Xoran Technologies, LLC

5210 S State Rd, Ann Arbor, Michigan, United States

FEI 3004198450

FDA has 7 inspections on file for Xoran Technologies, LLC in Ann Arbor, Michigan (FEI 3004198450), from 2009 to 2024. The latest ended on and was classified NAI, no action indicated. FDA issued a warning letter on (CDRH).

NAI means FDA found no objectionable conditions or practices, or found none that justified further action.

FDA published 4 citations from this inspection, all under 21 CFR Part 820, the device quality system rules. They are listed below. One FOIA request asked FDA for the 483 from a later inspection, ending Aug 24, 2026, that FDA's Data Dashboard doesn't list yet.

The most recent OAI inspection ended on May 14, 2012. No court action or import refusal is filed under this FEI.

17 regulations were cited at more than one inspection; they are listed below.

Sources:FDA firm profile ↗Warning letter ↗Synced Nov 23, 2026

Ask FDAgent about this site

Free account: 3 questions, no card needed. Each answer links to its FDA sources.

Email me when FDA posts a new inspection, Form 483 or warning letter for this site

Weekly, on a free account. Pause it any time.

Watch this facility

Comparing suppliers? Check this site beside 2 others, free

Latest inspection
NAI
Inspections on file
7, 2009–2024
3 OAI · 2 VAI · 2 NAI
Inspections with 483 observations shown
7 of 7
Form 483s posted by FDA
None held for this FEI
Requested under FOIA
1 request, latest Oct 15, 2026
Import refusals
None under this FEI
Import alerts
Not linked by FEI
FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗

Inspection history

7 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC

  1. NAINo objectionable conditions that need follow-upDevices · Radiation Control and Health Safety Act4 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Purchasing controls, Lack of or inadequate procedures

      Procedures to ensure that all purchased or otherwise received product and services conform to specified requirements have not been adequately established.

    2. Design validation - software validation not performed

      Validation of device software is inadequate.

    3. Design Validation - Risk analysis not performed/inadequate

      Risk analysis is inadequate.

    4. Lack of or inadequate complaint procedures

      Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been adequately established.

  2. NAINo objectionable conditions that need follow-upDevices · Radiation Control and Health Safety Act2 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Lack of or inadequate procedures

      Procedures for corrective and preventive action have not been established.

    2. Justification for not reporting

      A justification for not reporting the correction or removal action to FDA that included conclusions, follow-ups and reviews by a designated person was not included in the record.

  3. VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices9 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Providing adequate assembly information

      Instructions for assembly, installation, adjustment, and testing of components adequate to assure that the product will comply with the regulations, when assembled and tested as directed, was not provided to assemblers

    2. Training - Lack of or inadequate procedures

      Procedures for training and identifying training needs have not been adequately established.

    3. FDA not notified of defect or noncompliance

      You did not notify the Secretary, in accordance with 21 CFR 1003.20, of an electronic product which was defective which you produced, assembled, and distributed.

    4. Purchasing controls, Lack of or inadequate procedures

      Procedures to ensure that all purchased or otherwise received product and services conform to specified requirements have not been adequately established.

    5. Lack of or inadequate procedures

      Procedures for corrective and preventive action have not been established.

    Show 4 more citations
    1. Failure to report

      You did not immediately report to the Director, CDRH, FDA, an accidental radiation occurrence reported to or otherwise known to you. involving a product introduced or intended to be introduced into commerce by you.

    2. Lack of or inadequate complaint procedures

      Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been established.

    3. Quality audits - Lack of or inadequate procedures

      Procedures for quality audits have not been adequately established.

    4. Lack of or inadequate process validation

      A process whose results cannot be fully verified by subsequent inspection and test has not been adequately validated according to established procedures.

  4. OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Compliance: Devices19 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Lack of or inadequate procedures

      Procedures for corrective and preventive action have not been established.

    2. Dealers, distributors, purchasers not notified

      You did not notify the purchasers and subsequent transferees of an electronic product which was defective .

    3. Failure to report

      You did not immediately report to the Director, CDRH, FDA, an accidental radiation occurrence reported to or otherwise known to you. involving a product introduced or intended to be introduced into commerce by you.

    4. Product rework procedures, Lack of or inadequate procedures

      Procedures for rework of nonconforming product have not been established.

    5. Training records

      Personnel training is not documented.

    Show 14 more citations
    1. Documentation

      Acceptance activities were not adequately documented.

    2. Investigation of device failures

      Complaints involving the possible failure of a device to meet any of its specifications were not evaluated and investigated where necessary.

    3. Lack of or inadequate receiving acceptance procedures

      Procedures for acceptance of incoming product have not been established.

    4. Lack of or inadequate instructions

      Installation instructions have not been adequately established.

    5. FDA not notified of defect or noncompliance

      You did not notify the Secretary, in accordance with 21 CFR 1003.20, of an electronic product which was defective which you produced, assembled, and distributed.

    6. CAP not submitted

      In a situation where an electronic product you have manufactured has a defect subject to the notification requirements of 21 CFR 1003.10(b), you have not submitted a plan to the Secretary to without charge (including reimbursement of any transportation expenses), replace the product with a like or equivalent one which, or refund the cost of the product to the purchaser.

    7. Production processes

      Production processes were not conducted to ensure that a device conforms to its specifications.

    8. Lack of procedures, or not maintained

      Document control procedures have not been established.

    9. Lack/Inad procedure-Monitoring/Control of Validated Proces

      Procedures for monitoring and control of process parameters for a validated process have not been adequately established.

    10. Management ensuring quality policy is understood

      Management with executive responsibility has not ensured that the quality policy is understood, implemented and maintained at all levels of the organization.

    11. Servicing - Lack of or inadequate procedures

      Procedures or instructions for performing servicing activities and verifying that servicing meets specified requirements have not been adequately established.

    12. Lack of or inadequate process validation

      A process whose results cannot be fully verified by subsequent inspection and test has not been adequately validated according to established procedures.

    13. Certification not based on adequate test/testing program

      Certification was not based upon a testing program in accordance with good manufacturing practices.

    14. Purchasing controls, Lack of or inadequate procedures

      Procedures to ensure that all purchased or otherwise received product and services conform to specified requirements have not been established.

  5. OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Compliance: Devices31 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Management ensuring quality policy is understood

      Management with executive responsibility has not ensured that the quality policy is understood, implemented and maintained at all levels of the organization.

    2. Tube housing assembly mandatory information

      Statements for each tube housing assembly of the minimum filtration permanently in the useful beam expressed as millimeters of aluminum equivalent and the peak tube potential at which the aluminum equivalent was obtained were not provided to purchasers

    3. Failure to submit, distinct marking

      You did not submit a required product report prior to the introduction of the product into commerce.

    4. Compliance measurements--diagnostic source assembly

      Compliance for radiation leakage from the diagnostic source assembly was not determined by measurements averaged over an area of 100 square centimeters with no linear dimension greater than 20 centimeters.

    5. CTDI at 90, 180, 270 degrees

      The computed tomography dose index (CTDI) is not provided to users for locations along lines parallel to the axis of rotation and 1.0 cm interior to the surface of the phantom, at positions 90, 180, and 270 degrees from the position specified in 21 CFR 1020.33(c)(2)(i)(b).

    Show 26 more citations
    1. Calibration, Inspection, etc. Procedures Lack of or Inadequ

      Procedures to ensure equipment is routinely calibrated, inspected, checked, and maintained have not been established.

    2. Purchasing controls, Lack of or inadequate procedures

      Procedures to ensure that all purchased or otherwise received product and services conform to specified requirements have not been established.

    3. Failure to submit

      You did not submit an annual report for products requiring one.

    4. FDA not notified of defect or noncompliance

      You did not notify the Secretary, in accordance with 21 CFR 1003.20, of an electronic product which was defective which you produced, assembled, and distributed.

    5. Failure to report

      You did not immediately report to the Director, CDRH, FDA, an accidental radiation occurrence reported to or otherwise known to you. involving a product introduced or intended to be introduced into commerce by you.

    6. CAP not submitted

      In a situation where an electronic product you have manufactured has a defect subject to the notification requirements of 21 CFR 1003.10(b), you have not submitted a plan to the Secretary to remedy the defect without charge (including reimbursement of any transportation expenses), replace the product with a like or equivalent one which is free of defects relating to safety of its use, or refund the cost of the product to the purchaser.

    7. Performance data for CT conditions of operation - incomplete

      Performance data for the CT conditions of operation used to provide the information required by 21 CFR 1020.33(c)(2)(i) are incomplete.

    8. Generator rating and duty cycle

      A statement of the generator rating and duty cycle was not provided to purchasers.

    9. CTDI for conditions of operation which vary exposure

      The computed tomography dose index (CTDI) in the center location of the dosimetry phantom is not provided to users for each selectable condition of operation that varies either the rate or duration of x-ray exposure.

    10. Beam limiting devices--leakage technique factors

      Leakage technique factors for all combinations of tube housing assemblies and beam-limiting devices for which the beam-limiting device manufacturer states compatibility are not provided for each variable-aperture beam-limiting device to purchasers.

    11. Certification not based on adequate test/testing program

      Certification was not based upon a testing program in accordance with good manufacturing practices.

    12. leakage technique factors, beam limiting device

      A statement of the leakage technique factors for all combinations of tube housing assemblies and beam-limiting devices, for which you state compatibility, was not provided to the users or to others upon request at cost,

    13. Fixed technique factors--max deviation from nominal

      In the case of fixed technique factors, a statement of the maximum deviation from the nominal fixed value of each factor was not provided to purchasers.

    14. Maximum line current (x-ray control + generator)

      A statement of the maximum line current of the x-ray system based on the maximum input voltage and output current characteristics of the tube housing assembly compatible with rated output voltage and rated current characteristics of the x-ray control and associated high-voltage generator was not provided to purchasers.

    15. Measurement criteria for technique factors

      A statement describing the measurement criteria for all technique factors used was not provided to purchasers.

    16. CT conditions of operation indicated prior to scan

      The CT conditions of operation to be used during a scan or scan sequence are not indicated prior to initiation of the scan or scan sequence.

    17. Design changes - Lack of or Inadequate Procedures

      Procedures for design change have not been established.

    18. Dealers, distributors, purchasers not notified

      You did not notify all of the purchasers and subsequent transferees of an electronic product which was defective .

    19. Rated line voltage, range (x-ray control + generator)

      A statement of the rated line voltage and the range of line-voltage regulation for operation at maximum line current was not provided to purchasers.

    20. Lack of or inadequate procedures

      Procedures for corrective and preventive action have not been established.

    21. Lack of or inadequate receiving acceptance procedures

      Procedures for acceptance of incoming product have not been established.

    22. Rationale documented for no investigation

      Records for complaints where no investigation was made do not include required information.

    23. Lack of or inadequate process validation

      A process whose results cannot be fully verified by subsequent inspection and test has not been adequately validated according to established procedures.

    24. Investigation of device failures

      Complaints involving the possible failure of a device to meet any of its specifications were not evaluated and investigated where necessary.

    25. Combination of two or more components

      Prior written authorization to certify a combination of two or more components when installed into a diagnostic x-ray system was not obtained from CDRH.

    26. Report of assembly submitted within 15 days

      A completed report of assembly regarding certified components installed in a diagnostic x-ray system was not submitted to the Director, CDRH, the purchaser, and the State agency responsible for radiation protection within 15 days following the completion of the assembly.

  6. OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Postmarket Assurance: Devices8 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Evaluation of suppliers, contractors, etc., requirements

      Requirements that must be met by suppliers have not been established.

    2. Lack of Written MDR Procedures

      Written MDR procedures have not been implemented.

    3. Review and evaluation for investigation

      Not all complaints have been reviewed and evaluated to determine whether an investigation is necessary.

    4. Management ensuring quality policy is understood

      Management with executive responsibility has not ensured that the quality policy is understood, implemented and maintained at all levels of the organization.

    5. Lack of or inadequate procedures

      Procedures for corrective and preventive action have not been adequately established.

    Show 3 more citations
    1. Report of risk to health

      A correction or removal, conducted to reduce a risk to health posed by a device, was not reported in writing to FDA.

    2. FDA not notified of defect or noncompliance

      You did not notify the Secretary, in accordance with 21 CFR 1003.20, of an electronic product which was defective which you produced, assembled, and distributed.

    3. Quality audits - Lack of or inadequate procedures

      Procedures for quality audits have not been adequately established.

  7. VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices19 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Design output - review and approval

      The approval of design output (including the date and signature of the individual approving the output) was not documented.

    2. Design validation- Lack of or inadequate procedures

      Procedures for validating the device design were not complete.

    3. Approval of purchasing data {see also 820.40(a)}

      The purchasing data were not approved.

    4. Design verification procedures

      Procedures for verifying that design output meets design input were not complete.

    5. Design input approval documentation

      The approval of design input requirements (including the date and signature of the individual approving the requirements) was not documented.

    Show 14 more citations
    1. Management ensuring quality policy is understood

      Management with executive responsibility has not ensured that the quality policy has been fully implemented and maintained at all levels of the organization.

    2. DMR device specifications

      The device master record does not include or refer to the location of all, device, and software specifications.

    3. DMR - not or inadequately maintained

      The device master record does not include or refer to the location of installation procedures and methods.

    4. Approval of process

      Processes have not been approved.

    5. ID label, labeling

      The device history record does not include the primary identification label and labeling for each device.

    6. Design validation - software validation not performed

      Validation of device software is incomplete.

    7. Establish the design and development plan

      A design and development plan has not been documented.

    8. DHR documentation of label release {see also 820.184}

      The DHR does not include records of approval and release of device labeling, including date and signature of the examiner.

    9. Design review - documentation

      The design review results, including identification of the design, the date, and the individual performing the review, were not documented and filed in the design history file.

    10. Design changes - validation vs. verification

      Procedures were not documented to address when verification of design changes is sufficient in lieu of validation prior to their implementation.

    11. Not approved or obsolete document retrieval

      Documents that were not approved were observed at a location where they are being used.

    12. Design transfer - Lack of or inadequate procedures

      Procedures were not documented to ensure that the device design was correctly transferred into production.

    13. Nonconforming product evaluation/investigation

      The investigation of nonconforming product has not been documented.

    14. DHR content

      The device history record does not demonstrate the device is manufactured in accordance with the device master record.

Requested under FOIA

FDA FOIA logs, last synced Nov 23, 2026, 04:25 UTC

One request to FDA under the Freedom of Information Act named this site, on Oct 15, 2026. It asked for a Form 483.

  • Inspection ending Aug 24, 2026

    1 request. Not listed in FDA’s Data Dashboard yet.

    1. Oct 15, 2026FDA-FOIA-2026-10071Form 483· 1 of 5 sitesOpen

A request is listed when it names this site’s FEI number or one of its FDA inspection IDs, or, after review, its exact street address. Requester names are left out.

Regulations FDA cited here

The 29 sections of 21 CFR in FDA’s published citations for this facility’s 7 inspections, with how many of them cited each.

Show all 29

Warning letters and compliance actions

FDA Data Dashboard compliance actions

Go further than this page

FDAgent researches FDA’s records and links every answer to its source.

Free account: 3 questions, no card needed.

About this data. Every record here is filed under FEI 3004198450 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
  • Every record on this page is filed under FEI 3004198450. Records filed under another FEI or name, including other sites of the same company, are not shown here.
  • FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
  • Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
  • “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
  • Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
  • FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.