World Merchant Inc. dba Orthoextent
42882 Truro Parish Dr Ste 210, Broadlands, Virginia, United States
FEI 3011277653
FDA has 2 inspections on file for World Merchant Inc. dba Orthoextent in Broadlands, Virginia (FEI 3011277653), from 2018 to 2025. The latest ended on and was classified VAI, voluntary action indicated.
VAI means FDA investigators found objectionable conditions or practices, but FDA was not prepared to take or recommend regulatory action. The firm is expected to correct them voluntarily.
FDA published 9 citations from this inspection. They are listed below. One FOIA request has asked FDA for this inspection's 483.
No warning letter, court action or import refusal is filed under this FEI.
Seven regulations were cited at more than one inspection; they are listed below.
Sources:FDA firm profile ↗Synced Nov 23, 2026
Ask FDAgent about this site
Email me when FDA posts a new inspection, Form 483 or warning letter for this site
Weekly, on a free account. Pause it any time.
Comparing suppliers? Check this site beside 2 others, free
- Latest inspection
- VAI
- Inspections on file
- 2, 2018–2025
- 0 OAI · 2 VAI · 0 NAI
- Inspections with 483 observations shown
- 2 of 2
- Form 483s posted by FDA
- None held for this FEI
- Requested under FOIA
- 1 request, latest Jun 11, 2025
- Warning letters
- None on file
- Import refusals
- None under this FEI
- Import alerts
- Not linked by FEI
- FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗
Inspection history
2 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Postmarket Assurance: Devices9 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Label to bear a unique device identifier
The labels of the medical device(s) do not bear a unique device identifier (UDI) that meet the requirements of 801 Subpart B and 21 CFR 830.
Lack of or inadequate procedures for distribution
Procedures for control and distribution of finished devices have not been established.
Nonconforming product, Lack of or inadequate procedures
Procedures have not been established to control product that does not conform to specified requirements.
Lack of or inadequate procedures
Procedures for corrective and preventive action have not been established.
Lack of or inadequate complaint procedures
Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been established.
Show 4 more citations
Purchasing controls, Lack of or inadequate procedures
Procedures to ensure that all purchased or otherwise received product and services conform to specified requirements have not been established.
Devices subject to device identification GUDID data submission requirements.
The information required by the Global Unique Device Identification Database for each medical device version or model required to bear a unique device identifier (UDI) was not provided.
Lack of or inadequate procedures for storage.
Procedures for the control of storage areas and stock rooms have not been established.
Lack of Written MDR Procedures
Written MDR procedures have not been developed, maintained and implemented.
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices8 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Lack of or inadequate procedures for distribution
Procedures for control and distribution of finished devices have not been established.
Lack of or inadequate procedures for storage.
Procedures for the control of storage areas and stock rooms have not been established.
Nonconforming product, Lack of or inadequate procedures
Procedures have not been established to control product that does not conform to specified requirements.
Lack of or inadequate procedures for handling
Procedures for product handling have not been established.
Lack of or inadequate complaint procedures
Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been established.
Show 3 more citations
Lack of Written MDR Procedures
Written MDR procedures have not been developed.
Lack of or inadequate procedures
Procedures for corrective and preventive action have not been established.
Purchasing controls, Lack of or inadequate procedures
Procedures to ensure that all purchased or otherwise received product and services conform to specified requirements have not been established.
Requested under FOIA
FDA FOIA logs, last synced Nov 23, 2026, 04:25 UTC
One request to FDA under the Freedom of Information Act named this site, on Jun 11, 2025. It asked for a Form 483.
Inspection ending Apr 16, 2025VAI
1 request. In the inspection history above. FDA released records in answer to a request for its Form 483 on Aug 5, 2025. FDAgent holds no copy that FDA posted.
- Jun 11, 2025FDA-FOIA-2025-7443Form 483· 1 of 5 sitesRecords released
A request is listed when it names this site’s FEI number or one of its FDA inspection IDs, or, after review, its exact street address. Requester names are left out.
Regulations FDA cited here
The 10 sections of 21 CFR in FDA’s published citations for this facility’s 2 inspections, with how many of them cited each.
- 21 CFR 803.17What are the requirements for developing, maintaining, and implementing written MDR procedures that apply to me?2 of 2
- 21 CFR 820.50Purchasing controls2 of 2
- 21 CFR 820.90Nonconforming product2 of 2
- 21 CFR 820.100Corrective and preventive action2 of 2
- 21 CFR 820.150Storage2 of 2
- 21 CFR 820.160Distribution2 of 2
- 21 CFR 820.198Complaint files2 of 2
- 21 CFR 801.20Label to bear a unique device identifier1 of 2
Show all 10
- 21 CFR 820.140Handling1 of 2
- 21 CFR 830.300Devices subject to device identification data submission requirements1 of 2
Go further than this page
FDAgent researches FDA’s records and links every answer to its source.
About this data. Every record here is filed under FEI 3011277653 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
- Every record on this page is filed under FEI 3011277653. Records filed under another FEI or name, including other sites of the same company, are not shown here.
- FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
- Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
- “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
- Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
- FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.