LifeVac, LLC
110 Lake Ave S Ste 35, Nesconset, New York, United States
FEI 3011053282
FDA has 2 inspections on file for LifeVac, LLC in Nesconset, New York (FEI 3011053282), from 2018 to 2023. The latest ended on and was classified NAI, no action indicated. FDA issued a warning letter on (CDRH).
NAI means FDA found no objectionable conditions or practices, or found none that justified further action.
One of the three FOIA requests for this site has asked FDA for this inspection's records.
No court action or import refusal is filed under this FEI.
Sources:FDA firm profile ↗Warning letter ↗Synced Nov 23, 2026
Ask FDAgent about this site
Email me when FDA posts a new inspection, Form 483 or warning letter for this site
Weekly, on a free account. Pause it any time.
Comparing suppliers? Check this site beside 2 others, free
- Latest inspection
- NAI
- Inspections on file
- 2, 2018–2023
- 0 OAI · 1 VAI · 1 NAI
- Inspections with 483 observations shown
- 1 of 2
- Form 483s posted by FDA
- None held for this FEI
- Requested under FOIA
- 3 requests, latest FY2023
- Warning letters
- · CDRH ↗
- Import refusals
- None under this FEI
- Import alerts
- Not linked by FEI
- FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗
Inspection history
2 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC
- NAINo objectionable conditions that need follow-upDevices · Compliance: DevicesNone
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices16 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Lack of or inadequate procedures for handling
Procedures for product handling have not been established.
Design input - documentation
Design input requirements were not adequately documented.
Design output - review and approval
Design output was not reviewed and approved before release.
Calibration, Inspection, etc. Procedures Lack of or Inadequ
Procedures to ensure equipment is routinely calibrated have not been established.
Design validation- Lack of or inadequate procedures
Procedures for design validation have not been established.
Show 11 more citations
Design plans - Lack of or inadequate
Design plans that describe or reference the design and development activities and define responsibility for implementation have not been established.
Design validation acceptance criteria
Acceptance criteria were not established prior to the performance of validation activities.
Lack of or inadequate final acceptance procedures
Procedures for finished device acceptance have not been established.
Lack of or inadequate receiving acceptance procedures
Procedures for acceptance of incoming product have not been established.
Design input - Lack of or inadequate procedures
Procedures for design input have not been established.
Design input - review and approval
Design input requirements were not reviewed and approved by designated individual(s).
Quality Audit/Reaudit - conducted
Quality audits have not been performed.
Design output - Lack of or inadequate procedures
Procedures for design output have not been established.
Design review - Lack of or inadequate procedures
Procedures for design review have not been established.
Sampling methods - Lack of or inadequate procedures
Procedures to ensure sampling methods are adequate for their intended use have not been established.
Complaints
Complaint files are not adequately maintained.
Requested under FOIA
FDA FOIA logs, last synced Nov 23, 2026, 04:25 UTC
Three requests to FDA under the Freedom of Information Act named this site, most recently on FY2023.
Inspection ending May 25, 2023NAI
1 request. In the inspection history above. None of FDA’s log entries for these requests say records were released.
- Oct 30, 2024FDA-FOIA-2024-9614Inspection recordsClosed
Inspection date not given
2 requests. None of FDA’s log entries for these requests say records were released.
- FY2023FDA-FOIA-2023-4009EIRClosed
- FY2023FDA-FOIA-2023-2241EIRClosed
A request is listed when it names this site’s FEI number or one of its FDA inspection IDs, or, after review, its exact street address. Requester names are left out. “Closed” means FDA’s log gives no further reason.
Regulations FDA cited here
The 7 sections of 21 CFR in FDA’s published citations for 1 of this facility’s 2 inspections.
- 21 CFR 820.22Quality audit1 of 1
- 21 CFR 820.30Design controls1 of 1
- 21 CFR 820.72Inspection, measuring, and test equipment1 of 1
- 21 CFR 820.80Receiving, in-process, and finished device acceptance1 of 1
- 21 CFR 820.140Handling1 of 1
- 21 CFR 820.198Complaint files1 of 1
- 21 CFR 820.250Statistical techniques1 of 1
Warning letters and compliance actions
FDA Data Dashboard compliance actions
- Warning letterFormal notice of significant violationsCDRH · DevicesLetter on fda.govFDA record
Go further than this page
FDAgent researches FDA’s records and links every answer to its source.
About this data. Every record here is filed under FEI 3011053282 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
- Every record on this page is filed under FEI 3011053282. Records filed under another FEI or name, including other sites of the same company, are not shown here.
- FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
- Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
- “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
- Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
- FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.