Xoran Technologies, LLC
5210 S State Rd, Ann Arbor, Michigan, United States
FEI 3004198450
FDA has 7 inspections on file for Xoran Technologies, LLC in Ann Arbor, Michigan (FEI 3004198450), from 2009 to 2024. The latest ended on and was classified NAI, no action indicated. FDA issued a warning letter on (CDRH).
NAI means FDA found no objectionable conditions or practices, or found none that justified further action.
FDA published 4 citations from this inspection, all under 21 CFR Part 820, the device quality system rules. They are listed below. One FOIA request asked FDA for the 483 from a later inspection, ending Aug 24, 2026, that FDA's Data Dashboard doesn't list yet.
The most recent OAI inspection ended on May 14, 2012. No court action or import refusal is filed under this FEI.
17 regulations were cited at more than one inspection; they are listed below.
Sources:FDA firm profile ↗Warning letter ↗Synced Nov 23, 2026
Ask FDAgent about this site
Email me when FDA posts a new inspection, Form 483 or warning letter for this site
Weekly, on a free account. Pause it any time.
Comparing suppliers? Check this site beside 2 others, free
- Latest inspection
- NAI
- Inspections on file
- 7, 2009–2024
- 3 OAI · 2 VAI · 2 NAI
- Inspections with 483 observations shown
- 7 of 7
- Form 483s posted by FDA
- None held for this FEI
- Requested under FOIA
- 1 request, latest Oct 15, 2026
- Warning letters
- · CDRH ↗
- Import refusals
- None under this FEI
- Import alerts
- Not linked by FEI
- FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗
Inspection history
7 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC
- NAINo objectionable conditions that need follow-upDevices · Radiation Control and Health Safety Act4 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Purchasing controls, Lack of or inadequate procedures
Procedures to ensure that all purchased or otherwise received product and services conform to specified requirements have not been adequately established.
Design validation - software validation not performed
Validation of device software is inadequate.
Design Validation - Risk analysis not performed/inadequate
Risk analysis is inadequate.
Lack of or inadequate complaint procedures
Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been adequately established.
- NAINo objectionable conditions that need follow-upDevices · Radiation Control and Health Safety Act2 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Lack of or inadequate procedures
Procedures for corrective and preventive action have not been established.
Justification for not reporting
A justification for not reporting the correction or removal action to FDA that included conclusions, follow-ups and reviews by a designated person was not included in the record.
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices9 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Providing adequate assembly information
Instructions for assembly, installation, adjustment, and testing of components adequate to assure that the product will comply with the regulations, when assembled and tested as directed, was not provided to assemblers
Training - Lack of or inadequate procedures
Procedures for training and identifying training needs have not been adequately established.
FDA not notified of defect or noncompliance
You did not notify the Secretary, in accordance with 21 CFR 1003.20, of an electronic product which was defective which you produced, assembled, and distributed.
Purchasing controls, Lack of or inadequate procedures
Procedures to ensure that all purchased or otherwise received product and services conform to specified requirements have not been adequately established.
Lack of or inadequate procedures
Procedures for corrective and preventive action have not been established.
Show 4 more citations
Failure to report
You did not immediately report to the Director, CDRH, FDA, an accidental radiation occurrence reported to or otherwise known to you. involving a product introduced or intended to be introduced into commerce by you.
Lack of or inadequate complaint procedures
Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been established.
Quality audits - Lack of or inadequate procedures
Procedures for quality audits have not been adequately established.
Lack of or inadequate process validation
A process whose results cannot be fully verified by subsequent inspection and test has not been adequately validated according to established procedures.
- OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Compliance: Devices19 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Lack of or inadequate procedures
Procedures for corrective and preventive action have not been established.
Dealers, distributors, purchasers not notified
You did not notify the purchasers and subsequent transferees of an electronic product which was defective .
Failure to report
You did not immediately report to the Director, CDRH, FDA, an accidental radiation occurrence reported to or otherwise known to you. involving a product introduced or intended to be introduced into commerce by you.
Product rework procedures, Lack of or inadequate procedures
Procedures for rework of nonconforming product have not been established.
Training records
Personnel training is not documented.
Show 14 more citations
Documentation
Acceptance activities were not adequately documented.
Investigation of device failures
Complaints involving the possible failure of a device to meet any of its specifications were not evaluated and investigated where necessary.
Lack of or inadequate receiving acceptance procedures
Procedures for acceptance of incoming product have not been established.
Lack of or inadequate instructions
Installation instructions have not been adequately established.
FDA not notified of defect or noncompliance
You did not notify the Secretary, in accordance with 21 CFR 1003.20, of an electronic product which was defective which you produced, assembled, and distributed.
CAP not submitted
In a situation where an electronic product you have manufactured has a defect subject to the notification requirements of 21 CFR 1003.10(b), you have not submitted a plan to the Secretary to without charge (including reimbursement of any transportation expenses), replace the product with a like or equivalent one which, or refund the cost of the product to the purchaser.
Production processes
Production processes were not conducted to ensure that a device conforms to its specifications.
Lack of procedures, or not maintained
Document control procedures have not been established.
Lack/Inad procedure-Monitoring/Control of Validated Proces
Procedures for monitoring and control of process parameters for a validated process have not been adequately established.
Management ensuring quality policy is understood
Management with executive responsibility has not ensured that the quality policy is understood, implemented and maintained at all levels of the organization.
Servicing - Lack of or inadequate procedures
Procedures or instructions for performing servicing activities and verifying that servicing meets specified requirements have not been adequately established.
Lack of or inadequate process validation
A process whose results cannot be fully verified by subsequent inspection and test has not been adequately validated according to established procedures.
Certification not based on adequate test/testing program
Certification was not based upon a testing program in accordance with good manufacturing practices.
Purchasing controls, Lack of or inadequate procedures
Procedures to ensure that all purchased or otherwise received product and services conform to specified requirements have not been established.
- OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Compliance: Devices31 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Management ensuring quality policy is understood
Management with executive responsibility has not ensured that the quality policy is understood, implemented and maintained at all levels of the organization.
Tube housing assembly mandatory information
Statements for each tube housing assembly of the minimum filtration permanently in the useful beam expressed as millimeters of aluminum equivalent and the peak tube potential at which the aluminum equivalent was obtained were not provided to purchasers
Failure to submit, distinct marking
You did not submit a required product report prior to the introduction of the product into commerce.
Compliance measurements--diagnostic source assembly
Compliance for radiation leakage from the diagnostic source assembly was not determined by measurements averaged over an area of 100 square centimeters with no linear dimension greater than 20 centimeters.
CTDI at 90, 180, 270 degrees
The computed tomography dose index (CTDI) is not provided to users for locations along lines parallel to the axis of rotation and 1.0 cm interior to the surface of the phantom, at positions 90, 180, and 270 degrees from the position specified in 21 CFR 1020.33(c)(2)(i)(b).
Show 26 more citations
Calibration, Inspection, etc. Procedures Lack of or Inadequ
Procedures to ensure equipment is routinely calibrated, inspected, checked, and maintained have not been established.
Purchasing controls, Lack of or inadequate procedures
Procedures to ensure that all purchased or otherwise received product and services conform to specified requirements have not been established.
Failure to submit
You did not submit an annual report for products requiring one.
FDA not notified of defect or noncompliance
You did not notify the Secretary, in accordance with 21 CFR 1003.20, of an electronic product which was defective which you produced, assembled, and distributed.
Failure to report
You did not immediately report to the Director, CDRH, FDA, an accidental radiation occurrence reported to or otherwise known to you. involving a product introduced or intended to be introduced into commerce by you.
CAP not submitted
In a situation where an electronic product you have manufactured has a defect subject to the notification requirements of 21 CFR 1003.10(b), you have not submitted a plan to the Secretary to remedy the defect without charge (including reimbursement of any transportation expenses), replace the product with a like or equivalent one which is free of defects relating to safety of its use, or refund the cost of the product to the purchaser.
Performance data for CT conditions of operation - incomplete
Performance data for the CT conditions of operation used to provide the information required by 21 CFR 1020.33(c)(2)(i) are incomplete.
Generator rating and duty cycle
A statement of the generator rating and duty cycle was not provided to purchasers.
CTDI for conditions of operation which vary exposure
The computed tomography dose index (CTDI) in the center location of the dosimetry phantom is not provided to users for each selectable condition of operation that varies either the rate or duration of x-ray exposure.
Beam limiting devices--leakage technique factors
Leakage technique factors for all combinations of tube housing assemblies and beam-limiting devices for which the beam-limiting device manufacturer states compatibility are not provided for each variable-aperture beam-limiting device to purchasers.
Certification not based on adequate test/testing program
Certification was not based upon a testing program in accordance with good manufacturing practices.
leakage technique factors, beam limiting device
A statement of the leakage technique factors for all combinations of tube housing assemblies and beam-limiting devices, for which you state compatibility, was not provided to the users or to others upon request at cost,
Fixed technique factors--max deviation from nominal
In the case of fixed technique factors, a statement of the maximum deviation from the nominal fixed value of each factor was not provided to purchasers.
Maximum line current (x-ray control + generator)
A statement of the maximum line current of the x-ray system based on the maximum input voltage and output current characteristics of the tube housing assembly compatible with rated output voltage and rated current characteristics of the x-ray control and associated high-voltage generator was not provided to purchasers.
Measurement criteria for technique factors
A statement describing the measurement criteria for all technique factors used was not provided to purchasers.
CT conditions of operation indicated prior to scan
The CT conditions of operation to be used during a scan or scan sequence are not indicated prior to initiation of the scan or scan sequence.
Design changes - Lack of or Inadequate Procedures
Procedures for design change have not been established.
Dealers, distributors, purchasers not notified
You did not notify all of the purchasers and subsequent transferees of an electronic product which was defective .
Rated line voltage, range (x-ray control + generator)
A statement of the rated line voltage and the range of line-voltage regulation for operation at maximum line current was not provided to purchasers.
Lack of or inadequate procedures
Procedures for corrective and preventive action have not been established.
Lack of or inadequate receiving acceptance procedures
Procedures for acceptance of incoming product have not been established.
Rationale documented for no investigation
Records for complaints where no investigation was made do not include required information.
Lack of or inadequate process validation
A process whose results cannot be fully verified by subsequent inspection and test has not been adequately validated according to established procedures.
Investigation of device failures
Complaints involving the possible failure of a device to meet any of its specifications were not evaluated and investigated where necessary.
Combination of two or more components
Prior written authorization to certify a combination of two or more components when installed into a diagnostic x-ray system was not obtained from CDRH.
Report of assembly submitted within 15 days
A completed report of assembly regarding certified components installed in a diagnostic x-ray system was not submitted to the Director, CDRH, the purchaser, and the State agency responsible for radiation protection within 15 days following the completion of the assembly.
- OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Postmarket Assurance: Devices8 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Evaluation of suppliers, contractors, etc., requirements
Requirements that must be met by suppliers have not been established.
Lack of Written MDR Procedures
Written MDR procedures have not been implemented.
Review and evaluation for investigation
Not all complaints have been reviewed and evaluated to determine whether an investigation is necessary.
Management ensuring quality policy is understood
Management with executive responsibility has not ensured that the quality policy is understood, implemented and maintained at all levels of the organization.
Lack of or inadequate procedures
Procedures for corrective and preventive action have not been adequately established.
Show 3 more citations
Report of risk to health
A correction or removal, conducted to reduce a risk to health posed by a device, was not reported in writing to FDA.
FDA not notified of defect or noncompliance
You did not notify the Secretary, in accordance with 21 CFR 1003.20, of an electronic product which was defective which you produced, assembled, and distributed.
Quality audits - Lack of or inadequate procedures
Procedures for quality audits have not been adequately established.
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices19 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Design output - review and approval
The approval of design output (including the date and signature of the individual approving the output) was not documented.
Design validation- Lack of or inadequate procedures
Procedures for validating the device design were not complete.
Approval of purchasing data {see also 820.40(a)}
The purchasing data were not approved.
Design verification procedures
Procedures for verifying that design output meets design input were not complete.
Design input approval documentation
The approval of design input requirements (including the date and signature of the individual approving the requirements) was not documented.
Show 14 more citations
Management ensuring quality policy is understood
Management with executive responsibility has not ensured that the quality policy has been fully implemented and maintained at all levels of the organization.
DMR device specifications
The device master record does not include or refer to the location of all, device, and software specifications.
DMR - not or inadequately maintained
The device master record does not include or refer to the location of installation procedures and methods.
Approval of process
Processes have not been approved.
ID label, labeling
The device history record does not include the primary identification label and labeling for each device.
Design validation - software validation not performed
Validation of device software is incomplete.
Establish the design and development plan
A design and development plan has not been documented.
DHR documentation of label release {see also 820.184}
The DHR does not include records of approval and release of device labeling, including date and signature of the examiner.
Design review - documentation
The design review results, including identification of the design, the date, and the individual performing the review, were not documented and filed in the design history file.
Design changes - validation vs. verification
Procedures were not documented to address when verification of design changes is sufficient in lieu of validation prior to their implementation.
Not approved or obsolete document retrieval
Documents that were not approved were observed at a location where they are being used.
Design transfer - Lack of or inadequate procedures
Procedures were not documented to ensure that the device design was correctly transferred into production.
Nonconforming product evaluation/investigation
The investigation of nonconforming product has not been documented.
DHR content
The device history record does not demonstrate the device is manufactured in accordance with the device master record.
Requested under FOIA
FDA FOIA logs, last synced Nov 23, 2026, 04:25 UTC
One request to FDA under the Freedom of Information Act named this site, on Oct 15, 2026. It asked for a Form 483.
Inspection ending Aug 24, 2026
1 request. Not listed in FDA’s Data Dashboard yet.
- Oct 15, 2026FDA-FOIA-2026-10071Form 483· 1 of 5 sitesOpen
A request is listed when it names this site’s FEI number or one of its FDA inspection IDs, or, after review, its exact street address. Requester names are left out.
Regulations FDA cited here
The 29 sections of 21 CFR in FDA’s published citations for this facility’s 7 inspections, with how many of them cited each.
- 21 CFR 820.50Purchasing controls6 of 7
- 21 CFR 820.100Corrective and preventive action5 of 7
- 21 CFR 820.198Complaint files5 of 7
- 21 CFR 820.20Management responsibility4 of 7
- 21 CFR 1003.10Discovery of defect or failure of compliance by manufacturer; notice requirements4 of 7
- 21 CFR 820.30Design controls3 of 7
- 21 CFR 820.75Process validation3 of 7
- 21 CFR 1002.20Reporting of accidental radiation occurrences3 of 7
Show all 29
- 21 CFR 820.22Quality audit2 of 7
- 21 CFR 820.25Personnel2 of 7
- 21 CFR 820.40[Reserved]2 of 7
- 21 CFR 820.70Production and process controls2 of 7
- 21 CFR 820.80Receiving, in-process, and finished device acceptance2 of 7
- 21 CFR 820.90Nonconforming product2 of 7
- 21 CFR 1010.2Certification2 of 7
- 21 CFR 1020.30Diagnostic x-ray systems and their major components2 of 7
- 21 CFR 1004.1Manufacturer's obligation to repair, replace, or refund cost of electronic products2 of 7
- 21 CFR 803.17What are the requirements for developing, maintaining, and implementing written MDR procedures that apply to me?1 of 7
- 21 CFR 806.10Reports of corrections and removals1 of 7
- 21 CFR 806.20Records of corrections and removals not required to be reported1 of 7
- 21 CFR 820.72Inspection, measuring, and test equipment1 of 7
- 21 CFR 820.120Device labeling1 of 7
- 21 CFR 820.170Installation1 of 7
- 21 CFR 820.181Device master record1 of 7
- 21 CFR 820.184Device history record1 of 7
- 21 CFR 820.200Servicing1 of 7
- 21 CFR 1002.10Product reports1 of 7
- 21 CFR 1002.13Annual reports1 of 7
- 21 CFR 1020.33Computed tomography (CT) equipment1 of 7
Warning letters and compliance actions
FDA Data Dashboard compliance actions
- Warning letterFormal notice of significant violationsCDRH · DevicesLetter on fda.govFDA record
Go further than this page
FDAgent researches FDA’s records and links every answer to its source.
About this data. Every record here is filed under FEI 3004198450 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
- Every record on this page is filed under FEI 3004198450. Records filed under another FEI or name, including other sites of the same company, are not shown here.
- FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
- Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.