21 CFR 820.140: Handling
FDA's inspection data lists 49 Form 483 citations of 21 CFR 820.140 at 48 facilities, from October 2008 to March 20, 2025. 1 warning letter states a violation of it. eCFR shows this section removed on February 4, 2026.
Counts: FDA Data Dashboard citations, last synced Nov 23, 2026, 04:25 UTC, and FDA's warning letters, last synced Nov 23, 2026, 04:17 UTC. Regulation text: eCFR. FDAgent is independent of FDA.
- Form 483 citations
- 49
- At 49 inspections
- Facilities cited
- 48
- Distinct FEIs
- Most recent citation
- Mar 20, 2025
- MEDICREATIONS LLC
- Warning letters
- 1
- State a violation of it
The regulation
Removed from the CFR on 2026-02-04. The text shown is the last text in force before then (eCFR version of 2016-12-30).
§ 820.140 Handling.
Each manufacturer shall establish and maintain procedures to ensure that mixups, damage, deterioration, contamination, or other adverse effects to product do not occur during handling.
21 CFR Part 820: Quality Management System Regulation, Subpart LRead on eCFR
Citations by year
Form 483 citations of 21 CFR 820.140 by the year the inspection ended. FDA's citation data starts with inspections ending in October 2008, and the current year runs to the latest sync, so both are partial years (lighter bars).
Peak: 9 in 2011
| Year | Form 483 citations |
|---|---|
| 2008 | 1 |
| 2009 | 6 |
| 2010 | 4 |
| 2011 | 9 |
| 2012 | 2 |
| 2013 | 6 |
| 2014 | 3 |
| 2015 | 2 |
| 2016 | 3 |
| 2017 | 3 |
| 2018 | 2 |
| 2019 | 1 |
| 2021 | 2 |
| 2022 | 3 |
| 2023 | 1 |
| 2025 | 1 |
How FDA words this citation
FDA's short descriptions and citation text on these 49 citations, most used first. These are FDA's standard summaries in its citation data, not the text of any one Form 483.
Short descriptions
- Lack of or inadequate procedures for handling49 citations
Citation text
- “Procedures for product handling have not been adequately established.”27 citations
- “Procedures for product handling have not been established.”14 citations
- “Procedures have not been established to ensure that mix-ups, damage, or other adverse effects to product do not occur during handling.”3 citations
- “Procedures have not been implemented to ensure that mix-ups, damage, or other adverse effects to product do not occur during handling.”3 citations
Show 2 more
- “Procedures have not been defined and implemented to ensure that mix-ups, damage, or other adverse effects to product do not occur during handling.”1 citations
- “Procedures have not been established, defined, documented, complete, and implemented to ensure that mix-ups, damage, or other adverse effects to product do not occur during handling.”1 citations
Facilities cited most recently
The latest inspections where FDA cited 21 CFR 820.140, by inspection end date, with FDA's short description. Each links to the facility's FDA record on FDAgent.
Show 4 more
- Criticare Technologies, Inc.2019-10-29 · FEI 3012238587Lack of or inadequate procedures for handling
- World Merchant Inc. dba Orthoextent2018-01-09 · FEI 3011277653Lack of or inadequate procedures for handling
Warning letters stating a violation of 21 CFR 820.140
1 warning letter FDA posted states a violation of this section in a numbered violation sentence. The most recent are below, each linking to FDAgent's page for the letter.
“Failure to establish and maintain adequate procedures to ensure that mix-ups, damage, deterioration, contamination, or other adverse effects to product do not occur during handling as required by 21 CFR § 820.140.”
- USA Medical LLCIssued 2022-08-03
Often cited at the same inspection
Of the 49 inspections where FDA cited 21 CFR 820.140, the share where it also cited each of these sections.
- 21 CFR 820.100Corrective and preventive action30 (61%)
- 21 CFR 820.198Complaint files26 (53%)
- 21 CFR 820.70Production and process controls22 (45%)
- 21 CFR 820.30Design controls21 (43%)
- 21 CFR 820.50Purchasing controls19 (39%)
- 21 CFR 820.75Process validation18 (37%)
Ask FDAgent about 21 CFR 820.140
FDAgent reads FDA's Form 483 citations, observations and warning letters and answers in plain English, with links to the records.
Related
About this data
- Citation counts come from FDA's Data Dashboard, which lists the regulation behind each Form 483 observation for inspections since October 2008.
- Observation excerpts come from Form 483 PDFs in FDA's electronic reading room, tied to this section only by exact wording as described above.
- A warning letter counts for this section when one of its numbered violation sentences cites 21 CFR 820.140; FDAgent quotes that sentence exactly.