The Richline Group Inc

6701 Nob Hill Rd, Tamarac, Florida, United States

FEI 3006749547DUNS 962801937

FDA has 3 inspections on file for The Richline Group Inc in Tamarac, Florida (FEI 3006749547), from 2016 to 2025. The latest ended on and was classified OAI, official action indicated. FDA issued a warning letter on (CDRH).

OAI means FDA found significant objectionable conditions and recommended regulatory or administrative action. It is the most serious of FDA's three inspection outcomes.

FDA published 17 citations from this inspection: 15 under 21 CFR Part 820, the device quality system rules, 1 under 21 CFR Part 801, the device labeling rules, and 1 under 21 CFR Part 830, the unique device identification rules. They are listed below. Three FOIA requests have asked FDA for this inspection's 483.

No court action or import refusal is filed under this FEI.

Eight regulations were cited at more than one inspection; they are listed below.

As of Nov 21, 2026, the site is registered with FDA as a drug establishment for manufacturing and labeling, through Dec 31, 2026. Its registrant is The Richline Group, Inc. (D-U-N-S 962801937).

Sources:FDA firm profile ↗Warning letter ↗FDA drug establishment registrations ↗Synced Nov 23, 2026

Ask FDAgent about this site

Free account: 3 questions, no card needed. Each answer links to its FDA sources.

Email me when FDA posts a new inspection, Form 483 or warning letter for this site

Weekly, on a free account. Pause it any time.

Watch this facility

Comparing suppliers? Check this site beside 2 others, free

Latest inspection
OAI
Inspections on file
3, 2016–2025
2 OAI · 1 VAI · 0 NAI
Inspections with 483 observations shown
3 of 3
Form 483s posted by FDA
None held for this FEI
Requested under FOIA
3 requests, latest Apr 10, 2026
Import refusals
None under this FEI
Import alerts
Not linked by FEI
FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗
Drug establishment registration
Through

Inspection history

3 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC

  1. OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Compliance: Devices17 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Lack of or inadequate complaint procedures

      Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been adequately established.

    2. Incoming acceptance records, documentation

      Acceptance or rejection of incoming product was not documented.

    3. Sampling methods - Lack of or inadequate procedures

      Procedures to ensure sampling methods are adequate for their intended use have not been adequately established.

    4. DMR - not or inadequately maintained

      A device master record has not been adequately maintained.

    5. Nonconforming product control

      Products that do not conform to specifications are not adequately controlled.

    Show 12 more citations
    1. Evaluation of suppliers, contractors, etc., requirements

      Requirements that must be met by suppliers and consultants have not been adequately established.

    2. Calibration, Inspection, etc. Procedures Lack of or Inadequ

      Procedures to ensure equipment is routinely calibrated, inspected, checked and maintained have not been adequately established.

    3. Lack of or inadequate DHR procedures

      Procedures for device history records have not been adequately established.

    4. Lack of or inadequate process validation

      A process whose results cannot be fully verified by subsequent inspection and test has not been adequately validated according to established procedures.

    5. Design control - no procedures

      Procedures for design control have not been established.

    6. Identification procedures, Lack of or inadequate procedures

      Procedures for identifying product during all stages of receipt, production, distribution, and installation have not been adequately established.

    7. Contamination control, Lack of or inadequate procedures

      Procedures to prevent contamination of equipment or product by substances that may have an adverse effect on product quality have not been adequately established.

    8. Label to bear a unique device identifier

      The labels of the medical device(s) do not bear a unique device identifier (UDI) that meet the requirements of 801 Subpart B and 21 CFR 830.

    9. Training - Lack of or inadequate procedures

      Procedures for training and identifying training needs have not been adequately established.

    10. Lack of or inadequate procedures

      Procedures for corrective and preventive action have not been adequately established.

    11. Devices subject to device identification GUDID data submission requirements.

      The information required by the Global Unique Device Identification Database for each medical device version or model required to bear a unique device identifier (UDI) was not provided.

    12. Software validation for automated processes

      Software used as part of production has not been validated for its intended use according to an established protocol.

  2. OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Compliance: Devices16 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Contamination control, Lack of or inadequate procedures

      Procedures to prevent contamination of equipment or product by substances that may have an adverse effect on product quality have not been adequately established.

    2. Document review, approval documentation

      The documentation of approval of documents does not include the document approval date and the signature of the approving official.

    3. Software validation for automated processes

      Software used as part of production has not been validated for its intended use according to an established protocol. Electronic records are used, but they do not meet systems validation, system access limitation and audit trail requirements to ensure that they are trustworthy, reliable and generally equivalent to paper records.

    4. Lack of or inadequate DHR procedures

      Procedures for device history records have not been adequately established.

    5. Product rework documentation, DHR {see also 820.184}

      Rework and reevaluation activities have not been documented in the device history record.

    Show 11 more citations
    1. Supplier notification of changes

      There is no agreement with suppliers and contractors to notify you of changes in the product or service.

    2. DMR - not or inadequately maintained

      A device master record has not been adequately maintained.

    3. Management representative

      No management representative had been appointed to ensure that quality system requirements are met, and to report to management on the performance of the quality system.

    4. Management ensuring quality policy is understood

      Management with executive responsibility has not ensured that the quality policy is understood, implemented and maintained at all levels of the organization.

    5. Incoming acceptance records, documentation

      Acceptance or rejection of incoming product was not documented.

    6. Lack of or inadequate process validation

      A process whose results cannot be fully verified by subsequent inspection and test has not been adequately validated according to established procedures.

    7. Training - Lack of or inadequate procedures

      Procedures for training and identifying training needs have not been established.

    8. Calibration, Inspection, etc. Procedures Lack of or Inadequ

      Procedures to ensure equipment is routinely calibrated, inspected, checked and maintained have not been established.

    9. Maintenance schedule, Lack of or inadequate schedule

      Schedules for the adjustment, cleaning, and other maintenance of equipment have not been adequately established.

    10. Quality Audit/Reaudit - conducted

      Quality audits have not been performed.

    11. Management review - defined interval, sufficient frequency

      Management with executive responsibility has not reviewed the suitability and effectiveness of the quality system.

  3. VAIObjectionable conditions found; FDA took no regulatory actionDrugs · Drug Quality Assurance6 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Prepared for each batch, include complete information

      Batch production and control records do not include complete information relating to the production and control of each batch.

    2. Cleaning / Sanitizing / Maintenance

      Equipment and utensils are not cleaned, maintained and sanitized at appropriate intervals to prevent contamination that would alter the safety, identity, strength, quality or purity of the drug product.

    3. Procedures for non-sterile drug products

      Procedures designed to prevent objectionable microorganisms in drug products not required to be sterile are not established, written and followed.

    4. Written procedures followed

      Procedures for the preparation of master production and control records are not described in a written procedure and followed.

    5. Investigations of discrepancies, failures

      There is a failure to thoroughly review the failure of a batch or any of its components to meet any of its specifications whether or not the batch has been already distributed.

    Show 1 more citation
    1. Equipment Design, Size and Location

      Equipment used in the manufacture, processing, packing or holding of drug products is not of appropriate design to facilitate operations for its intended use and cleaning and maintenance.

Requested under FOIA

FDA FOIA logs, last synced Nov 23, 2026, 04:25 UTC

Three requests to FDA under the Freedom of Information Act named this site, most recently on Apr 10, 2026. Each one asked for a Form 483.

  • Inspection ending Apr 16, 2025OAI

    3 requests. In the inspection history above. FDA released records in answer to a request for its Form 483 on Jul 11, 2025. FDAgent holds no copy that FDA posted.

    1. Apr 10, 2026FDA-FOIA-2026-3627Form 483Records released
    2. Oct 15, 2025FDA-FOIA-2025-12015Form 483· 1 of 5 sitesOpen
    3. May 15, 2025FDA-FOIA-2025-4437Form 483· 1 of 5 sitesRecords released

A request is listed when it names this site’s FEI number or one of its FDA inspection IDs, or, after review, its exact street address. Requester names are left out.

Regulations FDA cited here

The 25 sections of 21 CFR in FDA’s published citations for this facility’s 3 inspections, with how many of them cited each.

Show all 25

Warning letters and compliance actions

FDA Data Dashboard compliance actions

FDA drug establishment registration

FDA registration search
Registered operations
Manufacturing, labeling
Registration runs through
Dec 31, 2026
Registrant
The Richline Group, Inc.D-U-N-S 962801937

From FDA’s Drug Establishments Current Registration Site, as of Nov 21, 2026. A registration says what a site told FDA it does. It is not an inspection result. FDA’s registration search looks firms up by name.

Go further than this page

FDAgent researches FDA’s records and links every answer to its source.

Free account: 3 questions, no card needed.

About this data. Every record here is filed under FEI 3006749547 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
  • Every record on this page is filed under FEI 3006749547. Records filed under another FEI or name, including other sites of the same company, are not shown here.
  • FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
  • Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
  • “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
  • Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
  • FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.