21 CFR 806.20: Records of corrections and removals not required to be reported
FDA's inspection data lists 193 Form 483 citations of 21 CFR 806.20 at 182 facilities, from October 2008 to July 31, 2026.
Counts: FDA Data Dashboard citations, last synced Nov 23, 2026, 04:25 UTC, and FDA's warning letters, last synced Nov 23, 2026, 04:17 UTC. Regulation text: eCFR. FDAgent is independent of FDA.
- Form 483 citations
- 193
- At 189 inspections
- Facilities cited
- 182
- Distinct FEIs
- Most recent citation
- Jul 31, 2026
- Lynn Medical Instrument Co.
- Warning letters
- 0
- State a violation of it
The regulation
The current text of 21 CFR 806.20 from the Electronic Code of Federal Regulations.
§ 806.20 Records of corrections and removals not required to be reported.
(a) Each device manufacturer or importer who initiates a correction or removal of a device that is not required to be reported to FDA under § 806.10 shall keep a record of such correction or removal.
(b) Records of corrections and removals not required to be reported to FDA under § 806.10 shall contain the following information:
(1) The brand name, common or usual name, classification, name and product code if known, and the intended use of the device.
(2) The unique device identifier (UDI) of the device, or the device identifier, universal product code (UPC), model, catalog, or code number of the device and the manufacturing lot or serial number of the device or other identification number.
(3) A description of the event(s) giving rise to the information reported and the corrective or removal action that has been, and is expected to be taken.
(4) Justification for not reporting the correction or removal action to FDA, which shall contain conclusions and any followups, and be reviewed and evaluated by a designated person.
(5) A copy of all communications regarding the correction or removal.
(c) The manufacturer or importer shall retain records required under this section for a period of 2 years beyond the expected life of the device, even if the manufacturer or importer has ceased to manufacture or import the device. Records required to be maintained under paragraph (b) of this section must be transferred to the new manufacturer or importer of the device and maintained for the required period of time.
21 CFR Part 806: Medical Devices; Reports of Corrections and Removals, Subpart BRead on eCFR
Citations by year
Form 483 citations of 21 CFR 806.20 by the year the inspection ended. FDA's citation data starts with inspections ending in October 2008, and the current year runs to the latest sync, so both are partial years (lighter bars).
Peak: 22 in 2012
| Year | Form 483 citations |
|---|---|
| 2008 | 3 |
| 2009 | 7 |
| 2010 | 20 |
| 2011 | 18 |
| 2012 | 22 |
| 2013 | 12 |
| 2014 | 13 |
| 2015 | 18 |
| 2016 | 12 |
| 2017 | 15 |
| 2018 | 10 |
| 2019 | 9 |
| 2020 | 3 |
| 2021 | 5 |
| 2022 | 6 |
| 2023 | 4 |
| 2024 | 6 |
| 2025 | 9 |
| 2026 | 1 |
How FDA words this citation
FDA's short descriptions and citation text on these 193 citations, most used first. These are FDA's standard summaries in its citation data, not the text of any one Form 483.
Short descriptions
- Justification for not reporting119 citations
- Records not kept52 citations
- Events9 citations
- Records of communications9 citations
- Device model identification3 citations
Show 1 more
- Device name, identification1 citations
Citation text
- “A justification for not reporting the correction or removal action to FDA that included conclusions, follow-ups and reviews by a designated person was not included in the record.”72 citations
- “There is no record maintained of a correction or removal action that was not required to be reported to FDA.”52 citations
- “A justification for not reporting the correction or removal action to FDA that included conclusions by a designated person was not included in the record.”22 citations
- “A justification for not reporting the correction or removal action to FDA that included conclusions and reviews by a designated person was not included in the record.”12 citations
Show 2 more
- “A copy of all communications regarding the correction or removal action was not contained in the record.”9 citations
- “A description of the events that led to the correction or removal actions was not contained in the record.”9 citations
Facilities cited most recently
The latest inspections where FDA cited 21 CFR 806.20, by inspection end date, with FDA's short description. Each links to the facility's FDA record on FDAgent.
Show 4 more
Often cited at the same inspection
Of the 189 inspections where FDA cited 21 CFR 806.20, the share where it also cited each of these sections.
- 21 CFR 820.100Corrective and preventive action127 (67%)
- 21 CFR 820.198Complaint files98 (52%)
- 21 CFR 820.30Design controls90 (48%)
- 21 CFR 820.50Purchasing controls49 (26%)
- 21 CFR 820.70Production and process controls38 (20%)
- 21 CFR 820.75Process validation38 (20%)
Ask FDAgent about 21 CFR 806.20
FDAgent reads FDA's Form 483 citations, observations and warning letters and answers in plain English, with links to the records.
Related
Topics that count this section:Device CAPA and complaints
About this data
- Citation counts come from FDA's Data Dashboard, which lists the regulation behind each Form 483 observation for inspections since October 2008.
- Observation excerpts come from Form 483 PDFs in FDA's electronic reading room, tied to this section only by exact wording as described above.
- A warning letter counts for this section when one of its numbered violation sentences cites 21 CFR 806.20; FDAgent quotes that sentence exactly.