Criticare Technologies, Inc.
333 Strawberry Field Rd Ste 11, Warwick, Rhode Island, United States
FEI 3012238587
FDA has 5 inspections on file for Criticare Technologies, Inc. in Warwick, Rhode Island (FEI 3012238587), from 2017 to 2025. The latest ended on and was classified OAI, official action indicated.
OAI means FDA found significant objectionable conditions and recommended regulatory or administrative action. It is the most serious of FDA's three inspection outcomes.
FDA published 8 citations from this inspection, all under 21 CFR Part 820, the device quality system rules. They are listed below. Two of the three FOIA requests for this site have asked FDA for this inspection's 483.
FDA has issued 2 warning letters under this FEI, the latest on (CDRH · Division of Medical Device and Radiological Health Operations East). 1 import refusal is filed under this FEI, on . No court action is filed under this FEI.
10 regulations were cited at more than one inspection; they are listed below.
Sources:FDA firm profile ↗Warning letter ↗Synced Nov 23, 2026
Ask FDAgent about this site
Email me when FDA posts a new inspection, Form 483 or warning letter for this site
Weekly, on a free account. Pause it any time.
Comparing suppliers? Check this site beside 2 others, free
- Latest inspection
- OAI
- Inspections on file
- 5, 2017–2025
- 4 OAI · 1 VAI · 0 NAI
- Inspections with 483 observations shown
- 5 of 5
- Form 483s posted by FDA
- None held for this FEI
- Requested under FOIA
- 3 requests, latest May 22, 2026
- Warning letters
- · CDRH · Division of Medical Device and Radiological Health Operations East ↗
- 2 on file
- Import refusals
- 1 refusal
- Import alerts
- Not linked by FEI
- FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗
Inspection history
5 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC
- OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Compliance: Devices8 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Lack of or inadequate complaint procedures
Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been adequately established.
Lack of or inadequate procedures
Procedures for corrective and preventive action have not been adequately established.
Purchasing controls, Lack of or inadequate procedures
Procedures to ensure that all purchased or otherwise received product and services conform to specified requirements have not been adequately established.
Design validation - Risk analysis
Results of the design risk analysis were not adequately documented.
Design validation- Lack of or inadequate procedures
Procedures for design validation have not been adequately established.
Show 3 more citations
Design verification - output does not meet input requirement
Design verification does not confirm that design output meets design input requirements.
Production processes
Production processes were not developed to ensure that a device conforms to its specifications.
Remedial action
When test/measurement equipment was found to not meet accuracy and precision limits, inadequate action was taken to evaluate whether there was any adverse effect on the device's quality.
- OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Compliance: Devices10 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Devices subject to device identification GUDID data submission requirements.
The information required by the Global Unique Device Identification Database for each medical device version or model required to bear a unique device identifier (UDI) was not provided.
Lack of or inadequate complaint procedures
Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been adequately established.
Training records
Personnel training is not documented.
Design validation- Lack of or inadequate procedures
Procedures for design validation have not been adequately established.
Design verification - output does not meet input requirement
Design verification does not confirm that design output meets design input requirements.
Show 5 more citations
Purchasing controls, Lack of or inadequate procedures
Procedures to ensure that all purchased or otherwise received product and services conform to specified requirements have not been adequately established.
Design Validation - Risk analysis not performed/inadequate
Risk analysis is inadequate.
Lack of or inadequate procedures - Acceptance activities
Procedures for acceptance activities have not been adequately established.
Lack of or inadequate procedures
Procedures for corrective and preventive action have not been adequately established.
Nonconforming product, Lack of or inadequate procedures
Procedures have not been adequately established to control product that does not conform to specified requirements.
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Postmarket Assurance: Devices10 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Sampling methods - Lack of or inadequate procedures
Procedures to ensure sampling methods are adequate for their intended use have not been adequately established.
Purchasing controls, Lack of or inadequate procedures
Procedures to ensure that all purchased or otherwise received product and services conform to specified requirements have not been established.
DHR content
The device history record does not demonstrate that the device was manufactured in accordance with the device master record and 21 CFR 820.
Lack of or inadequate procedures - Acceptance activities
Procedures for acceptance activities have not been established.
Design control - no procedures
Procedures for design control have not been established.
Show 5 more citations
Lack of procedures, or not maintained
Document control procedures have not been established.
Quality Audit/Reaudit - conducted
Quality audits have not been performed.
Investigation of device failures
Complaints involving the possible failure of labeling and packaging to meet any of its specifications were not evaluated and investigated where necessary.
Lack of or inadequate procedures for handling
Procedures for product handling have not been adequately established.
Lack of or inadequate procedures
Procedures for corrective and preventive action have not been established.
- OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Compliance: Devices5 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Servicing - Lack of or inadequate procedures
Procedures or instructions for performing servicing activities have not been adequately established.
Training - Lack of or inadequate procedures
Procedures for training and identifying training needs have not been adequately established.
Lack of or inadequate procedures
Procedures for corrective and preventive action have not been adequately established.
Investigation of device failures
Complaints involving the possible failure of a device to meet any of its specifications were not investigated where necessary.
Procedures for product review,disposition lack of/inadequate
Procedures that define the responsibility for review and the authority for the disposition of nonconforming product have not been adequately established.
- OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Postmarket Assurance: Devices13 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Lack of or inadequate procedures for labeling
Procedures to control labeling activities have not been established.
Training records
Personnel training is not documented.
Individual Report of Malfunction
An MDR report was not submitted within 30 days of receiving or otherwise becoming aware of information that reasonably suggests that a marketed device has malfunctioned and would be likely to cause or contribute to a death or serious injury if the malfunction were to recur.
Procedures not adequately established or maintained
Document control procedures have not been adequately established.
Calibration, Inspection, etc. Procedures Lack of or Inadequ
Procedures to ensure equipment is routinely calibrated, inspected, checked and maintained have not been adequately established.
Show 8 more citations
Purchasing controls, Lack of or inadequate procedures
Procedures to ensure that all purchased or otherwise received product and services conform to specified requirements have not been adequately established.
Identification procedures, Lack of or inadequate procedures
Procedures for identifying product during all stages of receipt, production, distribution, and installation have not been adequately established.
Documentation of software validation
Software validation activities and results for computers or automated data processing systems used as part of production have not been adequately documented.
Nonconforming product, Lack of or inadequate procedures
Procedures have not been adequately established to control product that does not conform to specified requirements.
Lack of or inadequate procedures - Acceptance activities
Procedures for acceptance activities have not been adequately established.
Lack of or inadequate procedures
Procedures for corrective and preventive action have not been adequately established.
Lack of Written MDR Procedures
Written MDR procedures have not been developed.
Lack of or inadequate complaint procedures
Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been established.
Requested under FOIA
FDA FOIA logs, last synced Nov 23, 2026, 04:25 UTC
Three requests to FDA under the Freedom of Information Act named this site, most recently on May 22, 2026. Each one asked for a Form 483.
Inspection ending Oct 28, 2025OAI
2 requests. In the inspection history above. FDA released records in answer to a request for its Form 483 on Mar 19, 2026. FDAgent holds no copy that FDA posted.
- May 22, 2026FDA-FOIA-2026-5217Form 483Records released
- Dec 9, 2025FDA-FOIA-2025-13791Form 483· 1 of 5 sitesRecords released
Inspection ending May 10, 2024OAI
1 request. In the inspection history above. None of FDA’s log entries for these requests say records were released.
- Oct 21, 2024FDA-FOIA-2024-9303Form 483· 1 of 2 sitesClosed
A request is listed when it names this site’s FEI number or one of its FDA inspection IDs, or, after review, its exact street address. Requester names are left out. “Closed” means FDA’s log gives no further reason.
Regulations FDA cited here
The 20 sections of 21 CFR in FDA’s published citations for this facility’s 5 inspections, with how many of them cited each.
- 21 CFR 820.100Corrective and preventive action5 of 5
- 21 CFR 820.198Complaint files5 of 5
- 21 CFR 820.50Purchasing controls4 of 5
- 21 CFR 820.25Personnel3 of 5
- 21 CFR 820.30Design controls3 of 5
- 21 CFR 820.80Receiving, in-process, and finished device acceptance3 of 5
- 21 CFR 820.90Nonconforming product3 of 5
- 21 CFR 820.40[Reserved]2 of 5
Show all 20
- 21 CFR 820.70Production and process controls2 of 5
- 21 CFR 820.72Inspection, measuring, and test equipment2 of 5
- 21 CFR 803.17What are the requirements for developing, maintaining, and implementing written MDR procedures that apply to me?1 of 5
- 21 CFR 803.50If I am a manufacturer, what reporting requirements apply to me?1 of 5
- 21 CFR 820.22Quality audit1 of 5
- 21 CFR 820.60Identification1 of 5
- 21 CFR 820.120Device labeling1 of 5
- 21 CFR 820.140Handling1 of 5
- 21 CFR 820.184Device history record1 of 5
- 21 CFR 820.200Servicing1 of 5
- 21 CFR 820.250Statistical techniques1 of 5
- 21 CFR 830.300Devices subject to device identification data submission requirements1 of 5
Warning letters and compliance actions
FDA Data Dashboard compliance actions
- Warning letterFormal notice of significant violationsCDRH · Division of Medical Device and Radiological Health Operations East · DevicesLetter on fda.govFDA record
- Warning letterFormal notice of significant violationsCDRH · DevicesFDA record
FDAgent links a letter’s text only when FDA’s compliance case number ties it to this FEI. No posted letter matched this one, so the link opens FDA’s record of the action.
Import refusals
FDA Data Dashboard, last synced Nov 23, 2026, 04:17 UTC. Shipments FDA refused entry into the US, filed under FEI 3012238587, one row per refused product line.
1 refusal on 2026-11-04 · charged NOT LISTED on 1, FRNMFGREG on 1 · no FDA sample analysis
FDA firm profile- 2026-11-04
MONITOR, PHYSIOLOGICAL, PATIENT74MHX
NOT LISTED 502(o), 801(a)(3); MISBRANDING
FRNMFGREG 502(o), 801(a)(3); MISBRANDING
REGISTERED 502(o), 801(a)(3); MISBRANDING
Division of West Coast Imports1FX-8088220-3/10001/1
Go further than this page
FDAgent researches FDA’s records and links every answer to its source.
About this data. Every record here is filed under FEI 3012238587 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
- Every record on this page is filed under FEI 3012238587. Records filed under another FEI or name, including other sites of the same company, are not shown here.
- FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
- Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
- “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
- Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
- FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.