July 2026 · 2026-07-01 to 2026-07-31
FDA enforcement, July 2026
In July 2026 (2026-07-01 to 2026-07-31), FDA posted 26 warning letters from its drug, biologics and device offices, and listed 138 Form 483s in its Electronic Reading Room for the first time. FDA's Enforcement Report added 381 drug and device recalls, 53 of them Class I. FDAgent recorded 6 new Official Action Indicated (OAI) classifications in FDA's inspection data. FDA published 31 firm entries on its import alert lists, 27 of them on a Red List.
Counts are FDA's records dated in July as of each source's last sync, which each section states; a late FDA update can still change them.
- Warning letters posted
- 26
- On 4 posting dates · 69 in June
- Form 483s listed
- 138
- From 138 facilities · 1 in June
- Inspections newly classified
- 285
- 6 OAI · 91 VAI · 188 NAI
- New OAI classifications
- 6
- Newly listed by FDA · 5 in June
- Recalled products
- 381
- 151 recall events · 53 Class I
- Import alert entries
- 31
- 27 Red List · 4 Green List
Ask FDAgent about this month
FDAgent reads the warning letters FDA posted this month and answers in plain English, with links to the letters.
Compared with June 2026 and July 2025
Each measure is counted the same way in every month. A year-over-year figure is shown only where FDAgent holds FDA's own dates for July 2025.
Warning letters posted26
Jun 2026: 69 · −43 (−62%)Jul 2025: 33 · −7 (−21%)
Form 483s listed in the reading room138
Jun 2026: 1 · +137Jul 2025: 4 · +134
Inspections newly classified285
Jun 2026: 293 · −8 (−3%)Jul 2025: —¹
New OAI classifications6
Jun 2026: 5 · +1Jul 2025: —¹
Recalled products (drug and device)381
Jun 2026: 366 · +15 (+4%)Jul 2025: 245 · +136 (+56%)
Recall events151
Jun 2026: 116 · +35 (+30%)Jul 2025: 107 · +44 (+41%)
Class I recall events16
Jun 2026: 17 · −1 (−6%)Jul 2025: 12 · +4 (+33%)
Import alert firm entries31
Jun 2026: 22 · +9 (+41%)Jul 2025: —²
¹ FDAgent has recorded when an inspection first appears in FDA's data only since late January 2026, so months before February 2026 have no count. ² FDA deletes an import alert entry when it removes a firm, and FDAgent keeps removed entries only from November 2026, so earlier months count only the entries FDA still lists, and the year before is not compared. Sources: FDA warning letters page, last synced Nov 22, 2026, 04:45 UTC; Electronic Reading Room, last synced Nov 22, 2026, 15:42 UTC; FDA Data Dashboard inspections, last synced Nov 22, 2026, 04:55 UTC; openFDA Enforcement Report, FDA last updated Nov 11, 2026; FDA import alert pages, last synced Nov 20, 2026, 03:55 UTC.
Inspections FDA newly classified
285 inspections FDAgent first found in FDA's inspection classification data in July, by FDA's classification. FDA gives no date for a classification, so each counts in the month FDAgent first recorded it. When FDA later removes an inspection from its Data Dashboard, it stops counting, so a finished month's count can fall. They ended a median of 47 days before FDAgent first recorded them, and FDA's Data Dashboard lists Form 483 citations for 62 of them. Inspection classifications last synced Nov 22, 2026, 04:55 UTC.
| Product type | OAI | VAI | NAI | Total |
|---|---|---|---|---|
| Drugs | 6 | 60 | 56 | 122 |
| Devices | 0 | 22 | 47 | 69 |
| Biologics | 0 | 9 | 85 | 94 |
| All | 6 | 91 | 188 | 285 |
The 6 OAI inspections are listed under New OAI classifications below.
Regulations cited most
The 21 CFR sections behind the Form 483 citations FDA's Data Dashboard lists for the inspections it newly classified in July (106 citations on 37 inspections), and the sections July's warning letters cite in their numbered violations (28 violations in 7 of the 26 letters). Each section links to its page. Inspection citations last synced Nov 22, 2026, 04:55 UTC.
| On inspections | Citations |
|---|---|
| 21 CFR 211.22Responsibilities of quality control unit | 10 |
| 21 CFR 211.100Written procedures; deviations | 9 |
| 21 CFR 211.160General requirements | 9 |
| 21 CFR 211.42Design and construction features | 8 |
| 21 CFR 211.67Equipment cleaning and maintenance | 8 |
| 21 CFR 211.113Control of microbiological contamination | 6 |
| 21 CFR 211.192Production record review | 5 |
| 21 CFR 211.68Automatic, mechanical, and electronic equipment | 4 |
| In warning letters | Letters |
|---|---|
| 21 CFR 211.100Written procedures; deviations | 3 |
| 21 CFR 211.192Production record review | 3 |
| 21 CFR 211.22Responsibilities of quality control unit | 2 |
| 21 CFR 211.160General requirements | 2 |
| 21 CFR 820.30Design controls (removed Feb 4, 2026) | 2 |
| 21 CFR 820.100Corrective and preventive action (removed Feb 4, 2026) | 2 |
| 21 CFR 820.198Complaint files (removed Feb 4, 2026) | 2 |
| 21 CFR 201.5Drugs; adequate directions for use | 1 |
Sections with the same count are ordered by section number. A section without its own page is shown without a link. 90 citations name no 21 CFR section and are not counted; device citations can name ISO 13485:2016 clauses instead. FDA can list an inspection's citations weeks after its classification, so these counts can grow.
Form 483s listed in the reading room
138 Form 483s FDA listed for the first time in its Electronic Reading Room in July, dated by the publish date FDA gives each, from 138 facilities. FDA sometimes lists a 483 days after that date, so a finished month can still gain 483s. FDA also re-posted 2 Form 483s it had listed before. Re-posts don't count as new 483s. FDA posts only some 483s, often long after the inspection: it listed these a median of 156 days after their record date. Each week's page lists every 483 with its PDF. Reading room last synced Nov 22, 2026, 15:42 UTC.
- 202682
- 202539
- 20243
- 20237
- 2022 and earlier7
The most FDA dated to one publish date was 19, on 2026-07-10.
Warning letters
26 letters FDA posted on its warning letters page in July, from its drug, biologics and device offices. 7 were issued in July and 19 earlier; FDA posted them a median of 14 days after issue. Warning letters last synced Nov 22, 2026, 04:45 UTC.
- CDER · Drugs
- 21
- CDRH · Devices
- 5
CGMP/QSR/Drug/Medical Devices/Adulterated
CDRH · issued 2026-05-27 · posted 2026-07-22
Almon Healthcare Private Limited
CGMP/Active Pharmaceutical Ingredient (API)/Adulterated
CDER · issued 2026-07-13 · posted 2026-07-21
CGMP/Finished Pharmaceuticals/Adulterated/Misbranded
CDER · issued 2026-07-13 · posted 2026-07-21
Island Kinetics, Inc. d.b.a. CoValence Laboratories
CGMP/Finished Pharmaceuticals/Unapproved New Drug/Misbranded/Adulterated
CDER · issued 2026-07-16 · posted 2026-07-21
Pratik Doshi, M.D. / Alembic Pharmaceuticals Ltd.
In Vivo Bioavailability-Bioequivalence Studies – Clinical
CDER · issued 2026-07-10 · posted 2026-07-21
CGMP/Active Pharmaceutical Ingredient (API)/Adulterated
CDER · issued 2026-07-13 · posted 2026-07-21
International Medication Systems Limited
CGMP/Finished Pharmaceuticals/Adulterated
CDER · issued 2026-07-02 · posted 2026-07-14
Nihon Kohden Digital Health Solutions, LLC
Medical Device/Adulterated/Misbranded/Lacks PMA and/or 510(k)
CDRH · issued 2026-06-12 · posted 2026-07-14
CGMP/Finished Pharmaceuticals/Adulterated
CDER · issued 2026-07-08 · posted 2026-07-14
All Ketamine HCL / www.allketaminehcl.com
Unapproved New Drugs/Misbranded
CDER · issued 2026-06-23 · posted 2026-07-07
Medical Device/Adulterated/Misbranded/Lacks PMA and/or 510(k)
CDRH · issued 2026-06-03 · posted 2026-07-07
Buy Keta Online / www.buyketaonline.com
Unapproved New Drugs/Misbranded
CDER · issued 2026-06-23 · posted 2026-07-07
Extra Medication / www.extramedication.com
Unapproved New Drugs/Misbranded
CDER · issued 2026-06-23 · posted 2026-07-07
Ket Plug / ketaminelab.org and buyketaminepowderonline.com
Unapproved New Drugs/Misbranded
CDER · issued 2026-06-23 · posted 2026-07-07
Keta Med Lab / www.ketamedlab.com
Unapproved New Drugs/Misbranded
CDER · issued 2026-06-23 · posted 2026-07-07
Ketacyn Pharmaceuticals / www.ketacynpharma.com
Unapproved New Drugs/Misbranded
CDER · issued 2026-06-23 · posted 2026-07-07
Ketamine Store / www.ketaminestore.org
Unapproved New Drugs/Misbranded
CDER · issued 2026-06-23 · posted 2026-07-07
Ketamine Troche Store / www.ketaminetrochestore.com
Unapproved New Drugs/Misbranded
CDER · issued 2026-06-23 · posted 2026-07-07
Show 6 more
Ketdaxandrugs / www.ketdaxandrugs.org
Unapproved New Drugs/Misbranded
CDER · issued 2026-06-23 · posted 2026-07-07
Legit Ketamine Suppliers / www.legitketaminesuppliers.com
Unapproved New Drugs/Misbranded
CDER · issued 2026-06-23 · posted 2026-07-07
Unapproved New Drugs/Misbranded
CDER · issued 2026-06-23 · posted 2026-07-07
Med Shop Inc / www.medshopinc.com
Unapproved New Drugs/Misbranded
CDER · issued 2026-06-23 · posted 2026-07-07
Pure Arylcyclohexylamine Store / www.arylcyclohexylamine.com
Unapproved New Drugs/Misbranded
CDER · issued 2026-06-23 · posted 2026-07-07
Medical Device/Adulterated/Misbranded/Lacks PMA and/or 510(k)
CDRH · issued 2026-05-19 · posted 2026-07-07
An FEI links to its facility page only when the letter names exactly one FEI and FDAgent holds FDA records for it.
Recalls by class
Drug and device recalls in FDA's Enforcement Report dated in July, from openFDA, which FDA last updated Nov 11, 2026. Each openFDA record is one recalled product; a recall event can cover many products, so events are counted too.
| Class | Drug products | Device products | EventsRecall events |
|---|---|---|---|
| Class I | 4 | 49 | 16 |
| Class II | 101 | 221 | 132 |
| Class III | 6 | 0 | 5 |
| All | 111 | 270 | 151 |
The largest Class I recall event, from Abiomed, Inc., covers 13 of the 53 Class I products. The Class I products are listed below. Class II and III recalls are listed on each week's page.
Class I53 recallsShow all
- D-0732-2026openFDA record
JB Chemicals and Pharmaceuticals Ltd
Cetirizine Hydrochloride Tablets USP 5 mg, 100-count bottle, Manufactured by: Unique Pharmaceuticals Labs, (A Div. of J.B. Chemicals & Pharmaceuticals, Ltd.), Mumbai 400 030, India. Distributed by: Rising Pharma Holdings, Inc., East Brunswick, NJ 08816, NDC 16571-401-10.
Reason: Cross Contamination with Other Products: Customer complaints for appearance of discolored tablets and red dots observed on Cetirizine Hydrochloride Tablets USP 5 mg. Determined to be Ranitidine.
Drug · initiated 2026-06-24 · Enforcement Report 2026-07-29 · Ongoing
- Z-2747-2026openFDA record
Medline Industries, LP
MEDLINE ANTERIOR HIP, Medline Kit SKU DYNJ909640G
Reason: Medline Industries, LP is issuing a recall for specific Medline kits containing B. Braun convenience kits that include Bupivacaine Hydrochloride in Dextrose Injection, USP. B. Braun Medical Inc. was notified by their supplier, Spectra Medical Devices, LLC of a product recall fro…
Device · initiated 2026-06-05 · Enforcement Report 2026-07-29 · Ongoing
- Z-2748-2026openFDA record
Medline Industries, LP
MEDLINE convenience kits labeled as: 1) C-SECTION, Medline SKU DYNJ910835; 2) C-SECTION, Medline SKU DYNJ910835A; medical convenience kits
Reason: Medline Industries, LP is issuing a recall for specific Medline kits containing B. Braun convenience kits that include Bupivacaine Hydrochloride in Dextrose Injection, USP. B. Braun Medical Inc. was notified by their supplier, Spectra Medical Devices, LLC of a product recall fro…
Device · initiated 2026-06-05 · Enforcement Report 2026-07-29 · Ongoing
- Z-2749-2026openFDA record
Medline Industries, LP
MEDLINE convenience kits labeled as: 1) TOTAL JOINT KIT, Medline Kit SKU DYKS1574A; 2) TOTAL KNEE BASIC, Medline Kit SKU DYNJ902378J; 3) TOTAL HIP, Medline Kit SKU DYNJ903492K; 4) ANTERIOR HIP, Medline Kit SKU DYNJ903493I; 5) ANTERIOR HIP, Medline Kit SKU DYNJ903493J; 6) ANTERIO…
Reason: Medline Industries, LP is issuing a recall for specific Medline kits containing B. Braun convenience kits that include Bupivacaine Hydrochloride in Dextrose Injection, USP. B. Braun Medical Inc. was notified by their supplier, Spectra Medical Devices, LLC of a product recall fro…
Device · initiated 2026-06-05 · Enforcement Report 2026-07-29 · Ongoing
- Z-2764-2026openFDA record
Medline Industries, LP
MEDLINE medical procedure convenience kit labeled as: HIP CDS-LF, Medline kit SKU CDS984556N
Reason: The procedure kits contain ICU Medical, Inc. pain management kits containing Bupivacaine Hydrochloride in Dextrose Injection, USP, which were recalled due to quality issues that could result in drug ineffectiveness.
Device · initiated 2026-06-08 · Enforcement Report 2026-07-29 · Ongoing
- Z-2765-2026openFDA record
Medline Industries, LP
MEDLINE medical procedure convenience kit labeled as: 1) TOTAL KNEE-LF, Medline kit SKU CDSOPT590T; 2) TOTAL KNEE-LF, Medline kit SKU CDSOPT590U; 3) TOTAL KNEE CDS, Medline kit SKU CDS983930J; 4) HIP CDS-LF, Medline kit SKU CDS984556L; 5) HIP CDS-LF, Medline kit SKU CDS984556M;…
Reason: The procedure kits contain ICU Medical, Inc. pain management kits containing Bupivacaine Hydrochloride in Dextrose Injection, USP, which were recalled due to quality issues that could result in drug ineffectiveness.
Device · initiated 2026-06-08 · Enforcement Report 2026-07-29 · Ongoing
- Z-2681-2026openFDA record
Abiomed, Inc.
ABIOMED Impella RP Flex with SmartAssist Set, Pump Set Code 1000323, containing Kit, 23Fr Introducer, 11cm, Sterile, 23 Fr, 11 cm, Product Code 2000342. heart pump and accessories
Reason: A potential for introducer sheath leakage from under the sheath cap and along the hub score lines, identified in 14Fr and 23Fr Introducers. These manufacturing issues are not visually detectable by the user, and 14Fr or 23Fr Introducers are provided in every pump set.
Device · initiated 2026-06-11 · Enforcement Report 2026-07-22 · Ongoing
- Z-2682-2026openFDA record
Abiomed, Inc.
ABIOMED Impella CP Set, Pump Set Code 0048-0003, containing BIOMED Impella Introducer, 14 Fr, 13 cm & 25 cm, Product Code 0052-3046. heart pump and accessories
Reason: A potential for introducer sheath leakage from under the sheath cap and along the hub score lines, identified in 14Fr and 23Fr Introducers. These manufacturing issues are not visually detectable by the user, and 14Fr or 23Fr Introducers are provided in every pump set.
Device · initiated 2026-06-11 · Enforcement Report 2026-07-22 · Ongoing
- Z-2683-2026openFDA record
Abiomed, Inc.
ABIOMED Impella CP Introducer Kit 14 Fr, 13 and 25 cm, Pump Set Code 1000550, containing Kit, 14Fr Introducer, 13cm&25cm, Sterile, Introducer Product Code 0052-3046. heart pump and accessories
Reason: A potential for introducer sheath leakage from under the sheath cap and along the hub score lines, identified in 14Fr and 23Fr Introducers. These manufacturing issues are not visually detectable by the user, and 14Fr or 23Fr Introducers are provided in every pump set.
Device · initiated 2026-06-11 · Enforcement Report 2026-07-22 · Ongoing
- Z-2684-2026openFDA record
Abiomed, Inc.
ABIOMED Introducer Kit 23 Fr, 11 cm, Pump Set Code 1000441, containing Kit, 23Fr Introducer, 11cm, Sterile, Introducer Product Code 2000342. heart pump and accessories
Reason: A potential for introducer sheath leakage from under the sheath cap and along the hub score lines, identified in 14Fr and 23Fr Introducers. These manufacturing issues are not visually detectable by the user, and 14Fr or 23Fr Introducers are provided in every pump set.
Device · initiated 2026-06-11 · Enforcement Report 2026-07-22 · Ongoing
- Z-2685-2026openFDA record
Abiomed, Inc.
ABIOMED Impella CP Introducer Kit 14 Fr, 13 cm, Pump Set Code 0052-0038, containing Kit, 14Fr Introducer, 13cm, Long Taper, Introducer Product Code 0052-3052. heart pump and accessories
Reason: A potential for introducer sheath leakage from under the sheath cap and along the hub score lines, identified in 14Fr and 23Fr Introducers. These manufacturing issues are not visually detectable by the user, and 14Fr or 23Fr Introducers are provided in every pump set.
Device · initiated 2026-06-11 · Enforcement Report 2026-07-22 · Ongoing
- Z-2686-2026openFDA record
Abiomed, Inc.
ABIOMED Impella CP Introducer Kit 14 Fr, 25 cm, Pump Set Code 0052-0039, containing Kit, 14Fr Introducer, 25cm, Long Taper, Introducer Product Code 0052-3053. heart pump and accessories
Reason: A potential for introducer sheath leakage from under the sheath cap and along the hub score lines, identified in 14Fr and 23Fr Introducers. These manufacturing issues are not visually detectable by the user, and 14Fr or 23Fr Introducers are provided in every pump set.
Device · initiated 2026-06-11 · Enforcement Report 2026-07-22 · Ongoing
- Z-2687-2026openFDA record
Abiomed, Inc.
ABIOMED Impella 5.5 Smart Assist Set, JP, Pump Set Codes 1000211, 1000212; containing Introducer Product Code 0052-3006-JP. heart pump and accessories
Reason: A potential for introducer sheath leakage from under the sheath cap and along the hub score lines, identified in 14Fr and 23Fr Introducers. These manufacturing issues are not visually detectable by the user, and 14Fr or 23Fr Introducers are provided in every pump set.
Device · initiated 2026-06-11 · Enforcement Report 2026-07-22 · Ongoing
- Z-2688-2026openFDA record
Abiomed, Inc.
ABIOMED Impella 5.5 SmartAssist S2 Set, JP, Pump Set Codes 1000459, 1000460; containing Introducer Product Code 0052-3006-JP. heart pump and accessories
Reason: A potential for introducer sheath leakage from under the sheath cap and along the hub score lines, identified in 14Fr and 23Fr Introducers. These manufacturing issues are not visually detectable by the user, and 14Fr or 23Fr Introducers are provided in every pump set.
Device · initiated 2026-06-11 · Enforcement Report 2026-07-22 · Ongoing
- Z-2689-2026openFDA record
Abiomed, Inc.
ABIOMED Impella CP Set, Pump Set Code 0048-0014; containing Introducer Product Code 0052-3046. heart pump and accessories
Reason: A potential for introducer sheath leakage from under the sheath cap and along the hub score lines, identified in 14Fr and 23Fr Introducers. These manufacturing issues are not visually detectable by the user, and 14Fr or 23Fr Introducers are provided in every pump set.
Device · initiated 2026-06-11 · Enforcement Report 2026-07-22 · Ongoing
- Z-2690-2026openFDA record
Abiomed, Inc.
ABIOMED Impella CP Smart Assist Set, JP, Pump Set Code 0048-0024-JP; containing Introducer Product Code 0052-3025-JP. heart pump and accessories
Reason: A potential for introducer sheath leakage from under the sheath cap and along the hub score lines, identified in 14Fr and 23Fr Introducers. These manufacturing issues are not visually detectable by the user, and 14Fr or 23Fr Introducers are provided in every pump set.
Device · initiated 2026-06-11 · Enforcement Report 2026-07-22 · Ongoing
- Z-2691-2026openFDA record
Abiomed, Inc.
ABIOMED Impella CP Set, Pump Set Code 0048-0044; containing Introducer Product Code 0052-3046. heart pump and accessories
Reason: A potential for introducer sheath leakage from under the sheath cap and along the hub score lines, identified in 14Fr and 23Fr Introducers. These manufacturing issues are not visually detectable by the user, and 14Fr or 23Fr Introducers are provided in every pump set.
Device · initiated 2026-06-11 · Enforcement Report 2026-07-22 · Ongoing
- Z-2692-2026openFDA record
Abiomed, Inc.
ABIOMED Insertion Kit, Pkgd, JP, Pump Set Code 0052-0011-JP; containing Introducer Product Code 0052-3006-JP. heart pump and accessories
Reason: A potential for introducer sheath leakage from under the sheath cap and along the hub score lines, identified in 14Fr and 23Fr Introducers. These manufacturing issues are not visually detectable by the user, and 14Fr or 23Fr Introducers are provided in every pump set.
Device · initiated 2026-06-11 · Enforcement Report 2026-07-22 · Ongoing
- Z-2693-2026openFDA record
Abiomed, Inc.
ABIOMED Intro Kit, 14Fr x 13 and 25 cm, PKGD, JP, Pump Set Code 0052-0023-JP; containing Introducer Product Code 0052-3025-JP. heart pump and accessories
Reason: A potential for introducer sheath leakage from under the sheath cap and along the hub score lines, identified in 14Fr and 23Fr Introducers. These manufacturing issues are not visually detectable by the user, and 14Fr or 23Fr Introducers are provided in every pump set.
Device · initiated 2026-06-11 · Enforcement Report 2026-07-22 · Ongoing
- Z-2704-2026openFDA record
Argon Medical Devices, Inc
Transvenous Pacing Electrode Tray w/ Bipolar Balloon Pacing Electrode 5F, REF: 008416A, 008566A
Reason: Kits containing Lidocaine HCL Injection USP 1% 10mg/mL 5mL ampules subject to another recall recalled due to not meeting release criteria for sterility.
Device · initiated 2026-06-01 · Enforcement Report 2026-07-22 · Ongoing
- Z-2705-2026openFDA record
Argon Medical Devices, Inc
One-Piece Introducer Tray, REF: 497455, 497434 cardiovascular
Reason: Kits containing Lidocaine HCL Injection USP 1% 10mg/mL 5mL ampules subject to another recall recalled due to not meeting release criteria for sterility.
Device · initiated 2026-06-01 · Enforcement Report 2026-07-22 · Ongoing
- Z-2706-2026openFDA record
Argon Medical Devices, Inc
General Biospy Tray, REF: GUTS1000
Reason: Kits containing Lidocaine HCL Injection USP 1% 10mg/mL 5mL ampules subject to another recall recalled due to not meeting release criteria for sterility.
Device · initiated 2026-06-01 · Enforcement Report 2026-07-22 · Ongoing
- Z-2707-2026openFDA record
Argon Medical Devices, Inc
T-Lok Bone Marrow Tray, REF: BMT1104STL; BMET1104TL; BMT1104CATL
Reason: Kits containing Lidocaine HCL Injection USP 1% 10mg/mL 5mL ampules subject to another recall recalled due to not meeting release criteria for sterility.
Device · initiated 2026-06-01 · Enforcement Report 2026-07-22 · Ongoing
- Z-2708-2026openFDA record
Argon Medical Devices, Inc
Arterial Line Kit 20ga x 6, REF: 400115A
Reason: Kits containing Lidocaine HCL Injection USP 1% 10mg/mL 5mL ampules subject to another recall recalled due to not meeting release criteria for sterility.
Device · initiated 2026-06-01 · Enforcement Report 2026-07-22 · Ongoing
- Z-2709-2026openFDA record
Argon Medical Devices, Inc
Bone Marrow Tray, REF: CB0033TL; CB0355TLA
Reason: Kits containing Lidocaine HCL Injection USP 1% 10mg/mL 5mL ampules subject to another recall recalled due to not meeting release criteria for sterility.
Device · initiated 2026-06-01 · Enforcement Report 2026-07-22 · Ongoing
- Z-2678-2026openFDA record
Oscor Inc.
Abiomed REF: 0052-3025, Introducer Kit for Impella Hemostatic Peel Away Introducer System with Infusion Sideport, 14F 13cm and 25cm. heart pump and accessories.
Reason: Due to manufacturing issues, there is a potential for the 14Fr and 23Fr introducer sheaths to leak from the sidearm, under the sheath cap, and along the hub score lines in the hub region.
Device · initiated 2026-06-10 · Enforcement Report 2026-07-22 · Ongoing
- Z-2679-2026openFDA record
Oscor Inc.
Abiomed REF: 0052-3021, Introducer Kit for Impella Hemostatic Peel Away Introducer System with Infusion Sideport, 14F, 23F. heart pump and accessories.
Reason: Due to manufacturing issues, there is a potential for the 14Fr and 23Fr introducer sheaths to leak from the sidearm, under the sheath cap, and along the hub score lines in the hub region.
Device · initiated 2026-06-10 · Enforcement Report 2026-07-22 · Ongoing
- Z-2680-2026openFDA record
Oscor Inc.
Abiomed REF: 0052-3006, Introducer Kit for Impella Hemostatic Peel Away Introducer System with Infusion Sideport, 23F. heart pump and accessories.
Reason: Due to manufacturing issues, there is a potential for the 14Fr and 23Fr introducer sheaths to leak from the sidearm, under the sheath cap, and along the hub score lines in the hub region.
Device · initiated 2026-06-10 · Enforcement Report 2026-07-22 · Ongoing
- Z-2548-2026openFDA record
Baxter Healthcare Corporation
Hillrom VOLARA System P.CIRCUIT KIT (Single Patient Use Circuit), Model/Catalog Number REF M08085; Oscillation Lung Expansion Therapy, positive pressure breathing device
Reason: There have been reports of air and medication leakage from the nebulizer cup during therapy of Volara system patient circuits. The leakage can lead to an oxygen level drop in the patient and ineffective nebulization, which may affect the delivery of the prescribed treatment. The…
Device · initiated 2026-05-21 · Enforcement Report 2026-07-08 · Ongoing
- Z-2549-2026openFDA record
Baxter Healthcare Corporation
Hillrom VOLARA P.CIRCUIT 5KIT (System Single Patient Use Circuit), Model/Catalog Number REF M08270; Oscillation Lung Expansion Therapy, positive pressure breathing device
Reason: There have been reports of air and medication leakage from the nebulizer cup during therapy of Volara system patient circuits. The leakage can lead to an oxygen level drop in the patient and ineffective nebulization, which may affect the delivery of the prescribed treatment. The…
Device · initiated 2026-05-21 · Enforcement Report 2026-07-08 · Ongoing
- Z-2550-2026openFDA record
Baxter Healthcare Corporation
Hillrom VOLARA System P.CIRCUIT KIT AC (Single Patient Use Circuit), Model/Catalog Number REF M08473; Oscillation Lung Expansion Therapy, positive pressure breathing device
Reason: There have been reports of air and medication leakage from the nebulizer cup during therapy of Volara system patient circuits. The leakage can lead to an oxygen level drop in the patient and ineffective nebulization, which may affect the delivery of the prescribed treatment. The…
Device · initiated 2026-05-21 · Enforcement Report 2026-07-08 · Ongoing
- Z-2551-2026openFDA record
Baxter Healthcare Corporation
Hillrom VOLARA System P.CIRCUIT 5KIT AC, (Single Patient Use Circuit), Model/Catalog Number REF M08474; Oscillation Lung Expansion Therapy, positive pressure breathing device
Reason: There have been reports of air and medication leakage from the nebulizer cup during therapy of Volara system patient circuits. The leakage can lead to an oxygen level drop in the patient and ineffective nebulization, which may affect the delivery of the prescribed treatment. The…
Device · initiated 2026-05-21 · Enforcement Report 2026-07-08 · Ongoing
- D-0692-2026openFDA record
Blaine Labs Inc
Revitaderm Wound Care Gel, Benzalkonium Chloride 0.1%, packaged in 1.0 FL OZ (37 mL) bottle, shorter twist cap, Manufactured By: Blaine Labs, Inc. 11037 Lockport Place Santa Fe Springs, CA 90670, NDC 63347-120-02.
Reason: Microbial contamination of Non-Sterile Product: samples identified the presence of Lysinibacillus fusiformis and other Bacillus spp
Drug · initiated 2026-04-08 · Enforcement Report 2026-07-08 · Ongoing
- Z-2463-2026openFDA record
Hamilton Medical AG
Breathing circuit set, coaxial: REF: 260127, 260128, 260167, 260168 used with HAMILTON-C1, HAMILTON-T1 and HAMILTON-MR1 ventilators.
Reason: Ventilator coaxial breathing circuit may have, expiratory valve sets with membranes sticking to valve body, not detectable during pre-operative test, which will result in "exhalation obstructed " alarm, which may result in inadequate ventilation, oxygen desaturation. Advising: P…
Device · initiated 2026-05-29 · Enforcement Report 2026-07-08 · Ongoing
- Z-2603-2026openFDA record
Insulet Corporation
Omnipod 5 Pods. Includes the below Model/REF Numbers: 1. POD-OMNI-I1-6720. 2. POD-OMNI-I1-6220. 3. OMNI-I1-6729 (10-Pack), OMNI-I1-6720 (Single Pod). 4. OMNI-I1-6220. 5. OMNI-I1-6720. 6. BLE-H1-529 (10-Pack), BLE-H1-520 (Single Pod).
Reason: External soft cannula damage during manufacturing that results in insulin leaking around the Pod instead of being delivered to the user regardless of basal or bolus delivery. The primary failure mode is pump under-delivery due to loss of insulin to an external leak. If insulin i…
Device · initiated 2026-05-20 · Enforcement Report 2026-07-08 · Ongoing
- Z-2604-2026openFDA record
Insulet Corporation
Omnipod DASH Pod. Includes the below Model/REF Numbers: 1. POD-BLE-I1-525 (5-Pack), POD-BLE-I1-520 (Single Pod). 2. POD-BLE-I1-529 (10-Pack), POD-BLE-I1-520 (Single Pod). 3. BLE-I1-529 (10-Pack), BLE-I1-520 (Single Pod). 4. BLE-C1-529 (10-Pack), BLE-C1-520 (Single Pod).
Reason: External soft cannula damage during manufacturing that results in insulin leaking around the Pod instead of being delivered to the user regardless of basal or bolus delivery. The primary failure mode is pump under-delivery due to loss of insulin to an external leak. If insulin i…
Device · initiated 2026-05-20 · Enforcement Report 2026-07-08 · Ongoing
- Z-2605-2026openFDA record
Insulet Corporation
Omnipod (Eros) Pod. Includes the below Model/REF Numbers: 1. POD-ZXP425 (10 Pack), POD- ZXP420 (Single Pod). 2. ZXR425 (10-Pack), ZXR420 (Single Pod). 3. ZXP425 (10 Pack), ZXP420 (Single Pod).
Reason: External soft cannula damage during manufacturing that results in insulin leaking around the Pod instead of being delivered to the user regardless of basal or bolus delivery. The primary failure mode is pump under-delivery due to loss of insulin to an external leak. If insulin i…
Device · initiated 2026-05-20 · Enforcement Report 2026-07-08 · Ongoing
- Z-2615-2026openFDA record
Medline Industries, LP
Reprocessed Electrophysiology Catheters and Ultrasound Catheters Item no. 401206RH, St. Jude Medical Response" Diagnostic Electrophysiology Catheter 4 Elec. 10mm; Item no. 401305RH, St. Jude Medical Response" Diagnostic Electrophysiology Catheter 10 Elec. 2-5-2mm.
Reason: Expansion of RES 98277. Specific lots of reprocessed catheters may contain small particles of residual material. If devices with residual particulates on patient-contacting surfaces are used, there is a potential risk of an inflammatory response or systemic infection. If particu…
Device · initiated 2026-05-15 · Enforcement Report 2026-07-08 · Ongoing
- Z-2624-2026openFDA record
Medtronic Heart Valves Division
Medtronic Harmony Delivery Catheter System
Reason: Devices containing inner shaft assemblies from specific manufacturing lots may have an increased potential for distal tip detachment from the delivery system under certain procedural or anatomical conditions due to a manufacturing issue.
Device · initiated 2026-05-22 · Enforcement Report 2026-07-08 · Ongoing
- Z-2619-2026openFDA record
Windstone Medical Packaging, Inc.
A M S, ALIGNED MEDICAL SOLUTIONS, DSAEK Pack-RX, Pack Number AMS12394B, surgical convenience kit
Reason: Kits contain Webcol Large Alcohol Prep Pad which were recalled by Cardinal Health following the discovery of a contaminant (Paenibacillus phoenicis) during a routine sterilization dose audit.
Device · initiated 2026-05-21 · Enforcement Report 2026-07-08 · Ongoing
- Z-2620-2026openFDA record
Windstone Medical Packaging, Inc.
A M S, ALIGNED MEDICAL SOLUTIONS, Thyroid FNA Pack-RX, Pack Number AMS17729, surgical convenience kit
Reason: Kits contain Webcol Large Alcohol Prep Pad which were recalled by Cardinal Health following the discovery of a contaminant (Paenibacillus phoenicis) during a routine sterilization dose audit.
Device · initiated 2026-05-21 · Enforcement Report 2026-07-08 · Ongoing
- Z-2621-2026openFDA record
Windstone Medical Packaging, Inc.
A M S, ALIGNED MEDICAL SOLUTIONS, Preop Kit-NS, Pack Number AMS18227, surgical convenience kit
Reason: Kits contain Webcol Large Alcohol Prep Pad which were recalled by Cardinal Health following the discovery of a contaminant (Paenibacillus phoenicis) during a routine sterilization dose audit.
Device · initiated 2026-05-21 · Enforcement Report 2026-07-08 · Ongoing
- Z-2622-2026openFDA record
Windstone Medical Packaging, Inc.
A M S, ALIGNED MEDICAL SOLUTIONS, Dr. Lewin Pack, Pack Number AMS7200D, surgical convenience kit
Reason: Kits contain Webcol Large Alcohol Prep Pad which were recalled by Cardinal Health following the discovery of a contaminant (Paenibacillus phoenicis) during a routine sterilization dose audit.
Device · initiated 2026-05-21 · Enforcement Report 2026-07-08 · Ongoing
- Z-2599-2026openFDA record
Abiomed, Inc.
Abiomed 14 Fr x 13 cm Low Profile Introducer Kit for Impella CP. Impella Set Product Code: 1000434. Introducer Product Code: 1000434.
Reason: Potential for thrombus formation during prolonged use of the introducer.
Device · initiated 2026-05-22 · Enforcement Report 2026-07-01 · Ongoing
- Z-2600-2026openFDA record
Abiomed, Inc.
Abiomed 14 Fr x 25 cm Low Profile Introducer Kit for Impella CP. Impella Set Product Code: 1000435. Introducer Product Code: 1000435.
Reason: Potential for thrombus formation during prolonged use of the introducer.
Device · initiated 2026-05-22 · Enforcement Report 2026-07-01 · Ongoing
- Z-2601-2026openFDA record
Abiomed, Inc.
Abiomed Impella CP Set with SmartAssist (10th Generation) containing affected 14Fr Low Profile Introducer Kits. Product Codes: 1000413, 1000834.
Reason: Potential for thrombus formation during prolonged use of the introducer.
Device · initiated 2026-05-22 · Enforcement Report 2026-07-01 · Ongoing
- Z-2602-2026openFDA record
Abiomed, Inc.
Abiomed Impella CP Set with SmartAssist (10th Generation) containing affected 14Fr Low Profile Introducer Kits 13 and 25 cm Japan. Product Code: 1000542.
Reason: Potential for thrombus formation during prolonged use of the introducer.
Device · initiated 2026-05-22 · Enforcement Report 2026-07-01 · Ongoing
- D-0624-2026openFDA record
CareFusion 213, LLC
BD ChloraPrep Clear (2% w/v chlorhexidine gluconate (CHG) and 70% v/v isopropyl alcohol (IPA)), 1 mL Applicator, 60 Applicators per inner Carton, Sterile Solution, Care Fusion 213, LLC, El Paso, TX 79912, subsidiary of Beckton, Dickinson and Co., NDC 54365-400-31.
Reason: Non-Sterility: Due to presence of Aspergillus penicillioides.
Drug · initiated 2026-05-28 · Enforcement Report 2026-07-01 · Ongoing
- D-0625-2026openFDA record
CareFusion 213, LLC
BD ChloraPrep FREPP Clear,(2% w/v chlorhexidine gluconate (CHG) and 70% v/v isopropyl alcohol (IPA)) 1.5 mL Applicator, 20 Applicators, Sterile Solution, Care Fusion 213, LLC, El Paso, TX 79912, subsidiary of Beckton, Dickinson and Co., NDC 54365-400-30.
Reason: Non-Sterility: Due to presence of Aspergillus penicillioides. And Lack of Assurance of Sterility: Due to wrinkles in the paper lidding which may breach the seal area.
Drug · initiated 2026-05-28 · Enforcement Report 2026-07-01 · Ongoing
- Z-2452-2026openFDA record
Medline Industries, LP
Medline Convenience Kits: 1) DYNJ905503F, Model Number: CV ANESTHESIA - ROOM SET UP; 2) DYNJ905503G, Model Number: CV ANESTHESIA - ROOM SET UP
Reason: The Medline Convenience Kits contain Arrow Kits and Sets that include Lidocaine Hydrochloride Injection, USP and Bupivacaine Hydrochloride in Dextrose Injection, USP which were recalled due to quality issues identified during a recent FDA inspection of the supplier.
Device · initiated 2026-05-06 · Enforcement Report 2026-07-01 · Ongoing
- Z-2453-2026openFDA record
Medline Industries, LP
Medline Convenience Kits: 1) OPEN HEART CDS, Model Number: CDS840144Q; 2) OPEN HEART, Model Number: CDS840150W; 3) OPEN HEART ADULT, Model Number: CDS840396AA; 4) CABG A CDS, Model Number: CDS982269N; 5) OPEN HEART ANESTHESIA PACK, Model Number: CDS983849G; 6) OPEN HEART ACCESSO…
Reason: The Medline Convenience Kits contain Arrow Kits and Sets that include Lidocaine Hydrochloride Injection, USP and Bupivacaine Hydrochloride in Dextrose Injection, USP which were recalled due to quality issues identified during a recent FDA inspection of the supplier.
Device · initiated 2026-05-06 · Enforcement Report 2026-07-01 · Ongoing
- Z-2454-2026openFDA record
Medline Industries, LP
Medline Convenience Kits: 1) DRAWER 6 CENTRAL LINE INSERTIO, Model Number: ACC011142; 2) CENTRAL AND ARTERIAL LINE TOTE, Model Number: DYNJ8407
Reason: The Medline Convenience Kits contain Arrow Kits and Sets that include Lidocaine Hydrochloride Injection, USP and Bupivacaine Hydrochloride in Dextrose Injection, USP which were recalled due to quality issues identified during a recent FDA inspection of the supplier.
Device · initiated 2026-05-06 · Enforcement Report 2026-07-01 · Ongoing
- Z-2455-2026openFDA record
Medline Industries, LP
Medline Convenience Kits: 1) DRAWER 6A ADULT CENTRAL LINE, Model Number: ACC010890; 2) OPEN HEART A OVERHEAD, Model Number: DYNJ910066; 3) OPEN HEART, Model Number: DYNJ9855016V; 4) OPEN HEART, Model Number: DYNJ9855016X
Reason: The Medline Convenience Kits contain Arrow Kits and Sets that include Lidocaine Hydrochloride Injection, USP and Bupivacaine Hydrochloride in Dextrose Injection, USP which were recalled due to quality issues identified during a recent FDA inspection of the supplier.
Device · initiated 2026-05-06 · Enforcement Report 2026-07-01 · Ongoing
Import alert entries
31 firm entries FDA published on its import alert pages this month, dated by each entry's date published: 27 on a Red List, 4 on a Green List. FDA lists firms by name and address; FDAgent links an entry to a facility only when FDA printed the firm's FEI. 2 entries whose name may be a person's rather than a firm's are not listed or counted here; signed-in users can see them in FDAgent. Import alerts last synced Nov 20, 2026, 03:55 UTC.
66-40Red List3 entriesAlert on fda.govShow allDetention Without Physical Examination of Drugs From Firms Which Have Not Met Drug GMPsAlert on fda.gov
- Published 2026-07-31
Zhejiang Gaorong Cosmetic Co., Ltd
CHINA
- Published 2026-07-21
Tianjin Kilo Pharmaceutical Sci-tech Co., Ltd.
CHINA
- Published 2026-07-09
Reliance Life Sciences Private Limited
INDIA
66-41Red List4 entriesAlert on fda.govShow allDetention Without Physical Examination of Unapproved New Drugs Promoted In The U.S.Alert on fda.gov
- Published 2026-07-31
LABORATORIOS SUZIOS
EL SALVADOR
- Published 2026-07-16
IDO PHARM
KOREA (THE REPUBLIC OF)
- Published 2026-07-13
PATANJALI AYURVED LIMITED
INDIA
- Published 2026-07-13
PATANJALI AYURVED LTD
INDIA
66-79Red List5 entriesAlert on fda.govShow allDetention Without Physical Examination of Drugs From Foreign Establishments Refusing FDA InspectionAlert on fda.gov
- Published 2026-07-31
Guangdong Bisutang Biotechnology Co., Ltd.
CHINA
- Published 2026-07-22
JUN DA BEAUTY-TEC CO., LTD.
TAIWAN
- Published 2026-07-15
Jiangxi Dentalbright Technology Co., Ltd
CHINA
- Published 2026-07-15
LANTERN HEALTH & BEAUTY SDN. BHD.
MALAYSIA
- Published 2026-07-13
Nantong Qinji Health Technology Co., Ltd.
CHINA
68-19Red List6 entriesAlert on fda.govShow allDETENTION WITHOUT PHYSICAL EXAMINATION OF UNAPPROVED FINISHED NEW ANIMAL DRUGSAlert on fda.gov
- Published 2026-07-27
Qingdao chongbeitian Pet Products Co., Ltd
CHINA
- Published 2026-07-16
Fatro Spa
ITALY
- Published 2026-07-16
Parfarm Sa
MEXICO
- Published 2026-07-16
Zirin Laboratories De
MEXICO
- Published 2026-07-02
AB SCIENCE
FRANCE
- Published 2026-07-02
Centre Specialites Pharmaceutiques
FRANCE
76-01Green List4 entriesAlert on fda.govShow allDetention Without Physical Examination Of Medical Instruments from PakistanAlert on fda.gov
- Published 2026-07-27
Sair International
PAKISTAN
- Published 2026-07-16
Zia & Company (Private) Limited
PAKISTAN
- Published 2026-07-02
LB Instruments PVT (LTD.)
PAKISTAN
- Published 2026-07-01
Glorious Mercantile Corp. Pvt. Ltd.
PAKISTAN
80-04Red List1 entryAlert on fda.govShow allSurveillance and Detention Without Physical Examination of Surgeon's and Patient Examination GlovesAlert on fda.gov
- Published 2026-07-16
KSG MEDICARE SDN BHD
MALAYSIA
89-04Red List1 entryAlert on fda.govShow allDetention Without Physical Examination of Devices from Firms that Have not met Device Quality System RequirementsAlert on fda.gov
- Published 2026-07-02
Huons Co., Ltd.
KOREA (THE REPUBLIC OF)
95-05Red List3 entriesAlert on fda.govShow allDetention Without Physical Examination of Electronic Products that Fail to Comply with Performance Standards or to Have Required CertificationAlert on fda.gov
- Published 2026-07-31
XUZHOU MAOBO MATERIALS TRADING CO., LTD
CHINA
- Published 2026-07-15
Shenzhen Lanlyn Technology Co., Ltd.
CHINA
- Published 2026-07-09
Shenzhen Guangyin Exploration Technology Co., Ltd
CHINA
99-34Red List4 entriesAlert on fda.govShow allDETENTION WITHOUT PHYSICAL EXAMINATION OF DRUGS OR MEDICAL DEVICES FROM FIRMS WITHOUT A VALID DRUG OR MEDICAL DEVICE REGISTRATION AND/OR LISTINGAlert on fda.gov
- Published 2026-07-06
(JLVN) JD Vietnam Investment Company LIMI
VIETNAM
- Published 2026-07-06
COTE DAZUR PHARMACY
FRANCE
- Published 2026-07-06
E FILLERS SA
GREECE
- Published 2026-07-06
FOREVER18 INTERNATIONAL CO., LTD
KOREA (THE REPUBLIC OF)
New OAI classifications
6 inspections classified Official Action Indicated that FDAgent first recorded from FDA's inspection classification data this month, after its reads of that data in the 8 days before had not found them. Older inspections FDAgent adds to its archive are left out. Dates are inspection end dates. Inspection classifications last synced Nov 22, 2026, 04:55 UTC.
- Inspection ended 2026-05-15FEI 1314625
Unither Manufacturing LLC.OAI
Rochester, New York, United States · Drugs · Drug Quality Assurance · recorded 2026-07-28
- Inspection ended 2026-04-21FEI 1000119582
Altaire Pharmaceuticals, Inc.OAI
Aquebogue, New York, United States · Drugs · Drug Quality Assurance · recorded 2026-07-14
- Inspection ended 2026-04-24FEI 3004886113
Hanlim Pharm Co. Ltd.OAI
Yongin, Gyeonggi-do, Korea (the Republic of) · Drugs · Drug Quality Assurance · recorded 2026-07-14
- Inspection ended 2026-04-24FEI 3003999190
Zhejiang Huahai Pharmaceutical Co., Ltd.OAI
Linhai, China · Drugs · Drug Quality Assurance · recorded 2026-07-14
- Inspection ended 2026-04-24FEI 3005949964
Catalent Indiana LLCOAI
Bloomington, Indiana, United States · Drugs · Drug Quality Assurance · recorded 2026-07-10
- Inspection ended 2026-04-14FEI 3008808745
Rohto-Mentholatum (Vietnam) Co., Ltd.OAI
Ho Chi Minh, Vietnam · Drugs · Drug Quality Assurance · recorded 2026-07-07
Week by week
Each week's page lists every Form 483 and recall. Weeks run Monday to Sunday, so the first and last can include days of the neighbouring months.
| Week | Dates | Letters | Form 483s | Recalls | New OAI |
|---|---|---|---|---|---|
| 2026-W31 | 2026-07-27 to 2026-08-02 | 0 letters | 32 Form 483s | 81 recalls | 1 new OAI |
| 2026-W30 | 2026-07-20 to 2026-07-26 | 6 letters | 24 Form 483s | 63 recalls | 0 new OAI |
| 2026-W29 | 2026-07-13 to 2026-07-19 | 4 letters | 56 Form 483s | 34 recalls | 3 new OAI |
| 2026-W28 | 2026-07-06 to 2026-07-12 | 16 letters | 24 Form 483s | 120 recalls | 2 new OAI |
| 2026-W27 | 2026-06-29 to 2026-07-05 | 7 letters | 2 Form 483s | 83 recalls | 0 new OAI |
Get each week’s page by email every Monday
A free weekly email with each week’s warning letters, new Form 483s, OAI classifications and Class I recalls, counted the same way as this page. Just your email address.
How this month is counted
Months are calendar months in UTC. Counts are read from FDAgent's copy of each FDA source, which refreshes daily, and FDAgent checks them against FDA's own sources every week. FDA posts some records late and removes others, so a finished month's counts can still change. FDAgent is independent of FDA.
- Warning letters posted
- Letters from FDA's drug, biologics and device offices, dated by the day FDA posted them, not the day it issued them.
- Form 483s listed
- Records FDA's Electronic Reading Room types as 483 or Amended 483, dated by FDA's publish date. A firm's 483 response is not counted. Each inspection's 483 counts once, when FDA first listed it: a later listing with the same FEI and record date is shown apart, as a re-post when FDA listed a 483 again under a new media id, or as an amended 483 when FDA types it Amended 483. FDA sometimes lists a 483 days after the publish date it gives it, so a finished month can still gain 483s.
- Inspections newly classified
- OAI, VAI and NAI inspections counted on the day FDAgent first found them in FDA's classification data, and only when its complete reads in the 8 days before had covered the inspection's end date without finding them. Inspections recorded before FDAgent kept a record of its reads, in November 2026, count when they ended in the 13 weeks before. Backfills and wider reads never count, nor do the Veterinary inspections of FDAgent's first read of them, in November 2026. FDA gives no date for a classification. When FDA removes an inspection from its Data Dashboard, it stops counting, so a finished month's count can fall.
- Regulations cited
- The FDA Data Dashboard's Form 483 citations on those inspections, by 21 CFR section. FDA often lists an inspection's citations weeks after its classification, so these counts grow after the month ends. For letters, the violation statements that cite a 21 CFR section, each section once per letter.
- Recalls
- openFDA's drug and device records by Enforcement Report date. Each record is one product; an event is FDA's recall event number.
- Import alert entries
- Firm entries by the date FDA published them on an import alert. Entries that may name a person are left out. FDA deletes an entry when it removes the firm; entries FDAgent recorded from November 2026 on keep counting in the month FDA dated them.
- Months in progress
- A month in progress counts the days so far and grows until it ends. Recalls for its latest days appear when openFDA updates.
- Comparisons
- Every month is counted the same way. A year-over-year figure appears only where FDAgent holds FDA's own dates for the earlier month. A percent change appears only when the earlier count is 10 or more. A month in progress shows no changes until it ends. A month that overlaps a federal government shutdown (Oct 1 to Nov 12, 2025) can have low counts, and month pages say so beside any comparison with one.
Cite this page
Quote a count with its date: the page refreshes daily, and a month's counts can change as FDA posts late records.
FDAgent. "FDA enforcement, July 2026." https://www.fdagent.app/enforcement/2026-07. Counts as of Nov 22, 2026.
About this data
- Weeks are ISO weeks, Monday to Sunday, in UTC. Each section is dated by its own FDA date: posting date for warning letters, reading-room publish date for Form 483s (each inspection's 483 counts once, when FDA first listed it; FDA's re-posts and amended 483s with the FEI and record date of one it listed before are shown apart), Enforcement Report date for recalls, and the date published FDA prints beside each firm on an import alert. FDA publishes no date for an inspection classification, so an OAI classification counts as new on the day FDAgent first recorded the inspection, and only when FDAgent's complete reads of FDA's inspection data in the 8 days before had covered the inspection's end date without finding it. Older inspections FDAgent adds to its archive, by a backfill or a wider read, never count as new, however late FDA posted them. Inspections FDAgent recorded before it began keeping a record of its reads, in November 2026, count only when they ended in the 13 weeks before, and the Veterinary inspections of FDAgent's first read of them, that same month, never count.
- Import alert entries come from FDA's import alert pages, where FDA lists firms by name and address, not by FEI. FDAgent links an entry to a facility only when FDA printed the firm's FEI, never by name. Earlier weeks list only entries still on FDA's pages, because FDA deletes an entry when it removes a firm, and FDAgent keeps removed entries only from the day it began recording them. FDAgent lists an entry here only when its name shows it is a firm, through a legal form such as Ltd, Inc or GmbH, a business word such as Laboratories or Trading, or a web address, or when FDA's inspection or drug registration records hold a firm under exactly that name; entries that may name a person are shown only to signed-in users.
- Import refusals are not listed by week: FDA reports a refusal weeks after it happens, so a week's count would keep changing after the week closed. Each facility page lists its own refusals.
- Warning letters are those issued by FDA's drug, biologics and device centers and their field and quality offices. Letters from FDA's food, tobacco, veterinary, cosmetics and dietary supplement offices are left out.
- FDA posts only some Form 483s to its Electronic Reading Room, so a missing 483 does not mean none was issued.
- FDA publishes an inspection's classification once it is final and updates its inspection data weekly. Source: FDA inspections FAQ.
- Recalls come from openFDA, which republishes FDA's weekly Enforcement Report; a report's recalls appear here once openFDA publishes it.
- FDAgent is independent of FDA. Every row links to its FDA source; check it before relying on a record.