May 2026 · 2026-05-01 to 2026-05-31
FDA enforcement, May 2026
In May 2026 (2026-05-01 to 2026-05-31), FDA posted 26 warning letters from its drug, biologics and device offices, and listed 14 Form 483s in its Electronic Reading Room for the first time. FDA's Enforcement Report added 326 drug and device recalls, 41 of them Class I. FDAgent recorded 11 new Official Action Indicated (OAI) classifications in FDA's inspection data. FDA published 27 firm entries on its import alert lists, 24 of them on a Red List.
Counts are FDA's records dated in May as of each source's last sync, which each section states; a late FDA update can still change them.
- Warning letters posted
- 26
- On 4 posting dates · 29 in April
- Form 483s listed
- 14
- From 14 facilities · 13 in April
- Inspections newly classified
- 275
- 11 OAI · 72 VAI · 192 NAI
- New OAI classifications
- 11
- Newly listed by FDA · 7 in April
- Recalled products
- 326
- 104 recall events · 41 Class I
- Import alert entries
- 27
- 24 Red List · 3 Green List
Ask FDAgent about this month
FDAgent reads the warning letters FDA posted this month and answers in plain English, with links to the letters.
Compared with April 2026 and May 2025
Each measure is counted the same way in every month. A year-over-year figure is shown only where FDAgent holds FDA's own dates for May 2025.
Warning letters posted26
Apr 2026: 29 · −3 (−10%)May 2025: 23 · +3 (+13%)
Form 483s listed in the reading room14
Apr 2026: 13 · +1 (+8%)May 2025: 10 · +4 (+40%)
Inspections newly classified275
Apr 2026: 319 · −44 (−14%)May 2025: —¹
New OAI classifications11
Apr 2026: 7 · +4May 2025: —¹
Recalled products (drug and device)326
Apr 2026: 451 · −125 (−28%)May 2025: 215 · +111 (+52%)
Recall events104
Apr 2026: 131 · −27 (−21%)May 2025: 86 · +18 (+21%)
Class I recall events15
Apr 2026: 12 · +3 (+25%)May 2025: 15 · 0 (0%)
Import alert firm entries27
Apr 2026: 15 · +12 (+80%)May 2025: —²
¹ FDAgent has recorded when an inspection first appears in FDA's data only since late January 2026, so months before February 2026 have no count. ² FDA deletes an import alert entry when it removes a firm, and FDAgent keeps removed entries only from November 2026, so earlier months count only the entries FDA still lists, and the year before is not compared. Sources: FDA warning letters page, last synced Nov 22, 2026, 04:45 UTC; Electronic Reading Room, last synced Nov 22, 2026, 15:42 UTC; FDA Data Dashboard inspections, last synced Nov 22, 2026, 04:55 UTC; openFDA Enforcement Report, FDA last updated Nov 11, 2026; FDA import alert pages, last synced Nov 20, 2026, 03:55 UTC.
Inspections FDA newly classified
275 inspections FDAgent first found in FDA's inspection classification data in May, by FDA's classification. FDA gives no date for a classification, so each counts in the month FDAgent first recorded it. When FDA later removes an inspection from its Data Dashboard, it stops counting, so a finished month's count can fall. They ended a median of 40 days before FDAgent first recorded them, and FDA's Data Dashboard lists Form 483 citations for 74 of them. Inspection classifications last synced Nov 22, 2026, 04:55 UTC.
| Product type | OAI | VAI | NAI | Total |
|---|---|---|---|---|
| Drugs | 9 | 44 | 36 | 89 |
| Devices | 2 | 25 | 69 | 96 |
| Biologics | 0 | 3 | 87 | 90 |
| All | 11 | 72 | 192 | 275 |
The 11 OAI inspections are listed under New OAI classifications below.
Regulations cited most
The 21 CFR sections behind the Form 483 citations FDA's Data Dashboard lists for the inspections it newly classified in May (144 citations on 53 inspections), and the sections May's warning letters cite in their numbered violations (33 violations in 10 of the 26 letters). Each section links to its page. Inspection citations last synced Nov 22, 2026, 04:55 UTC.
| On inspections | Citations |
|---|---|
| 21 CFR 211.22Responsibilities of quality control unit | 14 |
| 21 CFR 211.192Production record review | 12 |
| 21 CFR 211.84Testing and approval or rejection of components, drug product containers, and closures | 9 |
| 21 CFR 211.100Written procedures; deviations | 9 |
| 21 CFR 211.67Equipment cleaning and maintenance | 7 |
| 21 CFR 211.160General requirements | 7 |
| 21 CFR 211.68Automatic, mechanical, and electronic equipment | 6 |
| 21 CFR 211.42Design and construction features | 5 |
| In warning letters | Letters |
|---|---|
| 21 CFR 211.22Responsibilities of quality control unit | 5 |
| 21 CFR 211.84Testing and approval or rejection of components, drug product containers, and closures | 4 |
| 21 CFR 211.160General requirements | 4 |
| 21 CFR 211.100Written procedures; deviations | 3 |
| 21 CFR 211.165Testing and release for distribution | 3 |
| 21 CFR 211.166Stability testing | 3 |
| 21 CFR 211.192Production record review | 3 |
| 21 CFR 211.42Design and construction features | 2 |
Sections with the same count are ordered by section number. A section without its own page is shown without a link. 90 citations name no 21 CFR section and are not counted; device citations can name ISO 13485:2016 clauses instead. FDA can list an inspection's citations weeks after its classification, so these counts can grow.
Form 483s listed in the reading room
14 Form 483s FDA listed for the first time in its Electronic Reading Room in May, dated by the publish date FDA gives each, from 14 facilities. FDA sometimes lists a 483 days after that date, so a finished month can still gain 483s. FDA posts only some 483s, often long after the inspection: it listed these a median of 122 days after their record date. Each week's page lists every 483 with its PDF. Reading room last synced Nov 22, 2026, 15:42 UTC.
- 20268
- 20256
- 20240
- 20230
- 2022 and earlier0
The most FDA dated to one publish date was 4, on 2026-05-06.
Warning letters
26 letters FDA posted on its warning letters page in May, from its drug, biologics and device offices. 12 were issued in May and 14 earlier; FDA posted them a median of 13 days after issue. Warning letters last synced Nov 22, 2026, 04:45 UTC.
- CDER · Drugs
- 23
- CBER · Biologics
- 1
- CDRH · Devices
- 1
- Office of Compliance and Biologics Quality
- 1
Clinical Investigator/BIMO
Office of Compliance and Biologics Quality · issued 2026-03-12 · posted 2026-05-26
Alchymars ICM SM Private Limited
CGMP/Active Pharmaceutical Ingredient (API)/Adulterated
CDER · issued 2026-05-21 · posted 2026-05-26
CGMP/Finished Pharmaceuticals/Adulterated
CDER · issued 2026-05-14 · posted 2026-05-26
Hangzhou Yiqi Biotechnology Co., Ltd
CGMP/Active Pharmaceutical Ingredient (API)/Adulterated
CDER · issued 2026-04-15 · posted 2026-05-19
Harbin Jixianglong Biotech Co., Ltd.
CGMP/API/Adulterated and Misbranded Drugs
CDER · issued 2026-05-01 · posted 2026-05-19
CGMP/OTC Finished Pharmaceuticals/Adulterated
CDER · issued 2026-04-15 · posted 2026-05-19
Sourav K. Mishra, M.D. / All India Institute of Medical Sciences
In Vivo Bioavailability-Bioequivalence Studies – Clinical
CDER · issued 2026-04-29 · posted 2026-05-19
Investigational Device Exemptions (IDE)/Premarket Approval Application (PMA) Adulterated Device
CBER · issued 2026-04-22 · posted 2026-05-12
CGMP/OTC Finished Pharmaceuticals/Adulterated
CDER · issued 2026-05-04 · posted 2026-05-12
Mukesh B. Chandre, M.D., MBBS/Ashirwad Hospital & Research Centre
In Vivo Bioavailability-Bioequivalence Studies – Clinical
CDER · issued 2026-04-29 · posted 2026-05-12
CGMP/Finished Pharmaceuticals/Adulterated/Unapproved New Drug
CDER · issued 2026-05-04 · posted 2026-05-12
Active Cosmetics Manufacturing Inc.
CGMP/Finished Pharmaceuticals/Adulterated
CDER · issued 2026-04-22 · posted 2026-05-05
Show 6 more
CGMP/Finished Pharmaceuticals/Adulterated
CDER · issued 2026-04-30 · posted 2026-05-05
CGMP/OTC/Finished Pharmaceuticals/Adulterated
CDER · issued 2026-04-20 · posted 2026-05-05
An FEI links to its facility page only when the letter names exactly one FEI and FDAgent holds FDA records for it.
Recalls by class
Drug and device recalls in FDA's Enforcement Report dated in May, from openFDA, which FDA last updated Nov 11, 2026. Each openFDA record is one recalled product; a recall event can cover many products, so events are counted too.
| Class | Drug products | Device products | EventsRecall events |
|---|---|---|---|
| Class I | 2 | 39 | 15 |
| Class II | 37 | 237 | 81 |
| Class III | 9 | 2 | 8 |
| All | 48 | 278 | 104 |
The largest Class I recall event, from Medline Industries, LP, covers 11 of the 41 Class I products. The Class I products are listed below. Class II and III recalls are listed on each week's page.
Class I41 recallsShow all
- D-0553-2026openFDA record
Wisconsin Pharmacal Company
MG217 Multi-Symptom Treatment Cream & Skin Protectant (colloidal oatmeal 2%), NET WT 6 oz (170 g), Manufactured by: Pharmacal 1 Pharmacal Way, P.O. Box 198, Jackson, WI 53037.
Reason: Microbial Contamination of Non-Sterile Products: confirmed presence of Staphylococcus Aureus.
Drug · initiated 2026-04-24 · Enforcement Report 2026-05-27 · Ongoing
- Z-2156-2026openFDA record
ARROW INTERNATIONAL, LLC
Arrow Edge Hemodialysis Catheterization Product, REF: CS-15242-I, CS-15282-I, CS-15322-I
Reason: 16F dual-valved splittable sheath introducer, the subject of a supplier recall, was included in impacted hemodialysis kits & sets. Sheath introducer may not split as intended, and may result in withdrawal resistance, prolonged procedure time, pain, bleeding, hematoma, tissue inj…
Device · initiated 2026-04-10 · Enforcement Report 2026-05-20 · Ongoing
- Z-2157-2026openFDA record
ARROW INTERNATIONAL, LLC
Arrow-Clark VectorFlow Antegrade Chronic Hemodialysis Catheter: REF: ACS-15192-VFI, ACS-15232-VFI, ACS-15272-VFI, CS-15192-VFI, CS-15192-VFIE, CS-15232-VFI, CS-15232-VFIE, CS-15272-VFI, CS-15272-VFIE, CS-15312-VFI, CS-15312-VFIE, CS-15422-VFI, CS-15422-VFIE, CS-15552-VFI, CS-155…
Reason: 16F dual-valved splittable sheath introducer, the subject of a supplier recall, was included in impacted hemodialysis kits & sets. Sheath introducer may not split as intended, and may result in withdrawal resistance, prolonged procedure time, pain, bleeding, hematoma, tissue inj…
Device · initiated 2026-04-10 · Enforcement Report 2026-05-20 · Ongoing
- Z-2158-2026openFDA record
ARROW INTERNATIONAL, LLC
Cannon II Plus Hemodialysis Catheter, REF: CS-15242-VSP, CS-15282-VSP, CS-15322-VSP, CS-15362-VSP, CS-15552-VSP, CSD-15242-SP, CSD-15282-SP; Cannon II Plus Hemodialysis Catheter with Arrow Simplicity Micro-Introducer, REF: CS-15242-SPM, CS-15282-SPM, CS-15322-SPM, CS-15362-SPM
Reason: 16F dual-valved splittable sheath introducer, the subject of a supplier recall, was included in impacted hemodialysis kits & sets. Sheath introducer may not split as intended, and may result in withdrawal resistance, prolonged procedure time, pain, bleeding, hematoma, tissue inj…
Device · initiated 2026-04-10 · Enforcement Report 2026-05-20 · Ongoing
- Z-2159-2026openFDA record
ARROW INTERNATIONAL, LLC
NextStep Antegrade Chronic Hemodialysis Catheter, REF: AC-15192-SFX, AC-15232-SFX, CS-15192-SFX, CS-15232-SFX, CS-15272-SFX, CS-15312-SFX, CS-15422-SFX, CS-15502-SFX; NextStep Antegrade Hemodialysis Catheter with Arrow Simplicity Micro-Introducer, REF: CS-15192-SFXM, CS-15232-SF…
Reason: 16F dual-valved splittable sheath introducer, the subject of a supplier recall, was included in impacted hemodialysis kits & sets. Sheath introducer may not split as intended, and may result in withdrawal resistance, prolonged procedure time, pain, bleeding, hematoma, tissue inj…
Device · initiated 2026-04-10 · Enforcement Report 2026-05-20 · Ongoing
- Z-2138-2026openFDA record
Medline Industries, LP
Namic Angiographic Manifold, custom, Medline Product Number/SKU 64038200; Cardiopulmonary bypass adaptor, stopcock, manifold, or fitting.
Reason: Medline has identified the presence of particulate within the fluid path of the Manifolds.
Device · initiated 2026-03-24 · Enforcement Report 2026-05-20 · Ongoing
- Z-2139-2026openFDA record
Medline Industries, LP
Namic Angiographic Manifold, custom, Medline Product Number/SKU 64038301; Cardiopulmonary bypass adaptor, stopcock, manifold, or fitting.
Reason: Medline has identified the presence of particulate within the fluid path of the Manifolds.
Device · initiated 2026-03-24 · Enforcement Report 2026-05-20 · Ongoing
- Z-2140-2026openFDA record
Medline Industries, LP
Namic Angiographic Manifold, custom, Medline Product Number/SKU 64038402; Cardiopulmonary bypass adaptor, stopcock, manifold, or fitting.
Reason: Medline has identified the presence of particulate within the fluid path of the Manifolds.
Device · initiated 2026-03-24 · Enforcement Report 2026-05-20 · Ongoing
- Z-2141-2026openFDA record
Medline Industries, LP
Namic Preceptor Manifold, custom, Medline Product Number/SKU 64037107; Extravascular blood pressure transducer
Reason: Medline has identified the presence of particulate within the fluid path of the Manifolds.
Device · initiated 2026-03-24 · Enforcement Report 2026-05-20 · Ongoing
- Z-2142-2026openFDA record
Medline Industries, LP
Namic Preceptor Manifold, custom, Medline Product Number/SKU 64039307; Extravascular blood pressure transducer
Reason: Medline has identified the presence of particulate within the fluid path of the Manifolds.
Device · initiated 2026-03-24 · Enforcement Report 2026-05-20 · Ongoing
- Z-2143-2026openFDA record
Medline Industries, LP
Medline medical convenience kits, containing Namic Manifold, Labeled as: 1. ANGIOGRAPHY CATH LAB, Medline Kit Number/SKUDYNJ23532C; 2. CATH LAB PACK, Medline Kit Number/SKUDYNJ65185; 3. ANGIO KIT, Medline Kit Number/SKUVASC1057.
Reason: Medline has identified the presence of particulate within the fluid path of the Manifolds.
Device · initiated 2026-03-24 · Enforcement Report 2026-05-20 · Ongoing
- Z-2144-2026openFDA record
Medline Industries, LP
1. DBD-MTS,3V,MAN,KIT, Medline Kit Number/SKU 60010195; 2. MTS-LEFT HEART,KIT, Medline Kit Number/SKU 60033123; 3. MTS,3 VALVE,KIT, Medline Kit Number/SKU 60050183; 4. MTS,LEFT HEART,KIT, Medline Kit Number/SKU 60080349; 5. MTS,3 VALVE,LEFT HEART,KIT, Medline Kit Number/SKU 6010…
Reason: Medline has identified the presence of particulate within the fluid path of the Manifolds.
Device · initiated 2026-03-24 · Enforcement Report 2026-05-20 · Ongoing
- Z-2145-2026openFDA record
Medline Industries, LP
1. 3 VALVE MAN KIT, Medline Kit SKU Number: 65101544; 2. 3V MANIFOLD KIT, Medline Kit SKU Number: 65110351; 3. HEART CATH PACK-LF, Medline Kit SKU Number: DYNJ0373285M; 4. ANGIO DRAPE CARDIAC PACK, Medline Kit SKU Number: DYNJ51615P; 5. ANGIO DRAPE CARDIAC PACK, Medline Kit SKU…
Reason: Medline has identified the presence of particulate within the fluid path of the Manifolds.
Device · initiated 2026-03-24 · Enforcement Report 2026-05-20 · Ongoing
- Z-2146-2026openFDA record
Medline Industries, LP
Medline medical convenience kits, containing Namic Manifold, Labeled as: HEART CATH PACK-LF, Medline Kit Number/SKU DYNJ36478B
Reason: Medline has identified the presence of particulate within the fluid path of the Manifolds.
Device · initiated 2026-03-24 · Enforcement Report 2026-05-20 · Ongoing
- Z-2147-2026openFDA record
Medline Industries, LP
Medline medical convenience kits, containing Namic Manifold, Labeled as: 1. MTS,RT,HANDED/LFT,HRT,KIT, Medline Kit Number/SKU 60010508; 2. DBD-MTS,LEFT HEART,KIT, Medline Kit Number/SKU 60160075; 3. CARDIAC CATH PACK, Medline Kit Number/SKU DYNJ04636N; 4. CATH LAB TRAY, Medline…
Reason: Medline has identified the presence of particulate within the fluid path of the Manifolds.
Device · initiated 2026-03-24 · Enforcement Report 2026-05-20 · Ongoing
- Z-2148-2026openFDA record
Medline Industries, LP
Medline medical convenience kits, containing Namic Manifold, Labeled as: TAVI, Medline Kit Number/SKU DYNJ904981C
Reason: Medline has identified the presence of particulate within the fluid path of the Manifolds.
Device · initiated 2026-03-24 · Enforcement Report 2026-05-20 · Ongoing
- Z-2047-2026openFDA record
Omnicell, Inc.
Rolls of label stock used with the Omnicell i.v.Station. 1. Part Number 258920028, LABEL, STERILE, IVSTATION SYRINGE (40MMX40MM). 2. Part Number 258900029, LABEL, STERILE, IVSTATION SYRINGE (20MMX40MM). Model/Catalog Number: IVS-RBT-001. i.v.STATION is an automation device used…
Reason: Potential for mislabeled syringe produced by the i.v.STATION device.
Device · initiated 2026-03-25 · Enforcement Report 2026-05-20 · Ongoing
- Z-2066-2026openFDA record
Windstone Medical Packaging, Inc.
Custom Convenience Kit containing recalled Medline Namic Angiographic Control Syringes with Rotating Adapter (Namic RA Syringes) Angio Pack Part numbers AMS6908E, AMS6908F
Reason: Affected kits contain Medline Namic RA Syringes, which are recalled by their manufacturer due to risk of syringe rotating adapter unwinding during use, which may result in a loose connection or full disconnection between syringe and manifold. If unwinding occurs, there is a pote…
Device · initiated 2026-04-02 · Enforcement Report 2026-05-20 · Ongoing
- Z-1981-2026openFDA record
American Contract Systems Inc
Convenience kits containing Medline Namic Angiographic Control Syringes with Rotating Adaptor CORONARY ANGIO PACK Model numbers ANCA80AP ANCA80AQ
Reason: Impacted kits contain syringes recalled by Medline Industries. Syringe rotating adaptor may unwind during use, which may result in a loose connection or full disconnection between the syringe and manifold. Potential risk for biohazard exposure, blood loss, infection, or air embo…
Device · initiated 2026-03-19 · Enforcement Report 2026-05-13 · Ongoing
- Z-1970-2026openFDA record
AVID Medical, Inc.
Halyard CARDIAC CATH PACK kit. Model Number: EAMC1000-05, WAFB208-02.
Reason: Potential risk for Medline syringe rotating adaptor unwinding during use, which may result in a loose connection and/or full disconnection between the syringe and manifold.
Device · initiated 2026-03-13 · Enforcement Report 2026-05-13 · Ongoing
- Z-1971-2026openFDA record
AVID Medical, Inc.
Halyard CARDIAC CATH PACK SAN23CARDQ kit. Model Number: DRCC36.
Reason: Potential risk for Medline syringe rotating adaptor unwinding during use, which may result in a loose connection and/or full disconnection between the syringe and manifold.
Device · initiated 2026-03-13 · Enforcement Report 2026-05-13 · Ongoing
- Z-1972-2026openFDA record
AVID Medical, Inc.
Halyard CARDIAC CATH TRAY SHANDS JAX kit. Model Number: SHND401-04, SHND401-05.
Reason: Potential risk for Medline syringe rotating adaptor unwinding during use, which may result in a loose connection and/or full disconnection between the syringe and manifold.
Device · initiated 2026-03-13 · Enforcement Report 2026-05-13 · Ongoing
- Z-1973-2026openFDA record
AVID Medical, Inc.
Halyard CATH LAB kit. Model Number: SACL75AM.
Reason: Potential risk for Medline syringe rotating adaptor unwinding during use, which may result in a loose connection and/or full disconnection between the syringe and manifold.
Device · initiated 2026-03-13 · Enforcement Report 2026-05-13 · Ongoing
- Z-1974-2026openFDA record
AVID Medical, Inc.
Halyard CATH LAB kit. Model Number: SACL75-01.
Reason: Potential risk for Medline syringe rotating adaptor unwinding during use, which may result in a loose connection and/or full disconnection between the syringe and manifold.
Device · initiated 2026-03-13 · Enforcement Report 2026-05-13 · Ongoing
- Z-1975-2026openFDA record
AVID Medical, Inc.
Halyard HEART CATH, SELF REGIONAL kit. Model Number: SELF131-05.
Reason: Potential risk for Medline syringe rotating adaptor unwinding during use, which may result in a loose connection and/or full disconnection between the syringe and manifold.
Device · initiated 2026-03-13 · Enforcement Report 2026-05-13 · Ongoing
- Z-1976-2026openFDA record
AVID Medical, Inc.
Halyard NEURO IR PERIPHERAL (PS 64405) kit. Model Number: UINR44-01.
Reason: Potential risk for Medline syringe rotating adaptor unwinding during use, which may result in a loose connection and/or full disconnection between the syringe and manifold.
Device · initiated 2026-03-13 · Enforcement Report 2026-05-13 · Ongoing
- Z-1977-2026openFDA record
AVID Medical, Inc.
Halyard PERC TRAY kit. Model Numbers: SLPC34-01, SLPC34-02.
Reason: Potential risk for Medline syringe rotating adaptor unwinding during use, which may result in a loose connection and/or full disconnection between the syringe and manifold.
Device · initiated 2026-03-13 · Enforcement Report 2026-05-13 · Ongoing
- Z-1978-2026openFDA record
AVID Medical, Inc.
Halyard SAMMC ANGIOGRAPHY kit. Model Numbers: SAMM066-15.
Reason: Potential risk for Medline syringe rotating adaptor unwinding during use, which may result in a loose connection and/or full disconnection between the syringe and manifold.
Device · initiated 2026-03-13 · Enforcement Report 2026-05-13 · Ongoing
- Z-1979-2026openFDA record
AVID Medical, Inc.
Halyard VA DETROIT CATH LAB PACK kit. Model Numbers: VMCD029.
Reason: Potential risk for Medline syringe rotating adaptor unwinding during use, which may result in a loose connection and/or full disconnection between the syringe and manifold.
Device · initiated 2026-03-13 · Enforcement Report 2026-05-13 · Ongoing
- D-0540-2026openFDA record
B BRAUN MEDICAL INC
Lactated Ringer's Injection USP, 1000 mL container, Rx only, B. Braun Medical, Inc., Bethlehem, PA 18018-3524 USA, NDC 0264-7750-07.
Reason: Presence of Particulate Matter.
Drug · initiated 2026-04-28 · Enforcement Report 2026-05-13 · Ongoing
- Z-1934-2026openFDA record
Cook Incorporated
Centimeter Sizing Catheter RPN N5.0-35-100-P-10S-PIG-CSC-20 GPN G11916 Catheters for use in angiographic procedures are available in a variety of French sizes, endhole sizes, lengths, materials and designs (e.g., polyethylene or nylon, non-braided or braided with 1:1 torque). Th…
Reason: Marker bands may be at an increased risk of cracking/breakage. Potential adverse events that may result include increased procedural time, harms associated with device fragmentation/separation, and vessel injury. In a worst-case scenario, device fragmentation and separation may…
Device · initiated 2026-04-02 · Enforcement Report 2026-05-13 · Ongoing
- Z-1935-2026openFDA record
Cook Incorporated
Aurous Centimeter Sizing Catheter RPN N5.0-35-100-P-10S-PIG-CSC-20-01 GPN G47308 Sizing catheters are single-use, sterile catheters with gold marker bands. These catheters are for use in angiographic procedures and available in a variety of lengths (70 and 100 cm) and tip config…
Reason: Marker bands may be at an increased risk of cracking/breakage. Potential adverse events that may result include increased procedural time, harms associated with device fragmentation/separation, and vessel injury. In a worst-case scenario, device fragmentation and separation may…
Device · initiated 2026-04-02 · Enforcement Report 2026-05-13 · Ongoing
- Z-1936-2026openFDA record
Cook Incorporated
Beacon Tip Centimeter Sizing Catheter RPN/GPN NR5.0-35-70-P-10S-0-CSC-20 G31213 NR5.0-35-100-P-10S-0-CSC-20 G31214 NR5.0-35-100-P-10S-PIG-CSC-20 G31216 NR5.0-35-70-P-10S-VCF-CSC-20 G31220 NR5.0-35-100-P-10S-VCF-CSC-20 G31221 Sizing catheters are single-use, sterile catheters wit…
Reason: Marker bands may be at an increased risk of cracking/breakage. Potential adverse events that may result include increased procedural time, harms associated with device fragmentation/separation, and vessel injury. In a worst-case scenario, device fragmentation and separation may…
Device · initiated 2026-04-02 · Enforcement Report 2026-05-13 · Ongoing
- Z-1855-2026openFDA record
Intuitive Surgical, Inc.
Intuitive 8 mm SureForm 30 Gray Reloads Reference Numbers: 48230M-05 and 48230M-06
Reason: Reloads for curved-tipped stapler may potentially produce an incomplete staple line.
Device · initiated 2026-03-11 · Enforcement Report 2026-05-13 · Ongoing
- Z-2007-2026openFDA record
Medical Action Industries, Inc. 306
Medical Action Industries Inc. Pack Cath BHH, REF: BHCA49K, Sterile EO, Rx Only, QTY: 5/Case
Reason: Convenience kit contains a syringe in which the rotating adapter may unwind during use, resulting in a loose connection and/or full disconnection between the syringe and manifold"
Device · initiated 2026-03-17 · Enforcement Report 2026-05-13 · Ongoing
- Z-1942-2026openFDA record
Ventec Life Systems, Inc.
VOCSN V+Pro package, REF: PRT-01198-000; VOCSN V+Pro unit, REF: PRT-01185-000
Reason: Respiratory system intended to provide continuous/intermittent ventilator support, may not have been evaluated under all intended high-pressure conditions during production testing, so an oxygen leak condition could occur that may lead to reduced below intended levels of fractio…
Device · initiated 2026-03-25 · Enforcement Report 2026-05-13 · Ongoing
- Z-1770-2026openFDA record
Boston Scientific Corporation
Boston Scientific Pacemaker, Implantable Pulse Generator, Pacemaker (non- Cardiac Re-Synchronization Therapy) labeled as: 1. ESSENTIO SR SL, Model Number L100; 2. ESSENTIO DR SL Pacemaker, Model Number L101; 3. ESSENTIO SR SL MRI Pacemaker, Model Number L110; 4. ESSENTIO DR SL M…
Reason: Updated software, Brady software maintenance release 6 (SMR6), is now available for the ACCOLADE" family of pacemakers and cardiac resynchronization therapy pacemakers (CRT-Ps)1 and the advisory population is expanding to include all CRT-P and dual-chamber extended life (DR-EL)…
Device · initiated 2026-03-19 · Enforcement Report 2026-05-06 · Ongoing
- Z-1771-2026openFDA record
Boston Scientific Corporation
Boston Scientific Cardiac Resynchronization Therapy (CRT)labeled as: 1. VALITUDE CRT-P EL, Model Number U125; 2. VALITUDE CRT-P EL MRI, Model Number U128; 3. VISIONIST CRT-P EL, Model Number U225; 4. VISIONIST CRT-P EL, Model Number U226; 5. VISIONIST CRT-P EL MRI, Model Number…
Reason: Updated software, Brady software maintenance release 6 (SMR6), is now available for the ACCOLADE" family of pacemakers and cardiac resynchronization therapy pacemakers (CRT-Ps)1 and the advisory population is expanding to include all CRT-P and dual-chamber extended life (DR-EL)…
Device · initiated 2026-03-19 · Enforcement Report 2026-05-06 · Ongoing
- Z-1947-2026openFDA record
North American Rescue LLC.
AIDBAGs are first aid kits of convenience composed of individually labeled devices designed as specialty first aid kits designed for the appropriate level provider to evaluate and treat patients at the point-of-injury / point-of-illness. The original manufacturer labeling, inclu…
Reason: Kits contain the McKesson TRUE Metrix PRO Professional Monitoring Blood Glucose System that has been recalled for an issue with the software design of the E-5 Error Code where the meter displays an E-5 error code for a very high blood glucose event (> 600 mg/dL) or when there is…
Device · initiated 2026-03-27 · Enforcement Report 2026-05-06 · Ongoing
- Z-1826-2026openFDA record
WHILL, INC.
WHILL Model C2, with software versions before HMI: 2.22 / MC: 1.13; device is an indoor/outdoor battery-operated 2-wheel drive (rear-wheel drive) powered wheelchair.
Reason: Cyber vulnerability of BLE communication was found by CISA VDP program.
Device · initiated 2025-12-29 · Enforcement Report 2026-05-06 · Ongoing
- Z-1827-2026openFDA record
WHILL, INC.
WHILL Model F, with software versions before HMI: 2.23 / MC: 1.05; the device is an indoor/outdoor, foldable battery-operated 2-wheel drive (rear-wheel drive) powered wheelchair.
Reason: Cyber vulnerability of BLE communication was found by CISA VDP program.
Device · initiated 2025-12-29 · Enforcement Report 2026-05-06 · Ongoing
Import alert entries
27 firm entries FDA published on its import alert pages this month, dated by each entry's date published: 24 on a Red List, 3 on a Green List. FDA lists firms by name and address; FDAgent links an entry to a facility only when FDA printed the firm's FEI. 7 entries whose name may be a person's rather than a firm's are not listed or counted here; signed-in users can see them in FDAgent. Import alerts last synced Nov 20, 2026, 03:55 UTC.
57-22Red List7 entriesAlert on fda.govShow all“Detention Without Physical Examination of Unlicensed Biological Products Promoted in the U.S.”Alert on fda.gov
- Published 2026-05-29
Green Cross Wellbeing Co., Ltd
KOREA (THE REPUBLIC OF)
- Published 2026-05-29
GREEN CROSS WellBeing Corporation
KOREA (THE REPUBLIC OF)
- Published 2026-05-29
JAPAN BIO PRODUCTS CO LTD
KOREA (THE REPUBLIC OF)
- Published 2026-05-29
JAPAN BIO PRODUCTS CO LTD
JAPAN
- Published 2026-05-29
JAPAN BIO PRODUCTS CO., LTD.
JAPAN
- Published 2026-05-29
Japan Bio Products Co.,Ltd
JAPAN
- Published 2026-05-29
JBP GLOBAL USA
UNITED STATES
66-40Red List1 entryAlert on fda.govShow allDetention Without Physical Examination of Drugs From Firms Which Have Not Met Drug GMPsAlert on fda.gov
- Published 2026-05-19
Alchymars ICM SM Private Limited
INDIA
66-41Red List2 entriesAlert on fda.govShow allDetention Without Physical Examination of Unapproved New Drugs Promoted In The U.S.Alert on fda.gov
- Published 2026-05-19
Sarvotham Care Ltd
INDIA
- Published 2026-05-11
MOCHA ESSENCE EXPORTS LOCAL PRODUCTS
YEMEN
66-79Red List1 entryAlert on fda.govShow allDetention Without Physical Examination of Drugs From Foreign Establishments Refusing FDA InspectionAlert on fda.gov
- Published 2026-05-13
Iron Lab, S.A. de C.V.
MEXICO
68-19Red List2 entriesAlert on fda.govShow allDETENTION WITHOUT PHYSICAL EXAMINATION OF UNAPPROVED FINISHED NEW ANIMAL DRUGSAlert on fda.gov
- Published 2026-05-19
SHANDONG NATURAL MICRON PHARM TE
CHINA
- Published 2026-05-19
Shandong Natural Micron Pharm Tech Co.,Ltd
CHINA
76-01Green List3 entriesAlert on fda.govShow allDetention Without Physical Examination Of Medical Instruments from PakistanAlert on fda.gov
- Published 2026-05-22
AAYMED INSTRUMENTS
PAKISTAN
- Published 2026-05-15
Quality Concept (Pvt) Ltd.
PAKISTAN
- Published 2026-05-07
Healermed Instruments
PAKISTAN
89-08Red List4 entriesAlert on fda.govShow allDetention Without Physical Examination of Devices without Approved PMA's or IDE's and Other Devices Not Substantially Equivalent or Without a 510(k)Alert on fda.gov
- Published 2026-05-19
Spectrum Medical Ltd.
UNITED KINGDOM
- Published 2026-05-19
SPECTRUM MEDICAL SRL
ITALY
- Published 2026-05-08
Jiangsu Caina Medical Co., Ltd
CHINA
- Published 2026-05-07
BESTREAT SAFETY & FIRST AID SOLUTION CO., LTD
CHINA
95-04Red List1 entryAlert on fda.govShow allDetention Without Physical Examination Of Laser Pointers, Laser Gunsights, Laser Levels, Laser Light Shows, Laser Pointer Key Chains, & Similar Products That Fail To Comply With Applicable Performance Standards And Reporting Requirements.Alert on fda.gov
- Published 2026-05-15
KEAI ELECTRONIC TECHNOLOGY (GUANGZHOU) CO.,LTD
CHINA
95-05Red List4 entriesAlert on fda.govShow allDetention Without Physical Examination of Electronic Products that Fail to Comply with Performance Standards or to Have Required CertificationAlert on fda.gov
- Published 2026-05-19
JIANGSU BOFA ROBOT INTELLIGENT EQUIPMENT CO., LTD.
CHINA
- Published 2026-05-18
SHANDONG SANQIANG CNC EQUIPMENT CO., LTD.
CHINA
- Published 2026-05-14
DENER MAKINA SAN VE TIC AS
TURKEY
- Published 2026-05-14
DENER MAKINA SAN.TIC.LTD.STI.
TURKEY
99-34Red List2 entriesAlert on fda.govShow allDETENTION WITHOUT PHYSICAL EXAMINATION OF DRUGS OR MEDICAL DEVICES FROM FIRMS WITHOUT A VALID DRUG OR MEDICAL DEVICE REGISTRATION AND/OR LISTINGAlert on fda.gov
- Published 2026-05-05
NANJING CLEAN ENERGY ELECTRONIC TECHNOLOGY CO.,LTD.
CHINA
- Published 2026-05-05
NINGBO YIXIN INTERNATIONAL TRADING CO.,LTD
CHINA
New OAI classifications
11 inspections classified Official Action Indicated that FDAgent first recorded from FDA's inspection classification data this month, after its reads of that data in the 8 days before had not found them. Older inspections FDAgent adds to its archive are left out. Dates are inspection end dates. Inspection classifications last synced Nov 22, 2026, 04:55 UTC.
- Inspection ended 2026-03-19FEI 3009446038
Nanovis LLCOAI
Columbia City, Indiana, United States · Devices · Project Evaluation: Devices · recorded 2026-05-26
- Inspection ended 2026-02-27FEI 3004135088
Tan-Alize Kozmetik ve Temizlik Urunleri Sanayi ve Ticaret A.S.OAI
Cekmekoy, Turkey · Drugs · Drug Quality Assurance · recorded 2026-05-26
- Inspection ended 2026-02-27FEI 3009510975
Reliance Life Sciences Private LimitedOAI
Navi Mumbai, India · Drugs · Drug Quality Assurance · recorded 2026-05-26
- Inspection ended 2026-03-05FEI 3030456591
SURETEC INNOVATIONS LLCOAI
Gilbert, Arizona, United States · Drugs · Drug Quality Assurance · recorded 2026-05-26
- Inspection ended 2026-03-06FEI 1713747
Fresenius USA Manufacturing, Inc.OAI
Ogden, Utah, United States · Drugs · Drug Quality Assurance · recorded 2026-05-26
- Inspection ended 2026-02-23FEI 3008493151
SurgiSmoke SolutionsOAI
Saint Paul, Minnesota, United States · Devices · Compliance: Devices · recorded 2026-05-19
- Inspection ended 2026-02-26FEI 3011034132
Smith & Nephew Wound ManagementOAI
Fort Worth, Texas, United States · Drugs · Drug Quality Assurance · recorded 2026-05-19
- Inspection ended 2026-02-12FEI 1035794
Tropical Seas, Inc.OAI
Holly Hill, Florida, United States · Drugs · Drug Quality Assurance · recorded 2026-05-12
- Inspection ended 2026-02-13FEI 1519202
Woodbine Products Company Inc.OAI
Medina, Ohio, United States · Drugs · Drug Quality Assurance · recorded 2026-05-12
- Inspection ended 2026-03-05FEI 3012144557
Sharp Sterile Manufacturing, LLCOAI
Lee, Massachusetts, United States · Drugs · Drug Quality Assurance · recorded 2026-05-12
- Inspection ended 2026-02-20FEI 3017473850
Staska Pharmaceuticals, IncOAI
Bennet, Nebraska, United States · Drugs · Drug Quality Assurance · recorded 2026-05-09
Week by week
Each week's page lists every Form 483 and recall. Weeks run Monday to Sunday, so the first and last can include days of the neighbouring months.
| Week | Dates | Letters | Form 483s | Recalls | New OAI |
|---|---|---|---|---|---|
| 2026-W22 | 2026-05-25 to 2026-05-31 | 6 letters | 5 Form 483s | 36 recalls | 5 new OAI |
| 2026-W21 | 2026-05-18 to 2026-05-24 | 8 letters | 0 Form 483s | 73 recalls | 2 new OAI |
| 2026-W20 | 2026-05-11 to 2026-05-17 | 5 letters | 1 Form 483 | 107 recalls | 3 new OAI |
| 2026-W19 | 2026-05-04 to 2026-05-10 | 7 letters | 7 Form 483s | 110 recalls | 1 new OAI |
Get each week’s page by email every Monday
A free weekly email with each week’s warning letters, new Form 483s, OAI classifications and Class I recalls, counted the same way as this page. Just your email address.
How this month is counted
Months are calendar months in UTC. Counts are read from FDAgent's copy of each FDA source, which refreshes daily, and FDAgent checks them against FDA's own sources every week. FDA posts some records late and removes others, so a finished month's counts can still change. FDAgent is independent of FDA.
- Warning letters posted
- Letters from FDA's drug, biologics and device offices, dated by the day FDA posted them, not the day it issued them.
- Form 483s listed
- Records FDA's Electronic Reading Room types as 483 or Amended 483, dated by FDA's publish date. A firm's 483 response is not counted. Each inspection's 483 counts once, when FDA first listed it: a later listing with the same FEI and record date is shown apart, as a re-post when FDA listed a 483 again under a new media id, or as an amended 483 when FDA types it Amended 483. FDA sometimes lists a 483 days after the publish date it gives it, so a finished month can still gain 483s.
- Inspections newly classified
- OAI, VAI and NAI inspections counted on the day FDAgent first found them in FDA's classification data, and only when its complete reads in the 8 days before had covered the inspection's end date without finding them. Inspections recorded before FDAgent kept a record of its reads, in November 2026, count when they ended in the 13 weeks before. Backfills and wider reads never count, nor do the Veterinary inspections of FDAgent's first read of them, in November 2026. FDA gives no date for a classification. When FDA removes an inspection from its Data Dashboard, it stops counting, so a finished month's count can fall.
- Regulations cited
- The FDA Data Dashboard's Form 483 citations on those inspections, by 21 CFR section. FDA often lists an inspection's citations weeks after its classification, so these counts grow after the month ends. For letters, the violation statements that cite a 21 CFR section, each section once per letter.
- Recalls
- openFDA's drug and device records by Enforcement Report date. Each record is one product; an event is FDA's recall event number.
- Import alert entries
- Firm entries by the date FDA published them on an import alert. Entries that may name a person are left out. FDA deletes an entry when it removes the firm; entries FDAgent recorded from November 2026 on keep counting in the month FDA dated them.
- Months in progress
- A month in progress counts the days so far and grows until it ends. Recalls for its latest days appear when openFDA updates.
- Comparisons
- Every month is counted the same way. A year-over-year figure appears only where FDAgent holds FDA's own dates for the earlier month. A percent change appears only when the earlier count is 10 or more. A month in progress shows no changes until it ends. A month that overlaps a federal government shutdown (Oct 1 to Nov 12, 2025) can have low counts, and month pages say so beside any comparison with one.
Cite this page
Quote a count with its date: the page refreshes daily, and a month's counts can change as FDA posts late records.
FDAgent. "FDA enforcement, May 2026." https://www.fdagent.app/enforcement/2026-05. Counts as of Nov 22, 2026.
About this data
- Weeks are ISO weeks, Monday to Sunday, in UTC. Each section is dated by its own FDA date: posting date for warning letters, reading-room publish date for Form 483s (each inspection's 483 counts once, when FDA first listed it; FDA's re-posts and amended 483s with the FEI and record date of one it listed before are shown apart), Enforcement Report date for recalls, and the date published FDA prints beside each firm on an import alert. FDA publishes no date for an inspection classification, so an OAI classification counts as new on the day FDAgent first recorded the inspection, and only when FDAgent's complete reads of FDA's inspection data in the 8 days before had covered the inspection's end date without finding it. Older inspections FDAgent adds to its archive, by a backfill or a wider read, never count as new, however late FDA posted them. Inspections FDAgent recorded before it began keeping a record of its reads, in November 2026, count only when they ended in the 13 weeks before, and the Veterinary inspections of FDAgent's first read of them, that same month, never count.
- Import alert entries come from FDA's import alert pages, where FDA lists firms by name and address, not by FEI. FDAgent links an entry to a facility only when FDA printed the firm's FEI, never by name. Earlier weeks list only entries still on FDA's pages, because FDA deletes an entry when it removes a firm, and FDAgent keeps removed entries only from the day it began recording them. FDAgent lists an entry here only when its name shows it is a firm, through a legal form such as Ltd, Inc or GmbH, a business word such as Laboratories or Trading, or a web address, or when FDA's inspection or drug registration records hold a firm under exactly that name; entries that may name a person are shown only to signed-in users.
- Import refusals are not listed by week: FDA reports a refusal weeks after it happens, so a week's count would keep changing after the week closed. Each facility page lists its own refusals.
- Warning letters are those issued by FDA's drug, biologics and device centers and their field and quality offices. Letters from FDA's food, tobacco, veterinary, cosmetics and dietary supplement offices are left out.
- FDA posts only some Form 483s to its Electronic Reading Room, so a missing 483 does not mean none was issued.
- FDA publishes an inspection's classification once it is final and updates its inspection data weekly. Source: FDA inspections FAQ.
- Recalls come from openFDA, which republishes FDA's weekly Enforcement Report; a report's recalls appear here once openFDA publishes it.
- FDAgent is independent of FDA. Every row links to its FDA source; check it before relying on a record.