FDA has 7 inspections on file for Rockwell Medical, Inc in Wixom, Michigan (FEI 3001236991), from 2009 to 2025. The latest ended on and was classified VAI, voluntary action indicated.
VAI means FDA investigators found objectionable conditions or practices, but FDA was not prepared to take or recommend regulatory action. The firm is expected to correct them voluntarily.
FDA published 3 citations from this inspection: 2 under 21 CFR Part 820, the device quality system rules, and 1 under 21 CFR Part 806, the device correction and removal reporting rules. They are listed below. Two of the three FOIA requests for this site have asked FDA for this inspection's 483.
No warning letter, court action or import refusal is filed under this FEI.
Two regulations were cited at more than one inspection: 21 CFR 820.75(a) and 21 CFR 820.22.
Sources:FDA firm profile ↗Synced Nov 23, 2026
Ask FDAgent about this site
Email me when FDA posts a new inspection, Form 483 or warning letter for this site
Weekly, on a free account. Pause it any time.
Comparing suppliers? Check this site beside 2 others, free
- Latest inspection
- VAI
- Inspections on file
- 7, 2009–2025
- 0 OAI · 1 VAI · 6 NAI
- Inspections with 483 observations shown
- 4 of 7
- Form 483s posted by FDA
- None held for this FEI
- Requested under FOIA
- 3 requests, latest Jan 14, 2026
- Warning letters
- None on file
- Import refusals
- None under this FEI
- Import alerts
- Not linked by FEI
- FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗
Inspection history
7 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices3 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Lack of or inadequate procedures
Procedures for corrective and preventive action have not been adequately established.
Justification for not reporting
A justification for not reporting the correction or removal action to FDA that included conclusions by a designated person was not included in the record.
Remedial action
When test/measurement equipment was found to not meet accuracy and precision limits, inadequate action was taken to evaluate whether there was any adverse effect on the device's quality.
- NAINo objectionable conditions that need follow-upDevices · Postmarket Assurance: DevicesNone
- NAINo objectionable conditions that need follow-upDevices · Compliance: DevicesNone
- NAINo objectionable conditions that need follow-upDevices · Postmarket Assurance: Devices6 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Design validation - production units
The design was not validated under defined operating conditions and using initial production units, lots or batches or their equivalents.
Documentation
Process validation activities and results have not been adequately documented.
Design transfer - Lack of or inadequate procedures
Procedures for design transfer have not been adequately established.
Purchasing controls, Lack of or inadequate procedures
Procedures to ensure that all purchased or otherwise received product and services conform to specified requirements have not been adequately established.
Design review - Lack of or inadequate procedures
Procedures for design review have not been adequately established.
Show 1 more citation
Design verification - Lack of or inadequate procedures
Procedures for design verification have not been adequately established.
- NAINo objectionable conditions that need follow-upDevices · Compliance: DevicesNone
- NAINo objectionable conditions that need follow-upDevices · Postmarket Assurance: Devices4 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Packaging
Device packaging and/or shipping containers are not designed and constructed to protect the device from alteration or damage during processing, storage, handling, and distribution.
Design history file
The design history file does not demonstrate that the design was developed following the requirements of 21 CFR 820.
Documentation
Process validation activities and results have not been adequately documented and adequately approved.
Quality audits - auditor independence
Individuals who conduct quality audits have direct responsibility for the matters being audited.
- NAINo objectionable conditions that need follow-upDevices · Postmarket Assurance: Devices5 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Statistical techniques - Lack of or inadequate procedures
Procedures to identify valid statistical techniques needed for certain product analyses have not been established.
Equipment control activity documentation
Equipment calibrations have not been documented.
Production monitoring and control
During production, process parameters are not monitored and controlled.
Quality audits - Lack of or inadequate procedures
Procedures for conducting quality audits were not complete.
Report of violation of the Act (see 803.52(e)(9))
A violation of the FD&C Act involving a device which might present a risk to health was not reported to FDA.
Requested under FOIA
FDA FOIA logs, last synced Nov 23, 2026, 04:25 UTC
Three requests to FDA under the Freedom of Information Act named this site, most recently on Jan 14, 2026.
Inspection ending Nov 20, 2025VAI
2 requests. In the inspection history above. FDA released records in answer to a request for its Form 483 on Dec 31, 2025. FDAgent holds no copy that FDA posted.
- Jan 14, 2026FDA-FOIA-2026-496Form 483· 1 of 5 sitesRecords released
- Dec 18, 2025FDA-FOIA-2025-14165Form 483Records released
Inspection ending Sep 9, 2022NAI
1 request. In the inspection history above. FDA released records in answer to a request for its EIR on Jan 28, 2026. FDAgent holds no copy that FDA posted.
- Jun 24, 2025FDA-FOIA-2025-7968EIRRecords released
A request is listed when it names this site’s FEI number or one of its FDA inspection IDs, or, after review, its exact street address. Requester names are left out.
Regulations FDA cited here
The 11 sections of 21 CFR in FDA’s published citations for 4 of this facility’s 7 inspections, with how many of them cited each.
- 21 CFR 820.22Quality audit2 of 4
- 21 CFR 820.30Design controls2 of 4
- 21 CFR 820.72Inspection, measuring, and test equipment2 of 4
- 21 CFR 820.75Process validation2 of 4
- 21 CFR 806.10Reports of corrections and removals1 of 4
- 21 CFR 806.20Records of corrections and removals not required to be reported1 of 4
- 21 CFR 820.50Purchasing controls1 of 4
- 21 CFR 820.70Production and process controls1 of 4
Show all 11
- 21 CFR 820.100Corrective and preventive action1 of 4
- 21 CFR 820.130Device packaging1 of 4
- 21 CFR 820.250Statistical techniques1 of 4
Go further than this page
FDAgent researches FDA’s records and links every answer to its source.
About this data. Every record here is filed under FEI 3001236991 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
- Every record on this page is filed under FEI 3001236991. Records filed under another FEI or name, including other sites of the same company, are not shown here.
- FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
- Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
- “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
- Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
- FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.