Linemaster Switch Corporation
29 Plaine Hill Rd, Woodstock, Connecticut, United States
FEI 3001592626
FDA has 5 inspections on file for Linemaster Switch Corporation in Woodstock, Connecticut (FEI 3001592626), from 2012 to 2026. The latest ended on and was classified OAI, official action indicated. FDA issued a warning letter on (CDRH).
OAI means FDA found significant objectionable conditions and recommended regulatory or administrative action. It is the most serious of FDA's three inspection outcomes.
FDA posted the Form 483 from that inspection. FDA published 6 citations from this inspection. They are listed below. Four FOIA requests have asked FDA for this inspection's records.
No court action or import refusal is filed under this FEI.
One regulation was cited at more than one inspection: 21 CFR 820.198(a).
Sources:FDA firm profile ↗Form 483 PDF ↗Warning letter ↗Synced Nov 23, 2026
Ask FDAgent about this site
Email me when FDA posts a new inspection, Form 483 or warning letter for this site
Weekly, on a free account. Pause it any time.
Comparing suppliers? Check this site beside 2 others, free
- Latest inspection
- OAI
- Inspections on file
- 5, 2012–2026
- 1 OAI · 3 VAI · 1 NAI
- Inspections with 483 observations shown
- 3 of 5
- Form 483s posted by FDA
- ↗
- Requested under FOIA
- 4 requests, latest Aug 17, 2026
- Warning letters
- · CDRH ↗
- Import refusals
- None under this FEI
- Import alerts
- Not linked by FEI
- FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗
Inspection history
5 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC
- OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Compliance: DevicesPosted 483 · 6 citations
Read this Form 483 on its own page →Form 483 PDF on fda.gov
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
- ISO 13485:2016 Clause 8.3.1
Nonconforming product not identified and controlled
Product which does not conform to product requirements was not identified and controlled to prevent its unintended use or delivery.
- ISO 13485:2016 Clause 7.6
Software used for monitoring and measurement not validated
Software used for the monitoring and measurement of requirements was not adequately validated.
- ISO 13485:2016 Clause 6.2
Personnel competency not adequate
Personnel performing work that affects product quality are not adequately competent based on appropriate.
- ISO 13485:2016 Clause 8.5.2
Corrective action not taken
Corrective action was not taken to eliminate the cause of nonconformities in order to prevent recurrence.
- ISO 13485:2016 Clause 6.4.1
Work environment requirements and procedures not documented
Requirements for the work environment, and the procedures to monitor and control the work environment, were not adequately documented where conditions for the work environment can have an adverse effect on product quality.
Show 1 more citation
- ISO 13485:2016 Clause 7.1
Product realization risk management processes not documented
One or more processes for risk management in product realization have not been documented.
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: DevicesCitations not published
- NAINo objectionable conditions that need follow-upDevices · Postmarket Assurance: DevicesNone
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices2 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Lack of System for Determining MDR Events
The written MDR procedure does not include an internal system which provides for a standardized review process/procedure for determining when an event meets the criteria for reporting.
Lack of or inadequate complaint procedures
Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been adequately established.
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices5 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Documented evaluation
The evaluation of potential suppliers was not documented.
Lack of or inadequate complaint procedures
Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been established.
Design changes - Lack of or Inadequate Procedures
Procedures for design change have not been established.
Not approved or obsolete document retrieval
Documents that were obsolete were observed at a location where they are being used.
Design validation - documentation
The results of design validation, including identification of the design, method(s), the date, and the individual(s) performing validation, were not documented in the design history file.
Requested under FOIA
FDA FOIA logs, last synced Nov 23, 2026, 04:25 UTC
Four requests to FDA under the Freedom of Information Act named this site, most recently on Aug 17, 2026.
Inspection ending Mar 6, 2026OAI
4 requests. In the inspection history above. FDA has posted its Form 483.
- Aug 17, 2026FDA-FOIA-2026-7932Inspection recordsOpen
- Jun 16, 2026FDA-FOIA-2026-5849Form 483· 1 of 4 sitesRecords released
- Jun 8, 2026FDA-FOIA-2026-5614Form 483Records released
- May 6, 2026FDA-FOIA-2026-4583Form 483· 1 of 5 sitesRecords released
A request is listed when it names this site’s FEI number or one of its FDA inspection IDs, or, after review, its exact street address. Requester names are left out.
Regulations FDA cited here
The 5 sections of 21 CFR in FDA’s published citations for 3 of this facility’s 5 inspections, with how many of them cited each.
- 21 CFR 820.198Complaint files2 of 3
- 21 CFR 803.17What are the requirements for developing, maintaining, and implementing written MDR procedures that apply to me?1 of 3
- 21 CFR 820.30Design controls1 of 3
- 21 CFR 820.40[Reserved]1 of 3
- 21 CFR 820.50Purchasing controls1 of 3
Warning letters and compliance actions
FDA Data Dashboard compliance actions
- Warning letterFormal notice of significant violationsCDRH · DevicesLetter on fda.govFDA record
Go further than this page
FDAgent researches FDA’s records and links every answer to its source.
About this data. Every record here is filed under FEI 3001592626 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
- Every record on this page is filed under FEI 3001592626. Records filed under another FEI or name, including other sites of the same company, are not shown here.
- FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
- Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
- “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
- Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
- FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.