Analytical Control Systems, Inc

9058 Technology Dr, Fishers, Indiana, United States

FEI 1825740

FDA has 6 inspections on file for Analytical Control Systems, Inc in Fishers, Indiana (FEI 1825740), from 2010 to 2026. The latest ended on and was classified VAI, voluntary action indicated.

VAI means FDA investigators found objectionable conditions or practices, but FDA was not prepared to take or recommend regulatory action. The firm is expected to correct them voluntarily.

FDA posted the Form 483 from that inspection. FDA published 8 citations from this inspection. They are listed below. Two FOIA requests have asked FDA for this inspection's 483.

The most recent OAI inspection ended on Feb 29, 2016. FDA issued a warning letter on (CDRH). No court action or import refusal is filed under this FEI.

Six regulations were cited at more than one inspection; they are listed below.

Sources:FDA firm profile ↗Form 483 PDF ↗Synced Nov 23, 2026

Ask FDAgent about this site

Free account: 3 questions, no card needed. Each answer links to its FDA sources.

Email me when FDA posts a new inspection, Form 483 or warning letter for this site

Weekly, on a free account. Pause it any time.

Watch this facility

Comparing suppliers? Check this site beside 2 others, free

Latest inspection
VAI
Inspections on file
6, 2010–2026
2 OAI · 4 VAI · 0 NAI
Inspections with 483 observations shown
5 of 6
Form 483s posted by FDA
↗
Requested under FOIA
2 requests, latest May 13, 2026
Import refusals
None under this FEI
Import alerts
Not linked by FEI
FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗

Inspection history

6 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC

  1. VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: DevicesPosted 483 · 8 citations

    Read this Form 483 on its own page →Form 483 PDF on fda.gov

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. ISO 13485:2016 Clause 7.1

      Product realization risk management processes not documented

      One or more processes for risk management in product realization have not been adequately documented.

    2. ISO 13485:2016 Clause 8.5.2

      Corrective action procedures not documented

      Corrective action procedures defining the requirements for evaluating the need for action to ensure that nonconformities do not recur, planning and documenting action needed and implementing such action, verifying that the corrective action does not adversely affect the ability to meet applicable regulatory requirements or the safety and performance of the medical device and reviewing the effectiveness of corrective action taken have not been adequately documented.

    3. ISO 13485:2016 Clause 6.3

      Requirements for infrastructure not documented

      Infrastructure requirements needed to achieve conformity to product requirements have not been adequately documented.

    4. ISO 13485:2016 Clause 8.2.2

      Lack of documented justification for not investigating complaint(s)

      Justification for not investigating a complaint(s) was not adequately documented.

    5. ISO 13485:2016 Clause 8.2.4

      Internal audit procedures not documented

      Procedures to describe the responsibilities and requirements for planning and conducting internal audits and recording and reporting audit results were not adequately documented.

    Show 3 more citations
    1. ISO 13485:2016 Clause 7.5.6

      Procedures for validation of processes not documented

      Procedures for validation of processes were not adequately documented.

    2. ISO 13485:2016 Clause 5.6.1

      Management review procedures not documented

      Procedures for management review have not been adequately documented.

    3. ISO 13485:2016 Clause 4.1.2

      Risk-based approach not applied to QMS processes

      A risk-based approach to the control of the appropriate processes needed for the quality management system has not been adequately applied.

  2. VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices4 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Quality audits - Lack of or inadequate procedures

      Procedures for quality audits have not been adequately established.

    2. Lack of or inadequate process validation

      A process whose results cannot be fully verified by subsequent inspection and test has not been adequately validated according to established procedures.

    3. Management review - Lack of or inadequate procedures

      Procedures for management review have not been adequately established.

    4. Calibration, Inspection, etc. Procedures Lack of or Inadequ

      Procedures to ensure equipment is routinely checked and maintained have not been adequately established.

  3. VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: DevicesCitations not published
  4. OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Compliance: Devices8 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. DMR - not or inadequately maintained

      A device master record has not been maintained.

    2. Management review dates

      The results and/or dates of management reviews are not documented.

    3. Maintenance schedule, Lack of or inadequate schedule

      Schedules for the adjustment, cleaning, and other maintenance of equipment have not been established.

    4. Nonconforming product, Lack of or inadequate procedures

      Procedures have not been adequately established to control product that does not conform to specified requirements.

    5. Training - Lack of or inadequate procedures

      Procedures for training and identifying training needs have not been adequately established.

    Show 3 more citations
    1. Packaging

      Device packaging and/or shipping containers are not designed and constructed to protect the device from alteration or damage during processing, storage, handling, and distribution.

    2. Lack of or inadequate process validation

      A process whose results cannot be fully verified by subsequent inspection and test has not been validated according to established procedures.

    3. Quality Audit/Reaudit - conducted

      Quality audits and reaudits have not been performed.

  5. OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Compliance: Devices12 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Procedures not adequately established or maintained

      Document control procedures have not been adequately established and maintained.

    2. Management review - Lack of or inadequate procedures

      Procedures for management review have not been established.

    3. Lack of or inadequate procedures

      Procedures for corrective and preventive action have not been established.

    4. Nonconforming product, Lack of or inadequate procedures

      Procedures have not been adequately established to control product that does not conform to specified requirements.

    5. Report of violation of the Act (see 803.52(e)(9))

      A violation of the FD&C Act involving a device which might present a risk to health was not reported to FDA.

    Show 7 more citations
    1. Lack of or inadequate process validation

      A process whose results cannot be fully verified by subsequent inspection and test has not been validated according to established procedures.

    2. Design changes - Lack of or Inadequate Procedures

      Procedures for design change have not been established.

    3. Packaging

      Device packaging and/or shipping containers are not designed and constructed to protect the device from alteration or damage during processing, storage, handling, and distribution.

    4. Production and Process Change Procedures, lack of or Inad.

      Procedures for changes to a method, process, and procedure have not been adequately established.

    5. Quality audits - Lack of or inadequate procedures

      Procedures for quality audits have not been adequately established.

    6. Lack of Written MDR Procedures

      Written MDR procedures have not been developed, maintained, and implemented.

    7. Lack of or inadequate final acceptance procedures

      Procedures for finished device acceptance have not been adequately established.

  6. VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices3 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Lack of or inadequate final acceptance procedures

      Procedures for acceptance or rejection of finished device production runs, lots, or batches were not implemented.

    2. Procedure for analysis of data sources

      The corrective and preventive procedures addressing the analysis of sources of quality data to identify existing and potential causes of nonconforming product or other quality problems were not complete.

    3. Quality Audit/Reaudit - conducted

      Quality audits were not conducted at sufficient regular intervals, as prescribed by internal procedures to verify that the quality system is effective in fulfilling your quality system objectives.

Requested under FOIA

FDA FOIA logs, last synced Nov 23, 2026, 04:25 UTC

Two requests to FDA under the Freedom of Information Act named this site, most recently on May 13, 2026. Each one asked for a Form 483.

  • Inspection ending Mar 30, 2026VAI

    2 requests. In the inspection history above. FDA has posted its Form 483.

    1. May 13, 2026FDA-FOIA-2026-4917Form 483· 1 of 5 sitesRecords released
    2. Apr 28, 2026FDA-FOIA-2026-4217Form 483Records released

A request is listed when it names this site’s FEI number or one of its FDA inspection IDs, or, after review, its exact street address. Requester names are left out.

Regulations FDA cited here

The 15 sections of 21 CFR in FDA’s published citations for 5 of this facility’s 6 inspections, with how many of them cited each.

Show all 15

Warning letters and compliance actions

FDA Data Dashboard compliance actions

  • Warning letterFormal notice of significant violationsCDRH · DevicesFDA record

FDAgent links a letter’s text only when FDA’s compliance case number ties it to this FEI. No posted letter matched this one, so the link opens FDA’s record of the action.

Go further than this page

FDAgent researches FDA’s records and links every answer to its source.

Free account: 3 questions, no card needed.

About this data. Every record here is filed under FEI 1825740 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
  • Every record on this page is filed under FEI 1825740. Records filed under another FEI or name, including other sites of the same company, are not shown here.
  • FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
  • Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.