Skip to content

Analytical Control Systems, Inc. Form 483, March 2026

FDA issued this Form 483 to Analytical Control Systems, Inc. in Fishers, Indiana (FEI 1825740) at the end of a device inspection that ended March 30, 2026. FDA classified the inspection Voluntary Action Indicated (VAI).

A Form 483 lists the investigators' observations at the end of an inspection. It is not FDA's final determination.

Source: FDA's Electronic Reading Room, posted Nov 13, 2026. Form 483 PDF on fda.gov. FDAgent is independent of FDA.

Inspection ended
Mar 30, 2026
Devices · Compliance: Devices
FDA's outcome
VAI
Voluntary Action Indicated
Posted by FDA
Nov 13, 2026
228 days after the inspection
Facility
FEI 1825740
Fishers, Indiana

FDA's 8 citations from this inspection

FDA's citation titles and text from its Data Dashboard. These are FDA's summaries, not the Form 483 itself. FDAgent hasn't verified a reading of this PDF, so the Form 483's own words are in the PDF only.

  1. ISO 13485:2016 Clause 7.1

    Product realization risk management processes not documented

    One or more processes for risk management in product realization have not been adequately documented.

  2. ISO 13485:2016 Clause 8.5.2

    Corrective action procedures not documented

    Corrective action procedures defining the requirements for evaluating the need for action to ensure that nonconformities do not recur, planning and documenting action needed and implementing such action, verifying that the corrective action does not adversely affect the ability to meet applicable regulatory requirements or the safety and performance of the medical device and reviewing the effectiveness of corrective action taken have not been adequately documented.

  3. ISO 13485:2016 Clause 6.3

    Requirements for infrastructure not documented

    Infrastructure requirements needed to achieve conformity to product requirements have not been adequately documented.

  4. ISO 13485:2016 Clause 8.2.2

    Lack of documented justification for not investigating complaint(s)

    Justification for not investigating a complaint(s) was not adequately documented.

  5. ISO 13485:2016 Clause 8.2.4

    Internal audit procedures not documented

    Procedures to describe the responsibilities and requirements for planning and conducting internal audits and recording and reporting audit results were not adequately documented.

  6. ISO 13485:2016 Clause 7.5.6

    Procedures for validation of processes not documented

    Procedures for validation of processes were not adequately documented.

  7. ISO 13485:2016 Clause 5.6.1

    Management review procedures not documented

    Procedures for management review have not been adequately documented.

  8. ISO 13485:2016 Clause 4.1.2

    Risk-based approach not applied to QMS processes

    A risk-based approach to the control of the appropriate processes needed for the quality management system has not been adequately applied.

Ask FDAgent about this Form 483

FDAgent reads the 483, the facility's inspections, warning letters and FDA's citations and answers in plain English, with links to the records.

Edit it or ask as written. 3 questions free, no card needed.

About this data

  • The PDF and its dates come from FDA's Electronic Reading Room, last synced Nov 23, 2026, 04:33 UTC; the inspection, its outcome and FDA's citations from FDA's Data Dashboard.
  • FDAgent shows a Form 483’s own text only when its reading of that PDF passes every check.
  • FDAgent leaves out the names of the people the Form 483 is addressed to and of the investigators.
  • FDAgent is independent of FDA.