Prentke Romich Company DBA PRC-Saltillo
1022 Heyl Rd, Wooster, Ohio, United States
FEI 3010819851
FDA has 5 inspections on file for Prentke Romich Company DBA PRC-Saltillo in Wooster, Ohio (FEI 3010819851), from 2009 to 2025. The latest ended on and was classified NAI, no action indicated.
NAI means FDA found no objectionable conditions or practices, or found none that justified further action.
FDA published 5 citations from this inspection: 4 under 21 CFR Part 820, the device quality system rules, and 1 under 21 CFR Part 806, the device correction and removal reporting rules. They are listed below. One FOIA request has asked FDA for this inspection's 483.
No warning letter, court action or import refusal is filed under this FEI.
Three regulations were cited at more than one inspection: 21 CFR 820.198(a), 21 CFR 820.100(a) and 21 CFR 820.30(g).
Sources:FDA firm profile ↗Synced Nov 23, 2026
Ask FDAgent about this site
Email me when FDA posts a new inspection, Form 483 or warning letter for this site
Weekly, on a free account. Pause it any time.
Comparing suppliers? Check this site beside 2 others, free
- Latest inspection
- NAI
- Inspections on file
- 5, 2009–2025
- 0 OAI · 4 VAI · 1 NAI
- Inspections with 483 observations shown
- 4 of 5
- Form 483s posted by FDA
- None held for this FEI
- Requested under FOIA
- 1 request, latest Jul 10, 2025
- Warning letters
- None on file
- Import refusals
- None under this FEI
- Import alerts
- Not linked by FEI
- FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗
Inspection history
5 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC
- NAINo objectionable conditions that need follow-upDevices · Postmarket Assurance: Devices5 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Lack of or inadequate DHR procedures
Procedures for device history records have not been established.
Lack of or inadequate process validation
A process whose results cannot be fully verified by subsequent inspection and test has not been validated according to established procedures.
Records not kept
There is no record maintained of a correction or removal action that was not required to be reported to FDA.
Documentation
Corrective and preventive action activities and/or results have not been documented.
Lack of or inadequate complaint procedures
Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been established.
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices4 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Lack of or inadequate complaint procedures
Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been adequately established.
Lack of or inadequate procedures
Procedures for corrective and preventive action have not been adequately established.
Design validation - Risk analysis
Results of the design risk analysis were not adequately documented.
Design changes - Lack of or Inadequate Procedures
Procedures for design change have not been adequately established.
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices3 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Lack of or inadequate complaint procedures
Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been adequately established.
Design Validation - Risk analysis not performed/inadequate
Risk analysis is incomplete.
Lack of or inadequate procedures
Procedures for corrective and preventive action have not been adequately established.
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices3 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Design review - Lack of or inadequate procedures
Procedures for design review have not been adequately established.
Lack of or inadequate final acceptance procedures
Procedures for finished device acceptance have not been adequately established.
Lack of or inadequate procedures
Procedures for corrective and preventive action have not been established.
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: DevicesCitations not published
Requested under FOIA
FDA FOIA logs, last synced Nov 23, 2026, 04:25 UTC
One request to FDA under the Freedom of Information Act named this site, on Jul 10, 2025. It asked for a Form 483.
Inspection ending May 1, 2025NAI
1 request. In the inspection history above. FDA released records in answer to a request for its Form 483 on Jul 31, 2025. FDAgent holds no copy that FDA posted.
- Jul 10, 2025FDA-FOIA-2025-8502Form 483· 1 of 5 sitesRecords released
A request is listed when it names this site’s FEI number or one of its FDA inspection IDs, or, after review, its exact street address. Requester names are left out.
Regulations FDA cited here
The 7 sections of 21 CFR in FDA’s published citations for 4 of this facility’s 5 inspections, with how many of them cited each.
- 21 CFR 820.100Corrective and preventive action4 of 4
- 21 CFR 820.30Design controls3 of 4
- 21 CFR 820.198Complaint files3 of 4
- 21 CFR 806.20Records of corrections and removals not required to be reported1 of 4
- 21 CFR 820.75Process validation1 of 4
- 21 CFR 820.80Receiving, in-process, and finished device acceptance1 of 4
- 21 CFR 820.184Device history record1 of 4
Go further than this page
FDAgent researches FDA’s records and links every answer to its source.
About this data. Every record here is filed under FEI 3010819851 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
- Every record on this page is filed under FEI 3010819851. Records filed under another FEI or name, including other sites of the same company, are not shown here.
- FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
- Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
- “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
- Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
- FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.