reason: Blender knob shaft can loosen, which can affect the delivered oxygen concentration. The system can deliver an oxygen concentration that does not match the intended concentration set by the user, potentially resulting in hypoxia or hyperoxia in the patient.
Device · initiated 2026-05-08 · Enforcement Report 2026-06-24 · Ongoing
- Z-2359-2026openFDA record
Datex-Ohmeda Inc
Panda Freestanding Warmer, Infant Radiant Warmer Product Identifiers: 2063568-001 Impacted only if service blender M1091607-R was installed during the last Preventative Maintenance.
Reason: Blender knob shaft can loosen, which can affect the delivered oxygen concentration. The system can deliver an oxygen concentration that does not match the intended concentration set by the user, potentially resulting in hypoxia or hyperoxia in the patient.
Device · initiated 2026-05-08 · Enforcement Report 2026-06-24 · Ongoing
- D-0620-2026openFDA record
BEEKEEPER'S NATURALS USA INC.
BEEKEEPER'S NATURALS Saline Nasal Spray, Sinus Congestion Rinse, Made with Propolis + Xylitol, 1 FL OZ (30 mL) per bottle, Manufactured For: Beekeeper's Naturals USA Inc., Covina, CA 91789.
Reason: Microbial Contamination of Non-Sterile Products
Drug · initiated 2026-06-01 · Enforcement Report 2026-06-17 · Ongoing
- D-0595-2026openFDA record
Haleon US Holdings LLC
Gas-X Extra Strength, SIMETHICONE 125 mg/ANTIGAS, packaged in a) 120 SoftGels (UPC 3 00674 35041 9, b) 72 SoftGels (UPC 3 00439 00572 1), Distributed by: Haleon, Warren, NJ 07059.
Reason: Chemical Contamination: contamination with a diluted propylene glycol-based coolant from a machine leakage during the packaging process.
Drug · initiated 2026-06-04 · Enforcement Report 2026-06-17 · Ongoing
- D-0580-2026openFDA record
SUN PHARMACEUTICAL INDUSTRIES INC
DOXOrubin Hydrochloride Liposome injection, 50 mg/25 mL (2mg/mL), 25 mL single-dose vials, Sterile, Rx only, Manufactured for: Northstar Rx LLC., Memphis, TN 38141, Manufactured by: Sun Pharmaceutical Ind. Ltd., Halol-Baroda Highway, Halol, Gujarat, India, NDC 72603-200-01.
Reason: Presence of Particulate matter: Particulate matter identified as glass.
Drug · initiated 2026-05-12 · Enforcement Report 2026-06-17 · Ongoing
- Z-2276-2026openFDA record
B Braun Medical Inc
Spinal Needle procedure kits: Material Description (Material Number) 24GA. PENCAN SPINAL BUP. TRAY W/DURAPREP (333867); 25GA. PENCAN SPINAL BUP. TRAY W/DURAPREP (333862); GOVP24BK 24G PENCAN TRAY W/ BUP & DEXTROS (332607); P24BK 24G PENCAN TRAY W/BUP & DEXTROSE (333868); P25BKAY…
Reason: These procedure kits contain Bupivacaine Hydrochloride in Dextrose Injection, USP. The manufacturer of the Bupivacaine, Huons Co, Ltd., has issued a voluntary recall due to potential quality issues identified during a recent FDA manufacturing facility inspection.
Device · initiated 2026-05-06 · Enforcement Report 2026-06-10 · Ongoing
- Z-2277-2026openFDA record
B Braun Medical Inc
Spinocan Spinal Needle procedure kits: Material Description (Material Number): Spinocan 26 Ga. x 3-1/2 in. (90 mm), Bupivacaine 0.75% with Dextrose 8.25% Tray (Kit) (333201); Spinocan 25 Ga. x 3-1/2 in. (90 mm), Bupivacaine 0.75% with Dextrose 8.25% Tray (Kit) (333211); Spinocan…
Reason: These procedure kits contain Bupivacaine Hydrochloride in Dextrose Injection, USP. The manufacturer of the Bupivacaine, Huons Co, Ltd., has issued a voluntary recall due to potential quality issues identified during a recent FDA manufacturing facility inspection.
Device · initiated 2026-05-06 · Enforcement Report 2026-06-10 · Ongoing
- Z-2278-2026openFDA record
B Braun Medical Inc
Design Options Spinal Tray Spinal Needle Anesthesia procedure kits: Material Description (Material Number): SSK H M C 20061890 (560074), SSK S C H (560214), SSK BJC H S (ORIGINAL CUST) (560272), SSK CA P M C (ORIGINAL CUST) (560327), SSK T M H (560394), SSK S O 20070208 (560428)…
Reason: These procedure kits contain Bupivacaine Hydrochloride in Dextrose Injection, USP. The manufacturer of the Bupivacaine, Huons Co, Ltd., has issued a voluntary recall due to potential quality issues identified during a recent FDA manufacturing facility inspection.
Device · initiated 2026-05-06 · Enforcement Report 2026-06-10 · Ongoing
- Z-2248-2026openFDA record
Becton Dickinson & Company
BD¿Spinal Tray with BD¿Quincke Needle 22 G x 3.5 in.; lidocaine HCL (1%), 5 mL, bupivacaine HCL (0.75%) with dextrose (8.25%), 2 mL (marcaine). Catalog Number: 405609.
Reason: This is a sub-recall of a product recall from Huons Co, Ltd., the manufacturer of Bupivacaine Hydrochloride in Dextrose Injection, USP. The recall is due to quality issues identified during an FDA manufacturing facility inspection, citing specific quality deficiencies related to…
Device · initiated 2026-04-27 · Enforcement Report 2026-06-10 · Ongoing
- Z-2249-2026openFDA record
Becton Dickinson & Company
BD¿Spinal Tray with BD¿Quincke Needle 25 G x 3.5 in.; lidocaine HCL (1%), 5 mL, bupivacaine HCL (0.75%) with dextrose (8.25%), 2 mL (marcaine). Catalog Numbers: 405621, 405735.
Reason: This is a sub-recall of a product recall from Huons Co, Ltd., the manufacturer of Bupivacaine Hydrochloride in Dextrose Injection, USP. The recall is due to quality issues identified during an FDA manufacturing facility inspection, citing specific quality deficiencies related to…
Device · initiated 2026-04-27 · Enforcement Report 2026-06-10 · Ongoing
- Z-2250-2026openFDA record
Becton Dickinson & Company
BD¿Spinal Tray with BD¿Quincke Needle 26 G x 3.5 in.; lidocaine HCL (1%), 5 mL, bupivacaine HCL (0.75%) with dextrose (8.25%), 2 mL (marcaine). Catalog Number: 405632.
Reason: This is a sub-recall of a product recall from Huons Co, Ltd., the manufacturer of Bupivacaine Hydrochloride in Dextrose Injection, USP. The recall is due to quality issues identified during an FDA manufacturing facility inspection, citing specific quality deficiencies related to…
Device · initiated 2026-04-27 · Enforcement Report 2026-06-10 · Ongoing
- Z-2251-2026openFDA record
Becton Dickinson & Company
BD¿Spinal Tray with BD¿Whitacre Needle 22 G x 3.5 in.; lidocaine HCL (1%), 5 mL, bupivacaine HCL (0.75%) with dextrose (8.25%), 2 mL (marcaine). Catalog Number: 405645.
Reason: This is a sub-recall of a product recall from Huons Co, Ltd., the manufacturer of Bupivacaine Hydrochloride in Dextrose Injection, USP. The recall is due to quality issues identified during an FDA manufacturing facility inspection, citing specific quality deficiencies related to…
Device · initiated 2026-04-27 · Enforcement Report 2026-06-10 · Ongoing
- Z-2252-2026openFDA record
Becton Dickinson & Company
BD¿Spinal Tray with BD¿Whitacre Needle 24 G x 3.5 in.; lidocaine HCL (1%), 5 mL, bupivacaine HCL (0.75%) with dextrose (8.25%), 2 mL (marcaine). Catalog Numbers: 400868, 405652.
Reason: This is a sub-recall of a product recall from Huons Co, Ltd., the manufacturer of Bupivacaine Hydrochloride in Dextrose Injection, USP. The recall is due to quality issues identified during an FDA manufacturing facility inspection, citing specific quality deficiencies related to…
Device · initiated 2026-04-27 · Enforcement Report 2026-06-10 · Ongoing
- Z-2253-2026openFDA record
Becton Dickinson & Company
BD¿Spinal Tray with BD¿Whitacre Needle 25 G x 3.5 in., BD¿Quincke Needle 22 G x 3.5 in.; lidocaine HCL (1%), 5 mL, bupivacaine HCL (0.75%) with dextrose (8.25%), 2 mL (marcaine). Catalog Number: 405741.
Reason: This is a sub-recall of a product recall from Huons Co, Ltd., the manufacturer of Bupivacaine Hydrochloride in Dextrose Injection, USP. The recall is due to quality issues identified during an FDA manufacturing facility inspection, citing specific quality deficiencies related to…
Device · initiated 2026-04-27 · Enforcement Report 2026-06-10 · Ongoing
- Z-2254-2026openFDA record
Becton Dickinson & Company
BD¿Spinal Tray with BD¿Whitacre Needle 25 G x 3.5 in.; lidocaine HCL (1%), 2 mL, bupivacaine HCL (0.75%) with dextrose (8.25%), 2 mL (marcaine). Catalog Number: 405658.
Reason: This is a sub-recall of a product recall from Huons Co, Ltd., the manufacturer of Bupivacaine Hydrochloride in Dextrose Injection, USP. The recall is due to quality issues identified during an FDA manufacturing facility inspection, citing specific quality deficiencies related to…
Device · initiated 2026-04-27 · Enforcement Report 2026-06-10 · Ongoing
- Z-2255-2026openFDA record
Becton Dickinson & Company
BD¿Spinal Tray with BD¿Whitacre Needle 25 G x 3.5 in.; lidocaine HCL (1%), 5 mL, bupivacaine HCL (0.75%) with dextrose (8.25%), 2 mL (marcaine). Catalog Numbers: 400866, 405672, 405673, 405674, 405675, 405709, 405721, 405723, 406704.
Reason: This is a sub-recall of a product recall from Huons Co, Ltd., the manufacturer of Bupivacaine Hydrochloride in Dextrose Injection, USP. The recall is due to quality issues identified during an FDA manufacturing facility inspection, citing specific quality deficiencies related to…
Device · initiated 2026-04-27 · Enforcement Report 2026-06-10 · Ongoing
- Z-2256-2026openFDA record
Becton Dickinson & Company
BD¿Spinal Tray with BD¿Whitacre Needle 25 G x 3.5 in.; lidocaine HCL (1%), 5 mL, bupivacaine HCL (0.75%) with dextrose (8.25%), 2 mL (marcaine). Catalog Number: 405671.
Reason: This is a sub-recall of a product recall from Huons Co, Ltd., the manufacturer of Bupivacaine Hydrochloride in Dextrose Injection, USP. The recall is due to quality issues identified during an FDA manufacturing facility inspection, citing specific quality deficiencies related to…
Device · initiated 2026-04-27 · Enforcement Report 2026-06-10 · Ongoing
- Z-2257-2026openFDA record
Becton Dickinson & Company
BD¿Spinal Tray with BD¿Whitacre Needle 25 G x 3.5 in.; lidocaine HCL (1%), 5 mL, bupivacaine HCL (0.75%) with dextrose (8.25%), 2 mL (marcaine). Catalog Numbers: 405637, 405707.
Reason: This is a sub-recall of a product recall from Huons Co, Ltd., the manufacturer of Bupivacaine Hydrochloride in Dextrose Injection, USP. The recall is due to quality issues identified during an FDA manufacturing facility inspection, citing specific quality deficiencies related to…
Device · initiated 2026-04-27 · Enforcement Report 2026-06-10 · Ongoing
- Z-2207-2026openFDA record
Draeger, Inc.
Atlan A350. Model Number: 8211500. anesthesia workstation
Reason: The Atlan anesthesia workstation either indicated a failure of the piston ventilator before use or suffered a failure of the mechanical ventilation during use.
Device · initiated 2026-05-07 · Enforcement Report 2026-06-10 · Ongoing
- Z-2208-2026openFDA record
Draeger, Inc.
Atlan A350XL. Model Number: 8621600. anesthesia workstation
Reason: The Atlan anesthesia workstation either indicated a failure of the piston ventilator before use or suffered a failure of the mechanical ventilation during use.
Device · initiated 2026-05-07 · Enforcement Report 2026-06-10 · Ongoing
- Z-2223-2026openFDA record
Fresenius Kabi USA, LLC
LVP software of the Ivenix Infusion System (IIS); Product Code LVP-SW-0005.
Reason: Potential software anomaly that can cause the pump to report a false battery-health value (typically 69%) and trigger a battery-depletion alarm that will shut down the pump unless it is reconnected to AC mains power.
Device · initiated 2026-05-06 · Enforcement Report 2026-06-10 · Ongoing
- Z-2224-2026openFDA record
Fresenius Kabi USA, LLC
Ivenix Infusion System (IIS), Large Volume Pump, Model Number: LVP-0004
Reason: Fresenius Kabi has been made aware via customer complaints that dropping or severe jarring of the Ivenix LVP may cause damage to internal components that may not be apparent or readily visible and the devices therefore remained in service. More specifically, drops or jarring on…
Device · initiated 2026-05-06 · Enforcement Report 2026-06-10 · Ongoing
- Z-2212-2026openFDA record
KAYSERBETTEN GMBH & CO. KG
Pediatric care bed; Product Designation: KayserBett IDA;
Reason: If the adjustment functions of the hand control, as instructed in the Manual, are not locked with the provided key when the patient (child) is unattended, there is an increased risk that the patient (child) or third parties (children) will operate the hand control and trap thems…
Device · initiated 2026-05-04 · Enforcement Report 2026-06-10 · Ongoing
- Z-2211-2026openFDA record
Abiomed, Inc.
Automated Impella Controller (AIC); Product Code: 0042-0000-US;
Reason: A retrospective review identified that there are AIC units requiring specific hardware updates to address potential safety concerns. These hardware updates have been implemented through Abiomed s servicing process; therefore, AICs that have not received service are pending these…
Device · initiated 2026-04-20 · Enforcement Report 2026-06-03 · Ongoing
- Z-2238-2026openFDA record
Abiomed, Inc.
Automated Impella Controller (AIC) with the below product descriptions and corresponding Product Codes. 1. Impella Controller, Packaged, CA; Product Code: 0042-0000-CA. 2. Impella Controller, Packaged, EU; Product Code: 0042-0000-EU. 3. Impella Controller, Packaged, UK; Product…
Reason: Retrospective submission for following issues identified: 1. Alarm failures, power-path faults, electrical-short risks, and inaccurate motor-current sensing were identified in certain AIC consoles. 2. There is improper alignment between the purge cassette and the motor drive in…
Device · initiated 2011-01-17 · Enforcement Report 2026-06-03 · Ongoing
- Z-2160-2026openFDA record
Bolton Medical Inc.
RelayPro Thoracic Stent-Graft System (various sizes) Reference numbers: 28-N4-32-104-32U 28-N4-32-164-28U 28-N4-32-164-32U 28-N4-32-209-28U 28-N4-32-209-32U 28-N4-32-259-32U 28-N4-34-109-34U 28-N4-34-154-30U 28-N4-34-154-34U 28-N4-34-209-30U 28-N4-34-209-34U 28-N4-34-259-34U 28-…
Reason: Labeling update concerning proximal clasp disconnection from the outer control tube which may prevent stent graft from being released from the delivery system.
Device · initiated 2026-04-22 · Enforcement Report 2026-06-03 · Ongoing
- Z-2231-2026openFDA record
Medline Industries, LP
Medline kits containing Huons Co. Bupivacaine HCL in Dextrose Injection, USP, labeled as: 1) ANESTHESIA TRAY, Medline SKU DYNJRA1355A; 2) SPINAL ANESTHESIA TRAY, Medline SKU DYNJRA2151.
Reason: Medline kits were distributed containing Huons Bupivacaine Hydrochloride in Dextrose Injection, USP, which was subsequently recalled due to quality issues, including microbiology testing data integrity concerns and reported efficacy complaints.
Device · initiated 2026-04-10 · Enforcement Report 2026-06-03 · Ongoing
- Z-2232-2026openFDA record
Medline Industries, LP
Medline kits containing Huons Co. Bupivacaine HCL in Dextrose Injection, USP, labeled as: 1) PAIN TRAY, Medline Kit SKU DYNJRA1555; 2) SPINAL BLOCK TRAY WHITACRE 24G, Medline Kit SKU DYNJRA1817; 3) TRAY, 25G WHIT SPINAL W/PHARM, Medline Kit SKU DYNJRA1920; 4) ULTRASOUND GUIDED N…
Reason: Medline kits were distributed containing Huons Bupivacaine Hydrochloride in Dextrose Injection, USP, which was subsequently recalled due to quality issues, including microbiology testing data integrity concerns and reported efficacy complaints.
Device · initiated 2026-04-10 · Enforcement Report 2026-06-03 · Ongoing
- Z-2233-2026openFDA record
Medline Industries, LP
Medline kits containing Huons Co. Bupivacaine HCL in Dextrose Injection, USP, labeled as: 1) SPINAL TRAY W/PENCIL POINT NDL, Medline Kit SKU DYNJRA0245; 2) SPINAL TRAY W/PENCIL POINT NDL, Medline Kit SKU DYNJRA0555; 3) SPINAL TRAY, Medline Kit SKU DYNJRA0836A; 4) SPINAL TRAY, Me…
Reason: Medline kits were distributed containing Huons Bupivacaine Hydrochloride in Dextrose Injection, USP, which was subsequently recalled due to quality issues, including microbiology testing data integrity concerns and reported efficacy complaints.
Device · initiated 2026-04-10 · Enforcement Report 2026-06-03 · Ongoing
- Z-2234-2026openFDA record
Medline Industries, LP
Medline kits containing Huons Co. Bupivacaine HCL in Dextrose Injection, USP, labeled as: 1) BLOCK TRAY, Medline Kit SKU DYNJRA1181C; 2) SPINAL BLOCK 25G WHTCRE 5S BU, Medline Kit SKu DYNJRA1860.
Reason: Medline kits were distributed containing Huons Bupivacaine Hydrochloride in Dextrose Injection, USP, which was subsequently recalled due to quality issues, including microbiology testing data integrity concerns and reported efficacy complaints.
Device · initiated 2026-04-10 · Enforcement Report 2026-06-03 · Ongoing
- Z-2235-2026openFDA record
Medline Industries, LP
Medline kits containing Huons Co. Bupivacaine HCL in Dextrose Injection, USP, labeled as: SPINAL BLOCK 22G QUINCKE 5S BU, Medline Kit SKU DYNJRA9026
Reason: Medline kits were distributed containing Huons Bupivacaine Hydrochloride in Dextrose Injection, USP, which was subsequently recalled due to quality issues, including microbiology testing data integrity concerns and reported efficacy complaints.
Device · initiated 2026-04-10 · Enforcement Report 2026-06-03 · Ongoing
- Z-2236-2026openFDA record
Medline Industries, LP
Medline kits containing Huons Co. Bupivacaine HCL in Dextrose Injection, USP, labeled as: 1) SPINAL BLOCK TRAY, Medline Kit SKU DYNJRA1097; 2) TRAY, SPINAL PENCIL POINT 25G, Medline Kit SKU DYNJRA1154; 3) SPINAL TRAY, Medline Kit SKU DYNJRA1192A; 4) BLOCK TRAY, Medline Kit SKU D…
Reason: Medline kits were distributed containing Huons Bupivacaine Hydrochloride in Dextrose Injection, USP, which was subsequently recalled due to quality issues, including microbiology testing data integrity concerns and reported efficacy complaints.
Device · initiated 2026-04-10 · Enforcement Report 2026-06-03 · Ongoing
- Z-2237-2026openFDA record
Medline Industries, LP
Medline kits containing Huons Co. Bupivacaine HCL in Dextrose Injection, USP, labeled as: 1) SPINAL ANESTHESIA TRAY-LF, Medline Kit SKU DYNJRA0102A; 2) SPINAL TRAY-LF, Medline Kit SKU DYNJRA0143B; 3) SPINAL PREP TRAY-LF, Medline Kit SKU DYNJRA0205B; 4) SPINAL TRAY, Medline Kit S…
Reason: Medline kits were distributed containing Huons Bupivacaine Hydrochloride in Dextrose Injection, USP, which was subsequently recalled due to quality issues, including microbiology testing data integrity concerns and reported efficacy complaints.
Device · initiated 2026-04-10 · Enforcement Report 2026-06-03 · Ongoing
Import alert entries
22 firm entries FDA published on its import alert pages this month, dated by each entry's date published: 18 on a Red List, 4 on a Green List. FDA lists firms by name and address; FDAgent links an entry to a facility only when FDA printed the firm's FEI. 1 entry whose name may be a person's rather than a firm's is not listed or counted here; signed-in users can see them in FDAgent. Import alerts last synced Nov 20, 2026, 03:55 UTC.
66-40Red List8 entriesAlert on fda.govShow allDetention Without Physical Examination of Drugs From Firms Which Have Not Met Drug GMPsAlert on fda.gov
- Published 2026-06-22
BABIKIAN HEALTHCARE PRODUCTS
ARMENIA
- Published 2026-06-22
Babikian Healthcare Products, CJSC
ARMENIA
- Published 2026-06-22
Nanjing Han Xin Pharmaceutical Technology Co., Ltd. Branch 1
CHINA
- Published 2026-06-22
Nanjing Hanxin Pharmaceutical Technology Co., Ltd.
CHINA
- Published 2026-06-22
Tan-Alize Kozmetik ve Temizlik Urunleri Sanayi ve Ticaret A.S.
TURKEY
- Published 2026-06-09
Almon Healthcare Private Limited
INDIA
- Published 2026-06-05
Dabur India Limited
INDIA
- Published 2026-06-02
Shimoga Chemicals
INDIA
66-41Red List4 entriesAlert on fda.govShow allDetention Without Physical Examination of Unapproved New Drugs Promoted In The U.S.Alert on fda.gov
- Published 2026-06-23
Hebei Houde Hanfang Medical Devices Co., Ltd.
CHINA
- Published 2026-06-23
Kobayashi Pharmaceutical Co Ltd
JAPAN
- Published 2026-06-12
NAVEH PHARMA (1996) LTD
ISRAEL
- Published 2026-06-12
UNOLAB MANUFACTURING AVDA
SPAIN
66-79Red List2 entriesAlert on fda.govShow allDetention Without Physical Examination of Drugs From Foreign Establishments Refusing FDA InspectionAlert on fda.gov
- Published 2026-06-22
Shenyang Antibiotic Manufacturer
CHINA
- Published 2026-06-03
Yangzhou Hongshengding Chemical Co., Ltd.
CHINA
68-19Red List1 entryAlert on fda.govShow allDETENTION WITHOUT PHYSICAL EXAMINATION OF UNAPPROVED FINISHED NEW ANIMAL DRUGSAlert on fda.gov
- Published 2026-06-22
CHEMIPOWER OU
ESTONIA
76-01Green List4 entriesAlert on fda.govShow allDetention Without Physical Examination Of Medical Instruments from PakistanAlert on fda.gov
- Published 2026-06-10
Flozen International
PAKISTAN
- Published 2026-06-09
Miracle Instruments
PAKISTAN
- Published 2026-06-09
Next Edge Surgical
PAKISTAN
- Published 2026-06-03
Dibraj Surgical
PAKISTAN
89-04Red List1 entryAlert on fda.govShow allDetention Without Physical Examination of Devices from Firms that Have not met Device Quality System RequirementsAlert on fda.gov
- Published 2026-06-12
Heuer International (A division of GST Corporation Limited)
INDIA
99-34Red List1 entryAlert on fda.govShow allDETENTION WITHOUT PHYSICAL EXAMINATION OF DRUGS OR MEDICAL DEVICES FROM FIRMS WITHOUT A VALID DRUG OR MEDICAL DEVICE REGISTRATION AND/OR LISTINGAlert on fda.gov
- Published 2026-06-12
BRIGHT SUNWAY LIMITED
CHINA
99-49Red List1 entryAlert on fda.govShow allDetention Without Physical Examination of Devices Without a Unique Device Identifier and/or Matching GUDID RecordAlert on fda.gov
- Published 2026-06-22
BRIGHT SUNWAY LIMITED
CHINA
New OAI classifications
5 inspections classified Official Action Indicated that FDAgent first recorded from FDA's inspection classification data this month, after its reads of that data in the 8 days before had not found them. Older inspections FDAgent adds to its archive are left out. Dates are inspection end dates. Inspection classifications last synced Nov 22, 2026, 04:55 UTC.
- Inspection ended 2026-03-27FEI 2244819
Bentley Laboratories LLCOAI
Edison, New Jersey, United States · Drugs · Drug Quality Assurance · recorded 2026-06-23
- Inspection ended 2026-03-20FEI 3013559620
Eosera, Inc.OAI
Fort Worth, Texas, United States · Drugs · Drug Quality Assurance · recorded 2026-06-16
- Inspection ended 2026-04-02FEI 3011407349
Happy Farm Botanicals, Inc.OAI
Hyattsville, Maryland, United States · Drugs · Drug Quality Assurance · recorded 2026-06-16
- Inspection ended 2026-04-03FEI 1625647
Conroe Welding Supply, IncOAI
Conroe, Texas, United States · Drugs · Drug Quality Assurance · recorded 2026-06-09
- Inspection ended 2026-03-06FEI 3001592626
Linemaster Switch CorporationOAI
Woodstock, Connecticut, United States · Devices · Compliance: Devices · recorded 2026-06-02
Week by week
Each week's page lists every Form 483 and recall. Weeks run Monday to Sunday, so the first and last can include days of the neighbouring months.
| Week | Dates | Letters | Form 483s | Recalls | New OAI |
|---|---|---|---|---|---|
| 2026-W26 | 2026-06-22 to 2026-06-28 | 14 letters | 1 Form 483 | 171 recalls | 1 new OAI |
| 2026-W25 | 2026-06-15 to 2026-06-21 | 30 letters | 0 Form 483s | 74 recalls | 2 new OAI |
| 2026-W24 | 2026-06-08 to 2026-06-14 | 13 letters | 0 Form 483s | 96 recalls | 1 new OAI |
| 2026-W23 | 2026-06-01 to 2026-06-07 | 5 letters | 0 Form 483s | 25 recalls | 1 new OAI |
Get each week’s page by email every Monday
A free weekly email with each week’s warning letters, new Form 483s, OAI classifications and Class I recalls, counted the same way as this page. Just your email address.
How this month is counted
Months are calendar months in UTC. Counts are read from FDAgent's copy of each FDA source, which refreshes daily, and FDAgent checks them against FDA's own sources every week. FDA posts some records late and removes others, so a finished month's counts can still change. FDAgent is independent of FDA.
- Warning letters posted
- Letters from FDA's drug, biologics and device offices, dated by the day FDA posted them, not the day it issued them.
- Form 483s listed
- Records FDA's Electronic Reading Room types as 483 or Amended 483, dated by FDA's publish date. A firm's 483 response is not counted. Each inspection's 483 counts once, when FDA first listed it: a later listing with the same FEI and record date is shown apart, as a re-post when FDA listed a 483 again under a new media id, or as an amended 483 when FDA types it Amended 483. FDA sometimes lists a 483 days after the publish date it gives it, so a finished month can still gain 483s.
- Inspections newly classified
- OAI, VAI and NAI inspections counted on the day FDAgent first found them in FDA's classification data, and only when its complete reads in the 8 days before had covered the inspection's end date without finding them. Inspections recorded before FDAgent kept a record of its reads, in November 2026, count when they ended in the 13 weeks before. Backfills and wider reads never count, nor do the Veterinary inspections of FDAgent's first read of them, in November 2026. FDA gives no date for a classification. When FDA removes an inspection from its Data Dashboard, it stops counting, so a finished month's count can fall.
- Regulations cited
- The FDA Data Dashboard's Form 483 citations on those inspections, by 21 CFR section. FDA often lists an inspection's citations weeks after its classification, so these counts grow after the month ends. For letters, the violation statements that cite a 21 CFR section, each section once per letter.
- Recalls
- openFDA's drug and device records by Enforcement Report date. Each record is one product; an event is FDA's recall event number.
- Import alert entries
- Firm entries by the date FDA published them on an import alert. Entries that may name a person are left out. FDA deletes an entry when it removes the firm; entries FDAgent recorded from November 2026 on keep counting in the month FDA dated them.
- Months in progress
- A month in progress counts the days so far and grows until it ends. Recalls for its latest days appear when openFDA updates.
- Comparisons
- Every month is counted the same way. A year-over-year figure appears only where FDAgent holds FDA's own dates for the earlier month. A percent change appears only when the earlier count is 10 or more. A month in progress shows no changes until it ends. A month that overlaps a federal government shutdown (Oct 1 to Nov 12, 2025) can have low counts, and month pages say so beside any comparison with one.
Cite this page
Quote a count with its date: the page refreshes daily, and a month's counts can change as FDA posts late records.
FDAgent. "FDA enforcement, June 2026." https://www.fdagent.app/enforcement/2026-06. Counts as of Nov 22, 2026.
About this data
- Weeks are ISO weeks, Monday to Sunday, in UTC. Each section is dated by its own FDA date: posting date for warning letters, reading-room publish date for Form 483s (each inspection's 483 counts once, when FDA first listed it; FDA's re-posts and amended 483s with the FEI and record date of one it listed before are shown apart), Enforcement Report date for recalls, and the date published FDA prints beside each firm on an import alert. FDA publishes no date for an inspection classification, so an OAI classification counts as new on the day FDAgent first recorded the inspection, and only when FDAgent's complete reads of FDA's inspection data in the 8 days before had covered the inspection's end date without finding it. Older inspections FDAgent adds to its archive, by a backfill or a wider read, never count as new, however late FDA posted them. Inspections FDAgent recorded before it began keeping a record of its reads, in November 2026, count only when they ended in the 13 weeks before, and the Veterinary inspections of FDAgent's first read of them, that same month, never count.
- Import alert entries come from FDA's import alert pages, where FDA lists firms by name and address, not by FEI. FDAgent links an entry to a facility only when FDA printed the firm's FEI, never by name. Earlier weeks list only entries still on FDA's pages, because FDA deletes an entry when it removes a firm, and FDAgent keeps removed entries only from the day it began recording them. FDAgent lists an entry here only when its name shows it is a firm, through a legal form such as Ltd, Inc or GmbH, a business word such as Laboratories or Trading, or a web address, or when FDA's inspection or drug registration records hold a firm under exactly that name; entries that may name a person are shown only to signed-in users.
- Import refusals are not listed by week: FDA reports a refusal weeks after it happens, so a week's count would keep changing after the week closed. Each facility page lists its own refusals.
- Warning letters are those issued by FDA's drug, biologics and device centers and their field and quality offices. Letters from FDA's food, tobacco, veterinary, cosmetics and dietary supplement offices are left out.
- FDA posts only some Form 483s to its Electronic Reading Room, so a missing 483 does not mean none was issued.
- FDA publishes an inspection's classification once it is final and updates its inspection data weekly. Source: FDA inspections FAQ.
- Recalls come from openFDA, which republishes FDA's weekly Enforcement Report; a report's recalls appear here once openFDA publishes it.
- FDAgent is independent of FDA. Every row links to its FDA source; check it before relying on a record.