21 CFR 820.130: Device packaging
FDA's inspection data lists 28 Form 483 citations of 21 CFR 820.130 at 22 facilities, from July 2009 to November 26, 2024. eCFR shows this section removed on February 4, 2026.
Counts: FDA Data Dashboard citations, last synced Nov 23, 2026, 04:25 UTC, and FDA's warning letters, last synced Nov 23, 2026, 04:17 UTC. Regulation text: eCFR. FDAgent is independent of FDA.
- Form 483 citations
- 28
- At 28 inspections
- Facilities cited
- 22
- Distinct FEIs
- Most recent citation
- Nov 26, 2024
- BC3 TECHNOLOGIES
- Warning letters
- 0
- State a violation of it
The regulation
Removed from the CFR on 2026-02-04. The text shown is the last text in force before then (eCFR version of 2016-12-30).
§ 820.130 Device packaging.
Each manufacturer shall ensure that device packaging and shipping containers are designed and constructed to protect the device from alteration or damage during the customary conditions of processing, storage, handling, and distribution.
21 CFR Part 820: Quality Management System Regulation, Subpart KRead on eCFR
Citations by year
Form 483 citations of 21 CFR 820.130 by the year the inspection ended. FDA's citation data starts with inspections ending in October 2008, and the current year runs to the latest sync, so both are partial years (lighter bars).
Peak: 6 in 2013
| Year | Form 483 citations |
|---|---|
| 2009 | 1 |
| 2010 | 2 |
| 2011 | 2 |
| 2012 | 3 |
| 2013 | 6 |
| 2014 | 4 |
| 2015 | 4 |
| 2016 | 1 |
| 2017 | 2 |
| 2018 | 1 |
| 2021 | 1 |
| 2024 | 1 |
How FDA words this citation
FDA's short descriptions and citation text on these 28 citations, most used first. These are FDA's standard summaries in its citation data, not the text of any one Form 483.
Short descriptions
- Packaging28 citations
Citation text
- “Device packaging and/or shipping containers are not designed and constructed to protect the device from alteration or damage during processing, storage, handling, and distribution.”26 citations
- “Device packaging and/or shipping containers are not designed and constructed well enough to protect the device from damage during processing, storage, handling, and distribution.”2 citations
Facilities cited most recently
The latest inspections where FDA cited 21 CFR 820.130, by inspection end date, with FDA's short description. Each links to the facility's FDA record on FDAgent.
Show 4 more
Often cited at the same inspection
Of the 28 inspections where FDA cited 21 CFR 820.130, the share where it also cited each of these sections.
- 21 CFR 820.75Process validation19 (68%)
- 21 CFR 820.100Corrective and preventive action18 (64%)
- 21 CFR 820.198Complaint files18 (64%)
- 21 CFR 820.30Design controls16 (57%)
- 21 CFR 803.17What are the requirements for developing, maintaining, and implementing written MDR procedures that apply to me?11 (39%)
- 21 CFR 820.80Receiving, in-process, and finished device acceptance11 (39%)
Ask FDAgent about 21 CFR 820.130
FDAgent reads FDA's Form 483 citations, observations and warning letters and answers in plain English, with links to the records.
Related
About this data
- Citation counts come from FDA's Data Dashboard, which lists the regulation behind each Form 483 observation for inspections since October 2008.
- Observation excerpts come from Form 483 PDFs in FDA's electronic reading room, tied to this section only by exact wording as described above.
- A warning letter counts for this section when one of its numbered violation sentences cites 21 CFR 820.130; FDAgent quotes that sentence exactly.
- FDAgent rebuilds this page every week from FDA's public data and is independent of FDA.